The text of a manual · part 11 of 26
NCR 7870 Scanner/Scale Release 1.0 User's Guide: Chapter 5: Calibration, pages 87 to 95
Pages 87 to 95 of 228, from NCR 7870 Scanner/Scale Release 1.0 User's Guide
The original document is the PDF (1.3 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the NCR 7870
Page 87
Chapter 5: Calibration
General This chapter describes when and how to calibrate the NCR 7870 scanner/scale models and how to verify that the calibration is correct. Please note, this chapter pertains to the NCR 7870-2000 and 4500 ONLY. The NCR 7870-4000 does not require calibration. In the event calibration is needed, contact the NCR Service Center. The 7870-4000 has NO user serviceable scale parts or calibration procedure. During factory testing, the scale is calibrated one or more times to test the scale calibrating function. This calibration test is not sufficient to make the scale ready for weighing in commerce. To be in compliance with governmental Weights and Measures regulations, you must be certified to perform the scale calibration procedure and place the scale into service for trade. You MUST calibrate the scale when any of the following occurs. •
Initial installation of an NCR 7870 Scanner/Scale
•
When scale cannot be zeroed even after a power down/up sequence
•
When the diagnostics indicate a calibration error that persists after a power down/up sequence
•
When the Load Cell or Scale Board have been replaced
Note: Your scanner/scale may or may not have an Integrated Display or an NCR 7825 Remote Display when connected to some host terminals. If this is the case, a Field Service Calibration Display (NCR part number 250-0029750) is required unless your NCR 7870 has the optional Voice Messaging feature enabled. Calibration information is NOT sent to the host terminal.
Page 88
5-2
Chapter 5: Calibration
The calibration procedure sets the Load Cell and the electronics to interpret the weight of an item accurately. You can calibrate the scale after power has been supplied for 30 minutes if the ambient air conditions have been 20° C (68° F) for at least 24 hours. If the ambient air conditions have not been met, then the scale must be on for at least 6 hours before you can calibrate it. One or more certified weight sets are required to calibrate and certify the scale. •
Whole Pound Weight Set: NCR Part Number 998-0633009
•
Fractional Pound Weight Set: NCR Part Number 998-0633012
•
Kilogram Weight Set: Obtain locally
The EEPROM on the Scale Board maintains an audit trail of scale calibration and weigh parameter setting. The audit trail consists of two event counters. You can display the audit trail count by pressing and holding the Scale Zero button on the Operator Display Panel. The display alternates between Cal xxx and PAr xxx. The Cal value is the number of times the scale has been calibrated. The PAr value is associated with the weigh parameter setting and should never change. Note: The audit trail can be displayed only on units with the Integrated Display or with the NCR 7825 Remote Display.
Page 89
Chapter 5: Calibration
5-3
Calibration Procedure Section 1: Exercising the Scale Tools One of the following: Whole Pound Weight Set Fractional Pound Weight Set Kilogram Weight Set
(P/N 998-0633009) (P/N 998-0633012) Obtain locally
Instructions You must exercise the scale before performing a calibration. 1. Select appropriate Weight Feature for your Scanner/Scale.
Weight Feature
Total Weight on Top Plate
9.995 kg
0.0 kg
2.5 kg
5.0 kg
10.0 kg
5.0 kg
2.5 kg
0.0 kg
13.995 kg
0.0 kg
2.5 kg
5.0 kg
15.0 kg
5.0 kg
2.5 kg
0.0 kg
30.0 lb.
0.0 lb.
5.0 lb.
15.0 lb.
30.0 lb.
15.0 lb.
5.0 lb.
0.0 lb.
2. Add and remove weights to your scale in order given in chart. This is a building procedure in which you start with no weight (0) and sequentially add and remove weight to obtain the values shown in the chart. 3. Go through procedure four times. 4. Go to Section 2: Accessing the Calibration Switch.
Page 90
5-4
Chapter 5: Calibration
Section 2: Accessing the Calibration Switch Tools Screwdriver, flat blade Needle-nosed pliers (optional) The Scale Calibration Switch is located on the Scale Board. For security this switch is not readily accessible and may be secured with a lead/wire seal or seal sticker.
Instructions 1. Remove power from unit. 2. Remove Top Plate: a)
Press lift tab in.
b)
Grasp lift tab and lift Top Plate off unit.
Lift Tab
Top Plate Assembly
11041
3. Remove Subplate Assembly. a)
Pop cover off of Lift Strap Assembly.
Page 91
Chapter 5: Calibration b)
5-5
Remove screw securing Lift Strap to unit.
Cover/Lift Strap Assembly
Subplate Assembly
Tab Slot
11053 c)
Insert screwdriver blade into Tab Slot. Twist to free tab.
d)
Lift the front of Subplate Assembly into a nearly vertical position. Pull the back edge from under the Tower Cabinet and remove the Subplate Assembly.
Subplate Assembly
Horizontal Window
11039
Page 92
5-6
Chapter 5: Calibration
4. Remove the Scale Board Shield. a)
Break seal wire and seal sticker if present.
b)
Remove thumbscrew. Use needle-nosed pliers to loosen if thumbscrew is tight.
c)
Remove Shield.
Scale Board Shield
Calibration Switch (under shield)
Scale Hinge Assembly 11040
5. Set Top Plate on rubber bumpers of Scale Hinge Assembly. Make sure bumpers fit into corresponding sockets on bottom of Top Plate. 6. Apply power to 7870. 7. Does the unit have any of the following: an Integrated Display, a 7825 Remote Display, or the Voice Messaging Feature (must be enabled)? NO → Go to Section 3 - Connecting the Field Service Calibration Display YES → Go to Section 4 – Performing the Calibration
Page 93
Chapter 5: Calibration
5-7
Section 3: Connecting the Field Service Calibration Display Tools Field Service Calibration Display Philips screwdriver
(P/N 250-0029750)
The Field Service Calibration Display provides calibration information when no other display is available. To connect the Field Service Calibration Display, perform the following.
Instructions 1. Remove Tower Cabinet. Note: On some models, taking the Top Plate off first may make Tower Cabinet removal easier. a)
Remove four (4) screws from Tower Cabinet base.
b)
Carefully lift Tower Cabinet straight off unit. Check Indicator Panel harness to ensure it is still connected.
Tower Cabinet
Indicator Panel Harness 16828
Page 94
5-8
Chapter 5: Calibration c)
Connect the Field Service Calibration Display to J-4 on the LCDigital Board (J-5 on the old Digital Board).
J3
J4
DS1 J3
J5 J1
J2
Integrated Display Connector
Old Style Digital Board
LC Digital Board
11279 d)
Apply power to NCR 7870.
e)
Go to Section 4: Performing the Calibration
Page 95
Chapter 5: Calibration
5-9
Section 4: Performing the Calibration The scale firmware controls calibration. It waits for you to respond to the prompt before going to the next step. When you have placed the weight on the Top Plate and pressed the Scale Zero button, the firmware sounds a single tone and changes the display to the next prompt. If voice is enabled, the scale tells you which weight to put on next. You can end the procedure before you have completed the calibration function by turning the unit off. However, if you do this, you must still calibrate the scale before placing it into service. Note: the pound and kilogram weights used for calibration are not equivalent values. They are the actual weights the firmware needs to perform the calibration.
Instructions For the purpose of this procedure, a scale set-up to weigh pounds is used. Kilogram scales follow the same instructions but use the appropriate weights (shown in parentheses). 1. Press the Reset/Scale Zero button to display the Cal and PAr values. Record these values. 2. Remove Top Plate and press the Calibration Switch on the Scale Board. 3. Immediately reinstall Top Plate. Ensure that rubber bumpers of Scale Hinge Assembly are properly seated in sockets of Top Plate. 4. Display should show Ready C 0.00 lb (0.000 kg). 5. The calibration process consists of sequentially adding and then removing a series of weights. This sequence is shown in the following table.