Page 61

Verification

General Verification Output by ESP General verification output parameters allow the user choose specific output settings for Separator Character, ISO Grade Type, Symbol Type, and Symbol Dimension/Size as they appear in ISO/IEC 15415 and AS9132 verification output.

Note: The Alpha or Numeric parameter only affects ISO/IEC 15415 output.

Separator Character The separator you choose will appear between each field of data output.

Double-click on the character displayed in the control tree to bring up the separator character calculator. You can either type your choice of character in the text field, or, if you need to use a non-printable character, click the abbreviation that corresponds with that character.

ISO Grade Type This parameter only affects ISO/IEC 15415 verification output.

If you choose Alpha, ISO grades will be in alphabetical format (A, B, C, D, F). If you choose Numeric, ISO grades will be in numeric format.

Quadrus Verifier User’s Manual

3-17

Page 62

General Verification Output by ESP

Symbol Type When enabled, Symbol Type identifies the symbology of the mark being evaluated in the verification report output.

Symbol Dimension / Size When enabled, Symbol Dimensions states the dimensions (row value x column value) of the mark being evaluated in the verification report output.

3-18

Quadrus Verifier User’s Manual

Page 63

Verification

ISO/IEC 15415 Verification Setup This command allows the user to fine-tune lighting and camera settings to comply with ISO/IEC 15415’s optical requirements for 2D verification.

Aperture Definition:

The physical size of the synthetic aperture that will be applied to the captured symbol image by the Verifier’s software. This parameter is in units of 1/10000 of one inch, or 10x the mil size.

Serial Cmd:

<K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum>

Default:

50

Options:

10 to 160

Wavelength Definition:

Expresses the wavelength of LED illumination that will be directed at the candidate symbol during verification. Wavelength values are in nanometers (nm).

Serial Cmd:

<K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum>

Default:

660

Options:

400 to 700

Angle Definition:

The degree angle at which the candidate symbol will be illuminated in the lighting chamber during verification.

Serial Cmd:

<K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum>

Default:

45

Options:

30 to 90

Quadrus Verifier User’s Manual

3-19

Page 64

ISO/IEC 15415 Verification Setup

Reflectance Maximum Definition:

This setting represents the maximum reflectance value (percent) of the symbol used to calibrate the Verifier. The calibration symbol is described in Step 6 of the Quick Start procedure.

Serial Cmd:

<K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum>

Default:

85

Options:

0 to 100

Reflectance Minimum Definition:

This setting represents the minimum reflectance value (percent) of the symbol used to calibrate the Verifier. The calibration symbol is described in Step 6 of the Quick Start procedure.

Serial Cmd:

<K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum>

Default:

10

Options:

0 to 100

3-20

Quadrus Verifier User’s Manual

Page 65

Verification

ISO/IEC 15415 Verification Setup by ESP ISO/IEC 15415 Verification Setup allows the user to fine-tune lighting and camera settings to comply with ISO/IEC 15415’s optical requirements for 2D verification.

Aperture Aperture expresses the radius of the synthetic aperture that will be applied to the captured symbol image by the Verifier’s software. Radius values are in units of 1/10th of one pixel.

Wavelength Wavelength is the LED illumination wavelength that will be directed at the candidate symbol during verification. Wavelength values are in nanometers (nm).

Angle Angle is the degree angle at which the candidate symbol will be illuminated in the lighting chamber during verification.

Quadrus Verifier User’s Manual

3-21

Page 66

ISO/IEC 15415 Verification Setup by ESP

Reflectance Max Reflectance Max represents the maximum reflectance value (percent) of the symbol used to calibrate the Verifier.

Reflectance Min This setting represents the minimum reflectance value (percent) of the symbol used to calibrate the Verifier.

3-22

Quadrus Verifier User’s Manual

Page 67

Verification

ISO/IEC 15415 Serial Output This command allows the user to determine which ISO/IEC 15415 grades and/or values will be represented in the verification output string.

Grade Definition:

Note: The grade type (Alpha or Numeric) is determined by the “ISO Grade Type” field in the General Verification Serial Output command <K708>.

Serial Cmd:

<K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Default:

Disabled

Options:

0 = Disabled 1 = Grade

Output Example:

Overall Grade highlighted below (shown in numeric form).

symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

Aperture Value Definition:

Expresses the Synthetic Aperture Value in the verification output string.

Serial Cmd:

<K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Default:

Disabled

Options:

0 = Disabled 1 = Value

Output Example:

Synthetic Aperture Value highlighted below.

symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

Quadrus Verifier User’s Manual

3-23

Page 68

ISO/IEC 15415 Serial Output

Wavelength Value Definition:

When enabled, appends the LED illumination Wavelength Value to the verification output string. Wavelength Value expresses the peak wavelength of LED light output, measured in nanometers (nm).

Serial Cmd:

<K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Default:

Disabled

Options:

0 = Disabled 1 = Value

Output Example:

Wavelength Value highlighted below.

symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

Light Angle Value Definition:

Defines the angle of incidence of LED illumination. When enabled, appends the LED Light Angle Value (in degrees) to the verification output string.

Serial Cmd:

<K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Default:

Disabled

Options:

0 = Disabled 1 = Value

Output Example:

Light Angle Value highlighted below.

symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

3-24

Quadrus Verifier User’s Manual

Page 69

Verification

Decode Grade Definition:

To receive a passing Decode Grade, a symbol must be successfully decoded using the Verifier’s reference decode algorithm. When enabled, appends the symbol’s Decode Grade to the verification output string. A symbol will receive a 0 (F) if it cannot be decoded using the specified reference decode algorithm. It will receive a 4 (A) if it can be decoded using the specified reference decode algorithm.

Serial Cmd:

<K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Default:

Disabled

Options:

0 = Disabled 1 = Grade

Output Example:

Decode Grade highlighted below (shown in numeric form).

symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

Symbol Contrast Definition:

Measures the difference between light and dark symbol elements. This measurement also includes the symbol’s Quiet Zone. When enabled, appends the Symbol Contrast grade and/or value to the verification output string. Grading Scale: 4 (A) if > 70% 3 (B) if > 55% 2 (C) if > 40% 1 (D) if > 20% 0 (F) if < 20%

Serial Cmd:

<K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 2 = Value 3 = Grade and Value

Quadrus Verifier User’s Manual

3-25

Page 70

ISO/IEC 15415 Serial Output

Grade (Symbol Contrast) Output Example:

Symbol Contrast Grade highlighted below (shown in numeric form).

symbol_data,2,005,660,45,4,4,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

Value (Symbol Contrast) Output Example:

Symbol Contrast Value is a percentage, and is shown in the output string as a three-digit value from 000 to 100. For example, “075” = 75%.

symbol_data,2,005,660,45,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

Grade and Value (Symbol Contrast) Output Example:

Symbol Contrast Grade and Value highlighted below (grade shown in numeric form).

symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032

3-26