Page 51
Verification
AS9132 Evaluation Parameters Dot Center Offset A symbol’s Dot Center Offset value indicates the deviation of actual dot centers from theoretical or “ideal” dot centers.
x x x
Ideal dot center
Actual dot center
The difference between the ideal and actual dot centers is the Dot Center Offset value.
Cell Fill Cell Fill is the percentage of the ideal cell size that the module or element fills. The example at left shows dot peen elements that overfill the ideal cell size. The elements of the dots exceed the cell boundaries.
Dot Ovality Dot Ovality is the extent to which round elements deviate from a perfect circle. The example at left shows a symbol that would receive an unfavorable Dot Ovality evaluation.
D d Ideal dot shape
Module
If D - d > 20% of nominal module size, then dot ovality is out of spec with AS9132 requirements. Quadrus Verifier User’s Manual
3-7
Page 52
AS9132 Evaluation Parameters
Angle of Distortion 15°
X
The Angle of Distortion is the symbol’s deviation from a 90° relation between row and column. Y
90°
75°
90°
Ideal: 90° Row/Column
105°
15° Deviation from Ideal Row/Column
Symbol Contrast Symbol Contrast is the value difference between light and dark symbol elements, and between the quiet zone and perimeter elements. This example shows a low-contrast symbol. The dark elements (etched) and the light elements (the substrate’s surface) are too close in value, which undermines readability.
Quiet Zone The Quiet Zone is an unmarked space of at least one element in width surrounding the symbol, required for symbol readability. The red box in the example represents the outer perimeter of the minimum Quiet Zone requirement. The Quiet Zone can be any amount greater than one element in width, but any Quiet Zone width less than one element will make the symbol difficult or impossible to read.
3-8
Quadrus Verifier User’s Manual
Page 53
Verification
Pixels Per Element Pixels Per Element refers to the number of pixels in each individual symbol element.
11 pixels
This magnified symbol detail contains 4 elements, each with a width of 11 pixels.
Quadrus Verifier User’s Manual
3-9
Page 54
AIM DPM Evaluation Parameters
AIM DPM Evaluation Parameters Cell Contrast Cell Contrast is the value difference between light and dark symbol elements, and between the quiet zone and perimeter elements. This example shows a low-contrast direct part mark symbol. The light and dark elements are too close in value, which undermines readability.
Fixed Pattern Damage Fixed Pattern Damage refers to finder pattern and clock pattern damage. Notice the missing elements in the clock pattern and the damaged L-pattern in the example symbol.
Axial Non-Uniformity Axial Non-Uniformity is the amount of deviation along the symbol’s major axes. In this example, the symbol’s Y-axis dimension is clearly greater than its X-axis dimension.
X
Y
Grid Non-Uniformity Grid Non-Uniformity refers to a symbol’s cell deviation from the ideal grid of a theoretical “perfect symbol”. The Data Matrix reference decode algorithm is applied to a binarized image of the symbol, comparing its grid intersections to ideal grid intersections. The greatest distance from an actual to a theoretical grid intersection determines the Grid Non-Uniformity grade.
X
Y Actual grid intersection The reference decode algorithm plots the symbol’s grid intersections and compares them to an ideal grid.
Vector deviation Symbol Detail
3-10
Ideal grid intersection
The largest vector deviation on the grid determines the Grid Non-Uniformity grade.
Quadrus Verifier User’s Manual
Page 55
Verification
Cell Modulation Modulation refers to the reflectance uniformity of a symbol’s light and dark elements. In this example of a dot peen mark, notice that the light/dark values of some of the elements are inconsistent.
Unused Error Correction Unused Error Correction indicates the amount of available error correction in a symbol. 100% Unused Error Correction is ideal.
Print Growth Print Growth refers to the deviation (larger or smaller) of actual element size from intended element size due to printing problems. When a symbol is printed or chemical etched, the ink or etching agent may “bleed” when it comes in contact with the substrate, causing an Overprint. If there is not enough ink, or if there is some other problem with printing or etching equipment, the result may be an Underprint.
Overprint
Underprint
Pixels Per Element Pixels Per Element refers to the number of pixels in each individual symbol element.
11 pixels
This magnified symbol detail contains 4 elements, each with a width of 11 pixels.
Quadrus Verifier User’s Manual
3-11
Page 56
General Verification Serial Output
General Verification Serial Output This command allows the user to determine the specific output settings for Separator Character, ISO Grade Type, Symbol Type, and Symbol Dimension/Size as they appear in ISO/IEC 15415 and AS9132 verification output.
Separator Character Definition:
Inserts a separator between each field of verification report output.
Serial Cmd:
<K708,separator character,unused (0),ISO grade type,symbol type, symbol dimension/size>
Default:
, (comma)
Options:
Any ASCII character except NULL, < , or >.
ISO/IEC 15415 Output Example:
Default Separator Character (comma) highlighted below.
symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032 AS9132 Output Example:
Default Separator Character (comma) highlighted below.
symbol_data,P,F,045,057,P,002,001,F,-08.20,P,034,P,12.4,ECC200, 018x018
ISO Grade Type Definition:
Determines whether ISO/IEC 15415 grades are in alphabetical or numeric form. Note: This setting does not affect AS9132 output, because AS9132 symbol evaluations are on a pass/fail basis.
Serial Cmd:
<K708,separator character,unused (0),ISO grade type,symbol type,symbol dimension/size>
Default:
Alpha
Options:
0 = Alpha 1 = Numeric
3-12
Quadrus Verifier User’s Manual
Page 57
Verification
Alpha (ISO Grade Type) If alphabetical grading is chosen, ISO/IEC 15415 grades will be represented by the letters A (best), B, C, D, or F (fail). Output Example:
All Alpha grades highlighted below.
symbol_data,C,005,660,45,A,A,075,B,D,0.11,B,0.43,B,C,057,-0.82, 08.7,ECC200,032x032 Output
Command Field
2/C 005 660 45 4/A 4/A 075 3/B 1/D 0.11 3/B 0.43 3/B 2/C 057 -0.82 08.7 ECC200 032x032
Grade Aperture Value Wavelength Value Light Angle Value Decode Grade Symbol Contrast Grade Symbol Contrast Value Fixed Pattern Damage Grade Axial Non-Uniformity Grade Axial Non-Uniformity Value Grid Non-Uniformity Grade Grid Non-Uniformity Value Modulation Grade Unused Error Correction Capacity Grade Unused Error Correction Capacity Value Print Growth Value Pixels Per Element Value Symbol Type Symbol Dimensions
Quadrus Verifier User’s Manual
3-13
Page 58
General Verification Serial Output
Numeric (ISO Grade Type) If numeric grading is chosen, ISO/IEC 15415 grades will be represented by 0 (fail), 1, 2, 3, or 4 (best) for Single-Capture verification, or by the decimal values 0.0 to 4.0 for Multi-Capture verification. Output Example:
All Numeric grades highlighted below.
symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032 Output
Command Field
2/C 005 660 45 4/A 4/A 075 3/B 1/D 0.11 3/B 0.43 3/B 2/C 057 -0.82 08.7 ECC200 032x032
Grade Aperture Value Wavelength Value Light Angle Value Decode Grade Symbol Contrast Grade Symbol Contrast Value Fixed Pattern Damage Grade Axial Non-Uniformity Grade Axial Non-Uniformity Value Grid Non-Uniformity Grade Grid Non-Uniformity Value Modulation Grade Unused Error Correction Capacity Grade Unused Error Correction Capacity Value Print Growth Value Pixels Per Element Value Symbol Type Symbol Dimensions
3-14
Quadrus Verifier User’s Manual
Page 59
Verification
Symbol Type Definition:
When enabled, identifies the symbology of the mark being evaluated in the verification report output.
Serial Cmd:
<K708,separator character,unused (0),ISO grade type,symbol type,symbol dimension/size>
Default:
Disabled
Options:
0 = Disabled
ISO/IEC 15415 Output Example:
Symbol Type highlighted below.
1 = Enabled symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032 AS9132 Output Example:
Symbol Type highlighted below.
symbol_data,P,004,003,F,045,057,P,002,001,F,-08.20,P,034,P,12.4, ECC200,018x018 ISO/IEC 15415 Output
Command Field
2/C 005 660 45 4/A 4/A 075 3/B 1/D 0.11 3/B 0.43 3/B 2/C 057 -0.82 08.7 ECC200 032x032
Grade Aperture Value Wavelength Value Light Angle Value Decode Grade Symbol Contrast Grade Symbol Contrast Value Fixed Pattern Damage Grade Axial Non-Uniformity Grade Axial Non-Uniformity Value Grid Non-Uniformity Grade Grid Non-Uniformity Value Modulation Grade Unused ECC Grade Unused ECC Value Print Growth Value Pixels Per Element Value Symbol Type Symbol Dimensions
Quadrus Verifier User’s Manual
AS9132 Output
Command Field
P 004 003 F 045 057 P 002 001 F -08.20 P 034 P 12.4 ECC200 018x018
Dot Center Offset Grade Dot Center Offset Worst Case Value Dot Center Offset Average Value Cell Fill Grade Cell Fill Worst Case Value Cell Fill Average Value Dot Ovality Grade Dot Ovality Worst Case Value Dot Ovality Average Value Angle of Distortion Grade Angle of Distortion Value Symbol Contrast Grade Symbol Contrast Value Quiet Zone Grade Pixels Per Element Value Symbol Type Symbol Dimensions
3-15
Page 60
General Verification Serial Output
Symbol Dimension/Size Definition:
When enabled, states the dimensions (row value x column value) of the mark being evaluated in the verification report output.
Serial Cmd:
<K708,separator character,unused (0),ISO grade type,symbol type, symbol dimension/size>
Default:
Disabled
Options:
0 = Disabled 1 = Enabled
ISO/IEC 15415 Output Example:
Symbol Dimensions highlighted below.
symbol_data,2,005,660,45,4,4,075,3,1,0.11,3,0.43,3,2,057,-0.82, 08.7,ECC200,032x032 AS9132 Output Example:
Symbol Dimensions highlighted below.
symbol_data,P,004,003,F,045,057,P,002,001,F,-08.20,P,034,P,12.4, ECC200,018x018 ISO/IEC 15415 Output
Command Field
2/C 005 660 45 4/A 4/A 075 3/B 1/D 0.11 3/B 0.43 3/B 2/C 057 -0.82 08.7 ECC200 032x032
Grade Aperture Value Wavelength Value Light Angle Value Decode Grade Symbol Contrast Grade Symbol Contrast Value Fixed Pattern Damage Grade Axial Non-Uniformity Grade Axial Non-Uniformity Value Grid Non-Uniformity Grade Grid Non-Uniformity Value Modulation Grade Unused ECC Grade Unused ECC Value Print Growth Value Pixels Per Element Value Symbol Type Symbol Dimensions
3-16
AS9132 Output
Command Field
P 004 003 F 045 057 P 002 001 F -08.20 P 034 P 12.4 ECC200 018x018
Dot Center Offset Grade Dot Center Offset Worst Case Value Dot Center Offset Average Value Cell Fill Grade Cell Fill Worst Case Value Cell Fill Average Value Dot Ovality Grade Dot Ovality Worst Case Value Dot Ovality Average Value Angle of Distortion Grade Angle of Distortion Value Symbol Contrast Grade Symbol Contrast Value Quiet Zone Grade Pixels Per Element Value Symbol Type Symbol Dimensions
Quadrus Verifier User’s Manual