Page 41

Using ESP

View The View menu allows you to move quickly between interfaces without using the icon buttons on the App Mode toolbar. It also allows you to access the Bar Code Dialog.

Bar Code Dialog In the Bar Code Dialog you can create symbols by typing the text you wish to encode. This is a useful tool for creating configuration symbols, allowing you to configure your Verifier by reading the symbols you create.

Drag specific configuration values from the control tree directly into this field to encode new symbols. Choose a spatial orientation for the new symbol. Set a humanreadable caption for the symbol that matches the encoded data, or write your own caption.

The symbol you create will be displayed in the field at the bottom of the Bar Code Dialog.

Quadrus Verifier User’s Manual

2-11

Page 42

Navigating in ESP

Navigating in ESP To change Verifier settings, or to access the Configuration, Camera, Terminal, Utilities, or Output Format views, click the App Mode button.

To return to Verification, click the Verification button.

To make changes to configuration settings in the control trees: Open or close the full contents of a control tree by holding down the Alt key and single-clicking here.

The x denotes the default option setting. • Left click on the + to expand menu items.

• Double-click the desired parameter and singleclick in the selection box to view options. • Place your cursor in the selection box, scroll down to the setting you want to change, and single-click the setting. • Left click again on the open screen to complete the selection.

• Right click on the open screen and select Save to Reader to implement the command in the Verifier. You can send the command without saving it, or you can send and save the command simultaneously.

2-12

Quadrus Verifier User’s Manual

Page 43

Using ESP

Send/Receive Options To access Receive, Save and Default options, click the Send/Recv button. You can also access these options by right-clicking in any of the configuration views.

Receiving From the Send/Recv menu, select Receive Reader Settings. Caution: Do not select this option if you do not want to upload the Verifier’s settings. For example, if your ESP file has a number of custom settings that you want to maintain and download into the Verifier, these settings would be lost by choosing Yes. This is useful if you want to receive (upload) the Verifier’s settings and save them as a file for future use. For example, if your Verifier has settings that you do not want to change, choosing Yes would allow you to load those settings to ESP and save them in a ESP file for later retrieval. Receiving the Verifier’s settings will also assure that you will not be subsequently saving any unwanted changes that you or someone else has made previously in ESP.

Saving • Send, No Save (<A>) Saves ESP settings to current memory. • Send and Save (<Z>) Activates all changes in current memory and saves settings for power-on. • Send and Save as Customer Defaults (<Zc>) Saves your own default settings for quick retrieval with a <Zrc> command. This option will be visible only if you have checked Enable ‘Send and Save as Customer Defaults’ in the ESP Preferences dialog.

Quadrus Verifier User’s Manual

2-13

Page 44

Send/Receive Options

Defaulting When you select Default Current Menu Settings or Default all Settings, you are only defaulting the ESP settings.

Advanced Options Send Current View This is the same as Save to Reader > Send No Save except that only the commands in the current configuration tree are sent.

Send Current Command This is the same as Send Current View , except that it saves only the command that is currently selected.

Add / Remove Exception After you perform a Receive Reader Settings command1 and you click on the Add Exception option, you may see a list of serial commands. These are commands that may be in your Verifier’s firmware, but not included in, or different from, your current version of ESP. You can edit these commands by double-clicking on them and changing them as needed. It is important to note that these commands will be saved to your Verifier whenever you send a Save to Reader command, or an <A> or a <Z> command. Also, if there is a corresponding ESP menu item, the ESP Value column for that item will be blank following a Receive Reader Settings command.

1. From the Send/Recv button or by right-clicking from within the configuration trees.

2-14

Quadrus Verifier User’s Manual

Page 45

3 Verification Contents Verification Serial Commands ...................................................................................................... 3-2 Verification Operational Commands............................................................................................. 3-2 Overview of Verification................................................................................................................ 3-3 ISO/IEC 15415 Evaluation Parameters ........................................................................................ 3-5 AS9132 Evaluation Parameters ................................................................................................... 3-7 AIM DPM Evaluation Parameters............................................................................................... 3-10 General Verification Serial Output .............................................................................................. 3-12 General Verification Output by ESP ........................................................................................... 3-17 ISO/IEC 15415 Verification Setup .............................................................................................. 3-19 ISO/IEC 15415 Verification Setup by ESP ................................................................................. 3-21 ISO/IEC 15415 Serial Output ..................................................................................................... 3-23 ISO/IEC 15415 Output by ESP................................................................................................... 3-33 AIM DPM Verification Setup ....................................................................................................... 3-40 AIM DPM Verification Setup by ESP .......................................................................................... 3-42 ISO/IEC 15415 Verification by Serial Command ........................................................................ 3-44 AS9132/JES 131 Marking Method ............................................................................................. 3-49 AS9132/JES 131 Marking Method by ESP ................................................................................ 3-51 AS9132 Serial Output................................................................................................................. 3-53 AS9132 Output by ESP .............................................................................................................. 3-62 AS9132 Verification by Serial Command ................................................................................... 3-67 AIM DPM Verification by Serial Command................................................................................. 3-68 Verification by ESP..................................................................................................................... 3-70

This section describes the verification process, including specification requirements, software configuration, step-by-step procedures, and report output.

Quadrus Verifier User’s Manual

3-1

Page 46

Verification Serial Commands

Verification Serial Commands ISO/IEC 15415 Verification Setup AIM DPM Verification Setup General Verification Serial Output AS9132/JES 131 Marking Method AS9132 Serial Output

ISO/IEC 15415 Serial Output

<K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum> <K532,minimum element size,maximum element size,wavelength,lighting> <K708,separator character,unused (0),ISO grade type,symbol type,symbol dimension/size> <K711,element shape,marking method,JES 131> <K712,dot center offset,percentage failed offset, cell fill,percentage failed cell fill,dot ovality,percentage failed ovality,angle of distortion,symbol contrast,quiet zone grade,pixels per element value> <K756,grade,aperture value,wavelength value,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity,modulation grade,unused error correction,print growth value,pixels per element value>

Verification Operational Commands ISO/IEC 15415 Reflectance Calibration <@VER> ISO/IEC 15415 Single Capture Verification <V1> ISO/IEC 15415 Multi-Capture Verification <V2> AS9132 Verification <V3> AIM DPM Reflectance Calibration <@AIMDPM,R-max,R-min> AIM DPM Verification <V4>

3-2

Quadrus Verifier User’s Manual

Page 47

Verification

Overview of Verification The use of Data Matrix symbols in ID automation applications requires high-quality marks. The purpose of verification is to ensure reliability and consistency of symbols, based on the strict criteria outlined in the AS9132 and ISO/IEC 15415 standards and the AIM DPM quality guideline. The Quadrus Verifier is designed to evaluate marks based on the specific parameters in AS9132, ISO/IEC 15415, and AIM DPM.

AS9132 The AS9132 standard specifies uniform quality and technical requirements for direct part marking with Data Matrix symbols. Direct part marking can be achieved by a variety of means, including ink jet, dot peen, laser etch, and chemical etch. Note: AS9132 and AS9132A are used interchangeably throughout this documentation. “AS9132” is the name of the specification, and the suffix “A” denotes the current published version of the specification.

ISO/IEC 15415 The ISO/IEC 15415 standard specifies the methodologies for measuring, evaluating, and grading 2D symbol characteristics in order to provide an overall symbol grade.

AIM DPM The AIM DPM quality guideline assesses direct part mark quality for a number of parameters, including cell contrast, fixed pattern damage, axial and grid non-uniformity, cell modulation, unused error correction capacity, and minimum reflectance. Direct part marking can be achieved by a variety of means, including ink jet, dot peen, laser etch, and chemical etch. AIM DPM is called out in MIL-STD-130N as the preferred guide for ensuring symbol quality and reliability.

MIL-STD-130N MIL-STD-130N is a standard for implementing ID automation processes to track United States Department of Defense property. The DoD’s primary means of parts traceability is the IUID initiative. IUID, which stands for “Item Unique Identification”, is a system of establishing unique item identifiers (UIIs) by assigning a machine-readable character string or number to an item (a single hardware component or grouping of subassemblies), thereby distinguishing it from other items. Tracing items in this way requires the use of reliable symbols, whether in the form of printed labels or marks applied directly to parts. MIL-STD-130N calls out AIM DPM as the preferred standard for ensuring symbol quality and reliability.

Quadrus Verifier User’s Manual

3-3

Page 48

Overview of Verification

ISO/IEC 15426-2 Verifier Certification ISO/IEC 15426-2 is a verifier conformance standard that is referenced in the introduction to the ISO/IEC 15415 specification. Conformance to ISO/IEC 15426-2 is required for certification as a true Data Matrix verifier.

Why is Verifier re-calibration important? Like any measurement device, the Quadrus Verifier requires regular maintenance and calibration to ensure reliable and accurate operation over time.

How often does the Verifier need to be re-calibrated / re-certified? Microscan recommends that the Quadrus Verifier be re-calibrated and re-certified once every year. Important: Even if the Verifier has not been used during the first year of ownership, it should still be sent to Microscan for re-calibration one year from the date of original factory calibration.

What is the process for Verifier re-calibration / re-certification? Schedule a return under the SRO (Service Return Order) process. Contact the Microscan Service Department at 425.226.5700 to initiate the SRO process.

3-4

Quadrus Verifier User’s Manual

Page 49

Verification

ISO/IEC 15415 Evaluation Parameters Symbol Contrast Symbol Contrast is the value difference between light and dark symbol elements, and between the quiet zone and perimeter elements. This example shows a low-contrast symbol. The light and dark elements are too close in value, which undermines readability.

Fixed Pattern Damage Fixed Pattern Damage refers to finder pattern and clock pattern damage. Notice the missing elements in the clock pattern and the damaged L-pattern in the example symbol.

Axial Non-Uniformity Axial Non-Uniformity is the amount of deviation along the symbol’s major axes. In this example, the symbol’s Y-axis dimension is clearly greater than its X-axis dimension.

X

Y

Grid Non-Uniformity Grid Non-Uniformity refers to a symbol’s cell deviation from the ideal grid of a theoretical “perfect symbol”. The Data Matrix reference decode algorithm is applied to a binarized image of the symbol, comparing its grid intersections to ideal grid intersections. The greatest distance from an actual to a theoretical grid intersection determines the Grid Non-Uniformity grade.

X

Y Actual grid intersection The reference decode algorithm plots the symbol’s grid intersections and compares them to an ideal grid.

Vector deviation Symbol Detail

Quadrus Verifier User’s Manual

Ideal grid intersection

The largest vector deviation on the grid determines the Grid Non-Uniformity grade.

3-5

Page 50

ISO/IEC 15415 Evaluation Parameters

Modulation Modulation refers to the reflectance uniformity of a symbol’s light and dark elements. In this example, notice that the light/dark values of some elements are inconsistent.

Unused Error Correction Unused Error Correction indicates the amount of available error correction in a symbol. Error Correction is a method of reconstructing or replacing data that is lost through symbol damage. 100% Unused Error Correction is ideal. The example at left is an ECC 200 Data Matrix symbol in good condition. “ECC 200” indicates the error correction level of the symbology. A higher number indicates more robust error correction capacity.

Print Growth Print Growth refers to the deviation (larger or smaller) of actual element size from intended element size due to printing problems. When a symbol is printed, the ink may “bleed” when it comes in contact with the substrate, causing an Overprint. If there is not enough ink, or if there is some other problem with printing equipment, the result may be an Underprint.

Overprint

Underprint

Pixels Per Element Pixels Per Element refers to the number of pixels in each individual symbol element.

10 pixels

This magnified symbol detail contains 4 elements, each with a width of 10 pixels.

3-6

Quadrus Verifier User’s Manual