Ethernet
Step 3, Communicating in Ethernet By now either the Verifier should be configured with an IP Address, or the host computer should be configured to use the Verifier's default IP Address. Now ESP needs to be configured. 1. Select Settings… from ESP’s Connect dropdown menu. 2. Click the TCP/IP tab.
3. Type the Verifier's IP Address as it is shown on the Communications menu. 4. Click Connect. 5. After a few seconds, ESP should display CONNECTED at the bottom of the window.
Quadrus Verifier User’s Manual
16-7
Page 412
Ethernet Application
Step 4, Ethernet Application The Quadrus Verifier supports the Microscan communication protocol for Verifiers over two TCP/IP ports. This is the same protocol that is supported through the Verifier's RS-232 Host serial port. Note that on Ethernet, the Quadrus Verifier is a server device, which requires the host to establish the connection. Once the connection is established, however, the Verifier will send symbol data and diagnostic message data whenever it is generated.
Network Protocols Supported The Quadrus Verifier supports the following RFC-compliant protocols: IP ICMP, PING TCP Sockets ARP DHCP client TFTP server
RFC0791, RFC950 RFC0792 RFC0793 BSD v4.3 RFC0826 RFC0951, RFC1541, RFC2131, RFC2563 partial support Revision 2, RFC1350
Communication with the Quadrus Verifier can be established via a Telnet client (raw data only, no IAC command processing), or through a custom “sockets” application. The Quadrus Verifier becomes a “server” device. It listens for connection requests from the Host before communication can begin. All Quadrus Verifiers have a unique 48-bit hardware (MACID) address. This address is printed on the Verifier's product label.
16-8
Quadrus Verifier User’s Manual
Page 413
Ethernet
Microscan Protocol/Host RS-232 Supported • Configuration Commands (K commands) • ESP interface commands (<op> commands) • Utility commands <A> - <Z> • Host preamble <K141> and postamble <K142> strings are added to command responses and symbol data. • All data formatting is supported (<K740>, <K741>).
Differences from Other Protocols • Y-Modem is not supported. This includes the Firmware Download command <dy> and the Image Send command <uy>. • An immediate response to the Verifier’s Status command <?> is not sent. The response is queued for output between read cycles. Also, the response is formatted with host port preamble and postamble characters. • Aux port (RS-232) interaction is not supported for transferring data between Ethernet and the aux port (Transparent, Half Duplex, and Full Duplex). These modes only function with the Host RS-232 port, and are not affected by Ethernet. • Responses to “binary frame” commands from the Video TCP Port are sent, when connected. If not connected, the response will be sent to the Primary TCP Port.
Primary (Command) TCP Port This port is used for all command processing and data outputs (except “binary frame” command responses). The application protocol used is the same as the Host RS-232 port.
Quadrus Verifier User’s Manual
16-9
Page 414
Ethernet Application
Video TCP Port The Video TCP Port is transmit-only (output from the Verifier, read-only by the host), and is used for “binary frame” data outputs as follows: 1. Response to Image Send command <op,4>. 2. Response to Capture and Decode and Save command <op,5>. 3. Response to Symbol Information command <op,8>. 4. Response to Histogram command <op,14>. The primary purpose of this port is for access to a dedicated video stream, for ESP functions in particular. Because the format of the “binary frame” data is quite different from the Verifier's other command responses and outputs, sending this data on a dedicated, independent TCP port facilitates easier ESP implementation. Also, end users who wish to implement their own applications may find this useful since it has a consistent data type and provides access to a port whose only traffic is that which the host application has initiated. The Video TCP Port is read-only from the host's perspective. The Verifier does not receive any data or commands on this port. The Verifier only processes commands from the primary (Command) TCP port. Also, the Verifier only responds to the video port if the video port has a connection to the host and a command is received that requires a “binary frame” response. If the video port is unconnected, the Verifier will respond to the primary port. This allows hosts the option to operate via a single port.
Image File Transfers The Verifier's image files can be accessed with either “binary frame” commands <op,4> and <op,5>, or TFTP (Trivial File Transfer Protocol).
Binary Frames These are primarily intended for ESP's “near-real-time video” purposes. Although it is possible to get a full-size image from the Verifier using an <op,4> or <op,5> command, for quicker response, it is recommended that TFTP be used.
TFTP Server A TFTP client can access the Verifier's image files (the Verifier is a TFTP server). Since TFTP does not support a directory structure, the host application must know the Verifier's file naming convention, or request an image list from the Verifier through the primary TCP/IP port (or RS-232 port) with the <op,9> command.
Limitations The largest data size per packet (TCP MSS/MTU) the Verifier can receive and transmit is approximately 550 bytes. Since TCP/IP is used, this will not prevent larger data transmissions, but it will limit throughput. An MTU of 550 may be somewhat limiting for some networks, but this should not prevent successful operation.
16-10
Quadrus Verifier User’s Manual
Page 415
Appendices Appendix A General Specifications...............................................................................................A-2 Appendix B Electrical Specifications.............................................................................................A-4 Appendix C Connectivity Accessories ..........................................................................................A-8 Appendix D Serial Configuration Commands .............................................................................A-11 Appendix E ASCII Table .............................................................................................................A-17 Appendix F Data Matrix Symbology............................................................................................A-19 Appendix G Object Detector .......................................................................................................A-20 Appendix H Operational Tips ......................................................................................................A-21 Appendix I Embedded Menus.....................................................................................................A-22 Appendix J Interface Standards..................................................................................................A-23 Appendix K Glossary of Terms ...................................................................................................A-24
Quadrus Verifier User’s Manual
A-1
Page 416
General Specifications
Appendix A, General Specifications Mechanical Height: 9.88” (251mm) Width: 5.05” (128 mm) Depth: 4.75” (121 mm) Weight: ~60 oz. (~1,700 g)
Environmental Operating Temperature: 0° to 43°C (32° to 109°F) If mounted on non-metal surface, maximum operating temperature is 40°C (104°F) Storage Temperature: -50° to 75°C (-58° to 167°F) Humidity: Up to 90% (non-condensing)
Emissions/Immunity ITE Disturbances: IEC 55022:1998 (radiated and conducted) Class A General Immunity: IEC 55024:1998 (residential) Heavy Industrial Immunity: IEC 61000-6-2:1999 LED Radiation: IEC 60825-1
Light Source Type: High output LEDs External (45°, 30°): 660 nM
Light Collection CCD array: 656 x 496 pixels
Video Output Signal System: EIA (RS-170) Number of Scanning Lines: 525 lines/2:1 interlaced Output: Analog 1Vp-p/75 ohm
Communications Protocols Interface: RS-232, Ethernet
Symbology Data Matrix (ECC 0-200)
Symbol Verification Parameters Maximum Characters: 78 characters plus overhead characters at a 45° angle.
Verification Standards For Data Matrix: AIM DPM, ISO/IEC 15415, AS9132 Verifier Certification: ISO/IEC 15426-2
A-2
Verifier Dimensions
Quadrus Verifier User’s Manual
Page 417
Appendices
Indicators LEDs: Read Performance, Power, Read Status, Network Status Beeper: Read Performance, Power, Read Status, Network Status
Safety Certifications Designed for: FCC, CE
ISO Certification ISO 9001:2000 Issued by TUV USA Inc., Member of TUV NORD Group, Cert. No. 06-1080 Product specifications are given for typical performance at 25°C (77°F) using grade A symbols. Some performance characteristics may vary at high temperatures or other environmental extremes. ©2007 Microscan Systems, Inc. Specifications subject to change.
Quadrus Verifier User’s Manual
A-3
Page 418
Electrical Specifications
Appendix B, Electrical Specifications Power Requirements: Input, 10 to 28 VDC, 200 mV p-p max ripple, 333 mA at 24 VDC Trigger, New Master, Input 1: (Optoisolated) 5 to 28 VDC rated (12 mA at 24 VDC) Outputs 1/2: (Optoisolated) 1 to 28 VDC rated (ICE < 100 mA at 24 VDC, current limited by user) Output 3: Light control (Optoisolated) 1 to 28 VDC rated (ICE < 100 mA at 24 VDC, current limited by user)
25-Pin Connector
Quadrus Verifier Host 25-Pin Connector Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25
Host RS-232
Host and Aux RS-232 Chassis Ground
Ethernet
TxD RxD RTS CTS
Out In Out In Out
TxD RxD Output 1 (+) Signal Groundb Output 2 (+) Trigger (-) Trigger (+) Default Configurationc Input 1 (+) RxD (+) RxD (-) Light Control (+) TxD (-) Power Groundd Power + 10 to 28 VDC TxD (+) Output 1 (-) Output 2 (-) Light Control (-) Input 1 (-) New Master (-) New Master (+)
I/O a
Out In In In In In In Out Out In Out Out Out Out In In In a. Chassis Ground: Used to connect chassis body to earth ground only. Not to be used as power or signal return. b. Signal Ground: Used for communication and signal line grounds only. Not to be used as power or chassis return. c. The default is activated by connecting pin 11 to ground pin 7. d. Power Ground: Used for power return only. CAUTION: If using your own power supply, verify correct connection of power and ground lines. Incorrect connections or use of Chassis Ground, Power Ground, and Signal Ground lines could equipment failure or software failure.
A-4
Quadrus Verifier User’s Manual
Page 419
Appendices
Optoisolator Outputs Optoisolator circuits can transfer pulses between the Verifier and peripherals with no direct connection with the Verifier’s internal circuitry. However, not every optoisolator configuration provides complete isolation. The following diagrams show both fully optoisolated and non-optoisolated circuits. They are only examples and do not represent all the possible wiring configurations. Iload=5mA
Iload=50mA
VOut-On
0.5V
0.5V
1.0V tOn-Typ
5mS
0.7mS
0.8mS tOff-Typ
5µS
5µS
5µS
Scanner Output
Iload=100mA
Outputs (+) ILoad Outputs (–)
Output Circuit Examples Fully Optoisolated This circuit is fully optoisolated and the recommended configuration. It allows the user to apply 1 to 28 VDC to the circuit. Caution: The maximum current that can pass through the optoisolator is 100mA.
Host
Isolated Verifier Power (+)
1 to 28 V Scanner
Outputs (+)
Output
Input Outputs (–) Power ground
Not Optoisolated, Verifier Grounded In this diagram, power is applied externally, but the Verifier’s power ground is used to complete the circuit. This setup involves some risk to the optoisolator if excessive voltages are applied. Caution: The maximum current that can pass through the optoisolator is 100mA.
Non-Isolated Verifier
Host Power (+) 1 to 28 V
Scanner
Outputs (+)
Output
Input Outputs (–)
Power ground
Quadrus Verifier User’s Manual
A-5
Page 420
Electrical Specifications
Additional Isolated Output Circuit Examples Secondary Relay 1 to 28 V
Isolated Verifier Power +
RELAY
Isolated Verifier Power + Scanner
Scanner Output
Outputs (+)
V
PLC
Input
Outputs (+)
Outputs (–)
Output
PLC
PLC GND
Power ground Outputs (–) Power ground
A-6
Quadrus Verifier User’s Manual