Page 421

Appendices

Optoisolator Inputs All discrete inputs can be fully isolated pulses as PNP or NPN circuits. Inputs include Trigger, New Master, and Input 1. Generic Waveform Characteristics Minimum

Maximum

VIN-HIGH/IINHIGH

4.5V/4mA

28V/12mA

VIN-LOW/IINLOW

0 V/0mA

Pulse Width min.

48 µS

Input (+) I load Input (–)

2 V/2mA

Input Examples Fully Optoisolated PNP Source +V

Isolated Verifier V CC

NPN Source

Isolated Verifier V CC

+V Input (+)

Signal Input (+)

Input (–)

Input (–)

Signal

Not Optoisolated Non-Isolated Verifier V CC

PNP Source Power (+) Input (+)

+V

Non-Isolated Verifier V CC

Signal

NPN Source Power (+)

+V

Input (+) Input (–)

Input (–)

Power ground

Quadrus Verifier User’s Manual

Power ground

Signal

A-7

Page 422

Connectivity Accessories

Appendix C, Connectivity Accessories Three interface options are available for Verifier connectivity.

IB-150 Kit The IB-150 Interface Kit links the Verifier to the host, power supply, trigger, and aux port connections. In addition to multidrop and daisy chain hookups, the aux port can plug into the terminal strip interface. IB-150 Interface Kit T RIGGER POWER

HOST

IB-150 Interface

READER

AUX PORT

Port Connectors

Trigger 4-pin Connector

Pin 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25

Pin Function 1 Power + 10 to 28VDC (out)a 2 Trigger (–) (in)b 3 Power Ground 4 Trigger (+) (in)a

A-8

Verifier Host Aux Port Chassis ground Chassis ground Chassis ground TxD TxD RxD RxD RxD TxD RTS/Aux TxD RTS/Aux TxD NC CTS/Aux RxD CTS/Aux RxD NC Output 1 (+) Output 1 (+) Output 1 (+) Signal GND Signal GND Signal GND Output 2 (+) Output 2 (+) Output 2 (+) Trigger (–) Trigger (–) Trigger (–) Trigger (+) Trigger (+) Trigger (+) Default Default Default Input 1 (+) Input 1 (+) Input 1 (+) RxD (+) RxD (+) RxD (+) TxD (–) TxD (–) TxD (–) Light Control Light Control Light Control RxD (–) RxD (–) RxD (–) Power ground NC Power ground +10 to 28 VDC NC +10 to 28VDC TxD (+) TxD (+) TxD (+) Output 1 (–) Output 1 (–) Output 1 (–) Output 2 (–) Output 2 (–) Output 2 (–) Light Control Light Control Light Control Input 1 (–) Input 1 (–) Input 1 (–) New master (–) New master (–) New master (–) New master (+) New master (+) New master (+)

a. For NPN type, connect pins 1 and 4. b. For PNP type, connect pins 2 and 3.

Power 3-pin Connector Pin 1 2 3

Function Power Ground Chassis Ground Power + 10 to 28VDC (in)

Quadrus Verifier User’s Manual

Page 423

Appendices

Ethernet/USB Interface USB and standard Ethernet (RJ45) connections can be made through the Ethernet/USB Interface, which plugs directly into the host port of the Quadrus Verifier. The Ethernet/USB Interface has a second 25-pin port to pass through all but the RS-422/485 connections. QU ADRU S E Z ™

USB Connector (socket) To Quadrus Verifier

3 2

1

Ethernet (socket)

4

8

1

Category 5 cabling less than 50 m in length is recommended

1

25

Host Connector (plug)

USB

Ethernet Pin 1 2 3 4 5 6 7 8

Ethernet Port Ethernet TX (+) Ethernet TX (–) Ethernet RX (+) NC NC Ethernet RX (–) NC NC

Pin 1 2 3 4

USB Port +5VDC Data (–) Data (+) GND

Host Pin 13 14 16 19

Host Porta NC NC NC NC a. All other pins are as shown on the Verifier port.

Ethernet/USB Notes • Protection components are installed within the interface to address Ethernet susceptibility to line transients and electrostatic discharge (ESD) that can cause communication dropouts and connection termination. • The matching cable does not provide the proper matched impedance that the standard CAT 5 cable provides, thus allowing higher emission levels which exceed Class A CE limits and possible further signal degradation.

Quadrus Verifier User’s Manual

A-9

Page 424

Connectivity Accessories

Terminal Strip Interface Kit Custom terminal strip wiring can be done through this interface, which can either connect directly to the Verifier or indirectly by way of the aux port. Note: This kit includes a gender changer. The gender changer is only used for direct connection between the interface and the Verifier.

QUAD RUS E Z ™

TRIGGER POWER

HOST

IB-150 Interface

READER

AUX PORT

To Quadrus Verifier

Gender Changer

Note: Inside terminal assembly can be flipped over as required by application. 25

Steps for wiring:

(plug)

1

1. Remove two screws from top case at these locations. 2. Pry at this point to separate cases. 3. Connect external wires to terminal connections as marked.

12 11 10 9

8

7

6

5

4

25

22

A-10

2

1

13

24 23

3

14 21 20 19 18 17

15 16

Quadrus Verifier User’s Manual

Page 425

Appendices

Appendix D, Serial Configuration Commands Communications Host Port Connections Host Port Protocol Host 232/422 Status Auxiliary Port Daisy Chain Autoconfigure Daisy Chain ID Ethernet Configuration Preamble Postamble Response Timeout LRC Aux Port System Data Status Read Cycle Multisymbol Trigger Mode/Duration External Trigger State Serial Trigger Character Start Trigger Character Stop Trigger Character End of Read Cycle Active Camera Captures Capture Timing Dual Camera Operations Store No Read Image Symbologies Data Matrix QR Code Code 39 Code 128 BC412 Interleaved 2 of 5 UPC/EAN Quadrus Verifier User’s Manual

<K100,baud rate,parity,stop bits,data bits> <K140,protocol> <K102,host 422> <K101,aux port mode,baud rate,parity,stop bits,data bits,daisy chain ID status,daisy chain ID> <K150DAISY> <K151,daisy chain reader #,daisy chain reader ID> <K125,IP address,subnet address,gatewayaddress,IP address mode, primary TCP port,video TCP port> <K141,status,preamble characters> <K142,status,postamble characters> <K143,response timeout> <K145,status> <K146,aux port system data> <K222,number of symbols,multisymbol separator> <K200,trigger mode,trigger filter duration> <K202,extermal trigger state> <K201,serial trigger character> <K229,start character> <K230,stop character> <K220,end of read cycle,read cycle timeout> <K240,active camera> <K241,capture mode,number of captures,rapid capture mode> <K242,time before 1st capture,time between capture 1 and 2 ,,,,,,,,time between capture 7 and 8> <K243,switching mode,number of internal camera captures,number of external camera captures,internal camera timeout,external camera timeout> <K244,image storage type,image storage mode> <K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC100 status,ECC140 status,ECC 120 status,ECC 130 status> <K480,status> <K470, status,check digit status,check digit output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> <K474,status,fixed symbol length status,fixed symbol length> <K481, status,check digit output,fixed symbol length status,fixed symbol length> <K472,status,check digit status,check digit output,symbol length #1,symbol length #2,guard bar> <K473,UPC status,EAN status,supplementals status,separator status, separator character,supplemental type>

A-11

Page 426

Serial Configuration Commands Pharmacode RSS Expanded RSS Limited RSS-14 PDF417 Micro PDF417 Composite Narrow Margins/Symbology ID Background Color Unique Item Identifiers (UII) I/O Parameters Symbol Data Output No Read Message Bad Symbol Message No Symbol Message 1D/Stacked Symbology Qualification 2D Symbology Qualification

Read Duration Output LED Indicators Beeper LED Configuration Serial Verification Video Output Image Output Image Captioning Synchronous Trigger EZ Button EZ Button Modes Input 1 Output 1 Parameters Trend Analysis Output 1 Diagnostic Warnings to Output 1 ISO/IEC 15415 Verification Output 1 AS9132 Verification Output 1 Output 2 Parameters Trend Analysis Output 2

A-12

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar width status,direction,fixed threshold value> <K484,status,fixed symbol length status,fixed symbol length> <K483,status> <K482,status> <K476,status,[unused],fixed symbol length status,fixed symbol length,[unused],codeword collection> <K485,status,[unused],fixed symbol length status,fixed symbol length> <K453,mode,separator status,separator> <K450,narrow margins,symbology identifier status> <K451, background color> <K455,status,errstatus> <K705,symbol data output status,when to output> <K714,no read message status,no read message> <K715,[unused],message> <K716,[unused],message> <K717,minimum number of bars,minimum number of qualified scans,start/stop status> <K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value> <K706,status,separator> <K750,green flash mode,target pattern status,green flash duration> <K702,beeper status> <K737,LED mode,ISO/IEC 15415 grade,DPM grade> <K701,serial command echo status,serial command beep status,control/hex output> <K760,video output status,triggered mode,capture number> <K739,image output mode,comm port,file format,JPEG quality> <K762,mode> <K761,synchronous trigger mode> <K770,global status,default on power-on,load IP database,save for power-on> <K771,position 1 mode,position 2 mode,position 3 mode,position 4 mode> <K730,input mode,active state> <K810,output on,active state,pulse width,output mode> <K780,trend analysis mode,number of triggers,number to output on> <K790,over temp,service unit,external camera disconnect> <K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction,print growth value> <K840,dot center offset,cell fill, dot ovality,angle of distortion,symbol contrast,quiet zone> <K811,output on,active state,pulse width,output mode> <K781,trend analysis mode,trigger evaluation period,number to output on> Quadrus Verifier User’s Manual

Page 427

Appendices Diagnostic Warnings to Output 2 ISO/IEC 15415 Verification Output 2

AS9132 Verification Output 2 Output 3 Parameters Trend Analysis Output 3 Diagnostic Warnings to Output 3 ISO/IEC 15415 Verification Output 3 AS9132 Verification Output 3 Verification ISO/IEC 15415 Verification Setup AIM DPM Verification Setup General Verification Serial Output AS9132 Element Shape and Marking Method AS9132 Serial Output

ISO/IEC 15415 Serial Output

Matchcode Matchcode Type Sequence Step New Master Pin Number of Master Symbols Enter Master Symbol Data Read Next Symbol as Master Symbol Request Master Symbol Data Delete Master Symbol Data Diagnostics Power-on / Reset Counts External Camera Message Over Temperature Message Service Message

Quadrus Verifier User’s Manual

<K791,over temp,service unit,external camera disconnect> <K831,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction,print growth value> <K841,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast,quiet zone> <K812,output on,active state,pulse width,output mode> <K782,trend analysis mode,trigger evaluation period,number to output on> <K792,over temp,service unit,external camera disconnect> <K832,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction,print growth value> <K842,dot center offset,cell fill, dot ovality,angle of distortion,symbol contrast,quiet zone> <K531,aperture,wavelength,angle,reflectance maximum,reflectance minimum> <K532,minimum element size,maximum element size,wavelength,lighting> <K708,separator character,unused (0),ISO grade type,symbol type, symbol dimensions> <K711,element shape,marking method,JES 131> <K712,dot center offset,percentage failed offset, cell fill,percentage failed cell fill,dot ovality,percentage failed ovality,angle of distortion, symbol contrast,quiet zone grade,pixels per element value> <K756,grade,aperture,wavelength,light angle value,decode grade,symbol contrast,fixed pattern damage grade,axial non-uniformity,grid non-uniformity, modulation grade,unused error correction,print growth value,pixels per element value> <K223,type,sequential matching,match start position,match length, wildcard character,sequence on no read,sequence on mismatch> <K228,sequence step> <K225,status> <K224,number of master symbols> <K231,master symbol number,data> <Gmaster symbol number> <K231?,>[for all] or <K231?,master symbol number> <K231,master symbol number,> <K406,power-on,resets,power-on saves,customer default saves> <K410,disconnect msg status,disconnect message,connect message status, control message> <K402,over temperature status,warning message> <K409,status,service message,threshold,resolution>

A-13

Page 428

Serial Configuration Commands Camera Region of Interest Camera Illumination Source Thresholding Image Processing Mode Multiple Symbols in Fast Linear Mode Image Processing Timeout Hollow Mode Output Format Symbol Output Format Status Multisymbol Format Assignment Extraction Mode Insertion Mode Ordered Output Number of Filters Output Object Information Operational Commands ISO/IEC 15415 Reflectance Calibration ISO/IEC 15415 Single Capture Verification ISO/IEC 15415 Multi-Capture Verification AS9132 Verification AIM DPM Reflectance Calibration AIM DPM Verification

A-14

<K516,top,left,height,width> <K540,shutter speed,gain> <K535,illumination source> <K512,threshold mode,threshold value> <K513,processing mode> <K518,number of symbols> <K245,image processing timeout> <K517,hollow status> <K743,symbol output format status> <K742, symbol number,multisymbol status> <K740, output #,start position,length> <K741,output #,length,hex string> <K744,filter number,symbology,length,wildcard character,placeholder character,data,> <K745, number of filters> <K734,output frame #,output coordinates> <@VER> <V1> <V2> <V3> <@AIMDPM,Rmax,Rmin> <V4>

Quadrus Verifier User’s Manual

Page 429

Appendices

Serial Command Format Serial commands are of two types: utility and configuration.

Rules that apply to both utility and configuration commands • A less than < and greater than > characters enclose the commands. • Commands and data are “case sensitive.” That is, characters must be entered as upper or lower case, as specified.

Serial Utility Commands These are sent during operations and are not followed by a <A> or <Z>.

Serial Configuration K Commands These begin with a single ‘K’ character followed by a 3-digit numeric character, data fields, and an initializing command, as follows: <Knumeric parameter,data,data,...etc.><initializing command> An initializing command <A> or <Z> may follow the command. A <Z> initializes the Verifier’s memory and saves for power-on; an <A> initializes the Verifier’s memory but does not save for power-on. For example, to enable UPC and save the change for power-on, send <K473,1><Z>. To change Baud Rate and reset without saving changes for power-on, send <K100,3><A>.

Serial Configuration Command Conventions • All data fields (except the last) must be followed by a comma (without a space). • The following characters cannot be used: NULL, < , or >. • All fields preceding a modified field must be included. • If there is no change in preceding fields, then commas alone can be entered in these fields. For example, if only the last field in the following command is changing, <K100,4,1,0,0> can be entered as <K100,,,,0>. • All fields following a modified field can be omitted. For example, to change Baud Rate only, send <K100,3>.

Concatenating Configuration Commands Commands can be concatenated (added together) in a single string or data block. For example, <K145,1><K220,1><K450,1><A> enables LRC, sets End of Read Cycle mode to New Trigger, enables Narrow Margins, and resets the data buffers (without saving the changes for power-on).

Quadrus Verifier User’s Manual

A-15

Page 430

Serial Configuration Commands

Serial Command Status Request To ensure that any command was received and accepted, you can send the Show Verifier Status command: <?>. The status of a specific serial command can be requested by entering the command followed by a question mark. For example, send <K142?> to request the status of Postamble.

Entering Special Characters in Serial Commands To enter control characters within a serial command, hold down the control key while typing the desired character. Example: To enter a carriage return and line feed (^M^J), enter <K141,1,CNTL-m CNTL-j>.

Entering Special Characters in Embedded Menus Control Characters Control characters entered on the command line are displayed in the menu as mnemonic characters, such as: <CR><LF><NULL><NULL>. Press SP (the space bar) once, then enter the control character by holding down the control key and simultaneously pressing the desired character. For example to define a line feed, press SP, then Control and J simultaneously. It is displayed as ^J on the command line and as <LF> in the menu when the screen is refreshed.

To Define a Carriage Return as a Character Press SP, then CR. It is displayed as ^M on the command line and as <CR> in the menu when the screen is refreshed.

To Define a Space as a Character Press SP twice. It is displayed as a blank space in the menu when the screen is refreshed. While it appears that nothing has been assigned, the hex value 20 will be sent during data transmission.

To Select NUL as the Character Press SP, then a 0 (zero). It is displayed as NUL in the menu when the screen is refreshed.

A-16

Quadrus Verifier User’s Manual