The text of a manual · part 41 of 47

Quadrus Verifier User's Manual, pages 401 to 410

Pages 401 to 410 of 470

The original document is the PDF (4.1 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan Verifier

Page 401

Output Format

Length Definition:

Specifies the length of the decoded symbol allowed to occupy this location in multi-symbol output. Note: To filter or order a symbol, the symbol must meet all requirements of the selected filter index.

Serial Cmd:

<K744,filter number,symbology type,length,wildcard,placeholder,data, unused,database index>

Default:

0

Options:

0 to 64

Wildcard Definition:

This is the character to be used in the data output field when performing a data filter comparison. The wildcard character represents the end of matching, and allows for variable lengths of symbol output.

Serial Cmd:

<K744,filter number,symbology type,length,wildcard,placeholder,data, unused,database index>

Default:

“ * “ = 2A (hex)

Options:

Any ASCII input in the form of a pair of hex characters. Example: 2A = * 00 = disabled

Placeholder Definition:

The placeholder character requires a character to be present, but does not compare the data value.

Serial Cmd:

<K744,filter number,symbology type,length,wildcard,placeholder,data, unused,database index>

Default:

“ ? ” = 3F (hex)

Options:

Any ASCII input in the form of a pair of hex characters. Example: 3F = ? 00 = disabled

Quadrus Verifier User’s Manual

15-11

Page 402

Output Filter Configuration

Data Definition:

This is the data string to be used when comparing symbol data for output filtering and ordering. This data string may also contain wildcard and placeholder characters to facilitate matching. Remember that in order to filter or order symbol data, it must meet all the requirements of the selected filter index. Examples: • Filter data = “123*”. This will match data strings of “123”, “123456”, and “123ABC”, but not “12”. • Filter data = “123*AB?C”. This will be interpreted as “123*”. • Filter data = “123?”. This will match “1234” and “123A”, but not “123”, “12345”, or “1234C”. • Filter data = “123?A”. This will match “1234A” and “123BA”, but not “123”, “1234C”, or “1234ABCD”. • Filter data = “123?A?”. This will match “1234AB” and “123BAT”, but not “1234A” or “123BATS”. • Filter data = “12??*”. This will match “1234”, “123456”, and “123ABC”, but not “12” or “123”. • Filter data = “123?A*”. This will match “1234A”, “123BA”, and “123BATS”, but not “1234” or “1234C”.

Serial Cmd:

<K744,filter number,symbology type,length,wildcard,placeholder,data, unused,database index>

Default:

(00)

Options:

Any ASCII input in the form of a pair of hex characters. Examples: 41422A = AB* Data (00) = NULL represents string matching disabled.

Database Index Definition:

The index of the database entry that decodes a given symbol must equal this setting for filtering to occur. A setting of 0 allows any database index for this filter entry.

Serial Cmd:

<K744,filter number,symbology type,length,wildcard,placeholder,data, unused,database index>

Default:

0 (any index)

Options:

0 to 10

15-12

Quadrus Verifier User’s Manual

Page 403

Output Format

Ordered Output Filter Definition:

Number of Filters refers to the number of active output filters. 0 disables all output filters. Any non-zero numeral will enable filtering to be performed using the filter indexes covered by this value. For example, if the number of filters is 1, then only filter index # 1 will be applied. If the number of filters is 2, then only filter index # 1 and filter index # 2 will be applied, etc.

Serial Cmd:

<K745,number of filters>

Default:

0

Options:

0 to 10

Quadrus Verifier User’s Manual

15-13

Page 404

Ordered Output Filter

15-14

Quadrus Verifier User’s Manual

Page 405

16 Ethernet Contents Step 1 Setup............................................................................................................................... 16-2 Step 2 Preliminary Steps............................................................................................................ 16-3 Step 3 Communicating in Ethernet............................................................................................. 16-7 Step 4 Ethernet Application........................................................................................................ 16-8

While it is possible to connect the Quadrus Verifier directly to your host computer by Ethernet TCP/IP, typical usage is in a networked environment with either a switch or router.

Quadrus Verifier User’s Manual

16-1

Page 406

Setup

Step 1, Setup For Ethernet setup, you will need: 1. A Quadrus Verifier. 2. An Ethernet/USB interface. 3. An IB-150 Kit (interface and cable). 4. An Ethernet switch or router. RJ45 Ethernet cables (use a crossover cable if connecting directly to the host computer). 5. A laptop or desktop host computer with Ethernet connectivity to a network, Windows Vista, XP, or 2000, and Microscan’s ESP.

16-2

Quadrus Verifier User’s Manual

Page 407

Ethernet

Step 2, Preliminary Steps 1. Plug the Ethernet/USB interface into the Quadrus Verifier. 2. Plug the IB-150 Interface Kit cable into the Ethernet/USB interface and make the connection to the host computer and power supply. 3. Connect the Ethernet cables from the Ethernet/USB interface to the switch or router and from the switch or router to the host computer. The green NETWORK STATUS LED should illuminate when the Ethernet cable is connected. 4. Start ESP and establish communication with the host computer through the RS-232 host port. There are three ways the Quadrus Verifier can connect to a host by Ethernet: 1. Assigning a network address. 2. Using the Verifier’s default IP Address. 3. Using DHCP (Dynamic Host Configuration Protocol) assignment.

Quadrus Verifier User’s Manual

16-3

Page 408

Preliminary Steps

Assigning a Network Address This procedure will configure the Verifier to work with the host computer’s TCP/IP configuration. Note: This is only a temporary solution. The preferred method is to receive IP Address, Subnet, and Gateway (if necessary) from your IT department. If an IP Address has been assigned to the Verifier by your IT department (the preferred method), skip steps 1-3 below. 1. Determine the host computer's IP Address on the host's network. Go to the Windows Start menu, select Run, type “command”, and press Enter. a) On computers running a Windows2000 operating system, type “ipconfig” at the command prompt. 2. Add 1 to the last decimal value of the host computer’s IP Address. This will be the address that you assign to the Verifier. For example, if the host computer's address is 123.234.1.25, the Verifier's address will be 123.234.1.26. 3. To be sure that the new IP Address is available, go back to the command prompt and type: “ping [Verifier IP Address]”. For example: ping 123.234.1.26. If the ping program responds with a Request timed out message, then the IP Address can probably be used temporarily. If you receive a Reply message, then that address has already been assigned to another device and you need to contact your IT department for an available IP Address. 4. Send a Receive Reader Settings command. 5. From ESP’s Communications menu, enter your Verifier's new IP Address. For example, 123.234.1.26. Save this to the Verifier. The Verifier now has an IP Address that will allow communication with the host computer. In ESP’s Communications menu, you will see the view shown at right. 6. Go to Step 3, Communicating in Ethernet.

16-4

Quadrus Verifier User’s Manual

Page 409

Ethernet

Using the Verifier’s Default IP Address Note: This procedure is only used when the Verifier is directly connected to the computer. Also, this is only a temporary solution since the preferred method is to receive IP Address, Subnet, and Gateway (if necessary) from your IT department. 1. Determine the host computer's IP Address on the host's network. Go to the start menu, select Run, type “command”, and press Enter. a) On computers running a Windows 2000 operating system, type “ipconfig” at the command prompt. 2. Note the IP Address of the host computer. 3. At the Windows command prompt, type “route add 192.168.0.100 [host computer IP Address]” and press Enter. This adds the Verifier’s IP address to the computer’s routing table. 4. At the Windows command prompt, type “route print” and press Enter. 5. Look for 192.168.0.100 in the Network Destination column. If it does not appear in the Network Destination column, contact your IT department. 6. Go to Step 3, Communicating in Ethernet.

Quadrus Verifier User’s Manual

16-5

Page 410

Preliminary Steps

Using DHCP to Configure the Verifier This option assumes that the Quadrus Verifier is connected to a network with a DHCP server. RS-232 configuration is required to enable the Verifier's DHCP client (default IP Address Mode setting is Static, which disables DHCP). 1. Send a Receive Reader Settings command. 2. Under Network in ESP’s Communications menu, double-click on IP Address Mode and change Static to DHCP.

3. Right-click on the Communications menu and send Save to Reader > Send no Save. 4. Send a Receive Reader Settings command. 5. Notice that the network will have assigned new numbers to the IP Address, Subnet, and Gateway. (If the numbers have not changed, contact your IT department.) 6. Go to Step 3, Communicating in Ethernet.

16-6

Quadrus Verifier User’s Manual