Page 371
Utilities
Read Rate Read Rate by ESP You can access Read Rate from the Utilities menu. To see the number of decodes per second, click the Decodes/sec radio button and click the Start button. To see the percentage of decodes, click the Percent radio button and Start button. To end a read rate routine, click the Stop button (in the same position as the Start button).
Read Rate by Serial Command Enter Decodes/Second Test Sending <C> instructs the reader to transmit the decodes per second and symbol data (if any). The decode rate can vary dramatically due to the angle and location of the symbol in relation to the field of view. This test is very useful in aligning and positioning the reader during setup.
Enter Percent Test Sending <Cp> instructs the reader to transmit the percentage of decodes and any read symbol data.
Enable PDF Information Sending <a1> will cause PDF417 data to be prefaced with information consisting of error correction level (ECC Level n), number of rows (n Rows), number of columns (n Columns), number of informative code words (n Info Code Words) and the number of data characters (n Data Bytes). This feature can be disabled by re-sending <a1>.
End Read Rate Test Sending <J> ends both the Percent test and the Decodes/Second test.
Quadrus Verifier User’s Manual
14-3
Page 372
Counters
Counters Counter commands can be a numeric value from 00000 to 65,535. After reaching the maximum numeric limit of 65,535, an error message will be displayed and the counter will automatically roll-over and start counting again at 00000. To obtain the cumulative total of counts after the roll-over has occurred, add 65,536 per each roll-over (the reader does not keep track of the number of roll-overs) to the current count. Note: All counter values will be lost if power is recycled to the reader or the reader receives a reset or save command.
Counters by ESP You can access Counters from the Utilities menu. Click the Request button to display the appropriate count or Clear to set counter to zero.
14-4
Quadrus Verifier User’s Manual
Page 373
Utilities
Counters by Serial Command No Read Counter Sending <N> displays the total number of No Reads that have occurred since the last reset.
No Read Counter Reset Sending <O> sets No Read Counter to 00000.
Trigger Counter Sending <T> displays the total number of triggers since the last reset.
Trigger Counter Reset Sending <U> sets the trigger counter to 00000.
Good Read/Match Counter (or Good Read Counter) Sending <V> displays the total number of good reads matching the master symbol or, if Master Symbol is not enabled, the number of good reads since the last reset. This counter is always enabled, but will only work as a match count when Master Symbol is enabled. If Master Symbol is not enabled, this counter records the number of good reads. This count can be requested at any time.
Good Read/Match Counter Reset Sending <W> sets the Match Counter to 00000.
Mismatch Counter Sending <X> displays the number of decoded symbols since the last reset that do not match the master symbol.
Mismatch Counter Reset Sending <Y> sets the Mismatch Counter to zero.
Quadrus Verifier User’s Manual
14-5
Page 374
Device Control
Device Control Device Control by ESP
LED Light Control. May be configured in the same way as Output 1 and Output 2 if the Verifier is being used as a reader, and if the LED Light Control hardware is not connected.
Device Control By Serial Command Output 1 Pulse Sending <L1> pulses activates the link between Output 1(+) and Output 1(–) of the host connector (regardless of Master Symbol or Output 1 status).
Output 2 Pulse Sending <L2> pulses activates the link between Output 2(+) and Output 2(–) of the host connector (regardless of Master Symbol or Output 2 status).
Output 3 Pulse Important: Output 3 is used for the LED Light Control in the Verifier’s default hardware configuration. However, if the Verifier is being used as a reader and the LED Light Control is not connected, Output 3 can be configured in the same way as Output 1 and Output 2. Sending <L3> activates the link between Output 3(+) and Output 3(–) of the host connector (regardless of Master Symbol or Output 3 status).
Disable Reader Sending <I> will turn the reader OFF, end the current read cycle and not allow the reader to enter a read cycle until turned ON. This feature is useful during extended periods of time when no symbols are being scanned or the reader is being configured. Disabling the reader will not affect any downloaded commands to the reader.
Enable Reader Sending <H> will turn the reader ON and allow it to enter read cycles. 14-6
Quadrus Verifier User’s Manual
Page 375
Utilities
Differences from Default Clicking the Differences from Default button will cause ESP to check all stored configuration settings and compare them to default settings. All settings that are different than default will appear in the left column (shown below), and descriptions of those settings will appear in the right column. Click this button for a list of ESP configuration settings that are different than default settings.
Click Generate Barcode to bring up the Bar Code Dialog. Then create configuration symbols containing the configuration commands of your choice.
Send configuration settings to the reader without saving by clicking Send to Reader.
Click Save As to save the report as plain text or a tab-delimited text file.
Send configuration settings to the reader and save in ESP by clicking Send and Save.
• To create a symbol containing any of the command settings in the table, click Generate Barcode. This will bring up the Bar Code Dialog. • To save the Differences from Default report, either as plain text or as a tab-delimited text file, click Save As. • Click Send and Save to send the settings to the reader and save them, or Send to Reader to send the settings without saving them. Important: The use the Differences from Default feature, you must connect to the reader and Receive Reader Settings via the Send/Recv button on the toolbar.
Quadrus Verifier User’s Manual
14-7
Page 376
Master Database
Master Database Important: Master Database is used only for comparing entire symbols, when Sequential and Wild Card are NOT enabled, and Start Position is equal to 0.
Master Database Overview Usage:
Used where more than one master symbol is required, as in a Multisymbol setup, for matching and other matchcode operations.
Definition:
Allows you to define up to 10 master symbols as the master symbol database, which can be entered by keyboard, scanned in, displayed, or deleted by serial or ESP commands. 1. Click the Master Database tab. 2. Set the Master Symbol Database Size. 3. Select database index you want to enter the master symbol. 4. Do one of the following to enter master symbol data. a) Double-click the index row to type data directly into index. b) Click the Read Symbol into Selected Index to enter the next decoded symbol.
Set Master Symbol Database Size here. Loads the next symbol decoded into selected index.
Loads the reader’s saved master symbols into ESP.
Saves the displayed database to the reader.
Double-click row to enter data into popup dialog box.
Master Database
Important: Master Symbol Database is used only for comparing entire symbols, when Sequential and Wild Card are NOT enabled, and Start Position is equal to 0.
14-8
Quadrus Verifier User’s Manual
Page 377
Utilities
Master Symbol Database Size Definition:
Number of Master Symbols allows you to select 1 to 10 master symbols for the master symbol database.
Serial Cmd:
<K231,master symbol database size> Note: You must follow this command with a save command <A> or <Z>.
Default:
1
Options:
1 to 10
Use arrows to set Master Symbol Database Size.
Caution: Since the total number of characters available for the master symbol data base is 2710, changes to the Master Symbol Data Base Size will re-allocate the number of characters available for each master symbol and could cause existing master symbols to be deleted (except master symbol # 1 unless it also exceeds the size limitation). The table below specifies the maximum number of characters available to each symbol according to the number of master symbols defined, from 1 through 10. Maximum Characters for Master Symbol Master Symbol Number #1 #2 #3 #4 #5
Quadrus Verifier User’s Manual
Maximum Characters 2710 1355 903 677 542
Master Symbol Number #6 #7 #8 #9 #10
Maximum Characters 451 387 338 301 271
14-9
Page 378
Master Database
Enter Master Symbol Data Definition:
Allows you to enter master symbol data for a any enabled master symbol index number (1 to 10), provided the total number of characters does not exceed the maximum indicated in the Maximum Characters for Master Symbol table.
Serial Cmd:
<K231,master symbol number,master symbol data>
Options:
Enter data for 1 to 10 symbols (any combination of ASCII text up to the maximum indicated in Maximum Characters for Master Symbol). For example, to enter data for master symbol 5, after making certain that Master Symbol Database Size is enabled for 5 or more symbols (see Master Symbol Database Size table), send <K231,9,data>. Caution: If no data is entered, the existing data will be deleted.
ESP:
1. Open the Utilities menu. 2. Set the number of master symbols you want to create in Master Symbol Database Size. 3. Double-click on each symbol number you want to set up and copy or type your data in the popup dialog and click OK.
4. When all your data has been entered, click Send Database to the Reader.
14-10
Quadrus Verifier User’s Manual
Page 379
Utilities
Request Master Symbol Data Definition:
Returns master symbol data for any enabled master symbols from 1 to 10. For example, to request master symbol # 5, enter <K231?,5>. The reader transmits master symbol # 5 data in brackets in the following format: <5/ data>. If no master symbol data is available, the output will be: <5/>.
Serial Cmd:
<K231?,master symbol number> Caution: Be sure to add the ? or you will delete the master symbol. Returns the number of master symbols if no number is included.
ESP:
1. Click the Utilities button and the Master Database tab. 2. Click on the Receive Reader’s Database button.
Request All Master Symbol Data Serial Cmd:
<K231?> This command will return master symbol data for all symbols enabled (up to 10).
Read Next Symbol as Master Symbol Definition:
After you’ve set the size in the database, you can set the reader to read the next symbol as the master symbol for any given master symbol number.
Serial Cmd:
<G master symbol number> To store the next symbol read as master symbol #1, send: <G> or <G1>. To store next symbol read as the master symbol for any other master symbol database number, send: <Gmaster symbol number [1-10]>. For example, <G5> will cause the next symbol read to be entered as master symbol # 5.
ESP:
In the Master Database tab: 1. Select the master symbol index number in which you want to store the symbol data. 2. Click on Read Symbol into Selected Index. Caution: If you’ve selected an index which has existing data, that data will be copied over by scanned data when you use this command.
Quadrus Verifier User’s Manual
14-11
Page 380
Master Database
Request New Master Status Usage:
Informs the user when a new master symbol is pending and which position it is in.
Definition:
Returns the position in the master symbol database that will be loaded on the next read.
Serial Cmd:
<NEWM> The reader returns: <NEWM/next master to load> Once a symbol has been read and loaded, the status will be cleared and the response will be <NEWM/0>.
Delete Master Symbol Data Definition:
You can directly delete the master symbol data by serial command or ESP.
ESP:
1. Click the Utilities button to access the master symbol: 2. Click the Master Database tab and double-click the symbol number you want to delete. 3. Delete text and click OK.
Serial Cmd:
14-12
<K231,master symbol number,> To delete a master symbol, enter the database number and a comma, but leave the data field empty. For example, to delete master symbol # 5, send the following <K231,5,>. The command is entered with a blank master symbol data field which tells the reader to delete the selected master symbol from the database.
Quadrus Verifier User’s Manual