Page 361
IP Database
Request Selected Database Settings Definition:
Returns configuration settings for the selected database index.
Serial Cmd:
<K250?,database index>
Example: <K250?,5> This command phrase returns the configuration settings for database index 5.
Request All Database Settings Definition:
Returns configuration settings for all indexes in the database.
Serial Cmd:
<K250?>
Example: <K250?> This command phrase returns the configuration settings for all database indexes.
Quadrus Verifier User’s Manual
12-13
Page 362
Request All Database Settings
12-14
Quadrus Verifier User’s Manual
Page 363
13 Terminal Contents Terminal Window ........................................................................................................................13-2 Find ............................................................................................................................................13-3 Send............................................................................................................................................13-4 Macros ........................................................................................................................................13-5 Terminal Window Menus.............................................................................................................13-6
This section describes the Terminal window and macro functions. Note: You can learn the current setting of any parameter by inserting a question mark after the number, as in <K100?>. To see all K commands, send <K?>.
Quadrus Verifier User’s Manual
13-1
Page 364
Terminal Window
Terminal Window
To use the Terminal interface, click the Terminal button. You will see this window:
Terminal screen.
Copy, Paste, and Save.
Type text here to find in Terminal window. Scroll down to recall previous commands.
Type serial commands here, or directly in the terminal below.
Click on Macro arrow to Add Macro, Remove Macro, or Edit Macro.
Click on Macro to run.
Clear screen.
Macro list on this bar.
The Terminal screen allows you to enter serial commands from the macro box, by copying and pasting, or directly from your keyboard. The Terminal screen also displays symbol data or other information from the Verifier. You can also right click in the Terminal screen to bring up a menu of further options.
13-2
Quadrus Verifier User’s Manual
Page 365
Terminal
Find The Find function allows you to enter text strings to be searched for in the Terminal window. For example, a series of symbols have been scanned into the Terminal view and you want to determine if a particular symbol whose data begins with “ABC” has been read. 1. Type “ABC” into the Find box.
2. Press Enter. The first instance of “ABC” will be highlighted in the Terminal window. 3. Click the Find button to the right of the text field to locate additional instances of “ABC”.
Quadrus Verifier User’s Manual
13-3
Page 366
Send
Send The Send function allows you to enter serial commands and then send them to the Verifier. For example, you need to perform a single capture verification routine on a symbol. 1. Type the single capture command <V1> into the Send box.
2. Press Enter. Single capture verification will occur and a report will be generated in the Terminal window. 3. Click the Send button to the left of the text field to initiate additional single capture verification routines.
13-4
Quadrus Verifier User’s Manual
Page 367
Terminal
Macros Macros can be stored in a macro selection bar, edited in a separate window, and executed by clicking on the macro name.
Click on subsequent arrows to Edit macros. Click on Next Row to see the next row of macros.
Click on the first arrow here to see Add Macro or Default Macros. When you default macros, the entire macros set is restored to original commands.
When you click on the macro name, the macro is executed in the Terminal window. If this is a command, the command is also sent to the Verifier at the same time it is displayed.
Editing a Macro When you click the arrow next to a any macro and select Edit, the following appears:
You can edit an existing macro or type in the Macro Name text box and define it in the Macro Value text box. Click OK.
Quadrus Verifier User’s Manual
13-5
Page 368
Terminal Window Menus
Terminal Window Menus Right click on the Terminal window to display the following menu: • Copy selected text to clipboard. • Paste from Terminal or computer text. • Clear all text in Terminal window. • Select All text in the Terminal window. • Save... brings up a save as dialog box. • Change Font... of data text, brings up a text dialog. • Change Echo Font... to change typed in text or commands. • Change Background Color of Terminal window. • Default Settings changes all the above back to default settings.
Terminal Dropdown Menu The dropdown Terminal menu has Capture Text and Save Current Text functions, as well as the functions defined above. • Capture lets you append data in real time to a text file of your choice. While in operation, the text file cannot be opened. You can select Pause to interrupt the capture flow or Stop to end the flow and open the file. • Save Current Text... saves all text in the Terminal window to a text file of your choice.
13-6
Quadrus Verifier User’s Manual
Page 369
14 Utilities Contents Serial Utility Commands ............................................................................................................. 14-2 Read Rate .................................................................................................................................. 14-3 Counters ..................................................................................................................................... 14-4 Device Control ............................................................................................................................ 14-6 Differences from Default............................................................................................................. 14-7 Master Database ........................................................................................................................ 14-8 Firmware................................................................................................................................... 14-13 Bar Code Configuration. ........................................................................................................... 14-16 Defaulting/Saving/Resetting ..................................................................................................... 14-17 Status Requests ....................................................................................................................... 14-19 Other Operational Commands.................................................................................................. 14-20
Utility commands are generally commands that are performed during reader operations to check read rates, determine read rates or perform miscellaneous operations on reader hardware. Serial utility commands are not prefaced with a ‘K’ and a numeric code. Nor do they require an initialization command (<A> and <Z>). They can be entered from within any terminal program or from within ESP in the Terminal window or the window adjacent to the Utilities menu.
Quadrus Verifier User’s Manual
14-1
Page 370
Serial Utility Commands
Serial Utility Commands Read Rate
Counter Request and Clear
Master Database
Part Number and Checksum
Device Control
Default/Reset/Save
Status Commands
Operational Commands
14-2
<C> <Cp> <J> <N> <O> <T> <U> <V> <W> <X> <Y> <E> <F> <G> <M?> <#> <#a> <#b> <#d> <#f> <!> <!a> <!b> <!d> <!f> <L1> <L2> <L3> <A> <Ard> <Arp> <Arc> <Z> <Zc> <Zrd> <Zrc> <K?> <-> <?> <?1> <@VER> <V1> <V2> <V3> <@AIMDPM,R-max,R-min> <V4>
Enter Decode Rate Test Enter Percent Rate Test Exit Decode Rate and Percent Rate Test No Read Counter No Read Counter Reset Trigger Counter Trigger Counter Reset Good Read/Match Counter Good Read/Match Counter Reset Mismatch Counter Mismatch Counter Reset Enable Master Symbol * Disable Master Symbol* Store next Symbol read to Database. Request Master Symbol Information Display Software Code Part Numbers Display Application Code Part Number Display Boot Code Part Number Display DSP Part Number Display FPGA Part Number Display all 3 Checksums of Flash memory Display Application Code Checksum Display Boot Code Checksum Display DSP Code Checksum Display FPGA Code Checksum Programmable Output 1 Programmable Output 2 Programmable Output 3 Reset (does not save for power-on) Reset and recall Microscan defaults Reset and recall power-on parameters Reset and recall customer default parameters Save current settings for power-on Save current settings as customer default parameters Recall Microscan default parameters and save Recall customer default parameters and save All Configuration Commands Status Input Status Reader Status Extended Reader Status ISO/IEC 15415 Reflectance Calibration ISO/IEC 15415 Single Capture Verification ISO/IEC 15415 Multi-Capture Verification AS9132 Verification AIM DPM Reflectance Calibration AIM DPM Verification
Quadrus Verifier User’s Manual