Page 311

Matchcode

Wild Card Character Usage:

For example, with Wild Card Character defined as the default asterisk, defining CR*34 as the master symbol will result in matches for CR134, CR234, but not CR2345. Entering the wild card at the end of the master symbol as in CR* will result in matches for variable symbol lengths such as CR1, CR23, CR358, etc.

Definition:

Wild Card Character allows a user to define a wild card character as part of the master symbol.

Serial Cmd:

<K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on no read,sequence on mismatch>

Default:

* (asterisk)

Options:

Any valid ASCII character.

Quadrus Verifier User’s Manual

9-9

Page 312

Sequence On No Read

Sequence On No Read Usage:

Sequence On No Read is useful when the reader needs to stay in sequence even if no decode occurs.

Definition:

When Sequence On No Read is Enabled and Matchcode is set to Sequential, the reader sequences the master symbol on every match or No Read. When disabled, it does not sequence on a No Read.

Serial Cmd:

<K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on no read,sequence on mismatch>

Default:

Enabled

Options:

0 = Disabled

1 = Enabled

An example of Sequence on No Read Enabled: Master symbol 001 002 003 004 005 006 007

Decoded symbol 001 002 No Read 004 No Read No Read 007

Master symbol after decode 002 003 004 (sequenced on No Read) 005 006 (sequenced on No Read) 007 (sequenced on No Read) 008

An example of Sequence on No Read Disabled: Master symbol 001 002 003 003 004 004 004

9-10

Decoded symbol 001 002 No Read 003 No Read No Read 004

Master symbol after decode 002 003 003 (not sequenced) 004 004 (not sequenced) 004 (not sequenced) 005

Quadrus Verifier User’s Manual

Page 313

Matchcode

Sequence On Mismatch Note: Matchcode must be set to Sequential for this command to function. Usage:

Enable this parameter if every trigger event should have a decode and more than one consecutive mismatch may occur. Disable this parameter if every trigger event should have a decode but no more than one consecutive mismatch may occur.

Definition:

When set to Enabled, the master symbol sequences on every decode, match or mismatch. When set to Disabled, the master symbol will not sequence whenever consecutive mismatches occur.

Serial Cmd:

<K223,matchcode type,sequential matching,match start position,match length,wild card character,sequence on no read,sequence on mismatch>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

The reader will sequence the master to one more/less than the decoded symbol. An example of Sequence On Mismatch Enabled: Master symbol 001 002 003 004 005 006 007

Decoded symbol 001 002 abc 004 def ghi 007

Master symbol after decode 002 003 004 (sequenced on mismatch) 005 006 (sequenced on mismatch) 007 (sequenced on mismatch) 008

An example of Sequence On Mismatch Disabled: Master symbol 001 002 003 004 005 006 006

Decoded symbol 001 002 abc 004 def ghi 006

Quadrus Verifier User’s Manual

Master symbol after decode 002 003 004 (sequenced because of previous match) 005 006 (sequenced because of previous match) 006 (not sequenced because of previous mismatch) 007

9-11

Page 314

Sequence Step

Sequence Step Usage:

Useful in applications in which it is desirable to count by intervals other than 1.

Definition:

Sequencing in Matchcode operations can occur in steps from 1 to 32,768. Sequencing performs like a mechanical counter by displaying positive integers and a specific number of digits after roll-overs. For example, 000 – 3 = 997 (not –3) and 999 + 3 = 002 (not 1002).

Serial Cmd:

<K228,sequence step>

Default:

1

Options:

1 to 32768

Note: See <K223> command for more information. An example of Sequence Step, if Sequence Step is set to 3 and Sequential Matching is set to Increment: Master symbol 003 003 003 006 006 006

9-12

Decoded symbol 001 002 003 004 005 006

Master symbol after decode 003 003 006 006 006 009

Quadrus Verifier User’s Manual

Page 315

Matchcode

Match Replace Usage:

Provides a convenient shortcut for applications that need to output a pre-defined text string whenever a symbol matches a master symbol.

Definition:

Outputs a user-defined data string whenever a match occurs and Matchcode is enabled.

Serial Cmd:

<K735,status,replacement string>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Replacement String Definition:

User-defined string that when enabled will replace the matchcode data.

Serial Cmd:

<K735,status,replacement string>

Default:

MATCH

Options:

Any ASCII string up to 128 characters.

Quadrus Verifier User’s Manual

9-13

Page 316

Mismatch Replace

Mismatch Replace Usage:

Provides a convenient shortcut for applications that need to output a pre-defined text string whenever a symbol does not match a master symbol.

Definition:

Outputs a user-defined data string whenever a mismatch occurs and Matchcode is enabled.

Serial Cmd:

<K736,status,replacement string>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Replacement String Definition:

User-defined string that when enabled will be output whenever a mismatch occurs.

Serial Cmd:

<K736,status,replacement string>

Default:

MISMATCH

Options:

Any ASCII string up to 128 characters.

9-14

Quadrus Verifier User’s Manual

Page 317

Matchcode

New Master Pin Usage:

Some applications require the line worker to change the master symbol. This can be done by installing a switch at the location of the reader. It is very common to have a keyed switch so that accidental switching does not occur.

Definition:

After New Master Pin is enabled, a pulse can be received on the new master pin that will cause the reader to record the next decoded symbol(s) as the new master symbol(s). It is important to note that the enabling New Master Pin does not in itself cause a master symbol to be recorded. The master pin must then be activated momentarily (for a minimum of 10 ms) before a master symbol can be read into memory.

Serial Cmd:

<K225,status>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

After New Master Pin has been enabled and the pin activated, decodes will be saved in the master database beginning with master symbol #1. If the reader is configured for a multisymbol read cycle (Number of Symbols is greater than 1), the remaining decodes will be saved in each consecutive master symbol location. For example, if Number of Symbols is set to 3 and New Master Pin is then activated, at the end of the next read cycle, the decoded symbols will be saved as master symbols 1, 2, and 3.

Quadrus Verifier User’s Manual

9-15

Page 318

New Master Pin

9-16

Quadrus Verifier User’s Manual

Page 319

10 Diagnostics Contents Diagnostics by ESP.....................................................................................................................10-2 Diagnostics Serial Commands ....................................................................................................10-2 Counts (Read Only).....................................................................................................................10-3 External Camera Message..........................................................................................................10-4 Over Temperature Message .......................................................................................................10-6 Service Message.........................................................................................................................10-7

This section describes warning and operating messages and their settings.

Quadrus Verifier User’s Manual

10-1

Page 320

Diagnostics by ESP

Diagnostics by ESP To open the entire contents of the tree control, hold down the Alt key and single-click the +. To close the entire contents of the tree control, hold down the Alt key and single-click the - . Click this button to bring up the Diagnostics menu.

To open nested options, single-click the +.

To change a setting, double-click the setting and use your cursor to scroll through the options.

Diagnostics Serial Commands Power-on/Reset Counts External Camera Message Over Temperature Message Service Message

10-2

<K406,power-on,resets,power-on saves,customer default saves> <K410,disconnect msg status,disconnect message, connect msg status,control message> <K402,over temperature status,warning message> <K409,status,service message,threshold,resolution>

Quadrus Verifier User’s Manual