Page 291
I/O Parameters
Diagnostics (Output 2) Note: Output On under Output 2 Parameters must be set to Diagnostic Warning for this output to function.
Serial Cmd:
<K791,over temp,service unit,external camera disconnect>
See Diagnostics (Output 1) for command structure and options.
Quadrus Verifier User’s Manual
8-51
Page 292
Output 3 Parameters
Output 3 Parameters Note: The Quadrus Verifier does not have an Output 3 option by default. If your application requires an Output 3, contact your Microscan sales representative for more information. Output 3 has the same parameters and default settings as Output 1 and Output 2.
Serial Cmd:
<K812,output on,active state,pulse width,output mode>
Output On (Output 3)
Serial Cmd:
<K812,output on,output state,pulse width,output mode>
See Output On (Output 1) for command structure and options.
Output State (Output 3)
Serial Cmd:
<K812,output on,output state,pulse width,output mode>
See Output State (Output 1) for command structure and options. 8-52
Quadrus Verifier User’s Manual
Page 293
I/O Parameters
Pulse Width (Output 3)
Serial Cmd:
<K812,output on,output state,pulse width,output mode>
See Pulse Width (Output 1) for command structure and options.
Output Mode (Output 3)
Serial Cmd:
<K812,output on,output state,pulse width,output mode>
See Output Mode (Output 1) for command structure and options.
Trend Analysis (Output 3) Note: Output On under Output 3 Parameters must be set to Trend Analysis for this output to function.
Serial Cmd:
<K782,trend analysis mode,number of triggers,number to output on>
See Trend Analysis (Output 1) for command structure and options.
Quadrus Verifier User’s Manual
8-53
Page 294
Output 3 Parameters
ISO/IEC 15415 Symbol Quality (Output 3) Note: Output On under Output 3 Parameters must be set to Symbol Quality for this output to function.
All ISO/IEC 15415 Symbol Quality parameters except Print Growth feature the alphanumeric grade options shown below. If the decoded symbol’s grade falls below the selected threshold grade, the output will activate.
Serial Cmd:
<K832,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth value>
See ISO/IEC 15415 Symbol Quality (Output 1) for command structure and options.
AS9132 Symbol Quality (Output 3) Note: Output On under Output 3 Parameters must be set to Symbol Quality for this output to function.
All AS9132 Symbol Quality parameters feature the dropdown menu shown below. When set to Enabled, a failing grade will activate output. Note: The Symbol Contrast parameter only takes effect when Marking Method is set to Laser or Chemical Etch <K711,,1>.
Serial Cmd:
<K842,dot center offset,cell fill, dot ovality,angle of distortion,symbol contrast, quiet zone>
See AS9132 Symbol Quality (Output 1) for command structure and options. 8-54
Quadrus Verifier User’s Manual
Page 295
I/O Parameters
Diagnostics (Output 3) Note: Output On under Output 3 Parameters must be set to Diagnostic Warning for this output to function.
Serial Cmd:
<K792,over temp,service unit,external camera disconnect>
See Diagnostics (Output 1) for command structure and options.
Quadrus Verifier User’s Manual
8-55
Page 296
Configuring EZ Trax Output
Configuring EZ Trax Output If you intend to use the Quadrus Verifier with EZ Trax Software, you can configure the software in three different ways: using the EZ Trax interface, using the tree controls in ESP, and using serial commands. For detailed information about how to use the EZ Trax user interface for configuration, refer to the Help menu in EZ Trax software, or the EZ Trax Quick Start Guide, available on the Microscan Tools CD.
Configuring EZ Trax Output by ESP To configure EZ Trax using ESP, first click the I/O button in App Mode to display the I/O Parameters tree control.
8-56
Quadrus Verifier User’s Manual
Page 297
I/O Parameters
EZ Trax Output When EZ Trax output is enabled, the Imager will attempt to output all enabled EZ Trax options. EZ Trax will not function unless enabled.
Comm Port
Selects the communication port that will be used to transfer image files.
Image Mode
Determines which read cycle condition triggers image file output.
Image Format
Selects the format of the image output file.
JPEG Quality Sets the amount of JPEG image compression. A value of 1 represents the highest compression and lowest image quality, and 100 represents the lowest compression and highest image quality.
Quadrus Verifier User’s Manual
8-57
Page 298
Configuring EZ Trax Output
Object Info Output
If a symbol is decoded during an image capture, the symbol type and other supplementary symbol information is also included.
8-58
Quadrus Verifier User’s Manual
Page 299
I/O Parameters
Configuring EZ Trax Output by Serial Command Usage: Definition: Serial Cmd: Default: Options:
Useful for when you want to configure EZ Trax output using serial commands instead of the EZ Trax graphic interface. When EZ Trax output is enabled, the Imager will attempt to output all enabled EZ Trax options. EZ Trax will not function unless enabled. <K757,status,comm port,image mode,image format,jpeg quality,object info output> Disabled 0 = Disabled 1 = Enabled
Comm Port Definition: Serial Cmd: Default: Options:
Selects the communication port that will be used to transfer image files. <K757,status,comm port,image mode,image format,jpeg quality,object info output> Host 0 = Host 1 = Aux 2 = USB
Host Port Sends output using the current Host Port.
Aux Port Sends output using the Auxiliary Port.
USB Sends output using a USB connection.
Image Mode Definition: Serial Cmd: Default: Options:
Determines which read cycle condition triggers image file output. <K757,status,comm port,image mode,image format,jpeg quality,object info output> Disabled 0 = Disabled 1 = Good Read 2 = No Read 3 = No Read and Good Read 4 = Mismatch
Good Read Outputs the first Good Read image in the read cycle. The image file will immediately follow the symbol data.
Quadrus Verifier User’s Manual
8-59
Page 300
Configuring EZ Trax Output
No Read Outputs the first No Read image in the read cycle. The image file will immediately follow the symbol data. Note: If two symbols are present in the same image capture and one of them is successfully decoded, no image will be output.
No Read and Good Read Outputs the first No Read or Good Read image in the read cycle.
Mismatch Outputs the image of the first mismatch condition.
Image Format Definition: Serial Cmd: Default: Options:
Selects the format of the image output file. <K757,status,comm port,image mode,image format,jpeg quality,object info output> Compressed 0 = Lossless 1 = Compressed
Lossless A “lossless” image is one that is in a high resolution format, and that will not lose resolution even when manipulated and saved multiple times. When Lossless is enabled, the JPEG Quality setting has no effect.
Compressed Outputs the image in JPEG format. Image quality is determined by the JPEG Quality setting.
JPEG Quality Definition:
Serial Cmd: Default: Options:
8-60
Sets the amount of JPEG image compression. A value of 1 represents the highest compression and lowest image quality, and 100 represents the lowest compression and highest image quality. <K757,status,comm port,image mode,image format,jpeg quality,object info output> 90 1 to 100 (percent)
Quadrus Verifier User’s Manual