For the Microscan Verifier

Page 281

I/O Parameters

Number of Triggers (Trend Analysis) Definition:

The number of triggers to examine for the trend analysis condition.

Serial Cmd:

<K780,trend analysis mode,number of triggers,number to output on>

Default:

0

Options:

0 to 255

Number to Output On (Trend Analysis) Definition:

Sets the number of Trend Analysis Mode events (mismatches, No Reads or reads/trigger as configured by Trend Analysis Mode) to occur within the trigger evaluation time period before activating the associated output. For example, if Number to Output On is set to 3 and Trend Analysis Mode is set to No Read, then the output will not be activated until 3 No Reads have occurred.

Serial Cmd:

<K780,trend analysis mode,number of triggers,number to output on>

Default:

0

Options:

0 to 255

Quadrus Verifier User’s Manual

8-41

Page 282

Output 1 Parameters

ISO/IEC 15415 Symbol Quality (Output 1) Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function.

All ISO/IEC 15415 Symbol Quality parameters except Print Growth feature the alphanumeric grade options shown below. If the decoded symbol’s grade falls below the selected threshold grade, the output will activate.

Grade (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Decode (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

8-42

Quadrus Verifier User’s Manual

Page 283

I/O Parameters

Symbol Contrast (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Fixed Pattern Damage (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Axial Non-Uniformity (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Grid Non-Uniformity (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Quadrus Verifier User’s Manual

8-43

Page 284

Output 1 Parameters

Modulation (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Unused Error Correction Capacity (ISO/IEC 15415 Symbol Quality) Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

Disabled

Options:

0 = Disabled 1 = Grade 1 (D) 2 = Grade 2 (C) 3 = Grade 3 (B) 4 = Grade 4 (A)

Print Growth (ISO/IEC 15415 Symbol Quality)

The Print Growth dropdown menu shows percentage values of 1% - 9%. The output will activate if the absolute value calculated by the verification algorithm exceeds the selected percentage value parameter.

Serial Cmd:

<K830,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth>

Default:

0 (Disabled)

Options:

0 to 9 (0 = Disabled; 1% - 9%)

8-44

Quadrus Verifier User’s Manual

Page 285

I/O Parameters

AS9132 Symbol Quality (Output 1) Note: Output On under Output 1 Parameters must be set to Symbol Quality for this output to function.

All AS9132 Symbol Quality parameters feature the dropdown menu shown below. When set to Enabled, a failing grade will activate output. Note: The Symbol Contrast parameter only takes effect when Marking Method is set to Laser or Chemical Etch <K711,,1>.

Dot Center Offset (AS9132 Symbol Quality) Serial Cmd:

<K840,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast, quiet zone>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Cell Fill (AS9132 Symbol Quality) Serial Cmd:

<K840,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast, quiet zone>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Dot Ovality (AS9132 Symbol Quality) Serial Cmd:

<K840,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast, quiet zone>

Default:

Disabled

Options:

0 = Disabled

Quadrus Verifier User’s Manual

1 = Enabled

8-45

Page 286

Output 1 Parameters

Angle of Distortion (AS9132 Symbol Quality) Serial Cmd:

<K840,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast, quiet zone>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Symbol Contrast (AS9132 Symbol Quality) Serial Cmd:

<K840,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast, quiet zone>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Quiet Zone (AS9132 Symbol Quality) Serial Cmd:

<K840,dot center offset,cell fill,dot ovality,angle of distortion,symbol contrast, quiet zone>

Default:

Disabled

Options:

0 = Disabled

8-46

1 = Enabled

Quadrus Verifier User’s Manual

Page 287

I/O Parameters

Diagnostics (Output 1) Note: When Diagnostic Warning is enabled under Output 1 Parameters, Output Mode configuration has no effect. Note: Output On under Output 1 Parameters must be set to Diagnostic Warning for this output to function.

Usage:

Alerts user to critical conditions.

Definition:

Sets up specific warnings that will cause activation on output 1. The output will remain active as long as one of the diagnostic conditions is met. The output will go inactive as soon as it detects no active diagnostic warning.

Over Temperature Definition:

Sets the output to toggle to active when an over-temperature condition is detected.

Serial Cmd:

<K790,over temp,service unit,external camera disconnect>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Service Unit Definition:

Sets the output to toggle to active when the service timer has expired. This condition will only be held for one service timer tick.

Serial Cmd:

<K790,over temp,service unit,external camera disconnect>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

External Camera Disconnect Note: This feature cannot be used if in a Continuous Read mode. Definition:

Sets the output to toggle to active if the external camera goes off-line.

Serial Cmd:

<K790,over temp,service unit,external camera disconnect>

Default:

Disabled

Options:

0 = Disabled

Quadrus Verifier User’s Manual

1 = Enabled 8-47

Page 288

Output 2 Parameters

Output 2 Parameters Note: Output 2 has the same parameters and default settings as Output 1.

Output On (Output 2)

Serial Cmd:

<K811,output on,output state,pulse width,output mode>

See Output On (Output 1) for command structure and options.

Output State (Output 2)

Serial Cmd:

<K811,output on,output state,pulse width,output mode>

See Output State (Output 1) for command structure and options.

8-48

Quadrus Verifier User’s Manual

Page 289

I/O Parameters

Pulse Width (Output 2)

Serial Cmd:

<K811,output on,output state,pulse width,output mode>

See Pulse Width (Output 1) for command structure and options.

Output Mode (Output 2)

Serial Cmd:

<K811,output on,output state,pulse width,output mode>

See Output Mode (Output 1) for command structure and options.

Trend Analysis (Output 2) Note: Output On under Output 2 Parameters must be set to Trend Analysis for this output to function.

Serial Cmd:

<K781,trend analysis mode,number of triggers,number to output on>

See Trend Analysis (Output 1) for command structure and options.

Quadrus Verifier User’s Manual

8-49

Page 290

Output 2 Parameters

ISO/IEC 15415 Symbol Quality (Output 2) Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.

All ISO/IEC 15415 Symbol Quality parameters except Print Growth feature the alphanumeric grade options shown below. If the decoded symbol’s grade falls below the selected threshold grade, the output will activate.

Serial Cmd:

<K831,grade,decode,symbol contrast,fixed pattern damage,axial non-uniformity, grid non-uniformity,modulation,unused error correction capacity,print growth value>

See ISO/IEC 15415 Symbol Quality (Output 1) for command structure and options.

AS9132 Symbol Quality (Output 2) Note: Output On under Output 2 Parameters must be set to Symbol Quality for this output to function.

All AS9132 Symbol Quality parameters feature the dropdown menu shown below. When set to Enabled, a failing grade will activate output. Note: The Symbol Contrast parameter only takes effect when Marking Method is set to Laser or Chemical Etch <K711,,1>.

Serial Cmd:

<K841,dot center offset,cell fill, dot ovality,angle of distortion,symbol contrast, quiet zone>

See AS9132 Symbol Quality (Output 1) for command structure and options. 8-50