Page 261
I/O Parameters
Control/Hex Output Usage:
Useful for viewing settings with binary characters when using serial command on a terminal.
Definition:
Determines the response to an Serial Command Echo or status request command. When set to Control, two characters are transmitted to represent a nondisplayable character. For example, a carriage return will be shown as the two characters: ^M. When set to Hex, the output is the hex character.
Serial Cmd:
<K701,serial command echo status,serial command beep status,control/ hex output>
Default:
Control
Options:
0 = Control
Quadrus Verifier User’s Manual
1 = Hex
8-21
Page 262
Video Output
Video Output Usage:
Useful for remote visual confirmation and review of images.
Definition:
Configures the operation of video output of RS170 standard video monitors. Video source is configured via the Active Camera command <K240>. The video output can be confined to a single event such as a good read or can be fed as continuous live video.
Video Output
Video Output Status Serial Cmd:
<K760,video output status,trigger image mode,image frame>
Default:
Live (Real Time)
Options:
0 = Disabled
1 = Triggered
2 = Live (Real Time)
Disabled When selected the video output is disabled.
Triggered Usage:
Ideal for dynamic applications.
Definition:
Outputs specific capture to the video output port as per the setting in Trigger Image Mode.
Live (Real Time) Note: Whenever the Locate mode (target pattern) is activated by the EZ button, Live video will automatically be enabled. Usage:
Live output is useful during initial setup in locating symbols in the field of view in real time.
Definition:
The video output is configured for a real-time operating mode and is refreshed with every video frame. If used in conjunction with Dual Camera mode, video will only output from the internal camera.
Note: When Live video mode is active, in order to synchronize with the video format, a shutter time of 1/1000 is the slowest shutter speed setting that can be applied to the camera settings. Slower shutter speeds will disable the video output.
8-22
Quadrus Verifier User’s Manual
Page 263
I/O Parameters
Trigger Image Mode (Video Output) Definition:
Event which triggers a video output.
Serial Cmd:
<K760,video output status,trigger image mode,image frame>
Default:
Last Capture
Options:
0 = Last Capture 1 = Good Read 2 = No Read 3 = Slide Show 4 = Selected Capture
Last Capture At the end of a triggered read cycle, the video output will be the last capture.
Good Read At the end of a triggered read cycle, the video output will be the last good read capture. Note: If a good read does not occur, there is no video output.
No Read At the end of a triggered read cycle, the video output will be the last No Read capture.
Slide Show At the end of a triggered read cycle, all of the captures taken in the read cycle will be output at a rate of 350 mS between captures. The last capture will be held for 700 mS and then the cycle will be repeated. Note: This only functions in a triggered mode.
Selected Capture At the end of a triggered read cycle, the video output will be the capture specified in the Capture Number field.
Image Frame (Video Output) Definition:
Specifies the image frame that will be output at the end of the read cycle.
Serial Cmd:
<K760,video output status,trigger image mode,image frame>
Default:
1
Options:
1 to 5
Quadrus Verifier User’s Manual
8-23
Page 264
Image Output
Image Output Usage:
Useful for remote visual examination and review of images.
Definition:
Outputs an image file when a specified condition (good read or No Read) is met. The video output can be confined to a single event such as a good read or can be fed as continuous live video.
Serial Cmd:
<K739,image output status,com port,file format,JPEG quality>
Default:
Disabled
Options:
0 = Disabled 1 = Good Read 2 = No Read 3 = Good Read and No Read
Disabled No image file will be output at the end of the read cycle.
Good Read (Image Output) Usage:
Provides a visual record for comparison.
Definition:
If a Good Read condition is met, the image file of the first good read image will be output, immediately following the read cycle output + postamble. In order for a good read condition to occur all symbols must qualify in the read cycle.
No Read (Image Output) Usage:
Provides visual representation to identify quality issues with No Read images.
Definition:
If a No Read occurs, the image file of the first No Read image will be output, immediately following the read cycle output + postamble.
Conditions where an image will not be output: • The Verifier was expecting two symbols in the same image capture but only one was read. • The image had a good read present and therefore would not be considered a No Read image.
Good Read and No Read (Image Output) Usage:
Provides a visual record for comparison and allows the user to identify quality issues with No Read images.
Definition:
The image file of the first Good Read image and the first No Read image will be output immediately following the read cycle output + postamble.
8-24
Quadrus Verifier User’s Manual
Page 265
I/O Parameters
Communications Port (Image Output) Definition:
The communications port to which the image will be sent.
Serial Cmd:
<K739,image output mode,com port,file format,JPEG quality>
Default:
Host
Options:
0 = Host
1 = Aux
2 = Network
File Format (Image Output) Definition:
File format of the output image.
Serial Cmd:
<K739,image output mode,com port,file format,JPEG quality>
Default:
JPEG
Options:
0 = Bitmap
1 = JPEG
2 = Binary
Bitmap Outputs the image in a bitmap format.
JPEG Outputs the image in a JPEG format.
Binary Outputs the image in a raw binary format.
JPEG Quality (Image Output) Definition:
Determines the relative quality of the JPEG image sent, with 100 being the highest quality.
Serial Cmd:
<K739,image output mode,com port,file format,JPEG quality>
Default:
90
Options:
1 to 100 (where 100 is the highest quality)
Quadrus Verifier User’s Manual
8-25
Page 266
Image Captioning
Image Captioning Usage:
Useful in verifying data output visually in real time.
Definition:
Overlays text onto the specified image. The text displayed is dependent on the captioning mode that is enabled. The text captioning is overlaid onto the image frame that is output to the video port, and will be displayed on the uploaded image for that image frame as well. If more than one image frame is available in a read cycle, the image frame selected by the triggered video mode will be displayed. When the Slide Show video mode is enabled the text is overlaid onto every image frame prior to being output to the video port, so every image will have overlay text when uploaded.
Serial Cmd:
<K762,mode>
Default:
Disabled
Options:
0 = Disabled 1 = Read Cycle Results 2 = Statistic Mode 1 (counts) 3 = Statistic Mode (timing)
Important Notes: • If triggered video is not enabled this feature is disabled regardless of the mode setting. • Only 26 characters can be displayed per line, and only 2 lines of data will be displaye33d. If the data string is longer than this, it will be truncated. • 1 line of captioning takes approximately 125ms to complete overlay. Therefore, for 2 lines of overlay an additionally 250ms of overhead will have to be added to the read cycle duration.
Disabled Image captioning is disabled.
Read Cycle Results (Image Captioning) When enabled the results of the read cycle will be overlaid onto the triggered video image in the upper left hand corner. This will include Symbol data, No Read message(s) (if enabled), and any displayable formatting such as pre-amble, postamble, and Symbol ID.
8-26
Quadrus Verifier User’s Manual
Page 267
I/O Parameters
Statistic Mode 1 (counts) (Image Captioning) Outputs trigger count, image frame number, decode status, running read rate, good read/ match count, No Read count, and mismatch count. Count values are total number since reset. Format: DECODE XXX% F/X
T/XXXXX V/XXXXX N/XXXXX
Where: DECODE = Decode status: “DECODE” or “NOREAD” T/ = Trigger count
0 - 65535
(5 digits)
0 - 65535
(5 digits)
% = Read rate
0 - 100
(3 digits)
F/ = Image frame number
0-7
(1 digit)
N/ = No Read count
0 - 65535
(5 digits)
X/ = Mismatch count
0 - 65535
(5 digits)
V/ =
Good read / Match count
Important: • If Match Code is disabled, the Mismatch count will not be displayed. • Read rate is calculated as a running average.
Statistic Mode 2 (timing) (Image Captioning) Indicates timing information including processing time required for displayed image and total read cycle time. Also included is Decode status and image frame number. Format: TOTAL READ TIME = XXXXX ms (if decoded) or STATUS (if a No Read) F/X READ CYCLE = XXXXX ms Where: Total Read Time = Processing time required for displayed image (1ms resolution). 1 to 65535ms (up to 5 digits) Status = Decode status is a No Read F/ = Image frame number
0-7
(1 digit)
Read Cycle = Total read cycle time (1ms resolution). 1 to 65535ms (5 digits)
Quadrus Verifier User’s Manual
8-27
Page 268
Synchronous Trigger
Synchronous Trigger Usage:
Helps center the image on the video display.
Definition:
Configures the image capture trigger to be synchronous with the video frame to allow for stable video during image capture events.
Serial Cmd:
<K761,synchronous trigger mode>
Default:
Disabled
Options:
0 = Disabled 1 = Enabled
Disabled When disabled, the trigger event for an image capture is asynchronous with the video frame. This means that when a trigger occurs, the video timing is reset and an image capture occurs immediately. This causes flickering in an output monitor during image captures because video timing needs to resync every time the video signal is reset.
Enabled (Synchronous Trigger) When enabled, the trigger event for an image capture is synchronous with the video frame. This means that when a trigger occurs the image captured is not released until the start of the next video frame, which occurs every 16.68ms. The video timing never changes and no effect is seen during image captures on the video monitor.
Restrictions When the Synchronous Trigger mode is enabled, it will not take effect unless the following conditions are met: 1. Shutter selection is 1/1200 or faster. 2. Live Video Mode is in effect.
8-28
Quadrus Verifier User’s Manual
Page 269
I/O Parameters
Application Notes 1. If a Rapid Capture trigger mode is enabled, the timing on the trigger will change when Synchronous Trigger mode is enabled, and will probably not be sufficient for most applications. This is because the trigger can have up to a 16ms delta from the time the trigger is issued until the image capture occurs. Thus the trigger timing will not be stable and should not be used in a dynamic applications. 2. If the IP database is being used in a Continuous Capture mode, the capture sequence is sequential and no longer double buffered. Thus a capture cannot occur until the previous frame has been processed and up to a 16ms delay can be introduced between image captures. 3. In a Continuous trigger, Continuous Capture, or read rate mode, the decode speed should not be affected (while using the synchronous trigger mode) since a double buffer format is used for image capturing. The only time a slow down should be noticed in these modes is when the decode time is less than the capture time. Capture time = 16ms.
Quadrus Verifier User’s Manual
8-29
Page 270
EZ Button
EZ Button The EZ button has four positions, selectable by the length of time that the button is held down and indicated by one, two, three, and four beeps in succession. Each position can be programmed for any of eight options. Definition:
Serves as a master switch to enable/disable the EZ button status.
Serial Cmd:
<K770,global status,default on power-on,load IP database,save for power-on>
Default:
Enabled
Options:
0 = Disabled
1 = Enabled
2 = Trigger
Disabled When set to Disabled, the EZ button is disabled.
Enabled When selected, the EZ button is enabled and the function of each button position is selected by the EZ Button Mode command.
Trigger Mode (EZ Button) When selected, the EZ button acts as a trigger for the reader to start and end read cycles. All other button operations are inactive. In External Level:
The read cycle endures for as long as the EZ button is pressed, unless a timeout occurs and Timeout is enabled for End Of Read Cycle.
In External Edge:
As with Level, Edge allows a read cycle to be initiated by pressing the button, but unlike level mode, the read cycle ends with a good read output, a timeout, or a new trigger.
Default on Power-On (EZ Button) Definition:
When enabled, if the EZ button is held down on power-on the reader will default to customer defaults and saved for power-on. This is the same as sending a <Zrc> command.
Serial Cmd:
<K770,global status,default on power-on,load IP database,save for power-on>
Default:
Enabled
Options:
0 = Disabled
8-30
1 = Enabled
Quadrus Verifier User’s Manual