I/O Parameters
Symbol Size Mode (2D Symbology Qualification) Note: if Symbol Size 1 is larger than Symbol Size 2, it will be automatically reversed in the algorithm. Definition:
Specifies the outputs resulting from the results of searches for Symbol Size 1 and Symbol Size 2.
Serial Cmd:
<K718,finder pattern status,symbol size mode,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
Disabled
Options:
0 = Disabled 1 = Bad/No Symbol Output 2 = Object Qualification 3 = Enable Both
Disabled Decoded messages are output, but no attempt to qualify symbols or output Bad or No Symbol messages is undertaken.
Bad Symbol/No Symbol Output Checks first for decodes. If no symbol is decoded, checks for symbol size. If both Symbol Size 1 and Symbol Size 2 requirement are met, a Bad Symbol message is output. If both are not met, outputs a No Symbol message.
Object Qualification Checks first for matches to symbol size. If qualified, attempts to decode. If no decodes are found within the specified read cycle parameters, outputs a No Read message.
Enable Both Same as Object Qualification, except that if no objects qualify, then a No Symbol message is output; if at least one object qualifies, then a Bad Symbol message is output.
Symbol Size 1 (2D Symbology Qualification) Definition:
Checks for symbol sizes for a specified number of elements. In the case of rectangular symbols, checks the longer side.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
10
Options:
8 to 88 elements
Quadrus Verifier User’s Manual
8-11
Page 252
Bad Symbol/No Symbol Qualification
Symbol Size 2 (2D Symbology Qualification) Definition:
Checks for symbol sizes for a specified number of elements. In the case of rectangular symbols, checks the shorter side.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
10
Options:
8 to 88 elements
Symbol Size Tolerance (2D Symbology Qualification) Definition:
Sets the allowable deviation, up or down, for symbol sizes specified in Symbol Size 1 and Symbol Size 2.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
2
Options:
0 to 10
Dimension Mode (2D Symbology Qualification) Definition:
Enables the searches for symbol Dimension 1 and Dimension 2.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
Disabled
Options:
0 = Disabled 1 = Bad/no symbol output 2 = Object qualification 3 = Enable both
Disabled Decoded messages are output, but no attempt to qualify symbols or output Bad or No Symbol messages is undertaken.
Bad/No Symbol Output Checks first for decodes. If no symbol is decoded, checks for symbol dimension. If both Dimension 1 and Dimension 2 requirements are met, a Bad Symbol message is output. If both are not met, outputs a No Symbol message.
8-12
Quadrus Verifier User’s Manual
Page 253
I/O Parameters
Object Qualification Checks first for matches to symbol dimension. If qualified, attempts to decode. If no decodes are found within the specified read cycle parameters, outputs a No Read message.
Enable Both Same as Object Qualification, except that if neither Dimension 1 nor Dimension 2 qualifies, then a No Symbol message is output; if either Dimension 1 or Dimension 2 qualifies, then a Bad Symbol message is output.
Dimension 1 (2D Symbology Qualification) Definition:
Checks symbol dimension for a specified number of pixels. In the case of rectangular symbols, checks the longer side.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
0
Options:
0 to 656 pixels
Quadrus Verifier User’s Manual
8-13
Page 254
Bad Symbol/No Symbol Qualification
Dimension 2 (2D Symbology Qualification) Definition:
Checks symbol dimension for a specified number of pixels. In the case of rectangular symbols, checks the shorter side.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
0
Options:
0 to 656 pixels
Dimension Tolerance (2D Symbology Qualification) Definition:
Sets the allowable deviation, in percentage, for symbol sizes specified in Dimension 1 or Dimension 1.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
10
Options:
0 to 100
Orientation Mode (2D Symbology Qualification) Definition:
Sets the orientation value.
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
Disabled
Options:
0 = Disabled 1 = Bad/no symbol output 2 = Object qualification 3 = Enable both
Disabled Decoded messages are output, but no attempt to qualify symbols or output Bad or No Symbol messages is undertaken.
Bad/No Symbol Output Checks first for decodes. If no symbol is decoded, checks for symbol orientation. If Orientation Value is satisfied, a Bad Symbol message is output. If not satisfied, outputs a No Symbol message.
8-14
Quadrus Verifier User’s Manual
Page 255
I/O Parameters
Object Qualification Checks first for matches to symbol Orientation Value. If qualified, attempts to decode. If no decodes are found within the specified read cycle parameters, outputs a No Read message.
Enable Both Same as Object Qualification, except that if Orientation Value is satisfied but the symbol is not decoded, then a Bad Symbol message is output; if not satisfied, outputs a Bad Symbol message.
Orientation Value Usage:
Instructs the reader to ignore symbols not in the specified orientation.
Definition:
Sets the orientation of the symbol, in degrees clockwise from the default 0 orientation. For Data Matrix symbols, sets the finder “L” pattern, with the adjacent being and example of the default 0 “L” finder pattern orientation. (Data Matrix) For QR Code, sets the locator pattern (three squares), with the adjacent being an example of the default 0 orientation. The other position settings will rotate clockwise from the defaults shown above.
Three-square finder pattern (QR Code)
Serial Cmd:
<K718,finder pattern status,symbol size status,symbol size 1,symbol size 2,symbol size tolerance,dimension mode,dimension 1,dimension 2,dimension tolerance,orientation mode,orientation value>
Default:
0
Options:
0 to 359
Note: A symbol’s orientation can miss the precise Orientation Value by several degrees and still be qualified. To be certain of the extent of the allowable tolerance, you should experiment with symbols in various orientations. In no case can they be more than 45° from the assigned orientation value and still be qualified.
Quadrus Verifier User’s Manual
8-15
Page 256
Read Duration Output
Read Duration Output Usage:
Useful in evaluating actual read cycle timing results, especially when initially setting up an application to determine maximum line speed that can be obtained based on the spacing between symbols.
Definition:
When enabled, the duration of the read cycle (in milliseconds) is appended to the symbol data. The read duration is the time from the beginning of the read cycle until data is output.
Read Duration Output Mode Serial Cmd:
<K706,status,separator>
Default:
Disabled
Options:
0 = Disabled 1 = Enabled
Important: To measure the entire read cycle when in External Level trigger mode, set When to Output Symbol Data to End of Read Cycle. This output can measure over 49 days worth of duration; if exceeded, the OVERFLOW message will be output in place of the duration. Options:
0 = Disabled
1 = Enabled
Read Duration Output Separator Definition:
User-defined character that separates the symbol information from the Read Duration Output.
Serial Cmd:
<K706,status,separator>
Default:
[space character]
Options:
Any ASCII character
8-16
Quadrus Verifier User’s Manual
Page 257
I/O Parameters
Output Indicators The Quadrus Verifier has a beeper and LED arrays, arranged as follows.
Target Pattern Usage:
Assists users in positioning and locating symbols in the center of the reader’s FOV.
Definition:
The user can control when the targeting system is ON or OFF and save this condition for power-on.
Serial Cmd:
<K750,green flash mode,target pattern status,green flash duration>
Default:
Always OFF
Options:
0 = Always OFF 1 = ON when NOT in the read cycle 2 = ON when in the read cycle 3 = Always ON
Important: The targeting system actuated by the EZ button or operational command overrides this setting.
Always OFF The target pattern will remain OFF at all times unless overridden by the EZ button or operational command.
On When NOT In Read Cycle The target pattern is always ON except during the read cycle. If the EZ button or operational command overrides this setting, the target pattern will remain on at all times.
On When in the Read Cycle The target pattern will remain OFF except during the read cycle. If the EZ button or operational command overrides this setting, the target pattern will remain on at all times.
Always ON The target pattern is always ON.
Quadrus Verifier User’s Manual
8-17
Page 258
Output Indicators
Green Flash Mode Usage:
Used as a visual verification that a good read has occurred.
Definition:
An array of green LEDs in the front of the Quadrus Verifier can be programmed to flash briefly in response to user-defined conditions, or if in Static Presentation mode, illuminate for a set period of time.
Serial Cmd:
<K750,green flash mode,target pattern status,green flash duration>
Default:
Good Read
Options:
0 = Disabled 2 = Static Presentation 4 = Mismatch
1 = Good Read 3 = Match
Disabled Green flash LEDs are disabled.
Good Read (Green Flash) Green flash LEDs will flash for less than one second when a good read condition is meet or when matchcode is enabled and a match occurs.
Static Presentation (Green Flash) Static Presentation is used in conjunction with Continuous Read mode. When operating in Static Presentation mode, the red LEDs will illuminate while the Quadrus Verifier is searching for a symbol in Continuous Read mode. When a symbol is placed in the FOV and a good read occurs, the green LEDs will illuminate and stay on for the duration of time set in Green Flash Duration. Only one read will occur during that time unless more than one symbol is enabled in Number of Symbols. Note: If Static Presentation mode is selected but the reader is not in Continuous Read, the Green Flash will revert to Good Read mode.
To use Static Presentation: 1. Enable Continuous Read. 2. Select the number of symbols. 3. Enable Static Presentation in Green Flash Mode. 4. Select the read time in Green Flash Duration. Note: Green Flash Duration values only take effect in Static Presentation Mode.
Match (Green Flash) Green flash LEDs will flash for less than 1 second when a match condition is met. If multisymbol is enabled, then green flash LEDs will illuminate only if all symbols qualify as a match. If matchcode is disabled, then this mode will activate the LED's on a good read.
Mismatch (Green Flash) Same as On Match except it illuminates on a mismatch.
8-18
Quadrus Verifier User’s Manual
Page 259
I/O Parameters
Green Flash Duration Usage:
A visual verification that a good read has occurred.
Definition:
When a good read occurs, the Green flash LED's will illuminate and stay on for the time set for the Green Flash Duration value.
Serial Cmd:
<K750,green flash mode,target pattern status,green flash duration>
Default:
100 (1 second)
Options: 0 to 65535 (in 10mS increments) Note: Green Flash Duration values only take effect in Static Presentation Mode.
Beeper Usage:
An audible verification that either a good read or a No Read has occurred.
Definition:
A beep is emitted either after each good read or No Read.
Serial Cmd:
<K702,beeper status>
Default:
Good Read
Options:
0 = Disabled 1 = Good Read 2 = No Read
Note: Beeper will also sound if any of the following occur: • The reader is defaulted. • A send/save command from ESP or an Exit command from any embedded menu. • A calibration routine is completed. • A <Z>, <Zp>, <Zd>, or <K701,1> command is sent.
Quadrus Verifier User’s Manual
8-19
Page 260
Serial Verification
Serial Verification Allows the user to verify configuration command status.
Serial Command Echo Status Usage:
This command is useful in removing any doubt about the reader’s interpretation of any configuration command. For example, if the current preamble is “SOM” and <K701,1,START> is entered, the reader will echo back <K701, SOM> since the attempted entry “START” exceeds the four character limit for that command. Therefore it is rejected and the existing “SOM” message is echoed back and remains the preamble message.
Definition:
When enabled, a configuration command received from the host is echoed back to the host with the resultant settings.
Host
Verifier
Function:
If a command with multiple fields is processed, some of the fields may have been processed properly while others were not. The changes will appear in the string echoed back so that the user will know which fields did or did not change.
Serial Cmd:
<K701,serial command echo status,serial command beep status,control/ hex output>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Serial Command Beep Status Usage:
Used to audibly verify the acceptance and validity of a command.
Definition:
Causes the reader to beep once whenever a K command is entered to indicate that the command was accepted and processed.
Function:
If an invalid command is entered, the reader beeps 5 times to indicate an invalid entry. However, this does not necessarily mean that all data fields have been entered incorrectly. Only one bad field needs to be found in order to activate the 5 beep response.
Serial Cmd:
<K701,serial command echo status,serial command beep status,control/ hex output>
Default:
Disabled
Options:
0 = Disabled
8-20
1 = Enabled
Quadrus Verifier User’s Manual