Page 231

Symbologies

RSS Expanded Usage:

Used to encode primary and supplementary data in retail point-of-sale and other applications.

Definition:

RSS Expanded is a variable length symbology that can encode supplementary information in addition to the 14-digit EAN.UCC item identification number and is capable of encoding up to 74 numeric or 41 alphabetic characters.

Serial Cmd:

<K484,status,fixed symbol length status,fixed symbol length>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled (non-stacked) 2 = Enabled (stacked and non-stacked)

Where appropriate, use 1 (non-stacked) instead of 2 (stacked and non-stacked).

Fixed Symbol Length Status (RSS Expanded) Definition:

When enabled, the reader will check the symbol length against the symbol length field, minus the embedded check digit. If disabled, any length would be considered a valid symbol.

Serial Cmd:

<K484,status,fixed symbol length status,fixed symbol length>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length (RSS Expanded) Usage:

Fixed Symbol Length helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted.

Definition:

Specifies the exact number of characters that the reader will recognize (this does not include start and stop and check digit characters). The reader ignores any symbol not having the specified length.

Serial Cmd:

<K484,status,fixed symbol length status,fixed symbol length>

Default:

10

Options:

1 to 74

Quadrus Verifier User’s Manual

7-21

Page 232

RSS Limited

RSS Limited Usage:

RSS Limited is designed to be read by laser and CCD readers. It is not recommended for omnidirectional slot readers.

Definition:

Encodes a smaller 14-digit symbol (74 modules wide) that is not omnidirectional.

Serial Cmd:

<K483,status>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

7-22

Quadrus Verifier User’s Manual

Page 233

Symbologies

RSS-14 Usage:

Used in the grocery retail and prescription drug industries where full 14-digit EAN.UCC item identification may be needed.

Definition:

RSS-14 (Reduced Space Symbology) is a fixed length symbology that encodes 14-digits, including a 1 digit indicator digit and is 96 modules wide. It can be stacked into two rows, read omnidirectional if printed in full height, or not if height-truncated for small marking.

Serial Cmd:

<K482,status>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled (non-stacked) 2 = Enabled (stacked and non-stacked)

Where appropriate, use 1 (non-stacked) instead of 2 (stacked and non-stacked).

Quadrus Verifier User’s Manual

7-23

Page 234

PDF417

PDF417 Usage:

Used in applications where a large amount of information (over 32 characters) needs to be encoded within a symbol, typically where the symbol is transported from one facility to another. For example, an automobile assembly line might use a single symbol with multiple fields of information that will be read at several stations along the way, without reference to a database.

Definition:

A two-dimensional, multi-row (3 to 90), continuous, variable length symbology that has high data capacity for storing up to 2700 numeric characters, 1800 printable ASCII characters, or 1100 binary character per symbol. Each symbol character consists of 4 bars and 4 spaces in a 17-module structure.

Serial Cmd:

<K476,status,[unused],fixed symbol length status,fixed symbol length,[unused],codeword collection>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Note: Sending <a1> will cause PDF417 data to be prefaced with information consisting of error correction level (ECC Level n), number of rows (n Rows), number of columns (n Columns), number of informative code words (n Info Code Words) and the number of data characters (n Data Bytes). This feature can be disabled by re-sending <a1>.

Fixed Symbol Length Status (PDF417) Serial Cmd:

<K476,status,[unused],fixed symbol length status,fixed symbol length,[unused],codeword collection>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length (PDF417) Usage:

Used to increase data integrity by ensuring that only one symbol length will be accepted.

Definition:

When enabled, the PDF symbol must contain the same number of characters as the symbol length setting before it can be considered a good read. The reader will ignore any symbol not having the specified length.

Serial Cmd:

<K476,status,[unused],fixed symbol length status,fixed symbol length, [unused],codeword collection>

Default:

10

Options:

1 to 2710

Note: Fixed Symbol Length Status must be enabled for Fixed Symbol Length to take effect. 7-24

Quadrus Verifier User’s Manual

Page 235

Symbologies

Codeword Collection (PDF417) Usage:

Multiple Codeword Collection is useful in applications where portions of subsequent images can be read and pieced together so that no opportunity or time is lost to assemble codewords for decoding.

Definition:

When set to Multiple, PDF codewords is collected from multiple images and assembled throughout the read cycle until the read cycle ends or the symbol is fully decoded. It is important to note that when this feature is enabled, only one PDF symbol should be presented to the reader per read cycle.

Serial Cmd:

<K476,status,[unused],fixed symbol length status,fixed symbol length,[unused],codeword collection>

Default:

Single Image

Options:

0 = Single Image 1 = Multiple Image

Note: Fixed Symbol Length Status must be enabled for Fixed Symbol Length to take effect.

Quadrus Verifier User’s Manual

7-25

Page 236

MicroPDF417

MicroPDF417 Usage:

Used for labeling small items that need large capacity.

Definition:

A variant of PDF417, a very efficient and compact stacked symbology that can encode up to 250 alphanumeric characters or 366 numeric characters per symbol.

Serial Cmd:

<K485,status,[usused],fixed symbol length status,fixed symbol length>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length Status (MicroPDF417) Serial Cmd:

<K485,status,[unused],fixed symbol length status,fixed symbol length>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length (MicroPDF417) Usage:

Used to increase data integrity by ensuring that only one symbol length will be accepted.

Definition:

When enabled, the Micro PDF symbol must contain the same number of characters as the symbol length setting before it can be considered a good read. The reader will ignore any symbol not having the specified length.

Serial Cmd:

<K485,status,[unused],fixed symbol length status,fixed symbol length>

Default:

10

Options:

1 to 2710

Note: Fixed Symbol Length Status must be enabled for Fixed Symbol Length to take effect.

7-26

Quadrus Verifier User’s Manual

Page 237

Symbologies

Composite When set to Enabled or Required, will attempt to decode the composite component of a linear symbol. The linear symbol can be RSS14, RSS Expanded, RSS Limited, UCC/EAN128, UPC-A, EAN-13, EAN-8, and UPC-E. Usage:

Allows reading by both linear and 2D readers.

Definition:

Combines 2D and linear width modulated symbology on the same symbol where different messages can be read by each reader type.

Serial Cmd:

<K453,mode,separator status,separator>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Required

Enabled If set to Enabled it only decodes the linear component, it will output only the linear component.

Required If set to Required, it must decode both the linear and the composite components; otherwise it outputs a No Read.

Separator Status (Composite) Usage: Definition: Serial Cmd: Default: Options:

Allows user to distinguish between the main and Supplemental symbols. This separator separates the linear symbol and the composite component. <K453,mode,separator status,separator> Disabled 0 = Disabled 1 = Enabled

Separator Character (Composite) Note: If Separator Character has been changed to any other character and you wish to re-define the separator as a comma, you will need to use ESP or the embedded menu. Usage: Definition: Serial Cmd: Default: Options:

As required by the application. Allows the user to change the separator character from a comma to a new character. <K453,mode,separator status,separator> , (comma) Any ASCII character

Quadrus Verifier User’s Manual

7-27

Page 238

Narrow Margins

Narrow Margins Note: Narrow Margins is only used for linear symbology types. Do not use with PDF. Usage:

Used when the leading and trailing edges of the symbols are smaller than the standard margin or other objects encroach into the margins.

Definition:

Allows the reader to read symbols with quiet zones less than 8 times the width of the narrow bar element. “Quiet zone” is the space at the leading and trailing ends of a symbol. Each quiet zone can be as narrow as only five times the width of the narrow bar element when Narrow Margins is enabled.

Serial Cmd:

<K450,narrow margins,symbology identifier status>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Note: Do not use Narrow Margins with Large Intercharacter Gap enabled in Code 39.

7-28

Quadrus Verifier User’s Manual

Page 239

Symbologies

Symbology ID Usage:

Used when the symbology type and how it’s decoded needs to be known.

Definition:

Symbology ID is an ISO/IEC 15415 standard prefix set of characters that identifies the symbol type. When enabled, the reader analyzes and identifies the symbology and adds a three character identifying prefix to the data: 3. ] (closed bracket character) indicating the presence of a symbology ID. 4. A, C, E, I, L, Q, b, d, p A = Code 39; C = Code 128; E = UPC/EAN; I = I–2/5; L = PDF417; Q = QR Code; b = BC412; d = Data Matrix; p = Pharmacode 5. Modifier

Serial Cmd:

<K450, narrow margins,symbology identifier status>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Explanation of Modifiers for Code 39, Codabar, and Interleaved 2 of 5 • For Code 39, Codabar and I 2/5, the modifier indicates Check Digit and Check Digit Output status. • For Code 39 only, Full ASCII needs to be enabled to see modifiers 4, 5, and 7. Modifier

Check Digit

0 1 3 4 5 7

Disabled Enabled Enabled Disabled Enabled Enabled

Check Digit Output N/A Enabled Disabled N/A Enabled Disabled

Full ASCII conversion performed (Code 39 only) No No No Yes Yes Yes

Example: ]A5 indicates a Code 39 symbol with Check Digit and Check Digit Output enabled and Full ASCII conversion performed.

Explanation of Modifiers for Other Symbologies • For Code 128, a 1 indicates ECC/EAN-128; otherwise the modifier is a 0. • For QR Code, a 0 indicates Model 1; a 1 indicates Model 2. • For all other codes, the modifier is 0.

Quadrus Verifier User’s Manual

7-29

Page 240

Background Color

Background Color Usage:

If the background is darker than the symbol, then enable black background. Typically the background is white; but on PCBs for example, they can be black.

Definition:

Allows the user to choose which symbol background (white or black) the reader can read.

Serial Cmd:

<K451, background color>

Default:

White

Options:

0 = White

7-30

1 = Black

Quadrus Verifier User’s Manual