Page 211
7 Symbologies Contents Symbologies by ESP .................................................................................................................... 7-2 Symbologies Serial Commands ................................................................................................... 7-3 Data Matrix ................................................................................................................................... 7-4 QR Code....................................................................................................................................... 7-6 Code 39 ........................................................................................................................................ 7-7 Code 128 .................................................................................................................................... 7-10 BC412......................................................................................................................................... 7-11 Interleaved 2 of 5........................................................................................................................ 7-13 UPC/EAN.................................................................................................................................... 7-16 Pharmacode ............................................................................................................................... 7-19 RSS Expanded ........................................................................................................................... 7-21 RSS Limited................................................................................................................................ 7-22 RSS-14 ....................................................................................................................................... 7-23 PDF417 ...................................................................................................................................... 7-24 MicroPDF417.............................................................................................................................. 7-26 Composite .................................................................................................................................. 7-27 Narrow Margins .......................................................................................................................... 7-28 Symbology ID ............................................................................................................................. 7-29 Background Color....................................................................................................................... 7-30
The Quadrus Verifier conforms to ISO/IEC 15415 2D verification requirements (Data Matrix), AS9132 verification requirements (dot peen and laser-etch / chemical etch Data Matrix), and the AIM DPM Quality Guideline. This section describes the Data Matrix symbology, and also the other symbologies that can be read by the Quadrus Verifier when it is being used as a reader. See http://www.aimglobal.org/standards/aimpubs.asp for additional information about symbologies.
Quadrus Verifier User’s Manual
7-1
Page 212
Symbologies by ESP
Symbologies by ESP To open the entire contents of the tree control, hold down the Alt key and single-click the +. To close the entire contents of the tree control, hold down the Alt key and single-click the - . Click this button to bring up the Symbologies menu.
To open nested options, single-click the +.
To change a setting, double-click the setting and use your cursor to scroll through the options.
7-2
Quadrus Verifier User’s Manual
Page 213
Symbologies
Symbologies Serial Commands Data Matrix
QR Code Code 39
Code 128 BC412 Interleaved 2 of 5 UPC/EAN Pharmacode RSS Expanded RSS Limited RSS-14 PDF417 Micro PDF417 Composite Narrow Margins/ Symbology ID Background Color
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status,ECC100 status,ECC140 status,ECC 120 status,ECC 130 status> <K480,status> <K470, status,check digit status,check digit output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set> <K474,status,fixed symbol length status,fixed symbol length> <K481, status,check digit output,fixed symbol length status,fixed symbol length> <K472,status,check digit status,check digit output,symbol length #1,symbol length #2,guard bar> <K473,UPC status,EAN status,supplementals status,separator status,separator character,supplemental type> <K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar width status,direction,fixed threshold value> <K484,status,fixed symbol length status,fixed symbol length> <K483,status> <K482,status> <K476,status,[unused],fixed symbol length status,fixed symbol length,[unused],codeword collection> <K485,status,[usused],fixed symbol length status,fixed symbol length> <K453,mode,separator status,separator> <K450,narrow margins,symbology identifier status> <K451, background color>
Quadrus Verifier User’s Manual
7-3
Page 214
Data Matrix
Data Matrix The Quadrus Verifier is certified for verification of Data Matrix ECC 200 symbols, per the ISO/IEC 15415 and AS9132 standards. Usage:
Useful when information needs to be packed into a small area and/or when symbols need to be applied directly to the substrate with etching, dot peen, or other methods.
Definition:
Data Matrix is a type of Matrix symbology and has subsets ECC 000, 200. ECC 200 symbols have an even number of rows and an even number of columns. Most of the symbols are square with sizes from 10 x 10 to 144 x 144. Some symbols however are rectangular with sizes from 8 x 18 to 16 x 48. All ECC 200 symbols can be recognized by the upper right corner module being light (binary 0).
ECC 200 Definition:
When enabled, will decode ECC 200 symbols.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Enabled Note: This is the only symbol type enabled by default.
Options:
0 = Disabled
1 = Enabled
ECC 000 Definition:
When enabled, will decode ECC 000 symbols
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
ECC 050 Definition:
When enabled, will decode ECC 050 symbols.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
7-4
1 = Enabled
Quadrus Verifier User’s Manual
Page 215
Symbologies
ECC 080 Definition:
When enabled, will decode ECC 080 symbols.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
ECC 100 Definition:
When enabled, will decode ECC 100 symbols.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
ECC 140 Definition:
When enabled, will decode ECC 140 symbols.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
ECC 120 Definition:
ECC 120 is a legacy symbol and NOT part of the ISO/IEC 15415 standard.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
ECC 130 Definition:
ECC 130 is a legacy symbol and NOT part of the ISO/IEC 15415 standard.
Serial Cmd:
<K479,ECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status,ECC 120 status,ECC 130 status>
Default:
Disabled
Options:
0 = Disabled
Quadrus Verifier User’s Manual
1 = Enabled 7-5
Page 216
QR Code
QR Code Usage:
Widely implemented in the automotive industry in Japan and throughout their worldwide supply chain.
Definition:
The QR Code is capable of handling numeric, alphanumeric, byte data as well as Japanese kanji and kana characters. Up to 7,366 characters (numeric data) can be encoded using this symbol. Therefore, less space is required to encode the same amount of data in the QR Code than in a conventional symbol, helping to reduce the size of a symbol and lower the paper costs. Three position detection patterns in the symbol make omnidirectional, ultra-fast reading possible. QR Code has error correction capability. Data can be frequently be restored even if a part of the symbol has become dirty or been damaged.
Serial Cmd:
<K480,status>
Default:
Disabled
Options:
0 = Disabled
7-6
1 = Enabled
Quadrus Verifier User’s Manual
Page 217
Symbologies
Code 39 Usage:
Code 39 is considered the standard for non-retail symbology.
Definition:
An alphanumeric symbology with unique start/stop code patterns, composed of 9 black and white elements per character, of which 3 are wide.
Serial Cmd:
<K470,status,check digit status,check digit output status,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Check Digit Status (Code 39) Serial Cmd:
<K470,status,check digit status,check digit output,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Check Digit Output Status (Code 39) Usage:
Check digit Output, added to the symbol, provides additional security.
Definition:
When enabled, the check digit character is read and compared along with the symbol data. When disabled, symbol data is sent without the check digit. Note: With Check Digit Output and an External or Serial trigger option enabled, an invalid check digit calculation will cause a No Read message to be transmitted at the end of the read cycle.
Serial Cmd:
<K470,status,check digit status,check digit output,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
Disabled
Options:
0 = Disabled
Quadrus Verifier User’s Manual
1 = Enabled
7-7
Page 218
Code 39
Large Intercharacter Gap (Code 39) Usage:
Large Intercharacter Gap is helpful for reading symbols that are printed out of specification. Caution: Do not use Large Intercharacter Gap with Narrow Margins enabled since a large intercharacter gap (over 3x) could cause a narrow margins (5x) to be interpreted as an intercharacter gap.
Definition:
When enabled, the reader can read symbols with gaps between symbol characters that exceed three times (3x) the narrow element width.
Serial Cmd:
<K470,status,check digit status,check digit output,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Fixed Symbol Length Status (Code 39) Definition:
When enabled the reader will check the symbol length against the symbol length field. If disabled any length would be considered a valid symbol.
Serial Cmd:
<K470, status,check digit status,check digit output,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Fixed Symbol Length (Code 39) Usage:
Fixed Symbol Length helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted.
Definition:
Specifies the exact number of characters that the reader will recognize (this does not include start and stop and check digit characters). The reader ignores any symbology not having the specified length.
Serial Cmd:
<K470,status,check digit status,check digit output,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
10
Options:
1 to 128
7-8
Quadrus Verifier User’s Manual
Page 219
Symbologies
Full ASCII Set (Code 39) Usage:
Must be enabled when reading characters outside the standard character set (0-9, A-Z, etc.) User must know in advance whether or not to use Full ASCII Set option. Since Full ASCII Set requires two code words to encode one character, it is less efficient.
Definition:
Standard Code 39 encodes 43 characters; zero through nine, capital “A” through capital “Z”, minus symbol, plus symbol, forward slash, space, decimal point, dollar sign and percent symbol. When Full ASCII Set is enabled, the reader can read the full ASCII character set, from 0 to 255.
Serial Cmd:
<K470,status,check digit status,check digit output,large intercharacter gap,fixed symbol length status,fixed symbol length,full ASCII set>
Default:
Disabled
Options:
0 = Disabled
Quadrus Verifier User’s Manual
1 = Enabled
7-9
Page 220
Code 128
Code 128 Usage:
Code 128 is a smaller symbology useful in applications with tight spots and high security needs.
Definition:
A very dense alphanumeric symbology. It encodes all 128 ASCII characters, it is continuous, has variable length, and uses multiple element widths measured edge to edge.
Serial Cmd:
<K474,status,fixed symbol length status,fixed symbol length>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Fixed Symbol Length Status (Code 128) Definition:
When enabled the reader will check the symbol length against the symbol length field. If disabled any length would be considered a valid symbol.
Serial Cmd:
<K474, status,fixed symbol length status,fixed symbol length>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Fixed Symbol Length (Code 128) Usage:
Fixed Symbol Length helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted.
Definition:
It specifies the exact number of characters that the reader will recognize (this does not include start and stop and check digit characters). The reader ignores any symbol not having the specified length.
Serial Cmd:
<K474, status,fixed symbol length status,fixed symbol length>
Default:
10
Options:
1 to 128
Note: Fixed Symbol Length Status must be enabled for Fixed Symbol Length to take effect.
7-10
Quadrus Verifier User’s Manual