The text of a manual · part 20 of 47
Quadrus Verifier User's Manual, pages 191 to 200
Pages 191 to 200 of 470
The original document is the PDF (4.1 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan Verifier
Page 191
Read Cycle
Serial Trigger Usage:
Allows the user to define the trigger character and delimiters that start and stop the read cycle.
Definition:
A serial trigger is considered an online host command and requires the same command format as all host commands: that is, to be entered within angle brackets delimiters < and > or in the case of non-delimited triggers, define individual start and stop characters.
Serial Trigger Character (Delimited) Usage:
Allows the user to define the trigger character that initiates the read cycle.
Definition:
A single ASCII host serial trigger character that initiates the read cycle. A delimited trigger character is one that either starts or ends the read cycle and is enclosed by delimiters such as < and >.
Serial Cmd:
<K201,serial trigger character>
Default:
Spacebar (corresponds to <SP> displayed in the embedded menu)
Options:
Any single ASCII character, including control characters, except NULL (00), an existing host command character, or an on-line protocol character. Control characters entered on the command line are displayed in the menu as mnemonic characters. Note: Serial Data or Serial Data or Edge triggering mode must be enabled for Serial Trigger Character to take effect.
Start and Stop Trigger Characters (Non-Delimited) Usage:
It is useful in applications where different characters are required to start and end a read cycle.
Definition:
A non-delimited trigger character is one that either starts or ends the read cycle and is NOT enclosed by delimiters such as < and >. Both Start and Stop non-delimited characters can be defined and will function according to the trigger event, as follows: When defining Start and Stop trigger characters, the following rules apply: • In External Edge the reader looks only for the start trigger character and ignores any end trigger character that may be defined. • In External Level, the start trigger character begins the read cycle and end trigger character ends it. Note that even after a symbol has been decoded and the symbol data transmitted, the reader remains in External Level trigger read cycle until a Stop character is received. • In Serial Data and Edge trigger mode, command, either a start trigger character or a hardware trigger can start an edge trigger read cycle.
Quadrus Verifier User’s Manual
6-13
Page 192
Serial Trigger
Start Character (Non-Delimited) Definition:
A single ASCII host serial trigger character that starts the read cycle and is not enclosed by delimiters such as < and >.
Serial Cmd:
<K229,start character>
Default:
NULL (00) (disabled)
Options:
Two hex digits representing an ASCII character except <, >, XON, and XOFF.
Stop Character (Non-Delimited) Usage:
It is useful in applications where different characters are required to start and end a read cycle.
Definition:
A single ASCII host serial trigger character that ends the read cycle and is not enclosed by delimiters such as < and >.
Serial Cmd:
<K230,stop character>
Default:
NULL (00) (disabled)
Options:
Two hex digits representing an ASCII character except <, >, XON, and XOFF.
6-14
Quadrus Verifier User’s Manual
Page 193
Read Cycle
End of Read Cycle Definition:
The read cycle is the time during which the reader will attempt to capture and decode a symbol. A read cycle can be ended by a timeout, a new trigger, or by the last frame in a capture sequence or a combination of the above.
End of Read Cycle Mode Note: When operating in Continuous Read or Continuous Read 1 Output, the reader is always in the read cycle. Serial Cmd:
<K220,end of read cycle,read cycle timeout>
Default:
Last Frame
Options:
0 = Timeout 2 = Timeout or New Trigger 4 = Last Frame or New Trigger
1 = New Trigger 3 = Last Frame
Timeout Usage:
Typically used with Serial Data or Edge Trigger and Continuous One Output. Effective in highly controlled applications when the maximum length of time between objects can be predicted. It assures that a read cycle ends before the next bar-coded object appears, giving the system extra time to decode and transmit the data to the host.
Definition:
Timeout ends the read cycle, causing the reader to stop reading symbols and send the symbol data or No Read message when the time set in Timeout elapses (times out), if When to Output is set to End of Read Cycle. If in Continuous Read 1 Output, a timeout initiates a new read cycle and allows the same symbol to be read again. With either External Edge, Serial Data, or Serial Data and Edge enabled, a timeout ends the read cycle and symbol data or a No Read message is sent to the host. With External Level enabled, the read cycle does not end until the falling edge trigger or a timeout occurs. The next read cycle does not begin until the next rising edge trigger.
Quadrus Verifier User’s Manual
6-15
Page 194
End of Read Cycle
New Trigger Usage:
New Trigger is an effective way to end a read cycle when objects move past the reader at irregular intervals (not timing dependent).
Definition:
New Trigger ends the current read cycle and initiates a new one when a new trigger occurs. New Trigger refers only to a “rising edge” trigger. With either External Edge, Serial, or Serial or Edge enabled, an edge or serial trigger ends a read cycle and initiates the next read cycle. In the case of External Level, a falling edge trigger ends the read cycle but the next read cycle does not begin until the occurrence of the next rising edge trigger.
Timeout or New Trigger Usage:
Useful in applications that require an alternative way to end the read cycle. For example, if an assembly line should stop completely or the intervals between objects are highly irregular.
Definition:
Timeout or New Trigger is identical to Timeout except that a timeout or a new trigger (whichever occurs first) ends the read cycle.
Last Frame Usage:
Useful in applications in which the number of captures needed can be defined but the timeout duration varies.
Definition:
Last Frame only applies to Rapid Capture mode.
Last Frame or New Trigger Usage:
Useful in applications in which line speeds are irregular and a new labeled object could appear before the last frame in a Rapid Capture sequence.
Definition:
New Trigger or Last Frame is identical to New Trigger except that a new trigger or last frame (whichever occurs first) ends the read cycle.
Read Cycle Timeout Definition:
Read Cycle Timeout is the duration of the read cycle.
Serial Cmd:
<K220, end of read cycle,read cycle timeout>
Default:
100 (x 10mS)
Options:
0 to 65535
6-16
Quadrus Verifier User’s Manual
Page 195
Read Cycle
Active Camera Usage:
External and Dual options are useful where in applications where the reader cannot be located near the symbol or where both the internal and external cameras are required for products such as circuit boards that might have two symbols in different locations.
Definition:
The Quadrus Verifier has its own internal camera complete with illumination and led sensor circuitry for capturing static or moving symbols at various camera settings including shutter, contrast, etc. The Quadrus Verifier can also accept images from a remote (external) camera using the RS-170 communications protocol.
Serial Cmd:
<K240,active camera>
Default:
Internal
Options:
0 = Internal Camera 1 = External Camera 2 = Dual Camera
Important: The Quadrus Verifier does not control the timing of external cameras. Captures from external cameras are streamed into the Quadrus Verifier. To be sure to receive a complete first capture, allow an extra 33mS delay before the first external capture.
Quadrus Verifier User’s Manual
6-17
Page 196
Capture Mode
Capture Mode Definition:
Capture Mode relates to the way that multiple captures are processed by the Quadrus Verifier.
Serial Cmd:
<K241,capture mode,number of captures,rapid capture mode>
Default:
Rapid Capture
Options:
0 = Rapid Capture
1 = Continuous Capture
Number of Captures Usage:
Used to increase the opportunities for good reads and to “extend” the field of view in dynamic applications.
Definition:
Sets the total number of captures that are processed during a read cycle in the Rapid Capture mode when Switching Mode is set to Number of Captures.
Serial Cmd:
<K241,capture mode,number of captures,rapid capture mode>
Default:
1
Options:
0 to 5
Note: When processing in Rapid Capture mode and Dual Capture mode, the number of captures set for each camera will be limited by the number of captures set here. For example, if Number Of Captures is set to 4, and the individual camera captures are set for 2 and 3 respectively, the last capture will be omitted.
Rapid Capture Mode Usage:
Rapid Capture is useful in fast moving applications in which symbols are only in the field of view a short time or precise timing between captures is relevant or when dual camera mode is enabled. A single capture with Last Frame ending the read cycle is the same as “single shot”.
Definition:
In Rapid Capture, decoding occurs independent of and simultaneous with capturing, thus allowing precise timing or no delay at all between captures. The downside is that the number of captures is limited to 8 so that processing buffers are not overloaded.
Serial Cmd:
<K241,capture mode,number of captures,rapid capture mode>
Default:
Timed Capture
Options:
0 = Timed Capture
6-18
1 = Triggered Capture
Quadrus Verifier User’s Manual
Page 197
Read Cycle
Timed Capture Usage:
Timed Rapid Capture is useful in fast moving applications in which symbols are only in the field of view a short time and precise timing is required.
Definition:
In Timed Rapid Capture, decoding occurs independent of and simultaneous with capturing, thus allowing precise timing (Diagram B) or no delay (Diagram A) at all between captures. Also, consecutive captures are regarded as the same symbol if the output data is the same.
Start of Read Cycle Diagram A
No Time Delay Between Captures n
End of read cycle
Time Before First Capture Time Delay Between Captures End of read cycle
Diagram B
Processing
Rapid Capture Mode, Single Camera
Quadrus Verifier User’s Manual
6-19
Page 198
Capture Mode
Calculating Number of Captures in a Rapid Capture Application 1. First calculate the distance between multiple captures. For multiple captures in Rapid Capture mode, you can calculate the distance between successive symbols by multiplying the line speed by the capture time.
Distance traveled between captures
TRAVEL DISTANCE = Line Speed X 15mS Note: For Dual Camera operations, add 33mS to the overhead time for the first external capture. Example: A symbol moving at 10 ips (inches per second) past a CCD reader travels 0.010”/mS * 15mS = 0.15 inches between captures. 2. Next calculate number of captures. 0.78 inches FOV
Once the travel time is known, you can easily calculate the number of captures you can expect to occur inside a FOV by subtracting the symbol size from the FOV and dividing the result by the travel time. NUMBER OF CAPTURES = (FOV–Symbol Size)/Travel Distance
0.4 inches Symbol Size
Following up on the example from Step 1: 0.78" – 0.4"/0.15" = 2.5 captures If a required number of captures has been determined, you can also work the formula backwards and determine the minimum FOV by: FOV = (Number Of Captures * Travel Time) + Symbol Size
6-20
Quadrus Verifier User’s Manual
Page 199
Read Cycle
Triggered Capture Usage:
Useful in applications where delays between captures can be controlled by triggers rather than preset time delays.
Definition:
When enabled the reader will stay in the read cycle until either the userdefined number of captures has been met or, if configured, a timeout occurs.
Notes on Triggered Capture Mode 1. Trigger Mode <K200> must be set to Edge, Serial, or Serial and Edge. If set to Level, Triggered Capture Mode will operate the same as Timed Rapid Capture mode. 2. For End of Read Cycle settings <K220>: a) If set to Timeout or New Trigger and Timeout and a timeout occurs before number of capture have been met, it will abort the read cycle and disregard the remaining number of triggers. b) If set to New Trigger, the reader remains in the read cycle until the number of captures is satisfied and an additional trigger is received indicating both the end of the read cycle and the start and first capture of the next read cycle. First trigger starts the read cycle
Captures on every trigger
Capture
Processing
Read cycle ends on New Trigger or Timeout, as configured.
Triggered Captures, Typical
Quadrus Verifier User’s Manual
6-21
Page 200
Capture Mode
Continuous Capture Usage:
Continuous Capture is useful in applications with slower line speeds or where symbol spacing may be random or not time-dependent.
Definition:
In Continuous Capture a capture is followed sequentially by processing, as shown below. Since processing is completed before another capture can occur (this usually takes about 30mS), a large number of captures can take place throughout a read cycle. Capture Start of Read Cycle
End of read cycle Processing time
Continuous Capture
6-22
Quadrus Verifier User’s Manual