Page 181
Read Cycle
Read Cycle Serial Commands Multisymbol Trigger Mode/Duration External Trigger State Serial Trigger Character Start Trigger Character Stop Trigger Character End of Read Cycle Active Camera Captures Capture Timing Dual Camera Operations
Store No Read Image
Quadrus Verifier User’s Manual
<K222,number of symbols,multisymbol separator> <K200,trigger mode,trigger filter duration> <K202,external trigger state> <K201,serial trigger character> <K229,start character> <K230,stop character> <K220,end of read cycle,read cycle timeout> <K240,active camera> <K241,capture mode,number of captures,rapid capture mode> <K242,time before 1st capture,time between capture 1 and 2,,,,,,,,time between capture 7 and 8> <K243,switching mode,number of internal camera captures, number of external camera captures,internal camera timeout, external camera timeout> <K244,image storage type,image storage mode>
6-3
Page 182
Read Cycle Setup
Read Cycle Setup Based on your application, setting up read cycle and triggering parameters will involve a series of decisions, as follows: 1. Select the number of symbols to be read in a read cycle (must not exceed Number Of Captures). 2. Decide the trigger type to be used: if serial, the serial character; if external, Level or Edge. 3. Designate how the read cycle should end (Timeout, New Trigger, Last Frame). 4. Calculate the maximum and minimum field of view (FOV). 5. Consider the speed of the transport in inches per second. 6. Select Capture mode, Continuous or Rapid. 7. Select Number Of Captures. 8. Set the Time Before First Capture and Time Between Captures, if any. 9. If using an external camera, select the active camera(s). 10. Decide if you need to alternate between cameras, with Switching Mode. 11. Save settings to the IP Database. Note: Images can be captured at a rate of 60/second. Note: If you need to set up your Quadrus Verifier for use with EZ Trax software, see Setting Up the Verifier for EZ Trax.
6-4
Quadrus Verifier User’s Manual
Page 183
Read Cycle
Multisymbol Usage:
Multisymbol is commonly used in shipping applications where a shipping symbol contains individual symbols for part number, quantity, etc. This feature allows one trigger to pick up all the symbols.
Definition:
Multisymbol allows the user to define up to 6 symbols that can be read in a single read cycle.
Conditions:
The following conditions apply: 1. Each symbol must be different to be read. 2. The maximum number of characters in a read cycle is 32,520 for all symbols. 3. The maximum number of characters the reader can transmit is calculated by: Preamble + maximum number of symbols * (aux id + symbology id + maximum symbol length + ((number of insertion cells x cell length)+ separator) + postamble + LRC = 37,425. 4. All No Read messages are posted at the end of the data string. 5. If more than one symbol is within the field of view at the same time, symbol data may not be displayed in the order of appearance. 6. If Matchcode Type is set to Sequential or if Trigger is set to Continuous Read 1 Output, Number of Symbols will default to 1 (if set to any number greater than 1).
Number of Symbols Definition:
Number of Symbols is the number of different symbols that can be read in a single read cycle.
Serial Cmd:
<K222,number of symbols,multisymbol separator>
Default:
1
Options:
1 to 6
Quadrus Verifier User’s Manual
6-5
Page 184
Multisymbol
Multisymbol Separator Usage:
Used to delimit or separate data fields with a user defined character.
Definition:
Any valid ASCII character, inserted between each symbol read when Multisymbol is set to any number greater than 1.
Serial Cmd:
<K222,number of symbols,multisymbol separator> Note: If Multisymbol Separator has been changed to any other character than the default comma and you wish to re-define the separator as a comma, use ESP or the embedded menus.
Default:
, (comma)
Options:
Any available ASCII character, except < or > (if used as delimiters).
Note: If a NULL is entered for the multisymbol separator, the multisymbol separator output will be disabled. Note: If No Read messages are disabled and there are No Reads occurring, separators will only be inserted between symbol data outputs.
6-6
Quadrus Verifier User’s Manual
Page 185
Read Cycle
Trigger Definition:
The type of trigger event that will initiate the read cycle.
Note: When perofming calibration or read rate testing, the current trigger setting will be disregarded.
Trigger Mode Serial Cmd:
<K200,trigger mode,trigger filter duration>
Default:
External Edge
Options:
0 = Continuous Read 2 = External Level 4 = Serial Data
1 = Continuous Read 1 Output 3 = External Edge 5 = Serial Data and External Edge
Continuous Read Usage:
Continuous Read is useful in testing symbol readability or reader functions. It is not recommended for normal operations.
Definition:
In Continuous Read, trigger input options are disabled, the reader is always in the read cycle, and it will attempt to decode and transmit every capture. When To Output and No Read options have no affect on Continuous Read.
Serial Cmd:
<K200,0>
Quadrus Verifier User’s Manual
6-7
Page 186
Trigger
Continuous Read 1 Output Usage:
Continuous Read 1 Output can be useful in applications where it is not feasible to use a trigger and all succeeding symbols contain different information. It is also effective in applications where the objects are presented by hand.
Definition:
In Continuous Read 1 Output the reader self-triggers whenever it decodes a new symbol or a timeout occurs. If End Of Read Cycle is set to Timeout and the symbol doesn't change, the output is repeated at the end of each timeout period. For example, if Timeout is set to one second, the reader sends the symbol data immediately and repeats the output at intervals of one second for as long as the symbol continues to be captured. If End Of Read Cycle is set to New Trigger, the reader will send the current symbol data immediately, but send it only once. A new symbol appearing in the reader’s range will be read and sent immediately provided it is not identical to the previous symbol read.
Serial Cmd:
<K200,1>
Caution: In automated environments, Continuous Read 1 Output is not recommended because there is no one to verify that a symbol was missed. Note: If Trigger Mode is set to Continuous Read 1 Output, Number of Symbols will default to 1 (if set to any number greater than 1).
6-8
Quadrus Verifier User’s Manual
Page 187
Read Cycle
External Level Usage:
This mode is effective in an application where the speeds of the conveying apparatus are variable and the time the reader spends reading each object is not predictable. It also allows the user to determine if a No Read has occurred.
Definition:
External Trigger Level allows the read cycle (active state) to begin when a trigger (change of state) from an external sensing device is received. The read cycle endures until the object moves out of the sensor range and the active trigger state changes again.
Serial Cmd:
<K200,2>
Important: Level and Edge apply to the active logic state (Active Off (IOFF) or Active On (ION)) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object.
Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle.
Associated waveforms assume External Trigger State is set to Active On.
End Read Cycle: The same object, moving out of the detector beam, causes another change in the trigger state, ending the read cycle.
Trigger Level
Quadrus Verifier User’s Manual
6-9
Page 188
Trigger
External Edge Usage:
This mode is highly recommended in any application where conveying speed is constant or if spacing, object size, or timeouts are consistent.
Definition:
External Trigger Edge, as with Level, allows the read cycle (active state) to begin when a trigger (change of state) from an external sensing device is received. However, the passing of an object out of sensor range does not end the read cycle. The read cycle ends with a good read output or, depending on End of Read Cycle setting, a timeout or new trigger occurs.
Serial Cmd:
<K200,3>
Important: Level and Edge apply to the active logic state (Active Off (IOFF) or Active On (ION)) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object.
Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle.
Initiate Second Read Cycle: Object # 2, moving in front of the detector beam, causes another change in the trigger state. This signal initiates a new read cycle and ends the previous read cycle unless Timeout is enabled and a good read or timeout has not occurred.
Associated waveforms assume External Trigger State is set to Active On.
Trigger Edge
6-10
Quadrus Verifier User’s Manual
Page 189
Read Cycle
Serial Data Usage:
Serial Data is effective in a highly controlled environment where the host knows precisely when the object is in the field of view. It is also useful in determining if a No Read has occurred.
Definition:
In Serial Data, the reader accepts an ASCII character from the host or controlling device as a trigger to start a read cycle. A Serial Data trigger behaves the same as an External Edge trigger. Serial commands are entered inside corner brackets, such as <t>.
Serial Cmd:
<K200,4>
Note: In Serial Data, sending a non-delimited start serial character will start a read cycle; however a non-delimited stop serial character has no effect.
Serial Data or External Edge Usage:
Serial Data or External Edge is seldom used but can be useful in an application that primarily uses an external sensing device but occasionally needs to be manually triggered. An auxiliary terminal can be connected to the aux port so the user can send the serial trigger character through the reader to the host.
Definition:
In this mode the reader accepts either a serial ASCII character or an external trigger pulse to start the read cycle.
Serial Cmd:
<K200,5>
Note: In Serial Data or External Edge, sending a non-delimited start serial character will start a read cycle; however a non-delimited stop serial character has no effect.
Quadrus Verifier User’s Manual
6-11
Page 190
Trigger
Trigger Filter Duration Usage:
Trigger Filter Duration can be used to help the unit trigger more consistently with an unstable external trigger.
Definition:
Trigger Filter Duration can prevent trigger bounce from falsely triggering the reader by limiting the time during which trigger pulses can be received.
Serial Cmd:
<K200,trigger mode,trigger filter duration>
Default:
313 (x 32µs = 10.24mS)
Options:
0 to 65535 (corresponding to 0 to 2.097seconds in 32µs steps)
If the unit is in External Edge trigger mode, then Trigger Filter Duration specifies the time following an edge that the reader will not consider another edge as a valid trigger. If the unit is in External Level mode, then Trigger Filter Duration specifies the time following the active edge that the trigger must be sampled once again as active before considering it a valid level trigger.
External Trigger State Usage:
Allows users to select the trigger state that will operate with their systems. (If using an object detector, use Active Off.)
Definition:
When enabled for Active On (ION) the triggering device imposes a current on the optoisolator to activate the read cycle. When enabled for Active Off (IOFF) the triggering device interrupts the current to the optoisolator to activate the read cycle.
Serial Cmd:
<K202,external trigger state>
Default:
Active On
Options:
0 = Active Off
1 = Active On
Note: External Level, External Edge, or Serial Data or Edge triggering mode must be enabled for External Trigger to take effect.
6-12
Quadrus Verifier User’s Manual