The text of a manual · part 12 of 47
Quadrus Verifier User's Manual, pages 111 to 120
Pages 111 to 120 of 470
The original document is the PDF (4.1 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan Verifier
Page 111
Verification
AS9132 Verification by Serial Command Serial Cmd:
<V3>
AS9132 Verification Process • Place the candidate symbol as close to the center of the Verifier’s field of view as possible. • When the candidate symbol is in position, initiate the AS9132 Verification command <V3>. • The AS9132 Verification Report will appear, detailing the symbol’s adherence to AS9132 requirements.
AS9132 Verification Report The AS9132 results show data concerning marking method, element shape, quiet zone, contrast, cell fill, cell size, dot ovality, dot shape, dot center offset, dot spacing, angle of distortion, symbol type, matrix size, element size, and pixels per element. Symbol assessment is on a pass/fail basis.
Quadrus Verifier User’s Manual
3-67
Page 112
AIM DPM Verification by Serial Command
AIM DPM Verification by Serial Command Reflectance Calibration Definition:
The Reflectance Calibration command initiates a calibration process with the minimum and maximum reflectance values that are already configured in the Verifier. Note: If the minimum and maximum reflectance values configured in the Verifier correspond with a different calibration symbol, the results after calibration may be incorrect.
Serial Cmd:
<@AIMDPM,R-min,R-min>
• Place the calibration symbol provided in the approximate center of the Verifier’s field of view before entering the Reflectance Calibration command.
Place the Calibration Symbol as close to the center of the Verifier’s field of view as possible before initiating the Reflectance Calibration process.
IMPORTANT: Keep this card in a safe place! It must be saved and kept in good condition, as it is the Verifier’s most critical setup tool.
You can use the Terminal in ESP to send serial commands. Commands can either be entered in the Send text field, or directly in the terminal.
Important: After the Verifier is calibrated, you must allow 15 minutes of warmup time in Live Video Mode, <K760,2>, before starting a verification process. The LEDs must reach a steady output state for verification results to be valid.
3-68
Quadrus Verifier User’s Manual
Page 113
Verification
AIM DPM Verification Process Serial Cmd: <V4> • Once the reflectance calibration process is complete, place the candidate symbol as close to the center of the Verifier’s field of view as possible. • When the candidate symbol is in position, initiate the AIM DPM Verification command <V4>. • The AIM DPM Verification Report will then appear, detailing the symbol’s adherence to AIM DPM guidelines.
AIM DPM Verification Report The AIM DPM results show data concerning the reference decode algorithm, cell contrast, fixed pattern damage, axial and grid non-uniformity, cell modulation, unused error correction, reflectance, print growth, symbol type, symbol size, element size, and pixels per element. All but the last five parameters are given a numeric and alphabetical grade.
Quadrus Verifier User’s Manual
3-69
Page 114
Verification by ESP
Verification by ESP The Quadrus Verifier’s verification functionality can be controlled using ESP’s graphic interfaces and tree controls.
Report Tab The Report tab allows the user to initiate ISO/IEC 15415 Verification, AS9132 Verification, and AIM DPM Verification. Verification reports appear in the viewing area underneath the verification buttons. To save the report as a PDF file, an HTML file, a CSV file, or an RTF file, click the Save Report button.
The Verification user interface is divided into three sections: Report, Settings, and Preferences.
Locate and Calibrate buttons.
Use this dropdown menu to set LED illumination characteristics. See Illumination Source for more details.
The Verifier’s field of view is shown here. When a verification routine is completed, the initial report is also shown in this viewing area.
3-70
Quadrus Verifier User’s Manual
Page 115
Verification
Settings Tab The Settings tab allows users to view the candidate symbol in the Verifier’s field of view using live video; to calibrate the Verifier for ISO/IEC 15415, AS9132, and AIM DPM verification; to set the desired image capture number; and, if using a Triggered mode, to determine the method of triggering. Note: The Live (real time) setting is intended for applications in which an external video monitor is used. The video view in ESP is frame-by-frame, and is not “real-time” in the same sense as an external monitor.
To open the entire contents of the tree control, hold down the Alt key and single-click the + next to the first-level heading Symbol Verification. To close the entire contents of the tree control, hold down the Alt key and single-click the - .
Quadrus Verifier User’s Manual
3-71
Page 116
Verification by ESP
Preferences Tab The Preferences tab allows the user to set defaults that are automatically activated when verification processes are completed. For this reason, the user is advised to set preferences before performing verification.
This information is appended at the bottom of the verification report.
Choose the default report type to be generated when you click Save Report on the Report tab. Check Autofit Content to ensure that saved PDF report content will fit on a single page.
Typically you will be asked where you wish to store your verification report. If you do not want to be prompted to choose a file path, click Store Report without file path prompt.
Click the Include Image check box to output the captured symbol image with your saved report as a Bitmap or JPEG image file.
Returns all settings to default.
3-72
Returns all settings to most recent.
Quadrus Verifier User’s Manual
Page 117
Verification
ISO/IEC 15415 Single Capture Verification Single Capture Verification Process Reflectance Calibration The Quadrus Verifier comes with a calibration symbol on a card like the one shown below. A maximum and minimum reflectance value will be printed at the bottom of the card. These are the values you will enter in the Minimum and Maximum spin boxes in the Reflectance Calibration dialog. Place the Calibration Symbol as close to the center of the Verifier’s field of view as possible before initiating the Reflectance Calibration process.
IMPORTANT: Keep this card in a safe place! It is the Verifier’s most critical setup tool.
When the R-min and R-min values are entered, click Calibrate Reflectance.
Once reflectance calibration is complete, Single Capture Verification can be performed and a report generated.
Quadrus Verifier User’s Manual
3-73
Page 118
Verification by ESP
Single Capture Verification Report The ISO/IEC 15415 Single Capture results show data concerning the reference decode algorithm, symbol contrast, fixed pattern damage, axial and grid non-uniformity, modulation, unused error correction, print growth, symbol type, symbol size, element size, and pixels per element. All but the last five parameters are given a numeric and alphabetical grade.
3-74
Quadrus Verifier User’s Manual
Page 119
Verification
Save Single Capture Verification Report There are several possible ways to save your verification report, depending on what works best for your application.
Save Single Capture Verification Report as a PDF If you have already chosen PDF in the Preferences dialog, the report will be saved as a PDF file.
Verification results and grades.
Symbol data. Data for non-graded parameters.
Captured symbol image.
Reference data: Verifier firmware version, Verifier serial number, report creation date, ESP version, company name, and operator name.
Note: See Preferences Tab for information about how to “autofit” PDF reports to a single page. Quadrus Verifier User’s Manual
3-75
Page 120
Verification by ESP
Save Single Capture Verification Report as an HTML File
If you have already chosen HTML format in the Preferences dialog, the report will be saved as an HTML file.
Verification results and grades.
Symbol data.
Data for non-graded parameters.
Captured symbol image. Reference data: Verifier firmware version, Verifier serial number, report creation date, ESP version, company name, and operator name.
3-76
Quadrus Verifier User’s Manual