Page 121

Verification

Save Single Capture Verification Report as a CSV File

If you have already chosen CSV format in the Preferences dialog, the report will be saved as a CSV file.

CSV output is in a spreadsheet format, with all parameters listed in a horizontal orientation. Columns A - I: • Symbol Data • Reference Decode Grade • Symbol Contrast • Fixed Pattern Damage • Axial Non-Uniformity

Columns J- P: • Grid Non-Uniformity • Modulation • Unused ECC • Final Grade

Columns Q - W: • Print Growth • Symbol Type • Pixels Per Element • Aperture • Wavelength • Angle

Columns X - AC: • Firmware Version • Verifier Serial Number • Date/Time • ESP Version • Company • Operator

Quadrus Verifier User’s Manual

3-77

Page 122

Verification by ESP

Save Single Capture Verification Report as an RTF

If you have already chosen RTF in the Preferences dialog, the report will be saved as an RTF file.

Verification results and grades.

Non-graded parameters and symbol data.

Note: Image captures are output simultaneously with the RTF. The format (JPG or BMP) is determined in the Preferences dialog.

Reference data: Verifier firmware version, Verifier serial number, report creation date, ESP version, company name, and operator name.

3-78

Quadrus Verifier User’s Manual

Page 123

Verification

ISO/IEC 15415 Multi-Capture Verification Multi-Capture Verification Process Reflectance Calibration The Quadrus Verifier comes with a calibration symbol on a card like the one shown below. A maximum and minimum reflectance value will be printed at the bottom of the card. These are the values you will enter in the Minimum and Maximum spin boxes in the Reflectance Calibration dialog. Place the Calibration Symbol as close to the center of the Verifier’s field of view as possible before initiating the Reflectance Calibration process.

IMPORTANT: Keep this card in a safe place! It is the Verifier’s most critical setup tool.

When the R-min and R-min values are entered, click Calibrate Reflectance.

Once reflectance calibration is complete, Multi-Capture Verification can be performed and a report generated.

Quadrus Verifier User’s Manual

3-79

Page 124

Verification by ESP

Multi-Capture Sequence • Once the reflectance calibration process is complete, place the candidate symbol as close to the center of the verifier’s field of view as possible. Important: Multi-Capture Verification requires five captures at 72° intervals throughout a 360° rotation. When the symbol is placed in the Verifier’s field of view, its position should be considered its default position -- 0°. ISO/IEC 15415 requires that the symbol’s orientation for the first capture be 45°. This means 45° from the default symbol position, 0°. • When you see the first rotation prompt, trigger the first capture. The prompts that follow will guide you through the sequence.

Step 1: 45° Rotation Default Position: 0°

45°

Default Position: 0°

117°

Default Position: 0°

189°

Default Position: 0°

261°

Default Position: 0°

333°

Step 2: 117° Rotation

Step 3: 189° Rotation

Step 4: 261° Rotation

Step 1: 333° Rotation

3-80

Quadrus Verifier User’s Manual

Page 125

Verification

Multi-Capture Verification Report After the final rotation and trigger, the Verifier will output the ISO/IEC 15415 Multi-Capture Verification Report, shown below. The ISO/IEC Multi-Capture parameters are the same as those for ISO/IEC Single Capture, but Multi-Capture results are determined only after the symbol is read at five 72° intervals throughout a full 360° rotation. The overall symbol grade is based on an arithmetic mean of the results from the five reads.

Quadrus Verifier User’s Manual

3-81

Page 126

Verification by ESP

Save Multi-Capture Verification Report

Save Multi-Capture Verification Report as a PDF If you have already chosen PDF in the Preferences dialog, the report will be saved as a PDF file.

Verification results and mean values. Symbol data.

Data for non-graded parameters.

Captured symbol image.

Reference data: Verifier firmware version, Verifier serial number, report creation date, ESP version, company name, and operator name.

Note: See Preferences Tab for information about how to “autofit” PDF reports to a single page.

3-82

Quadrus Verifier User’s Manual

Page 127

Verification

Save Multi-Capture Verification Report as an HTML File

If you have already chosen HTML format in the Preferences dialog, the report will be saved as an HTML file.

Verification results and mean values.

Symbol data.

Data for non-graded parameters.

Captured symbol image.

Quadrus Verifier User’s Manual

Reference data: Verifier firmware version, Verifier serial number, report creation date, ESP version, company name, and operator name.

3-83

Page 128

Verification by ESP

Save Multi-Capture Verification Report as a CSV File CSV output is in a spreadsheet format, with all parameters listed in a horizontal orientation.

If you have already chosen CSV format in the Preferences dialog, the report will be saved as a CSV file. Columns A - M: • Symbol Data • Reference Decode Grade • Symbol Contrast

Columns N - Y: • Fixed Pattern Damage • Axial Non-Uniformity

Columns Z - AK: • Grid Non-Uniformity • Modulation

Columns AL - AW: • Unused ECC • Overall Grade • Final Grade

Columns AX - BC: • Print Growth • Symbol Type • Symbol Size • Pixels Per Element • Aperture • Wavelength Columns BD - BJ: • Angle • Firmware • Serial Number • Date/Time • ESP Version • Company • Operator

3-84

Quadrus Verifier User’s Manual

Page 129

Verification

Save Multi-Capture Verification Report as an RTF

If you have already chosen RTF in the Preferences dialog, the report will be saved as an RTF file.

Verification results, grades, and mean values.

Non-graded parameters and symbol data.

Note: Image captures are output simultaneously with the RTF. The format (JPG or BMP) is determined in the Preferences dialog.

Reference data: Verifier firmware version, Verifier serial number, report creation date, ESP version, company name, and operator name.

Quadrus Verifier User’s Manual

3-85

Page 130

Verification by ESP

AS9132 Verification AS9132 Calibration Select AS9132A Calibration on the Reflectance Calibration dialog and click Calibrate AS9132A.

Once calibration is complete, AS9132 Verification can be performed and a report generated.

AS9132 Verification Process • Be sure that the candidate symbol as close to the center of the verifier's field of view as possible. • When the symbol is in position, click the AS9132A Verification button.

• The AS9132A Verification Report will appear, detailing the symbol’s adherence to AS9132 requirements.

3-86