Page 321
Appendices
QX-1 Interface Device Mechanical Height: .75" (19.1 mm) Width: 2.50" (83.5 mm) Depth: 3.15" (80 mm) Weight: 7 oz. (200 g)
QX-1 Dimensions
QX Hawk Industrial Imager User Manual
A-5
Page 322
General Specifications
Read Ranges (Standard QX Hawk) Note: Ranges correspond to specific symbol element sizes, and range increases as element size increases.
CMOS Modular Zoom Optics: Inches (mm) Narrow Bar Width 1D
2D
0.0033 (0.08)
0.005 (0.13)
12° 0.0075 (0.19)
Read Range (using autofocus)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
3.4 to 6 (86 to 152)
0.9 (23)
1.42 (36)
0.2 (5)
0.4 (10)
0.010 (0.25)
3.3 to 12.2 (83 to 310)
0.9 (23)
2.62 (66)
0.4 (10)
1.5 (38)
0.0150 (0.38)
0.020 (0.51)
3.3 to 13 (82 to 330)
0.9 (23)
2.77 (70)
0.5 (13)
3.5 (89)
0.0350 (0.89)
0.050 (1.27)
4 to 16 (101 to 406)
1.03 (26)
3.34 (85)
0.7 (18)
6.5 (165)
Narrow Bar Width 1D
2D
0.0033 (0.08)
0.005 (0.13)
15° 0.0075 (0.19)
Read Range (using autofocus)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
1.9 to 5 (48 to 127)
0.75 (19)
1.53 (39)
0.3 (6)
0.4 (10)
0.010 (0.25)
1.8 to 8 (46 to 203)
0.72 (18)
2.28 (58)
0.4 (10)
1 (25)
0.0150 (0.38)
0.020 (0.51)
1.75 to 9 (44 to 229)
0.72 (18)
2.52 (64)
0.5 (13)
3 (76)
0.0350 (0.89)
0.050 (1.27)
3 to 10.8 (76 to 274)
1.03 (26)
2.96 (75)
0.6 (15)
4 (101)
Narrow Bar Width 1D
2D
0.0033 (0.08)
0.005 (0.13)
Read Range (using autofocus) 1 to 3 (25 to 76)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
0.83 (21)
1.8 (46)
0.4 (10)
0.8 (20)
30° 0.0075 (0.19)
0.010 (0.25)
1 to 6.5 (25 to 165)
0.83 (21)
3.5 (89)
0.8 (20)
1.8 (46)
0.0150 (0.38)
0.020 (0.51)
1 to 16 (25 to 406)
0.83 (21)
8.3 (211)
2 (51)
8.5 (216)
0.0350 (0.89)
0.050 (1.27)
2 to 32 (51 to 813)
1.3 (33)
16.4 (417)
3 (76)
21.5 (546)
Narrow Bar Width
45°
A-6
1D
2D
0.0075 (0.19)
0.010 (0.25)
0.015 (0.38) 0.035 (0.89)
Read Range (using autofocus)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
1 to 5 (25 to 127)
1.3 (33)
4.1 (104)
1.5 (38)
Outside Edge 2 (51)
0.020 (0.51)
1 to 9.5 (25 to 241)
1.3 (33)
7.5 (191)
2 (51)
6 (152)
0.050 (1.27)
1 to 23.5 (25 to 597)
1.3 (33)
18.3 (465)
3.8 (97)
19 (483)
QX Hawk Industrial Imager User Manual
Page 323
Appendices
CCD Modular Zoom Optics: Inches (mm) Narrow Bar Width 1D
2D
0.002 (0.05)
0.0033 (0.08)
12° 0.005 (0.13) 0.010 (0.25) 0.020 (0.51)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
3.5 to 7 (87 to 178)
0.9 (23)
1.62 (41)
0.1 (3)
0.3 (8)
0.0075 (0.19)
3.4 to 12 (86 to 304)
0.9 (23)
2.62 (66)
0.2 (5)
1 (25)
0.0150 (0.38)
3.3 to 13 (83 to 330)
0.9 (23)
2.77 (70)
0.4 (10)
2.5 (64)
0.0300 (0.76)
4 to 14 (101 to 355)
1.03 (26)
2.96 (75)
0.6 (15)
5 (127)
Narrow Bar Width 1D
2D
0.002 (0.05)
0.0033 (0.08)
15° 0.005 (0.13) 0.010 (0.25) 0.020 (0.51)
Read Range (using autofocus)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
2 to 5 (51 to 127)
0.78 (20)
1.53 (39)
0.2 (5)
0.4 (10)
0.0075 (0.19)
1.9 to 8.3 (48 to 210)
0.75 (19)
2.35 (60)
0.3 (8)
1 (25)
0.0150 (0.38)
1.9 to 9.5 (47 to 241)
0.75 (19)
2.65 (67)
0.5 (13)
3 (76)
0.0300 (0.76)
3 to 10.3 (76 to 261)
1.03 (26)
2.84 (72)
0.6 (15)
4.5 (114)
Narrow Bar Width 1D
2D
0.002 (0.05)
0.0033 (0.08)
30° 0.005 (0.13) 0.010 (0.25) 0.020 (0.51)
Read Range (using autofocus)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
1 to 3 (25 to 76)
0.83 (21)
1.8 (46)
0.4 (10)
0.8 (20)
0.0075 (0.19)
1 to 6.5 (25 to 165)
0.83 (21)
3.5 (89)
0.8 (20)
1.8 (46)
0.0150 (0.38)
1 to 16 (25 to 406)
0.83 (21)
8.3 (211)
2 (51)
8.5 (216)
0.0300 (0.76)
2 to 32 (51 to 813)
1.3 (33)
16.4 (417)
3 (76)
21.5 (546)
Narrow Bar Width
45°
Read Range (using autofocus)
1D
2D
0.005 (0.13)
0.0075 (0.19)
0.010 (0.25) 0.020 (0.51)
Read Range (using autofocus)
Field of View
Depth of Field
Inside Edge
Outside Edge
Inside Edge
Outside Edge
1 to 5 (25 to 127)
1.3 (33)
4.1 (104)
1.5 (38)
2 (51)
0.0150 (0.38)
1 to 9.5 (25 to 241)
1.3 (33)
7.5 (191)
2 (51)
6 (152)
0.0300 (0.76)
1 to 23.5 (25 to 597)
1.3 (33)
18.3 (465)
3.8 (97)
19 (483)
QX Hawk Industrial Imager User Manual
A-7
Page 324
General Specifications
Read Ranges (cont.)
FIS Options FIS-6801-0000G FIS-6801-0010G FIS-6801-0500G FIS-6801-0510G FIS-6801-0200G FIS-6801-0210G FIS-6801-0300G FIS-6801-0310G FIS-6801-0400G FIS-6801-0410G FIS-6801-1000G FIS-6801-1010G FIS-6801-1500G FIS-6801-1510G FIS-6801-1200G FIS-6801-1210G FIS-6801-1300G FIS-6801-1310G FIS-6801-1400G FIS-6801-1410G
A-8
QX Hawk, Industrial Imager, C-Mount Optic, Serial, CCD, SXGA QX Hawk, Industrial Imager, C-Mount Optic, Serial, CMOS, WVGA QX Hawk, Industrial Imager, 15 Degree Optic, Serial, CCD, SXGA QX Hawk, Industrial Imager, 15 Degree Optic, Serial, CMOS, WVGA QX Hawk, Industrial Imager, 30 Degree Optic, Serial, CCD, SXGA QX Hawk, Industrial Imager, 30 Degree Optic, Serial, CMOS, WVGA QX Hawk, Industrial Imager, 45 Degree Optic, Serial, CCD, SXGA QX Hawk, Industrial Imager, 45 Degree Optic, Serial, CMOS, WVGA QX Hawk, Industrial Imager, 12 Degree Optic, Serial, CCD, SXGA QX Hawk, Industrial Imager, 12 Degree Optic, Serial, CMOS, WVGA QX Hawk, Industrial Imager, C-Mount Optic, Ethernet, CCD, SXGA QX Hawk, Industrial Imager, C-Mount Optic, Ethernet, CMOS, WVGA QX Hawk, Industrial Imager, 15 Degree Optic, Ethernet, CCD, SXGA QX Hawk, Industrial Imager, 15 Degree Optic, Ethernet, CMOS, WVGA QX Hawk, Industrial Imager, 30 Degree Optic, Ethernet, CCD, SXGA QX Hawk, Industrial Imager, 30 Degree Optic, Ethernet, CMOS, WVGA QX Hawk, Industrial Imager, 45 Degree Optic, Ethernet, CCD, SXGA QX Hawk, Industrial Imager, 45 Degree Optic, Ethernet, CMOS, WVGA QX Hawk, Industrial Imager, 12 Degree Optic, Ethernet, CCD, SXGA QX Hawk, Industrial Imager, 12 Degree Optic, Ethernet, CMOS, WVGA
QX Hawk Industrial Imager User Manual
Page 325
Appendices
Accessories Power Supplies 97-000012-01 QX Power Supply, 100-240VAC, +24VDC, M12 12-pin Socket 97-000003-02 QX Power Supply, 24V M12 12-pin Plug, 1.3 m Cable, USA/Euro Cables, Communication Devices, Accessories QX Cordset, Common, M12 12-pin Socket (Screw-On) to M12 12-pin Plug (Screw-On), 3 m 61-000148-02 61-000162-02 QX Cordset, Common, M12 12-pin Socket (Screw-On) to M12 12-pin Plug (Screw-On), 1 m 61-000153-02 QX Cordset, Host, Serial, M12 12-pin Socket (Screw-On) to DB9 Socket, 1 m 61-000164-02 QX Cordset, Host, Serial, M12 12-pin Socket (Screw-On) to DB9 Socket, 3 m 61-000152-02 QX Cordset, Host, Serial, M12 12-pin Plug (Screw-On) to DB9 Socket, 1 m 61-000165-02 QX Cordset, Host, Serial, M12 12-pin Plug (Screw-On) to DB9 Socket, 3 m 61-000163-01 QX Cordset, Host, Ethernet, M12 8-pin Plug (Ultra-Lock) to RJ45, 3 m 61-000163-02 QX Cordset, Host, Ethernet, M12 8-pin Plug (Screw-On) to RJ45, 3 m 61-000160-01 QX Cordset, Host, Ethernet, M12 8-pin Plug (Ultra-Lock) to RJ45, 1 m 61-000160-02 QX Cordset, Host, Ethernet, M12 8-pin Plug (Screw-On) to RJ45, 1 m QX Cordset, M12 12-pin Plug and M12 12-pin Socket to MS-Connect 210 (RS-232), 2 m 61-000158-03 61-000158-04 QX Cordset, M12 12-pin Plug and M12 12-pin Socket to MS-Connect 210 (RS-422), 2 m 61-000166-02 QX Cordset, M12 12-pin Plug (Screw-On) to Flying Leads, 3 m 61-000167-02 QX Cordset, M12 12-pin Socket (Screw-On) to Flying Leads, 3 m 61-000207-01 QX Cordset, Smart Series Light to QX-1, Continuous ON/OFF 61-000204-01 QX Cordset, Smart Series Light to QX-1, Continuous Power 61-000218-01 QX Cordset, Smart Series Light to QX-1, Strobe, NPN 98-000103-02 QX-1 Interface Device FIS-0210-0001G MS-Connect 210, Connectivity Box with Display FIS-0210-0002G MS-Connect 210, Connectivity Box FIS-0210-0003G MS-Connect 210, Connectivity Box with Display and Ethernet FIS-0210-0004G MS-Connect 210, Connectivity Box with Ethernet 98-000143-01 Kit, Adapter Plate, QX Hawk 98-000144-01 Kit, Right Angle Mirror, QX Hawk 98-000146-01 Kit, Window Replacement, QX Hawk 98-000147-04 Kit, 12° Lens, QX Hawk 98-000147-01 Kit, 15° Lens, QX Hawk 98-000147-02 Kit, 30° Lens, QX Hawk 98-000147-03 Kit, 45° Lens, QX Hawk 98-000205-01 Kit, Glass Window with IR Coating, QX Hawk 98-000206-01 Kit, Glass Window, QX Hawk 98-000206-02 Kit, Lexan Window with Metal Bezel, QX Hawk 98-000149-01 Demo Kit, QX Hawk 98-000262-01 Kit, White LED Illumination, QX Hawk 98-000262-02 Kit, Blue LED Illumination, QX Hawk 98-000013-04 Relay Module, 120VAC, 3 Amp Output, Series 70, Type SM for MS-Connect 210 98-000013-05 Relay Module, 240VAC, 3 Amp Output, Series 70, Type SM for MS-Connect 210 98-000013-06 Relay Module, 24VDC, 3 Amp Output, Series 70, Type SM for MS-Connect 210 98-92800471 Lens Extension Tube, 5 mm (C-CS Mount Adapter)
QX Hawk Industrial Imager User Manual
A-9
Page 326
General Specifications 98-CO206 98-92800571 98-000257-01 98-000258-01 98-000259-01 98-92800575 98-92800576 98-92800577 98-92800311 98-92800371 98-92800372 98-000218-01 98-000226-01 20-610024-01 98-000037-01 98-000054-01 98-000016-01 99-000056-01 99-000058-01 99-000060-01 99-000061-01 99-000050-01 99-000052-01 99-000049-01 99-000051-01 98-000148-01 Object Detectors 99-000020-01 99-000020-02 Documentation 37-000010-01
A-10
Lens Extension Tube Set 0.5, 1, 5, 10, 20, 40 mm Lens 8mm F/1.4-16, Front Thread 25.5 mm P 0.5 mm, 2/3" C-Mount Lens 12.5mm F/1.4-close, Front Thread 40.5 mm P 0.5 mm, 1" C-Mount Lens 16mm F/1.4-16, Front Thread 27 mm P 0.5 mm, 2/3" C-Mount Lens 25mm F/1.4-16, Front Thread 27 mm P 0.5 mm, 2/3" C-Mount Lens 35mm F/2.1-22, Front Thread 25.5 mm P 0.5 mm, 2/3" C-Mount Lens 50mm F/2.8-22, Front Thread 25.5 mm P 0.5 mm, 2/3" C-Mount Lens 75mm F/3.9-32, Front Thread 25.5 mm P 0.5 mm, 2/3" C-Mount Lens, Skylight UV Filter, 25.5 mm Lens, Polarizing Filter, 25.5 mm, P0.5 Lens, Polarizing Filter, 27 mm, P0.5 Lens Protection Housing, Standard Length (up to 48 mm) Lens Protection Housing, Long (up to 72 mm) Trigger Connector, 4-pin Plug (Screw Terminal; Field-Wireable; Self-Wiring) Kit, Mounting Arm Extension, All Readers, 6 inch Kit, Mounting Stand Base Plate Kit, Mounting Arm/Adapter, QX Hawk, 6 inch Bracket, Smart Series DOAL 50 mm to QX Hawk Bracket, Smart Series DOAL 75 mm to QX Hawk Bracket, Smart Series DOAL 100 mm to QX Hawk Bracket, Smart Series DOAL 50 mm / 70 mm / 100 mm to C-Mount QX Hawk Bracket, Smart Series Ring 100 mm to QX Hawk Bracket, Smart Series Ring 60 mm / 70 mm to QX Hawk Bracket, Smart Series Ring 100 mm to C-Mount QX Hawk Bracket, Smart Series Ring 60 mm / 70 mm to C-Mount QX Hawk Kit, L-Bracket, QX Series QX Photo Sensor, M12 4-pin Plug, NPN, Dark Off, 2 m QX Photo Sensor, M12 4-pin Plug, NPN, Dark On, 2 m Microscan Tools Drive: Software, Documentation, Links to Microscan website
QX Hawk Industrial Imager User Manual
Page 327
Appendices
Appendix B, Electrical Specifications • CMOS Power Requirement: 5-28VDC, 200 mV p-p max ripple, 135 mA at 24VDC (typ.) • CCD Power Requirement: 5-28VDC, 200 mV p-p max ripple, 170 mA at 24VDC (typ.)
QX Hawk Connectors
B
A
A is a serial M12 12-pin plug. It is the same for all models of the QX Hawk. B is a serial M12 12-pin socket for serial models and an Ethernet M12 8-pin socket for Ethernet models. Output 2 Default
New Master
Power
Output 1
RS-232 (Host) TxD
Output 3
Trigger Ground RS-232 (Host) RxD
Input Common Output Common
A (All Models) M12 12-pin Plug Pin 1 2 3 4 5 6 7 8 9 10 11 12
Function Trigger Power Default New Master Output 1 Output 3 Ground Input Common RS-232 (Host) RxD RS-232 (Host) TxD Output 2 Output Common
Wire Color White Brown Green Yellow Gray Pink Blue Red Black Violet Gray/Pink Red/Blue
Continued next page
QX Hawk Industrial Imager User Manual
A-11
Page 328
Electrical Specifications
QX Hawk Connectors and Pinouts (continued) RS-422/485 RxD (–) Input Common
RS-232 TxD/ RS-232 RTS Trigger
Ground
RS-232 RxD/ RS-232 CTS
RS-422/485 RxD (+)
Power
RS-422/485 TxD (+)
Terminated
RS-422/485 TxD (–)
RX (+)
Function Trigger Power Terminated Input 1 Port 3 422/485 TxD (+) Port 3 422/485 RxD (+) Ground Input Common Port 2 TxD/Port 1 RTS Port 2 RxD/Port 1 CTS Port 3 422/485 TxD (–) Port 3 422/485 RxD (–)
TX (–) Terminated
B (Serial) M12 12-pin Socket
A-12
Terminated
RX (–)
Terminated Input 1
Pin 1 2 3 4 5 6 7 8 9 10 11 12
Terminated
TX (+)
Wire Color White Brown Green Yellow Gray Pink Blue Red Black Violet Gray/Pink Red/Blue
B (Ethernet) M12 8-pin Socket Pin 1 2 3 4 5 6 7 8
Function Terminated Terminated Terminated TX (–) RX (+) TX (+) Terminated RX (–)
QX Hawk Industrial Imager User Manual
Page 329
Appendices
QX-1 Connectors Connector T on the QX-1 Interface Device is the Trigger connector. Connectors 1, 2, and 3 can be used to bus power and data as required by the application. The connectors on the QX-1 interface device physically mirror those on the QX Hawk imager, but they do not have explicit pin assignments. The QX-1 connectors take on the communications and power roles that are assigned to them. Trigger
Ground
+10-28V
Trigger/New Master / Input 1 Common
QX-1 Trigger Connector 4-pin Socket
1 +10-28V 2 Trigger / New Master / Input 1 Common 3 Ground 4 Trigger
Power Switch Trigger Switch
QX-1 Wiring Diagram
Connectors 1 and 3 are 12-pin plugs, and Connector 2 is a 12-pin socket. All three connectors can be assigned to bus power and data. The two switches at the center of the device allow the user to route signals as needed. The simple diagram above (also shown on the base of the QX-1) illustrates how power, communications, I/O, and trigger signal can be routed through the QX-1 device. Power can be bussed between imagers and interface devices. At each location on a network where a new power supply is added, the Power switch on the QX-1 can be used to break power between Connector 2 and Connectors 1, 3, and T. The Trigger signal between Connector 2 and Connectors 1, 3, and T can be broken using the Trigger switch. This isolates trigger signals as required.
QX Hawk Industrial Imager User Manual
A-13
Page 330
Electrical Specifications
Optoisolated Outputs The imager has optoisolated outputs that can transfer signals from the imager to peripherals. Outputs can be configured as either NPN or PNP, but NPN and PNP cannot be mixed in a system, because the output common is shared by all outputs.
NPN Output for Host Input
NPN Output for External Load
A-14