Utilities
Static Validation Sending a <VAL> command from ESP’s Terminal will cause the imager to perform a static validation operation on a Data Matrix symbol and will output validation results to the Terminal.
Output Screen Data Matrix validation information is output in the format below in response to the <VAL> command. Note: If the imager fails to decode the Data Matrix symbol, a No Read message is displayed.
Validation Parameters Symbol Contrast Usage: Definition:
Lets the user know if contrast settings are less than acceptable. All the pixels that fall within the area of the test symbol, including its required zone, will be sorted by their reflectance values to select the darkest 10% and the lightest 10% of the pixels. The arithmetic mean of the darkest and the lightest pixels is calculated and the difference of the two means is the Symbol Contrast. (ANSI) Symbol Contrast grading is defined this way: A (4.0) if SC ? 70% B (3.0) if SC ? 55% C (2.0) if SC ? 40% D (1.0) if SC ? 20% F (0.0) if SC ? 20%
QX Hawk Industrial Imager User Manual
7-29
Page 312
Static Validation
Print Growth Definition:
The extent to which dark or light markings appropriately fill or exceed their module boundaries. These values are determined by counting pixels in the clock pattern of the binary digitized image, then comparing it to a nominal value and minimum and maximum values. The grade is defined in this way: A (4.0) if -.050 ? PG ? 0.50 B (3.0) if -.070 ? PG ? 0.70 C (2.0) if -.085 ? PG ? 0.85 D (1.0) if -.100 ? PG ? 1.00 F (0.0) if PG < -1.00 or PG > 1.00
Axial Non-Uniformity (Data Matrix Only) Definition:
Axial non-uniformity is a measure of how much the sampling point spacing differs from one axis to another, namely AN = abs (XAVG - YAVG) / ((XAVG + YAVG)/2) where abs ( ) yields the absolute value. If a symbology has more than two major axes, then AN is computed for those two average spacings which differ the most. The grade is defined in this way: A (4.0) if AN ? .06 B (2.0) if AN ? .08 C (2.0) if AN ? .10 D (1.0) if AN ? .12 F (0.0) if AN ? .12
Unused Error Correction (Data Matrix Only) Definition:
7-30
The correction capacity of Reed-Solomon decoding is expressed in the equation: e + 2d ? d - p where e is the number of erasures, t is the number of errors, d is the number of error correction codewords, and p is the number of codewords reserved for error detection. The Unused Error Correction parameter tests the extent to which regional or spot damage in the symbol has eroded the decoding safety margin that error correction provides. The grade is defined in this way: A (4.0) if UEC ? .62 B (3.0) if UEC ? .50 C (2.0) if UEC ? .37 D (1.0) if UEC ? .25 F (0.0) if UEC ? .25
QX Hawk Industrial Imager User Manual
Page 313
Utilities
Targeting Activate Targeting System Sending a <l1> command (lower-case L and numeral 1) from ESP’s Terminal activates the imager’s targeting system. Important: <l1> is not supported by the QX Hawk C-Mount Imager.
De-activate Targeting System Sending a <l0> command (lower-case L and numeral 0) de-activates the imager’s targeting system. Important: <l0> is not supported by the QX Hawk C-Mount Imager.
QX Hawk Industrial Imager User Manual
7-31
Page 314
Y-Modem Transfer Options
Y-Modem Transfer Options Y-Modem Upload <uy,path and image file name> Note: There are only three scales available from the imager. The three scales are full, quarter, and eighth. The Directory response will read out the image dimensions. To receive the image, request to download by the reported dimensions. If quarter-scaled or eighth-scaled images are required, then divide the dimensions by 4 or by 8 and request by the reduced dimensions. (1.) Directory (None assumes root) / (Root Directory) /saved (Saved Directory) (2.) Image Status Good/ (IP decoded a symbol) No Read/ (IP did not decode a symbol) Mismatch/ (IP decoded the wrong symbol) Pending/ (Not run through IP) (3.) Age 0 to n n = Max Images (8 ? Max) (4.) Image Name (Full-Scale) 1280 x 1024.bmp 1280 x 1024_gs.bmp 1280 x 1024.jpg 1280 x 1024_q(1-100).jpg Valid Scales (Depends on Full Dimensions) 1280 x 1024 (Full) 320 x 256 (Quarter - divide by 4) 160 x 128 (Eighth - divide by 8) (5.) Type .bmp (Bitmap image - no loss of resolution) .jpg (JPEG image - loss of resolution with compression) JPEG Compression Quality (1-100%) _q1.jpg _q100.jpg Examples: <uy,/saved/noread/2/320x256_gs.bmp> <uy,/saved/noread/2/320x256_q33.jpg>
7-32
QX Hawk Industrial Imager User Manual
Page 315
Utilities
Y-Modem Download (Code) <dy,key> Key app.bin dsp.bin ker.bin
Note: The “key” field only accepts app.bin, dsp.bin, or ker.bin.
QX Hawk Industrial Imager User Manual
7-33
Page 316
Temperature
Temperature Display Current Image Sensor Temperature Sending an <@TEMP> command from ESP’s Terminal will cause the imager to respond with the current temperature of the image sensor in degrees C. The output from this command is only updated after an image frame has been captured. If the command is requested without capturing an image, the temperature value will not change. Example Response: <TEMP/30.50c>
7-34
QX Hawk Industrial Imager User Manual
Page 317
Appendices Contents Appendix A General Specifications .............................................................................................. A-2 Appendix B Electrical Specifications .......................................................................................... A-11 Appendix C Ground and Shield Considerations......................................................................... A-20 Appendix D Flying Lead Cordset Pinouts................................................................................... A-22 Appendix E Serial Commands ................................................................................................... A-23 Appendix F Protocol Commands................................................................................................ A-85 Appendix G ASCII Table ............................................................................................................ A-94 Appendix H Configuring Ethernet TCP/IP .................................................................................. A-95 Appendix I Using EtherNet/IP..................................................................................................... A-98 Appendix J Allen-Bradley Version 16 PLC Setup..................................................................... A-107 Appendix K Allen-Bradley Version 20 PLC Setup .................................................................... A-120 Appendix L EIPScan Setup ...................................................................................................... A-132 Appendix M Interface Standards .............................................................................................. A-138 Appendix N Glossary of Terms ................................................................................................ A-139
QX Hawk Industrial Imager User Manual
A-1
Page 318
General Specifications
Appendix A, General Specifications Mechanical (Standard QX Hawk) Height: 1.59" (40.5 mm) Width: 2.27" (57.6 mm) Depth: 3.79" (96.3 mm) Weight: 10 oz. (280 g) Mechanical (QX Hawk C-Mount) Height: 4.03" (102.3 mm) Width: 2.27" (57.6 mm) Depth: 1.59" (40.5 mm) Weight: 11 oz. (320 g) Environmental Enclosure: Die-cast aluminum, IP65/67 rated Operating Temperature: CMOS: 0° to 50° C (32° to 122° F); CCD: 0° to 45° C (32° to 113° F) Storage Temperature: –29° to 70° C (–20° to 158° F) Humidity: up to 90% (non-condensing) Communication Interface Standard Interfaces: RS-232/422/485, Ethernet CE Mark General Immunity for Light Industry: EN 55024:1998 ITE Immunity Standard Radiated and Conducted Emissions of ITE Equipment: EN 55022:98 ITE Disturbances Light Source (Standard QX Hawk) Type: High-output LEDs
Symbologies 2D Symbologies: Data Matrix (ECC 0-200), QR Code, Micro QR Code, Aztec Code Stacked Symbologies: PDF417, MicroPDF417, GS1 DataBar (Composite and Stacked) Linear Symbologies: Code 39, Code 128, BC412, Interleaved 2 of 5, UCP/EAN, Codabar, Code 93, Pharmacode, Postal Symbologies Light Collection Options Progressive scan, square pixel Shutter: Software-adjustable 1/60 to 1/100,000 (CMOS); 1/30 to 1/100,000 (CCD) WVGA CMOS: 752 x 480 pixels SXGA CCD: 1280 x 960 pixels Indicators LEDs: Read performance, power, read status, network activity, I/O Beeper: Good Read, Match/Mismatch, No Read, serial command confirmation, On/Off Green Flash (Standard QX Hawk): Good Read Red X (Standard QX Hawk): Symbol locator
A-2
QX Hawk Dimensions Note: Nominal dimensions shown. Typical tolerances apply.
QX Hawk Industrial Imager User Manual
Page 319
Appendices Read Parameters Pitch: ±30° Skew: ±30° Tilt: ±360° CMOS Decode Rate: Up to 60 decodes per second CCD Decode Rate: Up to 20 decodes per second Laser Light (Standard QX Hawk) Type: Laser diode Output Wavelength: 655 nm nominal Operating Life: 50,000 hours @ 25° C Safety Class: Visible laser: Class 1
Protocols Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, Multidrop, Daisy Chain, User-Defined Multidrop, Ethernet TCP/IP, EtherNet/IP Electrical CMOS Power Requirement: 5-28VDC, 200 mV p-p max ripple, 135 mA at 24VDC (typ.) CCD Power Requirement: 5-28VDC, 200 mV p-p max ripple, 170 mA at 24VDC (typ.) Discrete I/O Input 1/Trigger/New Master: Bi-directional, Optoisolated, 4.5–28V rated (13 mA at 24VDC) Outputs (1, 2, 3): Bi-directional, Optoisolated, 1–28V rated (ICE < 100 mA at 24VDC, current limited by user)
QX Hawk C-Mount Dimensions Note: Nominal dimensions shown. Typical tolerances apply.
QX Hawk Industrial Imager User Manual
A-3
Page 320
General Specifications Safety Certifications CDRH, FCC, UL/cUL (Listed, UL60950 4K68), CE, CB, BSMI (compliant)
RoHS/WEEE Compliant
U.S. Patents 7,118,042 | 7,341,190 | 8,146,823 | 8,154,810 | 7,213,761 | 7,311,260 ©2015 Microscan Systems, Inc. All rights reserved. Specifications subject to change. Read Range and other performance data is determined using high quality Grade A symbols per ISO/IEC 15415 and ISO/IEC 15416 in a 25° C environment. For application-specific Read Range results, testing should be performed with symbols used in the actual application. Microscan Applications Engineering is available to assist with evaluations. Results may vary depending on symbol quality. Warranty–For current warranty nformation on this product, please visit www.microscan.com/warranty.
A-4
QX Hawk Industrial Imager User Manual