Page 31

Using ESP

Preferences > Advanced Tab

The Auto Sync options at the top of the Advanced tab allow the user to determine whether Auto Sync will be enabled automatically in sections of ESP where it is used, or if it will ask before it enables Auto Sync functions.

Always Ask Before Auto Sync Occurs If this option box is checked, specific Auto Sync functions can be enabled. Receive Settings from the Reader will automatically send the scanner’s settings to ESP when Auto Sync is enabled. Send ESP Settings to the Reader will automatically send all scanner configuration settings chosen in ESP to the scanner. Do Not Send or Receive Settings creates a condition in which Auto Sync will not automatically send scanner settings to ESP, or send ESP settings to the scanner.

Send XON with Autoconnect Sends an XON (Begin Transmission) command to the scanner before starting the Autoconnect routine.

QX-870 Industrial Raster Scanner User Manual

2-9

Page 32

Menu Toolbar

Preferences > Advanced Tab (cont.) Ask to Save ESP File when Quitting When enabled, prompts the user to save a .esp file when ending a session.

The .esp file will be saved in the location specified by the user.

Connect to Readers via TCP/IP When enabled, shows the TCP/IP Connection Wizard by default.

Use Default Storage Location When enabled, automatically stores data in ESP’s Application Data folder.

2-10

QX-870 Industrial Raster Scanner User Manual

Page 33

Using ESP

Document Memo The information entered in the Document Memo field will appear in a context-sensitive text box whenever the cursor hovers over the Document Memo item on the Options menu.

Model Memo Similar to Document Memo, the information entered in the Model Memo field will appear in a context-sensitive text box whenever the cursor hovers over the Model Memo item on the Options menu. Memos created in Model Memo are specific to the model enabled when the message was created.

Note: Memos must be saved in a .esp file to make them available in the next session. If the current session is not saved, any memos that have been entered during the session will be discarded, and will be unavailable in the next session.

QX-870 Industrial Raster Scanner User Manual

2-11

Page 34

Menu Toolbar

Connect The Connect dropdown menu allows the user to access the Connection Wizard, as well as the Autoconnect and Configure Multidrop dialogs. Connect and Disconnect can also be performed directly from the dropdown menu without opening a dialog.

Connection Wizard To connect using the Connection Wizard: • Click Connect on ESP’s menu toolbar, and then select Connection Wizard. • Select RS-232 or Ethernet to activate the appropriate display. • Configure RS-232 or Ethernet settings as required by the application, and click Connect.

Click the Search button to locate scanners on the network.

RS-232 Connection Wizard

Ethernet Connection Wizard

• When a connection is established, the green indicator in the status bar at the bottom right of the screen will be visible:

2-12

QX-870 Industrial Raster Scanner User Manual

Page 35

Using ESP

Autoconnect • If the RS-232 connection attempt fails, use Autoconnect to establish a connection between the scanner and the host.

• If the communication port is not the default COM1, use the dropdown menu to change the port.

• Once the correct port is chosen, click Start to connect.

• When a connection is established, the green indicator in the status bar at the bottom right of the screen will be visible.

QX-870 Industrial Raster Scanner User Manual

2-13

Page 36

Menu Toolbar

View The View menu allows the user to move quickly between the Setup, Terminal, and Utilities interfaces without using the icon buttons on the App Mode toolbar. It also allows the user to access the Bar Code Dialog, shown below.

Bar Code Dialog Symbols can be created in the Bar Code Dialog by typing the text to be encoded. This is a useful tool for creating configuration symbols, allowing the user to configure the scanner by reading the user-created symbols.

Drag specific configuration values from the control tree directly into this field to encode new symbols. Choose a spatial orientation for the new symbol. Create a caption for the symbol that matches or describes the encoded data.

The symbol will be displayed in the field at the bottom of the Bar Code Dialog.

2-14

QX-870 Industrial Raster Scanner User Manual

Page 37

Using ESP

Navigating in ESP To change scanner settings, or to access the Setup, Terminal, or Utilities views, click the App Mode button.

To return to EZ Mode, click the EZ Mode button.

To make changes to configuration settings in the tree controls:

1. Left-click on the + to expand menu items.

The X indicates that the setting is default.

2. Double-click the desired parameter and single-click in the selection box to view options. 3. Place the cursor in the selection box, scroll down 4. Left-click again on the open screen to complete the to the desired setting, and selection. single-click the setting. 5. Right-click on the open screen and select Save to Reader to implement the command in the scanner. The command can be sent without saving, or sent and saved simultaneously.

QX-870 Industrial Raster Scanner User Manual

2-15

Page 38

Send/Receive Options

Send/Receive Options To access Receive, Save, and Default options, click the Send/Recv button. These options can also be reached by right-clicking in any of the configuration views.

Receiving From the Send/Recv menu, select Receive Reader Settings. Caution: Selecting this option will upload the scanner’s settings. If the ESP file has a number of custom settings that must be maintained and downloaded into the scanner, these settings will be lost. This function is useful for receiving (uploading) the scanner’s settings and saving them as a file for future use. For example, if the scanner has settings that must not change, Receive Reader Settings would load those settings to ESP and save them in an ESP file for later retrieval. Receiving the scanner’s settings will also assure that any unwanted subsequent changes in ESP will not be saved.

Saving Send, No Save (<A>) Saves ESP settings to current memory.

Send and Save (<Z>) Activates all changes in current memory and saves to the scanner for power-on.

Send and Save as Customer Defaults (<Zc>) Saves default settings for quick retrieval. This option will be visible only if Enable ‘Send and Save as Customer Defaults’ is checked in ESP Preferences.

2-16

QX-870 Industrial Raster Scanner User Manual

Page 39

Using ESP

Defaulting When Default Current Menu Settings or Default all ESP Settings are selected, only the ESP settings are defaulted.

Advanced Options Send Current View This is the same as Save to Reader > Send No Save except that only the commands in the current configuration tree are sent.

Send Current Command This is the same as Send Current View, except that it saves only the command that is currently selected.

Add/Remove Exception After a Receive Reader Settings command is performed1 and the Add Exception option is selected, a list of serial commands may appear. These are commands that may be in the scanner’s firmware, but are not included in, or are different from, the current version of ESP. These commands can be edited by double-clicking them and changing them as needed. It is important to note that these commands will be saved to the scanner whenever a Save to Reader command is sent, or when an <A> or a <Z> command is sent. Also, if there is a corresponding ESP menu item, the ESP Value column for that item will be blank following a Receive Reader Settings command.

1. From the Send/Recv button or by right-clicking in any blank section of a tree control view.

QX-870 Industrial Raster Scanner User Manual

2-17

Page 40

Send/Receive Options

2-18

QX-870 Industrial Raster Scanner User Manual