Page 201
Calibration
Advanced Calibration Setup Embedded Menu The Calibration Setup menu allows you to change the starting focus position and the number of focus steps during the first pass. For example, if you are preparing to do calibration on symbols whose minimum distances from the scanner is 40 inches, there is no reason for calibration to cycle through all of the focal positions inside 40 inches. • To increase Starting Focus Position, type T. To decrease it, type R. • To increase the First Pass Focus Step, type G. To decrease it, type F. Increasing First Pass Focus Step will speed up the autofocus process, but can also cause autofocus to skip over a symbol and not decode it. Note: All single letter commands must be entered in upper case. • Auto Range is disabled by default. To enable, press A.
MS-890 Industrial Automation Scanner User’s Manual
9-21
Page 202
Calibration by Embedded Menu
9-22
MS-890 Industrial Automation Scanner User’s Manual
Page 203
10 Raster/Laser Contents Raster/Laser Setup by ESP ........................................................................................................10-2 Raster/Laser Serial Commands ..................................................................................................10-2 Raster Setup ...............................................................................................................................10-3 Laser Framing .............................................................................................................................10-7 Raster/Laser by Embedded Menu ..............................................................................................10-9
This section explains the setup and adjustments for the raster and laser. Note: The characters NUL, <, >, and , can only be entered through embedded menus, not through ESP or serial commands. Note: Raster setup commands in the ESP are accessed from the Read Cycle/Triggering menu or from the Raster/Laser or Calibration menus on the ESP toolbar.
MS-890 Industrial Automation Scanner User’s Manual
10-1
Page 204
Raster/Laser Setup by ESP
Raster/Laser Setup by ESP
Click this button to bring up the Raster/Laser menu. To open nested options, single-click the +.
To change a setting, double-click the setting and use your cursor to scroll through the options.
Raster/Laser Serial Commands Command Title Raster Laser On/Off
10-2
Format <K506,raster status,top offset,bottom offset,raster sweep rate,read cycle on/off> <K700,laser on/off status,laser framing status,laser on position,laser off position, laser power>
MS-890 Industrial Automation Scanner User’s Manual
Page 205
Raster/Laser
Raster Setup You can set up and make changes to the scanner’s raster parameters in ESP’s Read Cycle, Calibration, or Raster/Laser views (preferred). To access the Raster Setup menu, click the Raster/Laser button and the Raster Setup tab.
Usage:
Raster scanning is useful when symbols vary in placement and a single scan line cannot be counted on to cross the symbol, and for reading stacked symbologies such as PDF417.
Definition:
The user can enable raster scanning and adjust the raster sweep limits and the sweep rate.
Note: Changes made in the Raster Setup menu are dynamic; that is, they take effect without saving.
MS-890 Industrial Automation Scanner User’s Manual
10-3
Page 206
Raster Setup
Raster Status Definition:
When disabled, the raster mirror will park at the top of the raster sweep.
Serial Cmd:
<K506,raster status,top offset,bottom offset,raster sweep rate,read cycle on/off>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
Note: Raster sweeps are only counted during a read cycle (continuous, triggered, and read rate). Raster sweeps are NOT counted when the raster motor is parked, disabled, set to straight line operation (top and bottom are equal), or when performing a self-calibration.
Top Offset Definition:
Decreasing the Top Offset value causes the top of the raster pattern to move up.
Serial Cmd:
<K506,raster status,top offset,bottom offset,raster sweep rate,read cycle on/off>
Default:
127
Options:
0 to 255
10-4
MS-890 Industrial Automation Scanner User’s Manual
Page 207
Raster/Laser
Bottom Offset Definition:
Increasing the Bottom Offset value causes the bottom of the raster pattern to move down.
Serial Cmd:
<K506,raster status,top offset,bottom offset,raster sweep rate,read cycle on/off>
Default:
127
Options: 0 to 255 Note: For single scan line operation, set Top Offset equal to Bottom Offset or click on the Static Line option. Note: Assigning a larger value to Top Offset than Bottom Offset will park the raster mirror in its “home” (default) position.
Raster Sweep Rate Usage:
If your application allows it, a slower sweep rate can produce more scans per second and higher read rates. To determine the minimum read rate for your application, see Raster Picket Fence Calculation on page A-28.
Definition:
The raster sweep rate is the number of sweeps per second the raster mirror completes. A sweep is one pass of the raster, up or down. The maximum sweep rate is limited by the size of the arc the raster must move through.
Serial Cmd:
<K506,raster status,top offset,bottom offset,raster sweep rate,read cycle on/off>
Default:
10
Options:
0 to 30. The table below shows the maximum speeds at selected raster sweep arcs. To maximize the number of scans per symbol, use the lowest effective sweep rate required for the application. Raster Sweep Arc 0 to 10° 11 to 20° 21 to 36°
MS-890 Industrial Automation Scanner User’s Manual
Maximum Sweeps per Second 30 20 10
10-5
Page 208
Raster Setup
Read Cycle On/Off Usage:
This feature should only be used when the raster sweep rate is at least twice the required trigger rate (decoded outputs per second).
Definition:
When Read Cycle On/Off is enabled and Laser On/Off is also enabled, the raster mirror stops at top offset at the end of triggered read cycle.
Serial Cmd:
<K506,raster status,top offset,bottom offset,raster sweep rate,read cycle on/off>
Default:
Disabled
Options:
0 = Disabled
10-6
1 = Enabled
MS-890 Industrial Automation Scanner User’s Manual
Page 209
Raster/Laser
Laser Framing To access the Laser Framing menu, click the Raster/Laser button and the Laser Framing tab.
Laser On/Off Status Usage:
Laser On/Off extends the life of the laser. It is useful where there are significant time gaps between symbols. It also provides visual confirmation of a read cycle timeout and minimizes laser exposure to people.
Definition:
When enabled, the laser is ON only during the read cycle. When disabled, the laser operates continuously.
Serial Cmd:
<K700,laser on/off status,laser framing status,laser on position,laser off position,laser power>
Default:
Disabled
Options: 0 = Disabled 1 = Enabled Note: A serial or external trigger must be enabled for Laser On/Off to take effect. Note: Laser On/Off is not recommended when New Trigger is enabled, since the laser will effectively remain on continuously. MS-890 Industrial Automation Scanner User’s Manual
10-7
Page 210
Laser Framing
Laser Framing Status Usage:
Useful in avoiding highly reflective objects, for filtering out unwanted signals, or avoiding the wrong symbols.
Definition:
Sets the length of time the laser is ON during each scan so that only a selected portion of the scan width is effectively scanned.
Serial Cmd:
<K700,laser on/off status,laser framing status,laser on position,laser off position,laser power>
Default:
Disabled
Options:
0 = Disabled
1 = Enabled
To adjust laser framing, click the Raster button, click the Laser Setup tab, and click the Laser On and Laser Off arrows up or down. When Laser Framing is not enabled, the scan beam sweeps across the full scan width. When enabled, the laser remains OFF during the Laser Off Time beginning at the start of the scan beam sweep and turns ON for the duration of the Laser On Time.
Laser On Position Definition:
Measured in increments equal to 1/100th of total scan width.
Serial Cmd: <K700,laser on/off status,laser framing status,laser on position,laser off position, laser power> Default:
10
Options:
10 to 90
Laser Off Position Definition:
Measured in increments equal to 1/100th of total scan width.
Serial Cmd: <K700,laser on/off status,laser framing status,laser on position,laser off position, laser power> Default:
95
Options:
15 to 95
Laser Power Usage:
Used to optimize read performance at different distances. Typically higher power is used at greater distances.
Definition:
Allows the user or calibration routine to select laser power setting.
Serial Cmd: <K700,laser on/off status,laser framing status,laser on position,laser off position, laser power> Default: Options: 10-8
High Power (low or medium density scanner) 0 = Low (3.0 mW)
1 = Medium (3.25 mW)
2 = High (3.5 mW)
MS-890 Industrial Automation Scanner User’s Manual