Page 191

Calibration

Number of Active Database Settings Usage:

Can be useful when certain settings that are needed for one application are not needed in another and processing time is at a premium.

Definition:

This value sets the number of configuration database settings (in Database Index) that are active. If Control Trigger mode has been selected, this setting is ignored.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

1

Options:

Any number from 1 to 7.

Latch on Trigger Usage:

Useful when only one database index is required for the application.

Definition:

This feature is only used in Control Trigger mode. If enabled, the configuration database settings selected at the start of a read cycle will remain active during the entire read cycle. If Latch on Trigger is disabled, depending on the control trigger inputs, the scanner’s optical settings can change throughout the read cycle. Note: When Control Trigger #3 is enabled, Latch on Trigger is also enabled regardless of its previous setting.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

Enabled

Options:

0 = Disabled 1 = Enabled

Timer/Switch Value Usage:

Useful in slower applications where there is enough time to cycle through the database index.

Definition:

When Timer mode is enabled, Timer Value is the time interval in which database index settings are cycled.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

5

Options:

0 to 255 (increments of 10ms)

MS-890 Industrial Automation Scanner User’s Manual

9-11

Page 192

Optoelectric Control

Control Trigger 1 Status Definition:

When Control Trigger mode is enabled, Control Trigger #1 takes over Input 1 (see Input Mode on page 6-19) and uses it to select an optoelectric configuration from the configuration database. Active state is set by Input 1 serial command.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

Control Trigger 2 Status Definition:

New Master Pin is used for Control Trigger #2. When enabled and Mode is set to Control Trigger, this input is used to select an optoelectric configuration from Configuration Database. Active State is “Active Closed”.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

Control Trigger 3 Status Definition:

Trigger Input is used for Control Trigger #3. When enabled and Mode is set to Control Trigger, this input is used to select an optoelectric configuration from the Configuration Database. Active State is set by the External Trigger State serial command.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

9-12

MS-890 Industrial Automation Scanner User’s Manual

Page 193

Calibration

Database Cycles Definition:

Database Cycles refers to the number of times the scanner will load the series of active entries when it is set to Timer or Raster Sweep modes. When set to 0, the scanner will cycle repeatedly through all active database entries until the read cycle ends. When set to a value greater than 0, the scanner will only load the active database entries for the defined number of times before staying with the last active entry for the remainder of the read cycle.

Serial Cmd:

<K226,mode,# active database settings,latch on trigger,timer/switch value,control trigger 1 status,control trigger 2 status,control trigger 3 status, database cycles>

Default:

0

Options:

0 to 255

MS-890 Industrial Automation Scanner User’s Manual

9-13

Page 194

Advanced Calibration Features

Advanced Calibration Features In the Advanced tab in the Calibration menu you can constrain the scan width of the scan beam with Laser Framing and set up the rules the scanner will follow during calibration procedure. See Laser Framing on page 10-7 for information on laser framing.

Auto Range Usage:

Enabling Auto Range improves focus acquisition speed where extraneous objects are not mistaken for symbols.

Definition:

With Auto Range enabled, the scanner will focus outward and not begin the calibration routine until an object is detected by signal strength.

Serial Cmd:

<K510,auto range,starting focus position,focus step size>

Default:

Enabled

Options:

0 = Disabled 1 = Enabled

Using Auto Range 1. Be sure there are no extraneous objects between the symbol and the scanner. 2. Click on the Advanced tab in the Calibration menu. 3. If Auto Range is not enabled, click on it to enable it. 4. Click the Calibrate button to start calibration. 5. Examining the read rates in the Results graph following calibration.

9-14

MS-890 Industrial Automation Scanner User’s Manual

Page 195

Calibration

Starting Focus Position Usage:

Starting Focus Position can improve focus acquisition speed during calibration when the approximate range of the symbol is known.

Definition:

With Auto Range enabled, the scanner will begin the calibration routine at the focal range set in Starting Focus Position.

Serial Cmd:

<K510,auto range,starting focus position,focus step size>

Default:

0

Options:

0 to 40

After you’ve used calibration and observed the progress of the first and second passes in the graph and table in the Results tab, you should sense a relationship between focus position and the distance to the symbol.

Improving the speed and accuracy of focus position setup: 1. Click on the Advanced tab in the Calibration menu. 2. If Auto Range is enabled, click on it to disable. 3. From previous results, estimate the value of Starting Focus Position and drop down one or two positions. For example, if you estimate the distance to your symbol to correspond to a Starting Focus Position of 6, enter 5 or 4 to be certain of not missing the peak performance distance. 4. Click the Calibrate button to start calibration. 5. Examining the read rates in the Results graph following calibration.

MS-890 Industrial Automation Scanner User’s Manual

9-15

Page 196

Advanced Calibration Features

Focus Step Size Usage:

Increasing the Focus Step Size can improve focus acquisition speed for symbols.

Definition:

Focus Step Size is the number of focus increments that will be stepped through during the first calibration pass. The range is 2 to 5. The higher values will result in fewer readings during the first pass but will increase the risk of missing the peak performance position.

Serial Cmd:

<K510,auto range,starting focus position,focus step size>

Default:

5

Options:

2 to 5

Improving the speed and accuracy of focus step size setup: 1. Click on the Advanced tab in the Calibration menu. 2. If Auto Range is enabled, click on it to disable. 3. Enter a value in Focus Step Size. 4. Click the Calibrate button to start calibration. 5. Examine the read rates in the Results graph following calibration.

9-16

MS-890 Industrial Automation Scanner User’s Manual

Page 197

Calibration

Calibration by Embedded Menu To calibrate via the embedded Calibration menu, do the following: 1. Place the sample symbol at a fixed position in the center of the read range being tested. 2. Enter Calibration menu by sending an <@> command from a terminal.

3. If the application requires a change in scan rate, adjust the motor speed now with the M or N commands. Note: All single letter commands must be entered in upper case. 4. Start calibration by typing an uppercase A from your terminal while in the calibration menu. Note: A user can abort calibration at any time by pressing the ESC key on the terminal. No other keystroke has any effect while in calibration. 5. Wait for calibration to cycle through its settings. A Calibration in Progress message will be displayed. A new text line will appear below the menu displaying the progress of the calibration. 6. As with the serial command, repeat the routine at other positions in your scan range and width and save each position to the configuration database. 7. If not successful, a Calibration Failed message will be displayed. If successful, a Calibration Passed message will be displayed along with other settings. Note: Focus positions Inside, Outside, and Set posted near the bottom of the above menu are explained more fully in Calibration Results in Embedded Menus on page 9-20. Note: If you are using the 9-pin configuration port and doing calibration by serial command or embedded menus, you will be able to view decoded symbol data in the menu. However, after exiting the Calibration menu, you will not be able to see symbol data without sending the following command: <K146,,1>. If using ESP, this is not necessary.

MS-890 Industrial Automation Scanner User’s Manual

9-17

Page 198

Calibration by Embedded Menu

Scanner Settings by Embedded Menus Motor Speed, Gain Setting, and Tracking adjustments affect the read performance. Focus controls a stepper motor that physically slides a lens back and forth to control focal length. These settings are adjusted automatically in calibration. The status of these parameters is displayed on the bottom line of the menu under Motor, Gain, Track, and Focus. These settings are incremented or decremented by pressing the indicated key shown on the menu. PDF417 Decode, AGC Status, Laser Power, and Raster Status use single commands to toggle between states.

AGC Status Typing S toggles AGC between Enable/Disable, as displayed on a the status line near the bottom of the menu. With AGC disabled this value is simply the Gain setting. However, with AGC enabled, this value represents the gain the scanner has to set to maintain the desired signal amplitude. If a symbol is close to the scanner, the AGC level should be a small value because the scanner does not need much gain to achieve the desired signal amplitude. However, as a symbol gets farther away from the scanner, this value will increase to maintain the signal amplitude.

Laser Power Typing L toggles between High and Low laser settings.

PDF417 Decode Test Typing D toggles the PDF417 Decode Test between Enable/Disable, as displayed on a the status line near the bottom of the menu. Once set to Enabled, PDF417 only is enabled and a special test mode is enabled that displays the decodes per second of a full PDF417 symbol. While calibration is not performed in this mode, you can still adjust settings manually. This special read rate will remain enabled until the Calibration menu is exited or PDF417 is disabled with another D command. Note: This mode averages decode and capture time over a series of readings, thus assuring for a smooth read rate output, but at the cost of delayed reaction to symbol movement and user setting changes.

9-18

MS-890 Industrial Automation Scanner User’s Manual

Page 199

Calibration

Calibration Embedded Menu Options START Calibration Typing A initiates an calibration sequence. Once entered, no other keystrokes will be accepted with the exception of the ESC key, which terminates the routine.

RECALL Initial Settings Typing B restores the original settings that were present upon entering the Calibration Setup Menu. These settings include Motor Speed, Gain, Tracking, and Focus.

SAVE Settings to Database Typing C saves the current Gain, Tracking, and Focus settings to the Configuration Database. The database is made up of indexes 1 through 7. Current settings will be saved to the first open index.

EXIT Calibration Setup Menu To exit the Calibration Setup menu, press the ESC key. The following prompt will be displayed: “Do you want to save calibration settings for power on? (Y=yes N=no).” Typing Y will cause the current settings in the Calibration Setup menu to be saved to Flash Memory for recall when the scanner is powered up at a later date. These settings include Motor Speed, Gain, Tracking, Focus, AGC Status, Laser Power, and Raster Status. Typing N will cause the scanner to be reset without saving the settings to Flash memory.

Menu Selections Typing 1 will take you to the Calibration Results menu. Typing 2 will take you to the Raster/Laser Framing Setup menu. Typing 3 will take you to the Configuration Database menu. Typing 4 will take you to the Advanced Setup menu.

MS-890 Industrial Automation Scanner User’s Manual

9-19

Page 200

Calibration by Embedded Menu

Calibration Results in Embedded Menus To access Calibration Results menu from within the Calibration Setup menu, type the numeral 1.

The Calibration Results menu shows the results of the calibration process for Read Rate, Gain, and Tracking for both the first and second passes. Note that the first pass has only one reading for each parameter.

Focus Positions In the example above, the focus position selected from the first calibration pass (position 26) represents the optimal focus position for Read Rate, Gain, and Tracking. The second “fine tune” pass commences 3 focus positions in from this and steps out until the read rate drops off to zero.

Focal Range The algorithm selects the best focus position (SET) as the midpoint between those positions (IN and OUT) at which read rates drop off by no more than 6-2/3 percent of the maximum read rate. For example, in the above menu, the read rate for focus position 24 is 73. Since 73 is more than 6-2/3% below the maximum read rate of 100% shown elsewhere, the IN range falls under the next focus position, 25. A mirrored calculation is performed to locate the OUT focal range. Press the enter key (CR) to return to the main calibration menu or ESC to leave the Auto Calibration Setup menu entirely.

9-20