Page 121

Symbologies

Autodiscriminate Although Autodiscriminate is not a configuration command, but it is included here as a convenient tool for enabling most symbol types. Usage:

Commonly used for quick setup mode to detect symbol type. This is particularly useful for users who might be unfamiliar with their application’s symbology. Note: It does not alter individual fields such as Start/Stop, Fixed Length, etc. These need to be setup individually.

Definition:

Enables all available symbology types except PDF417, UPC/EAN, and Pharmacode. The user may also individually disable/enable each symbology type.

Serial Cmd:

<P>

Default:

Code 39 (only)

Options:

<P> Enables all codes except noted above. <Q> Enables Code 39 only <R> Enables Codabar only <S> Enables I 2/5 only

Note: If using an I 2/5 symbol, verify that the number of characters in the symbol being scanned matches the symbol length enabled for the I 2/5 symbol type (default is 10 and 6).

MS-890 Industrial Automation Scanner User’s Manual

5-27

Page 122

Autodiscriminate

5-28

MS-890 Industrial Automation Scanner User’s Manual

Page 123

6 I/O Parameters Contents I/O Parameters by ESP.................................................................................................................6-2 I/O Parameters Serial Commands ................................................................................................6-3 Symbol Data Output......................................................................................................................6-4 No Read Message ........................................................................................................................6-6 Bad Symbol Message ...................................................................................................................6-7 No Symbol Message .....................................................................................................................6-7 Read Duration Output ...................................................................................................................6-8 Output Indicators...........................................................................................................................6-9 Beeper.........................................................................................................................................6-12 Serial Verification ........................................................................................................................6-13 EZ Button ....................................................................................................................................6-15 EZ Button Modes ........................................................................................................................6-17 Input 1 .........................................................................................................................................6-19 Configurable Output 1 .................................................................................................................6-20 Trend Analysis Output 1..............................................................................................................6-24 Diagnostics Output 1...................................................................................................................6-27 Configurable Output 2 .................................................................................................................6-28 Trend Analysis Output 2..............................................................................................................6-28 Diagnostics Output 2...................................................................................................................6-28 Configurable Output 3 .................................................................................................................6-29 Trend Analysis Output 3..............................................................................................................6-29 Diagnostics Output 3...................................................................................................................6-29 Database Identifier Output ..........................................................................................................6-30 Quality Output .............................................................................................................................6-31

This section includes instructions on setting up conditions for changing input/output electrical transitions for control of the scanner’s internal and external devices. Note: The characters NUL, <, >, and , can only be entered through embedded menus, not through ESP or serial commands.

MS-890 Industrial Automation Scanner User’s Manual

6-1

Page 124

I/O Parameters by ESP

I/O Parameters by ESP

Click this button to bring up the I/O Parameters menu.

To change a setting, double-click the setting and use your cursor to scroll through the options.

To open nested options, single-click the +.

6-2

MS-890 Industrial Automation Scanner User’s Manual

Page 125

I/O Parameters

I/O Parameters Serial Commands Command Title Serial Verification Beeper Quality Output Symbol Data Output Read Duration Output No Read Message Bad Symbol Message No Symbol Message Input 1 Target/Flood LEDs Database Identifier Output EZ Button EZ Button Modes Trend Analysis Output 1 Trend Analysis Output 2 Trend Analysis Output 3 Diagnostic Output 1 Diagnostic Output 2 Diagnostic Output 3 Configurable Output 1 Configurable Output 2 Configurable Output 3

Command Format <K701,serial command echo status,serial command beep status, control/hex output> <K702,status> <K704,quality output separator,reads/trigger status,direction output> <K705,status,when to output> <K706,status,separator> <K714,status,message> <K715,status,bad symbol message> <K716,status,no symbol message> <K730,mode,active state> <K750,green flash LED status,green flash duration> <K759,status,separator> <K770,global mode,default on power-on status> <K771,one beep,two beeps,three beeps> <K780,trend analysis mode,number of triggers,number to output on,decodes/trigger threshold> <K781,trend analysis mode,number of triggers,number to output on,decodes/trigger threshold> <K782,trend analysis mode,number of triggers,number to output on,decodes/trigger threshold> <K790,over temperature,service threshold,[unused],laser current high,laser current low,low temperature> <K791,over temperature,service threshold,[unused],laser current high,laser current low,low temperature> <K792,over temperature,service threshold,[unused],laser current high,laser current low,low temperature> <K810,output on,output state,pulse width,output mode> <K811,output on,output state,pulse width,output mode> <K812,output on,output state,pulse width,output mode>

MS-890 Industrial Automation Scanner User’s Manual

6-3

Page 126

Symbol Data Output

Symbol Data Output Usage:

Useful when the host needs symbol data only under certain conditions.

Definition:

Defines the conditions under which decoded symbols are transmitted to the host.

Symbol Data Output Mode Serial Cmd:

<K705,output status,when to output>

Default:

Good Read

Options:

0 = Disabled 1 = Match 2 = Mismatch 3 = Good Read

Disabled Usage:

Disabled is useful when an application only needs to use the discrete outputs and can allow the scanner to do the decision-making. When set to Disabled, the host does not need the symbol data and the communication lines are used only for setup and status checks.

Definition:

When set to Disabled, the scanner will not transmit any data that is generated during a read cycle (symbols, No Reads, etc.).

Match Usage:

Match is typically used to route items with the same identifying data to the same location.

Definition:

With Match enabled, the scanner transmits symbol data whenever the symbol data information matches the master symbol.

Mismatch Usage:

Mismatch is typically used as a flag within the host system to prevent an item from being routed in the wrong container.

Definition:

With Mismatch enabled, the scanner transmits symbol data whenever the symbol data information does NOT match the master symbol. Note: A No Read can still be transmitted if enabled.

Good Read Usage:

Good Read is used when an application requires all symbol data to be transmitted. It’s typically used in tracking applications in which each object is uniquely identified.

Definition:

With Good Read enabled, the scanner transmits symbol data on any good read regardless of Matchcode Type setting. Note: A No Read can still be transmitted if enabled.

6-4

MS-890 Industrial Automation Scanner User’s Manual

Page 127

I/O Parameters

When to Output Symbol Data Definition:

This command allows the user to choose when symbol data can be sent to the host.

Serial Cmd:

<K705,symbol output status,when to output>

Default:

As Soon As Possible

Options:

0 = As Soon As Possible

1 = End of Read Cycle

As Soon As Possible Usage:

As Soon As Possible is useful in applications in which symbol data needs to be moved quickly to the host, typically when the host is making decisions based on symbol data.

Definition:

Enabling As Soon As Possible causes symbol data to be sent to the host immediately after a symbol has been successfully decoded. Note: More than one decode might in fact be required to qualify as a good decode, depending on how Decodes Before Output is set.

End of Read Cycle Note: See End of Read Cycle Mode on page 4-14 for options to end the read cycle. Usage:

End of Read Cycle is useful in timing-based systems in which the host is not ready to accept data at the time it is decoded.

Definition:

Enabling End of Read Cycle means that symbol data does not get sent to the host until the read cycle ends with a timeout or new trigger. Start of read cycle

End of read cycle Read Cycle Timeout Host expects output

Host activates trigger

Read Cycle

MS-890 Industrial Automation Scanner User’s Manual

6-5

Page 128

No Read Message

No Read Message Usage:

Used in applications where the host needs serial verification that a symbol has not been read and especially useful in new print verification.

Definition:

When enabled, and if no symbol has been decoded before timeout or the end of the read cycle, the No Read message will be transmitted to the host.

No Read Status Serial Cmd:

<K714,status,message>

Default:

Enabled

Options:

0 = Disabled

1 = Enabled

No Read Message Definition:

Any combination of ASCII characters can be defined as the No Read message.

Serial Cmd:

<K714,status,message>

Default:

NOREAD

Options: 1 to 128 ASCII characters. Note: No Read Message will only be transmitted if Symbol Output is set to Match, Mismatch or Good Read. No Read Message can be set to any ASCII characters except NUL, <, >, and , .

6-6

MS-890 Industrial Automation Scanner User’s Manual

Page 129

I/O Parameters

Bad Symbol Message Usage:

Definition: Serial Cmd: Default: Options:

Useful in determining if a symbol is present and if user defined requirements for that symbol are met. Can tell the user if a qualified symbol is present but not decodable. Sets the requirements that will qualify an object or a symbol before outputting a decode or message. When enabled, sends a message to the host whenever an object meets the qualifications setup but is not decoded. <K715,[unused],message> BAD_SYMBOL 1 to 128 ASCII characters

No Symbol Message Usage:

Definition: Serial Cmd: Default: Options:

Useful in determining if a symbol is present and if user defined requirements for that symbol are met. Can tell the user if an object does not qualify as a symbol. Sets the requirements that will qualify an object or a symbol before outputting a decode or message. When enabled, sends a message to the host whenever an object meets the qualifications setup but is not decoded. <K716,[unused],message> NO_SYMBOL 1 to 128 ASCII characters

MS-890 Industrial Automation Scanner User’s Manual

6-7

Page 130

Read Duration Output

Read Duration Output Usage:

Definition:

Useful in evaluating actual read cycle timing results, especially when initially setting up an application to determine maximum line speed (obtainable based on spacing between symbols.) When enabled the duration of the read cycle (in milliseconds) is appended to the symbol data. The read duration is the time from the beginning of the read cycle until data is output.

Read Duration Output Mode Serial Cmd: Default: Options:

<K706,status,separator> Disabled 0 = Disabled 1 = Enabled

Important: To measure the entire read cycle when in External Level trigger mode, set When to Output Symbol Data to End of Read Cycle. This output can measure over 49 days’ worth of duration; if exceeded, the “OVERFLOW” message will be output in place of the duration.

Read Duration Output Separator Definition:

User defined character that separates the symbol information from the Read Duration Output. <K706,status,separator>

Serial Cmd: Default:

, (comma)

Options:

Any ASCII character.

6-8