Page 111

Symbologies

Supplemental Status (UPC/EAN) Usage: Definition:

Serial Cmd: Default: Options:

Disabled: Enabled:

Required:

Reads Supplementals typically used in publications and documentation. A supplemental is a 2 or 5 digit symbol appended to the main symbol. When set to Enabled or Required, the scanner reads supplemental symbol data that has been appended to the standard UPC or EAN codes. <K473,UPC status,EAN status,supplemental status,separator status, separator character,supplemental type> Disabled 0 = Disabled 1 = Enabled 2 = Required UPC Supplementals will not be decoded. When enabled, the scanner will try to decode a main and a supplemental. However, if a supplemental is not decoded, at the end of the read cycle, the main will be sent by itself. The UPC main and supplemental symbols are considered to be one symbol and will be treated as such. When set to Required and either the main or supplemental symbol is not read, a single No Read condition results. The UPC main and supplemental symbols are treated as one symbol. For example, if Supplementals is set to Required, Separator is enabled, and an asterisk is defined as the UPC separator character, then the data will be displayed as: MAIN * SUPPLEMENTAL.

Note: Under no circumstances will supplemental symbol data be sent without a main symbol. Note: If additional symbols - other than the main or supplemental - will be read in the same read cycle, Number of Symbols should be set accordingly.

Separator Status (UPC/EAN) Usage: Definition:

Default:

Allows user to distinguish between the main and Supplemental symbols. Allows the user to insert a character between the standard UPC or UPC/EAN symbol and the supplemental symbol when Supplementals is set to Enabled or Required. <K473,UPC status,EAN status,supplemental status,separator status,separator character,supplemental type> Disabled

Options:

0 = Disabled

Serial Cmd:

1 = Enabled

MS-890 Industrial Automation Scanner User’s Manual

5-17

Page 112

UPC/EAN

Separator Character (UPC/EAN) Note: If Separator Character has been changed to any other character and you wish to redefine the separator as a comma, you will need to use ESP or the embedded menu. Note: Whenever Separator Character is defined as a comma (,) sending a <K473,?> command will return the current settings, including the separator character comma which appears after the separator status comma. Usage: Definition: Serial Cmd: Default: Options:

As required by the application. Allows the user to change the separator character from a comma to a new character. <K473,UPC status,EAN status,supplemental status,separator status, separator character,supplemental type> , (comma) Any ASCII character

Supplemental Type (UPC/EAN) Usage: Definition: Serial Cmd: Default: Options:

Both:

5-18

As required by symbology used in application. Allows the user to select 2 character or 5 character supplement, or both. <K473,UPC status,EAN status,supplemental status,separator status,separator character,supplemental type> Both 0 = Both 1 = 2 characters only 2 = 5 characters only Either 2 character or 5 character supplementals will be considered valid. 2 Char Only: Only two character supplementals will be considered valid. 5 Char Only: Only five character supplementals will be considered valid.

MS-890 Industrial Automation Scanner User’s Manual

Page 113

Symbologies

Code 93 Usage: Definition:

Used occasionally in the clinical industry. Code 93 is a variable length, continuous symbology employing four element widths. Each Code 93 character has nine modules that may be either black or white. Each character contains three bars and three spaces.

Code 93 Status Serial Cmd: Default:

<K475,status,fixed symbol length status,fixed symbol length> Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length Status (Code 93) Usage:

Serial Cmd: Default:

Helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted. When enabled, the scanner will check the symbol length against the symbol length field. When disabled, any length is considered a valid symbol. <K475,status,fixed symbol length status,fixed symbol length> Disabled

Options:

0 = Disabled

Definition:

1 = Enabled

Fixed Symbol Length (Code 93) Definition:

Serial Cmd: Default: Options:

Specifies the exact number of characters that the scanner will recognize (this does not include start and stop and check characters). The scanner ignores any symbol not having the specified length. <K475,status,fixed symbol length status,fixed symbol length> 10 1 to 128

MS-890 Industrial Automation Scanner User’s Manual

5-19

Page 114

Pharmacode

Pharmacode Usage: Definition:

Pharmacode is used mostly with packaging for the pharmaceuticals industry. The symbol encodes up to five different numbers, each with its own color which may be entered in decimal or “binary” format with a 1 represented by a thick bar and a 0 represented by a thin bar. Bar width is independent of height. In decimal format, each part can be up to 999999. In binary format, each input have up to 19 ones and zeros.

Note: It is recommended that you disable AGC before enabling Pharmacode.

Pharmacode Status Serial Cmd: Default:

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars, bar widths,decode direction,fixed threshold value> Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length Status (Pharmacode) Serial Cmd: Default:

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar widths,,decode direction,fixed threshold value> Disabled

Options:

0 = Disabled

1 = Enabled

Fixed Symbol Length (Pharmacode) Serial Cmd: Default: Options:

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar widths,decode direction,fixed threshold value> 10 4 to 16

Minimum Number of Bars (Pharmacode) Serial Cmd: Default: Options:

5-20

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar widths,decode direction,fixed threshold value> 4 4 to 16

MS-890 Industrial Automation Scanner User’s Manual

Page 115

Symbologies

Bar Widths (Pharmacode) Serial Cmd: Default: Options:

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars, bar widths,decode direction,fixed threshold value> Mixed 0 = Mixed 1 = Narrow 2 = Wide

Decode Direction (Pharmacode) Definition: Serial Cmd: Default:

Specifies the direction that a bar can be read. <K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar widths,decode direction,fixed threshold value> Forward

Options:

0 = Forward

1 = Reverse

Fixed Threshold Value (Pharmacode) Serial Cmd: Default:

<K477,status,fixed symbol length status,fixed symbol length,min. no. of bars,bar widths,decode direction,fixed threshold value> 400

Options:

1 to 65535

MS-890 Industrial Automation Scanner User’s Manual

5-21

Page 116

PDF417

PDF417 Note: Calibration does not function with PDF417 symbols. However, you can substitute a Code 128 symbol with the same mil size as your PDF417. When you do this, enable Code 128 and disable Raster scanning. After calibration, switch back to PDF417 and Raster scanning. Usage:

Definition:

Used in applications where a large amount of information (over 32 characters) needs to be encoded within a symbol, typically where the symbol is transported from one facility to another. For example, an automobile assembly line might use a single symbol with multiple fields of information that will be read at several stations along the way, without reference to a database. A two-dimensional, multi-row (3 to 90), continuous, variable length symbology that has high data capacity for storing up to 2700 numeric characters, 1,800 printable ASCII characters, or 1100 binary character per symbol. Each symbol character consists of 4 bars and 4 spaces in a 17-module structure.

PDF417 Status Serial Cmd:

<K476,status,raster sweeps before decode attempt,fixed symbol length status,fixed symbol length,decode at end of read>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Raster Sweeps Before Decode Attempt (PDF417) Usage:

Might be useful in dynamic applications where symbols are close together and need to be decoded more quickly than the default setting allows.

Definition:

Specifies the number of raster sweeps the scanner will execute on PDF symbols before decoding. A value of 0 effectively disables this feature, which allows the decoder to make a “best guess” as to when enough symbol words have been captured.

Serial Cmd:

<K476,status,raster sweeps before decode attempt,fixed symbol length status,fixed symbol length,decode at end of read>

Default:

0 (disabled)

Options:

0 to 4

5-22

MS-890 Industrial Automation Scanner User’s Manual

Page 117

Symbologies

Fixed Symbol Length Status (PDF417) Usage:

Used to increase data integrity by ensuring that only one symbol length will be accepted.

Definition:

When enabled, the PDF symbol must contain the same number of characters as the symbol length setting before it can be considered a good decode. The scanner will ignore any symbol not having the specified length.

Serial Cmd:

<K476,status,raster sweeps before decode attempt,fixed symbol length status,fixed symbol length,decode at end of read>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

Fixed Symbol Length (PDF417) Definition:

Specifies the exact number of characters the scanner will recognize.

Serial Cmd:

<K476,status,raster sweeps before decode attempt,fixed symbol length status, fixed symbol length,decode at end of read>

Default:

10

Options:

1 to 2710 Note: Fixed Symbol Length Status must be enabled for Fixed Symbol Length to take effect.

Decode at End of Read (PDF417) Definition:

Allows the user to specify that the decoding of the PDF417 symbol take place after the requirements of End of Read Cycle have been satisfied. This means that the scanner will accumulate information about the PDF417 candidate symbol but will not attempt to error-correct or translate the information until the end of the read cycle.

Serial Cmd:

<K476,status,raster sweeps before decode attempt,fixed symbol length status, fixed symbol length,decode at end of read>

Default:

Disabled

Options:

0 = Disabled 1 = Enabled

MS-890 Industrial Automation Scanner User’s Manual

5-23

Page 118

Narrow Margins/Symbology ID

Narrow Margins/Symbology ID Usage:

Used when the leading and trailing edges of the symbols are smaller than the standard margin or when other objects encroach into the margins.

Definition:

Allows the scanner to read symbols with quiet zones less than 8 times the width of the narrow bar element. The “quiet zone” is the space at the leading and trailing ends of a symbol. Each quiet zone can be as narrow as only five times the width of the narrow bar element when Narrow Margins is enabled.

Serial Cmd:

<K450,narrow margins status,symbology identifier status>

Default:

Disabled

Options:

0 = Disabled

1 = Enabled

Note: Do not use Narrow Margins with Large Intercharacter Gap enabled in Code 39 or Codabar.

Symbology ID Usage:

Used when the symbology type and how it’s decoded needs to be known.

Definition:

Symbology ID is an AIM standard prefix set of characters that identify the symbol type. When enabled, the scanner analyzes and identifies the symbology and adds a three character identifying prefix to the data: 1. ] (close bracket character) indicating the presence of a symbology ID 2. A, C, E, I, L, Q, b, d, p, P, M A = Code 39, C = Code 128 or EAN-128, E = UPC/EAN, I = I–2/5, F = Codabar; G = Code 93, L = PDF417, a = Pharmacode, e = Composite/RSS, L = PDF417 and Micro PDF417 3. Modifier (see Explanation of Modifiers below)

Serial Cmd:

<K450,narrow margins,symbology identifier status>

Default:

Disabled

Options:

0 = Disabled

5-24

1 = Enabled

MS-890 Industrial Automation Scanner User’s Manual

Page 119

Symbologies

Explanation of Modifiers for Code 39, Codabar, and Interleaved 2/5 • For Code 39, Codabar and I–2/5, the modifier indicates Check Character and Check Character Output Status. • For Code 39 only, Full ASCII needs to be enabled to see modifiers 4, 5, and 7. Modifier

Check Character

0 1 3 4 5 7

Disabled Enabled Enabled Disabled Enabled Enabled

Check Character Output N/A Enabled Disabled N/A Enabled Disabled

Full ASCII conversion performed (Code 39 only) No No No Yes Yes Yes

Example:]A5 indicates a Code 39 symbol with Check Character and Check Character Output Status enabled and Full ASCII conversion performed.

Explanation of Modifiers for Other Symbologies • For Code 128, a 1 indicates EAN-128; otherwise the modifier is a 0. • For all other codes, the modifier is 0.

MS-890 Industrial Automation Scanner User’s Manual

5-25

Page 120

Background Color

Background Color Usage:

Serial Cmd: Default:

If the background is darker than the symbol, then enable black background. Typically the background is white; but on PCBs for example, they can be black. Allows the user to choose which symbol background (white or black) the scanner can read. <K451,background color> White

Options:

0 = White

Definition:

5-26

1 = Black

MS-890 Industrial Automation Scanner User’s Manual