For the Microscan MS-7100 and Microscan MS-7180

Page 91

Calibration

Other Menu Parameters The Calibration Setup menu allows you to monitor read rate and other parameters and to adjust internal scanner functions by software commands to improve read rate performance. It is particularly useful for adjusting to poorer quality labels or labels placed on the fringes or beyond the recommended read ranges. The Calibration Setup menu is launched from the terminal keyboardwith a <t> command (see Chapter 5, “Operational Commands”) to the scanner.

xx

80

Figure 4-1 Calibration Setup Menu

Motor Speed Pot, Gain Pot, and Tracking Pot adjustments control potentiometers (variable resistor circuits) that are incremented or decremented by pressing the indicated key one stroke at a time to change slowly, or by holding down the key to change quickly. The status of these three parameters is displayed on the bottom line of the menu under Gain, Motor, and Track. Bandwidth and Clock Speed use single commands (B and C respectively) to toggle between states. Big Bar and Small Bar measure bar widths and are not adjustable. Scan Rate is the number of scans per second and will change in response to changes in Motor Speed. To the immediate right of Scan Rate is the read rate, that is shown below as 66%. To the right of this is the current label data, in this example, 1234567890.

MS-7100/7180 Industrial Scanner User’s Manual

4-5

4–Calibration

Bandwidth and Clock status, along with Laser and Motor are shown on a single line in the bottom half of the menu as A_Bdwth, Clk, Laser, and Motor.

Page 92

Chapter 4

Calibration

80

Values displayed on the bottom line of the Calibration Setup menu, from Gain through Small Bar, are updated during operation, without refreshing the screen.

Recalling Factory Settings The original factory settings for the five parameters (Motor Speed Pot through Clock Speed) of the Calibration Setup menu can be restored by typing an upper case D,1 or by defaulting the scanner, recalling settings for both the calibration and the configuration menus.

Saving Current Settings Current Calibration settings can be saved to non-volatile RAM for power-up by typing Y upon exiting the Calibration Setup menu. To save changes when exiting the Calibration Setup menu, 1. Press the ESC key. You will see the following prompt on the bottom line:

2. Press Y to save current settings for power on or N to exit without saving. Note: Choosing Y will only save current settings that have been made in the Calibration menu and not affect the Configuration menu settings.

4–Calibration

Motor Speed Motor Speed directly effects scan rate (scans per second). The scanner is factory-set at the upper rate of 1000 scans per second, although it can be set to rates as low as 600. A slower scan speed may improve small bar acquisition (see “Small Bar” on page 4-9), and allow greater label range. However, before changing the motor speed, first determine if the number of scans for your application will be sufficient at the new speed. (See Step 5, “Calculate Number of Scans,” on page 1-6.)

1. Commands issued from inside the Calibration Setup menu do not require any brackets.

4-6

MS-7100/7180 Industrial Scanner User’s Manual

Page 93

Calibration To adjust Motor Speed: 1. Place the label to be used in your application at the desired range from a scanner in a fixed position. 2. From your terminal enter<t>. 3. From your terminal, enter a series of upper case Ns to decrease scan speed or upper case Ms to increase scan speed. This will change both Motor and Scan Rate values on the bottom line of the Calibration Setup menu.1 Motor values range from 1 to 80 (or higher) corresponding to 600 to 1000 scans per second (displayed under Scan Rate). 4. Recalculate the number of scans. 5. When the desired number of scans is reached, do “Autocalibration” starting with 5 on page 4-3.

Gain Adjusts the gain potentiometer for an optimum signal when testing a specific label and range. Gain is set during Autocalibration but can be user-adjusted if circumstances warrant. To manually optimize Gain, 1. Position label in read range and enter the Calibration Setup menu by typing <t>. 2. Observe the read rate on the right side of the bottom line of your screen.

Displayed values for Gain range from 0 to 255. Note: If electronic AGC is enabled (default), Gain may have little or no effect on read rate.

1. Scan Rate setting requires some time to stabilize after Motor adjustments are made. MS-7100/7180 Industrial Scanner User’s Manual

4-7

4–Calibration

3. Press G to increment or F to decrement Gain while observing the read rate and relative values displayed under Gain on the bottom line of the Calibration menu.

Page 94

Chapter 4

Calibration

Tracking Adjusts tracking potentiometer of analog-to-digital converter that affects the accuracy of the digitized signal. As with Gain, Tracking is set in Autocalibration but can be user-adjusted if circumstances warrant. To manually optimize Tracking, 1. Follow the same steps described for Gain except press T to increment and R to decrement. Values for Tracking can range from 1 to 255. 2. Observe the read rate and relative values displayed under “Track” on the bottom line of the Calibration menu.

Bandwidth Controls the range of frequencies (energy transitions) that are allowed to pass through bandpass filters during signal processing. Generally, Bandwidth should remain in the factory-set HIGH position for maximum resolution. However, with low density labels at close range or poorly printed labels, changing Bandwidth to LOW may improve readability by accepting a narrower range of frequencies. To change Bandwidth, 1. Place the label to be used in your application at the desired range from the scanner in a fixed position. 2. From your terminal send a <t> (Enter Calibration Setup command) to the scanner.

4–Calibration

3. Press B to toggle between LOW and HIGH. The resulting change will be displayed on the status line of the Calibration Setup menu. 4. Do “Autocalibration” starting with 5 on page 4-3.

Clock Speed Relates to internal decoder timing relative to individual bars and spaces. Note: Clock speed has no relationship to scan speed. Clock Speed is either FAST (factory set) or SLOW and displayed on the status line about one quarter distance from the bottom of the Calibration Setup menu. The default FAST Clock Speed allows better acquisition on narrower bar-width labels, or labels at greater distances from the scanner; a

4-8

MS-7100/7180 Industrial Scanner User’s Manual

Page 95

Calibration SLOW Clock Speed may in some instances improve acquisition on wider bar-width labels or labels at closer distances. To change Clock Speed, 1. Place the label to be used in your application at the desired range from the scanner in a fixed position. 2. From your terminal send a <t> (Enter Calibration Setup command) to the scanner. 3. Press C to toggle between SLOW and FAST. The resulting change will be displayed on the status line of the Calibration Setup menu. 4. Do “Autocalibration” starting with 5 on page 4-3.

Bar Size Measurements Measures relative bar sizes in terms of scan duration (clock cycles). Big Bar and Small Bar cannot be directly adjusted but will change in response to changes in narrow-bar-width type, range, and motor speed. They are updated continuously in response to sensor input.

Big Bar Measures large bar widths in logarithmic values corresponding to clock cycles. Maximum Big Bar value will never exceed 255 (1024 clock cycles).

Small Bar

If a consistent minimum read of 75 is displayed, try decreasing motor speed (see “Motor Speed” on page 4-6) or label range until the small bar display rises slightly above 75.

Scan Rate Scan Rate is the scans per second of the laser beam that the scanner produces in operation, with laser scanning enabled. The standard range for Scan Rate is 600 to 1000 scans per second. Changes to Scan Rate take place in response to changes in Motor Speed. (See “Motor Speed” on page 4-6 for information on changing Scan Rate.

MS-7100/7180 Industrial Scanner User’s Manual

4-9

4–Calibration

Measures small bar widths in logarithmic values corresponding to clock cycles. Minimum Small Bar value will not fall below 75 (approximately 8 clock cycles).

Page 96

Calibration

4–Calibration

Chapter 4

4-10

MS-7100/7180 Industrial Scanner User’s Manual

Page 97

Chapter

5

Operational Commands

Chapter Contents Summary of Operational Commands................................................... 5-2 Program Management Commands...................................................... 5-3 Device Control Commands .................................................................. 5-3 Code Type Commands ........................................................................ 5-4 Counter Commands............................................................................. 5-4 Test Commands................................................................................... 5-5 Status Commands ............................................................................... 5-5 Master Label Commands..................................................................... 5-5

This chapter describes the scanner’s serial operational commands and their functions. On-line serial operational commands are sent from the host to the scanner to carry out routine operations “on the fly” as distinguished from serial configuration commands that are generally used in initial setup.1 Operational commands are always preceded by a < left angle bracket symbol (unless redefined by Command Start Character command) and followed by a > right angle bracket symbol.2

MS-7100/7180 Industrial Scanner User’s Manual

5-1

5–Operations

1. The scanner will only recognize a <D> serial command (Enter Menu Configuration Program) from an auxiliary terminal. 2. Command start character by default is a left angle bracket, <. It may be redefined by menu or serial command. However, the end character, a right angle bracket, >, cannot be changed.

Page 98

Chapter 5

Operational Commands

Summary of Operational Commands Table 5-1 Operational Commands

Command Type

Program Management

Command

Result

<A> <D> <Z> <t>

Software Reset (does not save for power-on) Enter Configuration Mode Save Configuration for Power-on Enter Calibration Setup menu Autocalibration initiation Autocalibration factory default Save autocalibration Echo Data to Aux Monitor Enable Laser Scanning Disable Laser Scanning Host Relay Driver Autodiscriminate All Codes Enable Code 39 Only* Enable Codabar Only* Enable I 2 Of 5 Only* Noread Counter Noread Counter Reset Trigger Counter Trigger Counter Reset Match Code (or Good Read) Counter Match Code Counter Reset Mismatch Counter Mismatch Counter Reset Enter Read Rate Test Enter Scan Rate Test Exit Read Rate or Scan Rate Test Display Software Part Number Display Checksum of EPROM Enable Match Code Option* Disable Match Code Option* Store Next Label Scanned as Master Label Download Master Label Information Request Master Label Information Delete Master Label Information

<KT> <KTD> <KTZ>

Device Control

Code Type

Counter

Test

5–Operations

Status

Master Label

<Bdata> <H> <I> <L> <P> <Q> <R> <S> <N> <O> <T> <U> <V> <W> <X> <Y> <C> <m> <J> <#> <!> <E> <F> <G> <)XXXX)> <)> <))>

* Can also be accomplished in configuration menu and serial configuration command.

5-2

MS-7100/7180 Industrial Scanner User’s Manual

Page 99

Operational Commands

Program Management Commands <A> Software Reset. Initializes all serial <A> Initialize configuration commands in RAM and resets all counters and operating parameters. RAM Changes for Baud Rate, Stop Bits, Code ROM NOVRAM Length, etc. do not take effect until this command is sent. Note: Software Reset will cause the numeric counters in use to lose their count; record all data that you wish to save prior to sending this command. <D> Enter Configuration Menu. Enters the menu configuration program. (Can be entered from the auxiliary port.) <Z> Save Configuration for Power-on. Initialize <Z> Saves the current configuration to nonvolaRAM tile memory for availability on power-on. The values of numeric counters are not ROM NOVRAM saved by this command. Note: The <Z> Save Configuration command can be executed 10,000 times. In normal usage this will exceed the life of the decoder. If frequent changes to the operating parameters are required, the <Z> command should be used only when the current configuration has been changed and the changes are to be permanent. <t> Calibration (Setup Menu). Allows you to monitor the scanner’s optical parameters, optimize performance, and make custom adjustments for special applications. (Can be entered from the auxiliary port.)1 <KT> Autocalibration Initiate. Initiates autocalibration changes and internal scanner settings. (See <KTD> Autocalibration Default. Resets autocalibration and motor speed settings to factory defaults. <KTZ> Autocalibration Save. Saves the settings changed in autocalibration for power up.

Device Control Commands

<H> Enable Laser Scanning. Normal Operation. 1. The <D> and <t> commands are the only commands that the MS-7100/7180 scanner will recognize from an auxiliary terminal. MS-7100/7180 Industrial Scanner User’s Manual

5-3

5–Operations

<Bdata> Echo Data to Aux Monitor. Echoes data from host to auxiliary monitor (see Appendix I, “Auxiliary Monitor,” on page A-16 for more detail).

Page 100

Chapter 5

Operational Commands <I> Disable Laser Scanning. This feature is useful during extended periods of time when no bar code labels are being scanned. Disabling laser scanning will not affect any downloaded commands to the scanner. <L> Host Relay Driver. Allows you to send a pulse (at any time regardless of Match Code or Relay Settings status) to internal relay contacts to make or break a connection between pins 7 and 9 of both the power and trigger connectors.

Code Type Commands <P> Autodiscriminate All Codes. Enables the scanner to decode all available bar code types without changing scanner configuration settings Note: For maximum scanning speed, enable only those bar code symbologies used in the application. <Q> Enable Code 39 Only. Allows only Code 39 labels to be read. <R> Enable Codabar Only. Allows only Codabar labels to be read. <S> Enable I 2 of 5 Only. Allows only Interleaved 2 of 5 labels to be read.

Counter Commands The Xs in all counter commands denote a numeric value from 00000 to 65,535. After reaching the maximum numeric limit of 65,535, you will receive an error message and the counter will automatically rollover and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65,536 per each rollover (the decoder does not keep track of the number of rollovers) to the current count. Note: You will lose all counter values if power to the scanner is cycled, an <A> command is sent, or upon entering and exiting Configuration or Calibration Setup menus. Note: If you activate the counter command during a read cycle, the scanner will not output the count until the read cycle ends. <N> Noread Counter. The message N/XXXXX displays the total number of noreads that have occurred since power-on or the last Noread Counter Reset command. <O> Noread Counter Reset. Sets Noread Counter to 00000.

5–Operations

<T> Trigger Counter. The message T/XXXXX displays the total number of triggers since power-on or the last Trigger Counter Reset command. <U> Trigger Counter Reset. Sets the trigger counter to 00000. <V> Match Counter (or Good Read Counter). The message V/XXXXX displays the total number of good reads matching the master label since power-on 5-4

MS-7100/7180 Industrial Scanner User’s Manual