The text of a manual · part 6 of 11

MS-610 Scanner User's Manual, pages 51 to 60

Pages 51 to 60 of 102

The original document is the PDF (723 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

For the Microscan MS-610 and Microscan MS-615

Page 51

User Outputs Menu

Beeper Volume Default: Level 4 Options: Level 4, Level 5, Level 1, Level 2, Level 3

Beeper Speed

Full Screens Default: Enabled Options: Enabled, Disabled Allows you to display either the full menu screen or just the command line. When Full Screens is disabled, only the command line will be displayed.

New Master Pin Default: Disabled Options: Disabled, Enabled Note: Match Code (page 2-17) and a triggered mode (page 2-11) must be enabled for New Master Pin to take effect. Allows you to momentarily connect pin 5 to ground (pin 10) in order to clear any existing master label information from memory and allows the next label to be read as the new master label information. Note: Even if it is enabled, New Master Pin will not load the new master label information if RS-422, RS-485 or RTS/CTS is enabled because it uses the CTS pin.

Laser On/Off Default: Disabled Options: Disabled, Enabled Note: A serial or external trigger (see “Triggering Mode” on page 2-11) must be enabled for Laser On/Off to take effect. When enabled, the laser is ON only during the read cycle. When disabled, the laser operates continuously. Note: Laser On/Off does not relate to the <H> (Enable Laser Scanning) or <I> (Disable Laser Scanning) operational commands on page 4-3.

MS-610 Scanner User’s Manual

2-27

2–Menu Config.

Default: Fast Options: Fast, Slow Allows you to set the beeper to accommodate your application. The beeper takes approximately 80 ms to sound when Slow is enabled. If your application speed is faster than 80 ms (approximately), enable Fast.

Page 52

Chapter 2

Menu Configuration

Clock Speed

2–Menu Config.

Default: Slow Options: Slow, Fast Provides a means to set internal clock speed according to the label range and bar code label density. Generally, the slow setting is used with lower density labels (20 mil narrow-bar-width or wider) and with labels placed closer to the scanner; while the fast setting is used with higher density labels (about 5 mil narrow-bar-width) and with labels placed further from the scanner.

Relay Driver Default: Good Read Options: Good Read, Noread, Match, Mismatch, Mismatch or Noread Note: If Relay Driver is set to Mismatch or Noread, or Mismatch, a relay pulse will not output unless Match Code (page 2-17) is enabled and a master label is downloaded into memory (page 2-27). Allows you to set the conditions that will output a relay pulse of 5V to pin 7. To see your options for enabling Match Code and downloading a master label into memory, see figure 2-6, “ Match Code Logic Diagram ,” on page 2-18. If Match Code (page 2-17) is enabled, Relay Driver must be set to an option other than Good Read. If Relay Driver is set to Good Read, there will be no output to the host or a relay pulse. Selecting:

2-28

Has this effect:

Good Read

Sends a pulse when the label’s data is accurately scanned and decoded.

Noread

Sends a pulse when the label’s data is not decoded before the end of the read cycle.

Match

Sends a pulse when the decoded label matches the master label. (If Match Code is disabled, a good read will have the same effect.)

Mismatch

Sends a pulse when the label’s data does not match that of the master label.

Mismatch or Noread

Sends a pulse when the label's data does not match that of the master label or the label is not decoded before the end of the read cycle.

MS-610 Scanner User’s Manual

Page 53

User Outputs Menu

Relay Polarity Default: Positive Options: Positive, Negative Allows you to choose between positive and negative output signals at pin 7.

Relay Pulse Width (in 10 ms incs)

Command Start Character Default: < Options: Any ASCII character Allows you to define a new ASCII start character in a serial command.

MS-610 Scanner User’s Manual

2-29

2–Menu Config.

Default: 5 (.05 seconds). Corresponds to 50 ms displayed in the menu. Options: 0 to 255 (0 to 2.55 seconds). Divide the number entered on the command line by 100 for time in seconds. Allows you to set the duration of the good match/noread output signals at pin 7.

Page 54 read from a scan by OCR; it may hold reading errors

Chapter 2 Menu Configuration

2 = c je} O =] 7 o + N

2-30 MS-610 Scanner User’s Manual

Page 55

Chapter

3

Serial Configuration Chapter Contents

Most of the configuration changes to the scanner that can be made in the menu can also be accomplished by command strings from the host. Serial command strings are entered from an ASCII terminal. As with menu configuration commands, serial configuration commands relate to the initial scanner setup. To quickly reference the serial configuration commands, see table 3-1 on page 3-2.

MS-610 Scanner User’s Manual

3-1

3–Serial Config.

Summary of Serial Configuration Commands...................................... 3-2 Serial Configuration Command Format ............................................... 3-3 Concatenating Serial Commands ........................................................ 3-4 Loss of Communications...................................................................... 3-4 Trigger Filter Timing Value................................................................... 3-4 Communications Commands............................................................... 3-5 Protocol................................................................................................ 3-5 Operations Commands ........................................................................ 3-8 Code Types Commands .................................................................... 3-10 User Outputs Commands .................................................................. 3-12

Page 56

The text of this page cannot be shown here; see it in the PDF.

Page 57

Table 3-1 Summary of Serial Configuration Commands (continued)

Function

User Outputs

Parameter

Command

Format

Ku

Beeper

<Kustatus,volume,speed>

KN

New Master Pin

<KNstatus>

KC

Laser On/Off

<KCstatus>

Kz

Clock Speed

<Kzspeed>

Kv

Relay Driver

<Kvstatus>

Kw

Relay Polarity

<Kwstatus>

Kx

Relay Pulse Width

<Kxnumber>

KE

Command Start Character

<KEASCII character>

Serial Configuration Command Format The format for a serial configuration command is, <Kparameterdata,data,...etc.><initializing command>

When using serial configuration commands, note also the following conventions: • Parameters and data are “case sensitive.” That is, characters must be entered as upper or lower case, as specified. • All data fields (except the last) must be followed by a comma (without a space). • If there is no change in a given field, then commas can be entered alone, or with the existing data (for example, <Ka,,,0> or <Ka4,1,0,0>).

1. Command start character by default is a left angle bracket, <. It may be redefined by menu or serial command. However, the end character, a right angle bracket, >, cannot be changed. 2. See “Initializing serial configuration commands” on page 3-4 for definitions and examples.

MS-610 Scanner User’s Manual

3-3

3–Serial Config.

Where: • Less than < and greater than > symbols are included as part of the commands. 1 • “parameter,” as used here, are those character(s) that precede the data. • The “initializing command” <A> or <Z> is sent after configuration is complete. <Z> resets and saves for power up. <A> initializes the change to RAM.2 (For more information, see page 4-3.) For example, the following command enables UPC and saves the change for power-on: <Ks1><Z>.

Page 58

Chapter 3

Serial Configuration • All fields preceding the modified field must be included. For example, in Host Port, to change Data Bits to Eight without changing any other field, enter either: <Ka,,,1> or <Ka4,1,0,1>. • All fields following the modified field can be left out. For example, in the RS-232 port, to change Baud Rate to 4800, enter <Ka3>. (See examples on following pages.)

Initializing serial configuration commands To ensure that a serial configuration command will take effect, you need to follow it with one of the operational commands below: <A> To reset but not save changes for power-on <Z> To reset and save changes for power-on For example, to change Baud Rate and reset without saving changes for power-up, enter <Ka3><A>. To change Baud Rate and reset, saving the changes to NOVRAM, enter <Ka3><Z>.

3–Serial Config.

Concatenating Serial Commands Commands can be concatenated (added together) to a maximum of 64 characters in a single string or data block. Additional data blocks of 64 or less characters can be sent provided there is at least a 10 ms pause between blocks. For example, <Kc1><Kh1><Ko1><A> enables LRC, sets End of Read Cycle mode to New Trigger, enables Narrow Margins and resets the data buffers (without saving the changes for power-on).

Loss of Communications Making changes to communications parameters such as Baud Rate, Parity, Stop Bits, LRC, etc. without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, default the scanner. See Appendix D, “Defaulting the Scanner ,” on page A-8.

Trigger Filter Timing Value Format: <KF?> Returns the current trigger filter timing value in milliseconds. Format: <KFone tenth trigger filter timing value > Allows you to set trigger filter timing. Divide the desired number of milliseconds by 10 and enter the quotient.

3-4

MS-610 Scanner User’s Manual

Page 59

Communications Commands

Communications Commands Protocol If selecting one of the options form 0 to 4 (Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, or Polling Mode D), use this format: Format: <Kfprotocol> protocol: 0 = Point-to-Point 1 = Point-to-Point with RTS/CTS 2 = Point-to-Point with XON/XOFF 3 = Point-to-Point with RTS/CTS & XON/XOFF

4 = Polling Mode D 5 = Multidrop (requires address) 6 = User Defined 7 = User Defined Multidrop

Example: To change the Protocol to Polling Mode D, enter <Kf4>.

Format: <Kf5protocol,address> protocol: 5 = Multidrop address: Any number from 1 to 50

Example: To change the Protocol to Multidrop with an address of 33, enter <Kf5,33>. If selecting User Defined (6) or User Defined Multidrop (7), complete the format by either choosing new parameters or concatenating unchanged data fields (separate by commas). Tip: For User Defined, first select Point-to-Point <Kf0> and then User Defined <Kf6...>. For user Defined Multidrop, first select Multidrop <Kf5>, then User Defined Multidrop <Kf7...>.

MS-610 Scanner User’s Manual

3-5

3–Serial Config.

If selecting Multidrop (5), you must define an address and append it to the command string.

Page 60

Chapter 3

Serial Configuration Format: <Kfprotocol,RES,address,REQ,EOT,STX,ETX,ACK,NAK> Protocol: 6 = User Defined 7 = User Defined Multidrop Note: Address, can be assigned any ASCII character except a null. Control characters are used to define RES through NAK (except Address). Table 3-2 lists the control characters used for these data fields. (Refer to Table A-5, “ASCII Table with Control Characters,” on page A-5 for more information.)

Example: To select an unpolled ACK/NAK User Defined protocol with LRC disabled, enter <Kf0><Kf6,,,,,,,^F,^U><Kc0>.1 ACK and NAK will be displayed in the menu. 2 Table 3-2 Protocol Commands

Protocol Command

Control Characters

3–Serial Config.

(Mnemonic displayed on (Entered in menu or serial Microscan menu) command) RES ^D

Effect of Command Reset

REQ

^E

Request

EOT

^D

Reset

STX

^B

Start of Text

ETX

^C

End of Text

ACK

^F

Acknowledge

NAK

^U

Negative Acknowledge

Preamble Format: <Kdstatus,preamble character(s) > status: 0 = Disabled 1 = Enabled preamble character(s): Enter one or two preamble characters except a null (00H). Default is ^M.

Example: To enable Preamble with just one character, an FF (form feed), enter <Kd1,^L>.

1. <Kf0> nulls the address and <Kc0> disables LRC. 2. A control character, although conventionally represented here and in the ASCII table on page A-5 as two characters (^F or ^U, etc.), is actually a single ASCII character that is entered on the keyboard by holding down the control key while pressing the desired letter.

3-6

MS-610 Scanner User’s Manual