The text of a manual · part 4 of 11
MS-610 Scanner User's Manual, pages 31 to 40
Pages 31 to 40 of 102
The original document is the PDF (723 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Microscan MS-610 and Microscan MS-615
Page 31
Communications Menu Selecting:
Has this effect:
Multidropa
Used only with RS-232 or RS-422. ASCII characters Note: A specific ASCII can be assigned as an address and as protocol character must not be commands (RES, REQ, EOT, STX, ETX, ACK, and NAK). User Defined is necessary when a new protocol assigned more than must be defined to match a specific host protocol. once. When User Defined is selected, the displayed protocol commands match those of the previously selected protocol. User Defined is considered to be in a polled mode only if an address has been assigned. The address can be any ASCII character from the ASCII in appendix B, except NUL. b
User Defined
User Defined Multidrop
Used when connecting to a concentrator or other device that does not match standard Multidrop protocol. Note: A specific ASCII Any single character (01 hex to 7E hex) in the ASCII table character must not be can be assigned as the address character. The character assigned more than chosen is used as the poll character and the subsequent once. ASCII character becomes the select character. For example, if a ^A (01 hex) is selected as the address, ^B (02 hex) becomes the select address that the host will use in sending host select commands. (See Table A-7, “Multidrop Address Characters,” on page A-18.)
a. Once the scanner is configured for Multidrop, a profile card, a terminal connected to the auxiliary RS-232 pins, or a default procedure must be used to access the configuration menus again (although serial commands will continue to function). b. For example a simple ACK/NAK protocol can be developed by first selecting Point-to-Point, then User Defined, and then assigning characters to ACK and NAK commands. First scroll to the following command: HOST PROTOCOL --> PROTOCOL --> USER DEFINED --> ACK = --> Enter a ^F by holding down the Control key while pressing the F key, and then press CR to see the following: HOST PROTOCOL --> PROTOCOL --> USER DEFINED --> ACK = ^F The mnemonics ACK and NAK replace the default NULs in the menu.
MS-610 Scanner User’s Manual
2-7
2–Menu Config.
Similar to Polling Mode D except that a unique poll address and select address are required for each Note: Scanners intended to link up to a multidrop device, and only one host port connection is needed for up to 50 devices. (For Multidrop poll and Microscan MS-5000 select characters, see Table A-7, “Multidrop Address multidrop concentrator can only be config- Characters,” on page A-18.) ured in standard Requires a concentrator or controller using RS-485 Multidrop protocol. communications. When Multidrop is selected, the protocol characters for RES, REQ, etc. are assigned automatically. (See “Appendix I, Multidrop Communications” on page A-15 for poll and select sequences.)
Page 32
Chapter 2
Menu Configuration
Preamble ^M (and a null). Corresponds to <CR><NUL> (carriage return/null) displayed in the menu. Options: Any ASCII character, including control characters. Control characters entered on the command line are displayed in the menu as mnemonic characters. See “ Defining Special Characters ” on page 2-4 and Table A-5, “ASCII Table with Control Characters ,” on page A-5. Allows you to define a one or two character data string that can be added to the front of the decoded data. For example, a carriage return and line feed would display each decoded message on its own line.
2–Menu Config.
Default:
Preamble (enable/disable) Default: Disabled Options: Disabled, Enabled (within any protocol) Allows you to enable or disable the preamble character(s).
Postamble Default:
^M^J. Corresponds to <CR><LF> (carriage return/line feed) displayed in the menu. Options: Any ASCII character, including control characters. Control characters entered on the command line are displayed in the menu as mnemonic characters. See “ Defining Special Characters ” on page 2-4 and Table A-5, “ASCII Table with Control Characters ,” on page A-5. Allows you to define a one or two character data string that can be added after the decoded message.
Postamble (enable/disable) Default: Disabled Options: Disabled, Enabled (within any protocol) Allows you to enable or disable the Postamble character(s).
Longitudinal Redundancy Check Default: Disabled Options: Disabled, Enabled An error-checking routine that verifies the accuracy of transmissions. It is the exclusive OR of all characters following the SOM (start of message) up to and including the EOM (end of message). 2-8
MS-610 Scanner User’s Manual
Page 33
Communications Menu
Response Timeout Default: 12 ms Options: 0 to 65,000 ms. A zero (0) will cause an indefinite wait. Allows you to set the time the scanner will wait before timing out if ACK, NAK, and ETX are enabled, and a host response is expected.
Intercharacter Delay
Host Port Allows you to set parameters for RS-232 communication whenever an RS232 protocol has been enabled.
Baud Rate Default: 9600 Options: 9600, 19.2K, 300, 600, 1200, 2400, 4800 Allows you to set the number of bits transmitted per second.
Parity Default: Even Options: Even, Odd, None Allows you to select an error detection routine in which one data bit in each character is set to 1 or 0 so that the total number of 1 bits in the data field is even or odd.
Stop Bits Default: One Options: One, Two Allows you to select the last one or two bits in each character to indicate the end of the character.
MS-610 Scanner User’s Manual
2-9
2–Menu Config.
Default: 0. Corresponds to 0 ms displayed in the menu. Options: 0 to 255. Zero (0) causes no delay between characters. Allows you to set the time interval in milliseconds between individual characters transmitted from the scanner to the host. A high setting will significantly slow down communications. For example, a 200 setting will result in a 1/5 second delay between each character that is transmitted.
Page 34
Chapter 2
Menu Configuration
Data Bits Default: Seven Options: Seven, Eight Allows you to establish the total number of bits in each character.
RS-422 2–Menu Config.
Default: Disabled Options: Disabled, Enabled Used only in Point-to-Point protocol, and not with RTS or CTS. Whenever RS-422 is disabled, RS-232 is enabled in the background. However, an exception occurs whenever Multidrop has been selected. In this case the functioning protocol will be RS-485 regardless of the displayed status of RS-422 in the menus. See also Appendix H, “ Interface Standards ,” on page A-14.
2-10
MS-610 Scanner User’s Manual
Page 35
Operations Menu
Operations Menu The Operations menu allows you to set the operations parameters for the scanner. To help visualize the menu’s organization and to locate the page number where each topic is described, see figure 2-3. – Bold text represents default settings.
Operations
Triggering Mode – Continuous
Read
(page 2-14)
End of Read Cycle – Timeout
– Continuous
Read 1 Output
– External Level – External Edge
– New Trigger – Timeout &
New Trigger
(page 2-14)
Timeout in 10 ms incs * – 100
– User
Definable (ASCII char.)
(page 2-15)
Serial Trigger Character – ^]
– User
Definable (ASCII char.)
– Serial Data
– Serial Data &
Edge
(page 2-15)
(page 2-16)
– Positive
– NOREAD
External Trigger Level
– Negative
(page 2-16)
When to Output
– As Soon As
Possible – End Of Read Cycle
Noread Message 7 digits
(page 2-16)
(page 2-16)
– Enabled
– Enabled
Noread Message
– Disabled
Bar Code Output
– Disabled
– User Definable
(ASCII char.)
(page 2-17)
Good Decode Reads – 1
(page 2-17)
Match Code – Disabled – Enabled
– User Definable
(1 to 31)
* Divide the number entered on the command line by 100 for time in seconds.
Figure 2-3 Operations Menu Structure
Triggering Mode Default: Continuous Read Options: Continuous Read, Continuous Read 1 Output, External Level, External Edge, Serial Data, Serial Data & External Edge MS-610 Scanner User’s Manual
2-11
2–Menu Config.
(page 2-11)
Page 36
Chapter 2
Menu Configuration
2–Menu Config.
Allows you to establish the type of trigger event that will initiate or end the read cycle. (See “End of Read Cycle” on page 2-14.) Selecting:
Has this effect:
Continuous Read
Trigger input options are disabled and the scanner is always in the read cycle. Bar code data is decoded, and label information is transmitted repeatedly, as long as the label is in the read range of the scanner. When to Output options have no affect on Continuous Read. Continuous Read is useful in testing label or scanner functions. Note: If both Continuous Read and Match Code are enabled the scanner defaults to Continuous Read 1 Output mode.
2-12
Continuous Read 1 Output
Label data is immediately transmitted once every time new label data is placed in front of the scanner. With Timeout enabled for End of Read Cycle, if the label doesn't change, the scanner will repeat the output at the end of each subsequent timeout period. For example, if Timeout were set to one second, the scanner would output the label data immediately, and then repeat the output at intervals of one second, for as long as the label continued to be scanned. With Timeout disabled (that is, End of Read Cycle set to New Trigger), the scanner will output the current label data immediately, but output it only once. A new label appearing at any time in the scan range will produce a new read output as long as the new label is not identical to the previous label.
External Level
Allows a read cycle to be initiated by a trigger signal from an object detector when an object appears within the detector’s range. The read cycle exists as long as the detector “sees” the object and ends when the object moves out of the detector’s range (figure 2-4 on page 2-13). a
External Edge
As with Level, Edge allows a read cycle to be initiated by a trigger signal from an object detector when it detects the appearance of an object (rising edge). But unlike Level mode, the removal of an object (falling edge) does not end the read cycle. With Edge enabled, the read cycle ends with a good read output, a timeout, or a new trigger (figure 2-5 on page 2-13).
Serial Data
The scanner accepts an ASCII character from the host or controlling device as a trigger to start a read cycle. Serial data trigger behaves the same as External Edge.
Serial Data & External Edge
The scanner accepts either an external trigger or a serial ASCII command to start a read cycle. MS-610 Scanner User’s Manual
Page 37
Operations Menu a. Level and Edge apply to the active logic state (positive or negative) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object.
2–Menu Config.
First trigger An object, moving in front of the detector beam, causes the first trigger (rising edge trigger) signal to be sent to the scanner initiating the read cycle.
1
Object detector
1 Second trigger The same object, moving out of the detector's beam, causes a second trigger (falling edge trigger) signal to be sent to the scanner ending the read cycle.
Object detector
Figure 2-4 External Level Trigger Signals
1
First rising edge trigger Object #1, moving in front of the detector’s beam, causes the first trigger (rising edge trigger) signal to be sent to the scanner, initiating the read cycle.
Object detector
Subsequent rising edge trigger 2
Object detector
Object #2, moving in front of the detector's beam, causes a second trigger (rising edge trigger) signal to be sent to the scanner. This signal initiates a new read cycle and ends the previous read cycle provided that New Trigger or Timeout and New Trigger is enabled and a good read or timeout has not already occurred.
Figure 2-5 External Edge Trigger Signals MS-610 Scanner User’s Manual
2-13
Page 38
Chapter 2
Menu Configuration
End of Read Cycle
2–Menu Config.
Default: Timeout Options: Timeout, New Trigger, Timeout & New Trigger Allows you to choose the circumstances that will end the read cycle. The read cycle is the time during which the scanner will receive and process label data. When the Triggering Mode option is set in an External or Serial mode of operation, the trigger event initiates the read cycle. Note: When operating in Continuous Read or Continuous Read 1 Output, the scanner is always in the read cycle. Selecting:
2-14
Has this effect:
Timeout
Can end the read cycle after a specified period of time, and if no label has been read, causes a noread message, if enabled, to be transmitted. With either External Edge, Serial Data, or Serial Data & Edge enabled, a timeout ends the read cycle. With External Level enabled, the read cycle does not end until the falling edge trigger occurs, and the next read cycle does not begin until the next rising edge trigger. With Continuous Read 1 Output enabled, a timeout initiates a new read cycle and allows the same label to be read again.
New Trigger
Ends the read cycle at the occurrence of a new trigger event, and if no label has been read, causes a noread message, if enabled, to be transmitted at the occurrence of the new trigger event. With either External Edge, Serial Data, or Serial Data & Edge enabled, an edge or serial trigger ends a read cycle and initiates the next read cycle. With External Level enabled, a falling edge trigger ends a read cycle. However, the next read cycle does not begin until the occurrence of the next rising edge trigger.
Timeout & New Trigger
Ends the read cycle after a specified period of time or at the occurrence of new trigger event, and if no label has been read, causes a noread message, if enabled, to be transmitted. With either External Edge, Serial Data, or Serial Data & Edge enabled, a timeout, or an edge or serial trigger, whichever comes first, ends the read cycle. With External Level enabled, the read cycle does not end until the occurrence of a falling edge, and the next read cycle does not begin until the next rising edge trigger. MS-610 Scanner User’s Manual
Page 39
Operations Menu
Timeout (in 10 ms incs) Default:
100 (one second). Corresponds to 1000 ms displayed in the menu. Options:0 to 65535. Divide the number entered on the command line by 100 for time in seconds. Note: Timeout or Timeout & New Trigger under End of Read Cycle (page 2-14) must be enabled for Timeout (in 10 ms incs) to take effect. Allows you to define the duration of the timeout period. Default: ^]. Corresponds to <GS> displayed in the menu. Options: Any single ASCII character , including control characters, except NUL (00H), an existing host command character, 1 or an on-line protocol character. Control characters entered on the command line are displayed in the menu as mnemonic characters. Note: See “Defining Special Characters” on page 2-4 and Table A-5, “ASCII Table with Control Characters,” on page A-5. Note: Serial Data (page 2-12) or Serial Data & Edge (page 2-12) must be enabled for Serial Trigger Character to take effect. “N/A” is displayed in the menu when all other triggering modes are enabled. \Allows you to define a single ASCII character as the host serial trigger character that initiates the read cycle. The serial trigger is considered an online host command and requires the same command format as all host commands (that is, to be entered within the < > brackets).
External Trigger Level Default: Positive Options: Positive, Negative Note: External Level (page 2-12), External Edge (page 2-12), or Serial Data & Edge (page 2-12) must be enabled for External Trigger Level to take effect. “N/ A” is displayed in the menu when all other triggering modes are enabled. Allows you to determine whether a positive or negative transition will initiate the read cycle. Note: If using the Microscan object detector (P/N 99-440001-03), use positive trigger level. 1. For example, assigning an upper case D would nullify the <D> (Enter Menu Configuration) command. For a list of operational commands used by the scanner, see Chapter 4, “Operational Commands.” MS-610 Scanner User’s Manual
2-15
2–Menu Config.
Serial Trigger Character
Page 40
Chapter 2
Menu Configuration
Noread Message Default: NOREAD Options: Up to seven ASCII characters (except NUL) Allows you to define any combination of ASCII characters (except NUL) up to seven characters as the noread message. When enabled, the noread message will be transmitted to the host at a timeout or the end of a read cycle if the bar code label has not been decoded.
2–Menu Config.
Noread Message (enable/disable) Default: Enabled Options: Enabled, Disabled Note: If Noread Output is enabled, the noread message will only output if Bar Code Output (page 2-16) is also enabled. Allows you to enable or disable the noread message.
Bar Code Output Default: Enabled Options: Enabled, Disabled Allows you to choose whether or not to send label data (or noread messages) to the host. When disabled, a label is decoded and the read cycle transpires as usual, but neither label data nor the noread message is transmitted to the host. All scanner counters are updated, and the number of good reads or noreads can be obtained via operational commands.
When to Output (bar code data) Default: As Soon As Possible Options: As Soon As Possible, End of Read Cycle Allows you to choose when bar code data is sent to the host.
2-16
Selecting:
Has this effect:
As Soon As Possible
Causes bar code data (good reads) to be transmitted immediately upon a good decode.
End of Read Cycle
Causes bar code data output to be delayed until the end of the read cycle.
MS-610 Scanner User’s Manual