Page 21
Install Scanner
The MS-610 scanner can be mounted directly to a mounting surface of your choice, or indirectly, via a mounting plate. To permanently mount the scanner: 1. Position the scanner in a dry place, devoid of sunlight, bright lights, or laser light from other sources. b. Before mounting, ensure sufficient clearance at the rear of the scanner to allow for the 15-pin connector and cable. c. If not using the mounting plate, use the measurements provided in figure 1-9 to locate centers of mounting holes and drill four 5/32 inch (4 mm) holes, or If using the mounting plate (see appendix D), use the four perimeter holes to mount the plate and mount scanner directly to the four inner holes on the plate. d. Secure scanner with four 6-32 screws. 2.25" (57.15 mm) 1.40" (35.56 mm)
.425" (10.8 mm)
1.0" (25.4 mm)
2.29" (74.17 mm) 1.50" (38.10 mm)
Caution: Do not allow the mounting screws to penetrate into the scanner case more than 0.175 in. (4.4 mm) or damage to the scanner may result. Caution: To prevent twisting or distortion of the scanner housing, ensure that the mounting surface or bracket is flat.
6-32 threads (4 places) Front
Figure 1-9 Bottom Mounting Holes (not full size)
MS-610 Scanner User’s Manual
1-13
1–Setup and Inst.
9 Install Scanner
Page 22
1–Setup and Inst.
Chapter 1
Setup and Installation
10 Position Object Detector
If you are not using an object detector, continue with step 11. In a typical operation, a scanner will wait for bar code data only during a triggered read cycle. A read cycle is initiated by a “trigger” that can be in the form of a serial command from the host (internal trigger) or a signal from an object detector (external trigger). When an object detector is used, it is set up so that its beam will bounce off the approaching object and the resulting pulse will be sent to the scanner to begin the read cycle. Typically, a detector is positioned so that it will detect the presence of an object before the object’s label can be scanned by the scanner. An object detector is mounted in almost any position relative to the object as long as (1) the object passes within range of the detector and (2) direct or reflected light from the detector does not interfere with the scanner’s reception. As the item continues to move down the line, its label moves into the scanner beam and is read by the scanner. Figure 1-10 shows a picket fence label being scanned with the detector and scanner at right angles to each other. Figure 1-11 shows a label in the ladder direction with the object detector on the same side as the scanner, but positioned so as not to cause interference.
Label
Detector Scan beam
Direction of label travel
MS-610 Scanner
Figure 1-10 Detector Right-Angle Orientation Direction of label travel Detector
MS-610 Scanner
Alternative detector position Label
Figure 1-11 Detector Side Orientation
1-14
MS-610 Scanner User’s Manual
Page 23
Operational Tips
Do: • Check inputs (label speed, length, height, etc.) to ensure the desired number of scans per label. • Avoid excessive tilt, pitch, and skew of the bar code label. • Check the label for readability by doing a read rate test. If there is any question about the label's readability, contact your Microscan representative. • After changing any parameter that might affect read rate, repeat read rate test. • Clean the scanner window with a clean, dry Q-tip or cotton cloth on a regular basis. Do Not: • Aim the scanner into direct light or sunlight. • Aim the scanner into an external object detector or other light-emitting device. • Obstruct the scanner window with mounting hardware or other objects. • Connect chassis of scanner and host to different ground potentials (see Ground and Shield Considerations ” on page 1-16). • Operate the scanner in excessive temperature environments (see “ Environment” on page A-2).
MS-610 Scanner User’s Manual
1-15
1–Setup and Inst.
Operational Tips
Page 24
1–Setup and Inst.
Chapter 1
Setup and Installation
Ground and Shield Considerations Proper grounding is necessary for operator safety, noise reduction, and the protection of equipment from voltage transients. Buildings, including any steelwork, all circuits, and all junction boxes must be grounded directly to an earth ground in compliance with local and national electrical codes. RS-232 signals have a common signal ground (pin 4). Pin 4 is normally connected to pin 13 (chassis ground) in the scanner; however, under certain conditions (e.g., when potential differences exist between power outlet grounds) signal and chassis grounds can be isolated from each other inside the scanner by Microscan technicians.
Noise Interference Noise interference can be minimized if cabling subject to HOST noise interference is twisted and/or Shielding shielded or encased Communications Cable in grounded conduit, and the conduit or Earth Ground shielding (“drain” line) is grounded only to earth ground SCANNER at the host, as shown in figure Figure 1-12 Grounding Diagram, Scanner-Host 1-12. You might need to examine and if necessary cut the shielding connection at or near the scanner cable connector.
Ground Loops Ground loops, signal degradation due to different ground potentials in communicating devices, can be eliminated or minimized by ensuring that both the host, scanner, and their power supplies are connected to a common earth ground.
1-16
MS-610 Scanner User’s Manual
Page 25
Chapter
2
Menu Configuration Entering the Menu Configuration Program ......................................... 2-2 Using the Menu Configuration Program .............................................. 2-2 Saving Menu Changes ........................................................................ 2-4 Loss of Communications ..................................................................... 2-4 Defining Special Characters ............................................................... 2-4 Communications Menu ........................................................................ 2-5 Operations Menu ............................................................................... 2-11 Code Types Menu.............................................................................. 2-19 User Outputs Menu............................................................................ 2-26
This chapter describes how to configure the MS-610 scanner with on screen menu commands from a host or an auxiliary terminal. All keystrokes are in bold typeface. Default parameters in the menu structures are also in bold typeface.
Communicating with an ASCII Terminal The MS-610 scanner communicates in full duplex, terminal mode with no handshake. It also recognizes carriage returns and line feeds. The host or ASCII terminal must match the following default settings before any communication can take place: 9600 Baud Rate, Seven Data Bits, Even Parity, and One Stop Bit. A PC computer can be used as an ASCII terminal if connected as shown under “Host Connector” and running a communications program set to the above defaults. See your computer manual for communication’s port pinouts.
MS-610 Scanner User’s Manual
2-1
2–Menu Config.
Chapter Contents
Page 26
Chapter 2
Menu Configuration
Entering the Menu Configuration Program To see the Main menu (figure 2-1), from an ASCII terminal that is connected to the scanner, enter the operational command <D> (enter the < > brackets as well as the upper case D).1
2–Menu Config.
MICROSCAN SYSTEMS, INC. CONFIGURATION PROGRAM MAIN MENU 35-216001-xx
1) 2) 3) 4)
TOPICS
DESCRIPTIONS
COMMUNICATIONS OPERATIONS CODE TYPES USER OUTPUTS
HOST PROTOCOL, HOST PORT TRIGGERING, TIMEOUTS, ETC. CODE SELECTIONS, CODE LENGTHS, ETC. BEEPER, DATA LINE POLARITY, ETC.
ESC = MAIN MENU OR EXIT M = PREVIOUS MENU B = PREVIOUS ITEM
N = NEXT ITEM SP = NEXT ITEM CR = THIS ITEM
MAIN- - > COMMUNICATIONS
Figure 2-1 Configuration Program - Main Menu
Using the Menu Configuration Program The bottom line on the screen is called the command line. The command line identifies your place in the menu program, shows current status and allows you to review and change options. Use the designated keys 2 to scroll to and select the parameter you wish to change; press SP (space bar key) or N to scroll ahead, B to scroll back, CR (carriage return key) to select, and M to return to the previous higher level menu. To return to the Main menu at any time, press ESC (escape key) and M.
1. Command start character by default is a left angle bracket, <. It may be redefined by menu or serial command. However, the end character, a right angle bracket, >, cannot be changed. 2. The menu navigational keys are displayed in each menu.
2-2
MS-610 Scanner User’s Manual
Page 27
For example, to enable LRC (see figure 2-2, “ Communications Menu Structure,” on page 2-5 and “Longitudinal Redundancy Check ” on page 2-8), you would use the following command line path: From the Main menu, press CR at the Communications prompt (this is the first prompt displayed in the Menu Configuration Program) to access the Communications menu.
Communications, > Host Protocol Communications, > Host Port
Since LRC is a subtopic of Host Protocol, press CR to access the Host Protocol parameters.
Host Protocol, > Protocol Host Protocol, > Preamble = ^M Host Protocol, > Preamble = Disabled Host Protocol, > Postamble = ^M^J Host Protocol, > Postamble = Disabled Host Protocol, > LRC = Disabled Host Protocol, > Response Timeout in 1ms incs = 12 Host Protocol, > Intercharacter Delay in 1ms incs = 0 Host Protocol, > LRC Host Protocol, > LRC
= Disabled = Enabled
Protocol is the first parameter under Host Protocol. Press SP until you reach LRC, then press CR.
To enable LRC, press CR, SP, and CR.
To view LRC’s new status in the menu, press M to refresh the screen. To return to the Main menu, press M again. You can make additional changes within another menu before exiting the program. Simply follow the same method of scrolling to and selecting each main topic, then its subtopics, until you reach the parameter you want to change. Remember, to return to the Main menu at any time, press ESC (escape key) and M. Some parameters are user defined, in which case they prompt you with an arrow for data, such as:
At the prompt, redefine the parameter within the allowable range, and press CR to enable.
MS-610 Scanner User’s Manual
2-3
2–Menu Config.
Main, > Communications Main, > Operations Main, > Code Types Main, > User Outputs
Page 28
Chapter 2
Menu Configuration
Saving Menu Changes Press ESC (escape key) to see the following on the command line: Press M to return to the Main menu, or press E to exit the Menu Configuration program. If E is pressed, the following question will appear:
2–Menu Config.
Press N to exit without saving changes, or press Y to retain the current settings for power-on. If Y is selected, a beep will indicate the save has been carried out.
Loss of Communications Making changes to communications parameters such as Baud Rate, Parity, Stop Bits, LRC, etc. without corresponding changes in linked device(s) can result in the loss of menu access. If this should occur, default the scanner. See Appendix D, “Defaulting the Scanner ,” on page A-8.
Defining Special Characters To define any control character from the ASCII table: Press SP once, then enter the control character by holding down the control key and simultaneously pressing the desired character. For example to define a line feed, press SP, then Control and J simultaneously. It is displayed as ^J on the command line and as <LF> in the menu when the screen is refreshed. To define CR as a character: Press SP, then CR. It is displayed as ^M on the command line and as <CR> in the menu when the screen is refreshed. To define a space as a character : Press SP twice. It is displayed as a blank space in the menu when the screen is refreshed. While it appears that nothing has been assigned, the hex value 20 will be sent during data transmission. To select NUL as the character: Press SP, then a 0 (zero). It is displayed as <NUL> in the menu when the screen is refreshed.
2-4
MS-610 Scanner User’s Manual
Page 29
Communications Menu
Communications Menu The Communications menu allows you to set the communication protocols between the scanner and the host. Communications
– Bold text represents default settings.
Host Protocol (page 2-6)
(page 2-8)
(page 2-8)
Preamble
– Point-to-Point – Point-to-Point
(page 2-8)
Preamble
– ^M
Postamble
– Disabled
– User
– ^M ^J
– Enabled w/RTS/CTS Definable (ASCII char.) – Point-to-Point w/XON/XOFF – Point-to-Point w/RTS/CTS & XON/XOFF – Polling Mode D – Multidrop – Address = 1 – User Definable – User Defined
– User
Definable (ASCII
– RES – REQ – STX – Address = ^A – EOT – User Defined Multidrop
– ACK – ETX
– NAK
– RES – REQ – STX – Address = ^A – EOT
– ACK – ETX
– NAK
–
–
(page 2-8)
LRC
– Disabled – Enabled
(page 2-8)
Postamble – Disabled – Enabled
(page 2-9)
(page 2-9)
Response Timeout
Interchar Delay – 0
– 12 ms
– User Definable
– User Definable (0 to 255)
Host Port (page 2-9)
Baud Rate – 9600
– 600
– 19.2K – 1200 – 300
– 4800
(page 2-9)
(page 2-9)
(page 2-10)
(page 2-10)
– Even
– One
– Seven
– Disabled
Parity
– Odd
– None
Stop Bits – Two
Data Bits – Eight
RS-422
– Enabled
Figure 2-2 Communications Menu Structure
Note: Be sure that all new settings match your system; otherwise you risk losing communications with the configuration terminal when you exit the menu program (whether or not changes are saved for power-on).
MS-610 Scanner User’s Manual
2-5
2–Menu Config.
Protocol
Page 30
Chapter 2
Menu Configuration
Host Protocol Protocol
2–Menu Config.
Default: Point-to-Point Options: Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS & XON/XOFF, Polling Mode D, User Defined, User Defined Multidrop Protocols define the sequence and format in which information is transferred between devices.
2-6
Selecting:
Has this effect:
Point-to-Point
Has no address and sends data to the host (RS-232) whenever it is available and without any request or handshake from the host.
Point-to-Point with RTS/CTS (Requestto-Send/Clear-toSend)
Used only with RS-232. This is a simple handshaking protocol that allows a device to initiate data transfers to the host with an RTS (request-to-send) transmission. The host, when ready, responds with a CTS (clear-tosend) and the data is transmitted. CTS and RTS signals are transmitted over two dedicated wires (pins 6 and 10) as defined in the RS-232 standard.
Point-to-Point with XON/XOFF (Transmitter On/Off)
Used only with RS-232. This selection enables the host to send a single byte transmission command of start (XON) or stop (XOFF). If an XOFF has been received from the host, data will not be sent to the host until the host sends an XON. During the XOFF phase, the host is free to carry on other chores and accept data from other devices.
Point-to-Point with RTS/CTS & XON/ XOFF
Used only with RS-232. It is a combination of Point-toPoint with RTS/CTS and Point-to-Point with XON/ XOFF.
Polling Mode D
Like Point-to-Point, Polling Mode D requires a separate channel to the host but unlike Point-to-Point, it requires a device to have an address and to wait for a poll from the host before sending data. When in Polling Mode D, an address 1 is automatically displayed on the configuration screen. However, during transmission, a 1C hex poll address (FS) and a 1D hex select address (GS) are substituted for the 1.
MS-610 Scanner User’s Manual