Page 61

Read Cycle

No Read Message Note: A No Read message can only be transmitted when in Single symbol mode and no decodes occur. Usage: Definition:

Used in applications where the host needs serial verification that a symbol has not been read. When enabled, and if no symbol has been decoded before timeout or the end of the read cycle, the No Read message will be transmitted to the host.

No Read Status Serial Cmd: Default: Options:

<K714,status,No Read message> Enabled 0 = Disabled 1 = Enabled

No Read Message Definition: Serial Cmd: Default: Options:

Any combination of ASCII characters can be defined as the No Read message. <K714,status,No Read message> No Read 1 to 20 ASCII characters.

Note: A No Read message is not output when the reader is disabled (when an <I> command has been sent).

MS-2 CCD Reader User’s Manual

4-15

Page 62

Reader Setup

Reader Setup Reader Setup allows shutter speed to be controlled automatically, set to a specific value, or defined within a fixed range.

Shutter Speed Important: Shutter Speed settings should not be changed unless they are being increased to accomodate a fast line speed in a dynamic application. Usage: Definition: Serial Cmd: Default: Options:

Faster shutter speeds may be necessary in faster-moving applications. Slower shutter speeds are useful in slower, lower-contrast applications. Shutter speed is the amount of time that the sensor is exposed to light. <K522,shutter speed,minimum speed,maximum speed> 0 (Automatic) 70 to 65535

Minimum Speed Usage: Definition: Serial Cmd: Default: Options:

Useful in applications with variable speeds, variable symbol sizes, and variable distances. The slowest shutter speed and longest amount of time that the sensor is exposed to light. <K522,shutter speed,minimum speed,maximum speed> 70 70 to 65535

Maximum Speed Usage: Definition: Serial Cmd: Default: Options:

4-16

Useful in applications with variable speeds, symbol sizes, and distances. The fastest shutter speed and shortest amount of time that the sensor is exposed to light. <K522,shutter speed,minimum speed,maximum speed> 65535 70 to 65535

MS-2 CCD Reader User’s Manual

Page 63

5 Symbologies Contents Symbologies by ESP.................................................................................................................... 5-2 Symbologies Serial Commands ................................................................................................... 5-3 Code 39........................................................................................................................................ 5-4 Codabar........................................................................................................................................ 5-6 Interleaved 2 of 5.......................................................................................................................... 5-8 Code 128/EAN-128 .................................................................................................................... 5-10 Code 93...................................................................................................................................... 5-11 GS1 DataBar (RSS) ................................................................................................................... 5-12 MSI Code ................................................................................................................................... 5-13 Plessey Code ............................................................................................................................. 5-14 Industrial 2 of 5........................................................................................................................... 5-15 UPC/EAN ................................................................................................................................... 5-17 Symbology Identifier................................................................................................................... 5-20

This section describes the various symbol types that can be decoded by the MS-2 CCD Reader. See the following sites for additional symbology information: http://www.aimglobal.org/standards/aimpubs.asp http://barcodes.gs1us.org/dnn_bcec/Default.aspx?tabid=82

MS-2 CCD Reader User’s Manual

5-1

Page 64

Symbologies by ESP

Symbologies by ESP

Click this button to bring up the App Mode view.

To open nested options, single-click the +.

Click this button to bring up the Symbologies tree control. To change a setting, double-click the setting and use your cursor to scroll through the options.

5-2

MS-2 CCD Reader User’s Manual

Page 65

Symbologies

Symbologies Serial Commands Symbology Identifier Code 39 Codabar Interleaved 2 of 5 Code 128/EAN 128 Code 93 DataBar-14 (RSS-14) DataBar Limited (RSS Limited) DataBar Expanded (RSS Expanded) MSI Code Plessey Code Industrial 2 of 5 UPC/EAN

MS-2 CCD Reader User’s Manual

<K450,0,symbology identifier status> <K470,status,check character status,check character output status, 0,0,minimum symbol length,full ASCII set> <K471,status,0,start/stop match output status,0,symbol length status, symbol length,check character status,check character output status> <K472,status,check character status,check character output status, symbol length #1,symbol length #2> <K474,status,0,minimum symbol length,1,1,0,application record separator character> <K475,status,0,minimum symbol length> <K482,status> <K483,status> <K484,status> <K486,status,check character mode,check character output status, 0,minimum symbol length> <K487,status,check character output status,0,minimum symbol length> <K488,status,check character status,check character output status, format,symbol length #1,symbol length #2> <K489,UPC-A status,UPC-E status,EAN-8 status,EAN-13 status, supplementals required,supplemental 2 status,supplemental 5 status, UPC-A as EAN-13,UPC-E as UPC-A,EAN-8 as EAN-13>

5-3

Page 66

Code 39

Code 39 Usage: Definition: Serial Cmd: Default: Options:

Code 39 is the most widely used non-retail 1D symbology. An alphanumeric symbology with unique start/stop code patterns, composed of 9 black and white elements per character, 3 of which are wide. <K470,status,check character status,check character output status,0,0, minimum symbol length,full ASCII set> Enabled 0 = Disabled 1 = Enabled

Check Character Status (Code 39) Definition: Serial Cmd: Default: Options:

When enabled, the reader will verify a Modulus 10 check character in order to identify the decode as successful. <K470,status,check character status,check character output status, 0,0,minimum symbol length,full ASCII set> Disabled 0 = Disabled 1 = Enabled

Check Character Output Status (Code 39) Usage: Definition:

Serial Cmd: Default: Options:

5-4

Check Character Output Status, when added to the symbol, provides additional data security. When enabled, the check character is read and output along with the symbol data. When disabled, symbol data is output without the check character. Note: With Check Character Output Status and an External or Serial trigger option enabled, an invalid check character calculation will cause a No Read message to be transmitted at the end of the read cycle. <K470,status,check character status,check character output status, 0,0,minimum symbol length,full ASCII set> Disabled 0 = Disabled 1 = Enabled

MS-2 CCD Reader User’s Manual

Page 67

Symbologies

Minimum Symbol Length (Code 39) Usage: Definition:

Serial Cmd: Default: Options:

Minimum Symbol Length helps prevent truncations and increases data integrity by ensuring that only one symbol length will be accepted. Specifies the exact number of characters that the reader will recognize (this does not include start and stop characters and check characters). The reader ignores any symbology that is less than the specified length. <K470,status,check character status,check character output status,0,0,minimum symbol length,full ASCII set> 0 0 to 32

Full ASCII Set (Code 39) Usage:

Definition:

Serial Cmd: Default: Options:

Must be enabled when reading characters outside the standard character set (0-9, A-Z, etc.) The user must know in advance whether or not to use the Full ASCII Set option. Since Full ASCII Set requires two code words to encode one character, it is less efficient. Standard Code 39 encodes 43 characters; zero through nine, capital “A” through capital “Z”, minus symbol, plus symbol, forward slash, space, decimal point, dollar sign, and percent symbol. When Full ASCII Set is enabled, the reader can read the full ASCII character set, from 0 to 255. <K470,status,check character status,check character output status,0,0,minimum symbol length,full ASCII set> Disabled 0 = Disabled 1 = Enabled

MS-2 CCD Reader User’s Manual

5-5

Page 68

Codabar

Codabar Usage: Definition: Serial Cmd: Default: Options:

Used in photo-finishing and library applications. Previously used in medical applications, but not typically used in newer medical applications. Codabar is a 16-bit character set (0 through 9, and the characters $, :, /, ., +, and –) with start/stop codes and at least two distinctly different bar widths. <K471,status,0,start/stop output status,0,symbol length status,symbol length,check character status,check character output> Disabled 0 = Disabled 1 = Enabled

Start/Stop Output Status (Codabar) Definition:

Serial Cmd: Default: Options:

When enabled, the start and stop characters will be present in the data output of the decoded symbol. When disabled, the start and stop characters will not be present in the data output of the decoded symbol. Note: Because the start and stop characters are included as part of the data, the characters must be included as part of the length in a fixed length mode of operation. <K471,status,0,start/stop output status,0,symbol length status,symbol length,check character status,check character output> Enabled 0 = Disabled 1 = Enabled

Symbol Length Status (Codabar) Definition:

Serial Cmd: Default: Options:

5-6

When set to Any/Minimum, any symbol length is considered valid. When set to Fixed, the reader will check the symbol length against the Symbol Length Status parameter. <K471,status,0,start/stop output status,0,symbol length status,symbol length,check character status,check character output> Any/Minimum 0 = Any/Minimum 1 = Fixed

MS-2 CCD Reader User’s Manual

Page 69

Symbologies

Symbol Length (Codabar) Definition:

Serial Cmd: Default: Options:

Specifies the number of characters that the reader will recognize, including start and stop check characters. The Note: When setting Symbol Length to any value other than 0, add 2 to account for reserved spaces. For example, if your symbol has 3 characters, set Symbol Length to 5. Note: If Symbol Length is set to 0 and Symbol Length Status is set to Any/ Minimum, all symbol lengths are decoded. <K471,status,0,start/stop output status,0,symbol length status,symbol length,check character status,check character output> 6 0 to 32

Check Character Status (Codabar) Definition:

Serial Cmd: Default: Options:

When disabled, the reader will not perform any character checking calculations on decoded Codabar symbols. When set to Modulus 16, the reader will perform a Modulus 16 check character calculation on the symbol. If the symbol does not pass this calculation, it will not be decoded. <K471,status,0,start/stop output status,0,symbol length status,symbol length,check character status,check character output> Disabled 0 = Disabled 1 = Enabled (Modulus 16)

Check Character Output (Codabar) Definition:

Serial Cmd: Default: Options:

When this field is disabled and a check character calculation is enabled, the reader will strip the verified check character from the symbol data output. This condition must be accounted for if a minimum length is also being used. When enabled, the reader will output the check character as part of the symbol data. This condition must be accounted for if a minimum length is also being used. <K471,status,0,start/stop output status,0,symbol length status,symbol length,check character status,check character output> Disabled 0 = Disabled 1 = Enabled

MS-2 CCD Reader User’s Manual

5-7

Page 70

Interleaved 2 of 5

Interleaved 2 of 5 Usage:

Definition:

Serial Cmd: Default: Options:

Interleaved 2 of 5 has been popular because it is the most dense symbology for printing numeric characters less than 10 characters in length; however, Microscan does not recommend this symbology for any new applications because of inherent problems such as truncation. A dense, contimuous, self-checking, numeric symbology. Characters are paired together so that each character has five elements, two wide and three narrow, representing numbers 0 through 9, with the bars representing the first character and the interleaved spaces representing the second character. (A check character is highly recommended). <K472,status,check character status,check character output status,symbol length #1,symbol length #2> Disabled 0 = Disabled 1 = Enabled

Check Character Status (Interleaved 2 of 5) Usage: Definition: Serial Cmd: Default: Options:

This option is not typically used, but it can be enabled for additional security in applications where the host requires redundant check character verification. An error correcting routine in which the check character is added. <K472,status,check character status,check character output status,symbol length #1,symbol length #2> Disabled 0 = Disabled 1 = Enabled

Check Character Output Status (Interleaved 2 of 5) Definition: Serial Cmd: Default: Options:

5-8

When enabled, a check character is sent along with the symbol data for added data security. <K472,status,check character status,check character output status,symbol length #1,symbol length #2> Disabled 0 = Disabled 1 = Enabled

MS-2 CCD Reader User’s Manual