For the Microscan MS-2
Page 51
Read Cycle
Number of Symbols Usage: Definition: Conditions:
Commonly used in shipping applications where an object contains individual symbols for part number, quantity, and other values. This feature allows one trigger event to decode all symbols present. The following conditions apply: • When Number of Symbols is set to Single, the read cycle ends when one symbol is decoded and output. • When Number of Symbols is set to Multiple, all decoded symbol data (for both the same and different symbols) is output while the read cycle is active and/or no timeout has occurred. • If set to Multiple and Time Between Identical Decodes is less than the decode time (typically about 4ms), the same symbol data can be output multiple times.
Number of Symbols Definition: Serial Cmd: Default: Options:
Number of Symbols is the number of different symbols that can be read in a single read cycle. <K222,number of symbols> 1 0 = Multiple 1 = Single
MS-2 CCD Reader User’s Manual
4-5
Page 52
Time Between Identical Decodes
Time Between Identical Decodes Usage:
Definition: Serial Cmd: Default: Options:
4-6
Can be used to prevent repetition of data output. Also useful for emulating Multiple symbol operations. When used in conjunction with Number of Symbols set to Multiple, Time Between Identical Decodes can be set so that more than one symbol can be read within a read cycle. Determines the length of time that the reader will wait before it outputs the same symbol’s data a second time. <K503,time between identical decodes> 0 0 to 2550ms
MS-2 CCD Reader User’s Manual
Page 53
Read Cycle
Trigger Mode Definition: Serial Cmd: Default: Options:
The Trigger is the event that initiates a read cycle. <K200,trigger mode> Continuous Read 0 = Continuous Read 2 = External Level 3 = External Edge 4 = Serial Data 5 = Serial Data and Edge
Continuous Read Usage: Definition:
Serial Cmd:
Continuous Read is useful in testing symbol readability or reader functions. It is not recommended for normal operations. In Continuous Read, trigger input options are disabled and the reader is always in the read cycle. If a single symbol stays within read range for multiple read cycles, its data will be transmitted repeatedly until it leaves the read range. The reader sends replies to serial commands that require responses when symbol data is transmitted. Note: No Read options have no affect on Continuous Read. <K200,0>
External Level Initiate Read Cycle: Object #1, moving in front of the detector beam, causes a change in the trigger state, which initiates the read cycle. End Read Cycle: The same object, moving out of the detector beam, causes another change in the trigger state, which ends the read cycle.
Usage:
Definition:
Serial Cmd:
This mode is effective in an application where the speeds of the conveying apparatus are variable and the time the imager spends reading each object is not predictable. It also allows the user to determine if a No Read has occurred. External Level allows the read cycle (active state) to begin when a trigger (change of state) from an external sensing device is received. The read cycle persists until the object moves out of the sensor range and the active trigger state changes again. <K200,2>
MS-2 CCD Reader User’s Manual
4-7
Page 54
Trigger Mode
External Edge Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, which initiates the read cycle. Initiate Second Read Cycle: Object # 2, moving in front of the detector beam, causes another change in the trigger state. This signal initiates a new read cycle and ends the previous read cycle unless Timeout is enabled and a good read or timeout has not occured.
Usage: Definition:
Serial Cmd:
This mode is highly recommended in any application where conveying speed is constant, or if spacing, object size, or read cycle timeouts are consistent. External Edge, as with Level, allows the read cycle (active state) to begin when a trigger (change of state) from an external sensing device is received. However, the passing of an object out of sensor range does not end the read cycle. The read cycle ends with a good read output, or, depending on the End of Read Cycle setting, a timeout or new trigger occurs. <K200,3>
Serial Data Usage:
Serial Data is effective in a highly controlled environment where the host knows precisely when the object is in the field of view. It is also useful in determining if a No Read has occurred. Definition: In Serial Data, the reader accepts an ASCII character from the host or controlling device as a trigger to start a read cycle. A Serial Data trigger behaves the same as an External Edge trigger. Serial commands are entered inside angle brackets, as shown here: <t>. Serial Cmd: <K200,4> Note: In Serial Data, sending a non-delimited start serial character will start a read cycle; however, a non-delimited stop serial character has no effect.
Serial Data and Edge Usage:
Serial Data and Edge is seldom used but can be useful in an application that primarily uses an external sensing device but occasionally needs to be triggered manually. An auxiliary terminal can be connected to the auxiliary port so the user can send the serial trigger character through the reader to the host. Definition: In this mode the reader accepts either a serial ASCII character or an external trigger pulse to start the read cycle. Serial Cmd: <K200,5> Note: In Serial Data, sending a non-delimited start serial character will start a read cycle; however, a non-delimited stop serial character has no effect. 4-8
MS-2 CCD Reader User’s Manual
Page 55
Read Cycle
Continuous Read 1 Output Emulation Usage:
Definition:
Continuous Read 1 Output can be useful in applications where it is not feasible to use a trigger and all succeeding symbols contain different information. It is also effective in applications where the objects are presented by hand. The MS-2 does not offer this feature, but it can be emulated by following the steps described below. In Continuous Read 1 Output Emulation, the trigger must be active (grounded) for as long as the MS-2 needs to be scanning symbols. The MS-2 will only output symbol data once while that symbol remains in the MS-2’s field of view. If the symbol is removed and then presented again, it must be removed from the field of view for the amount of time specified in the Time Between Identical Decodes value before the MS-2 will decode the symbol again. However, if a new symbol is presented, the MS-2 will decode it and output symbol data immediately.
Setting Up Continuous Read 1 Output Emulation 1. Set Number of Symbols to Multiple.
Serial Cmd:
<K222,0>
2. Set Time Between Identical Decodes to 500 (ms).
Serial Cmd:
<K503,500>
3. Set Trigger Mode to External Level.
Serial Cmd:
<K200,2>
Important: Continuous Read 1 Output Emulation is not recommended for automated environments, because there is typically no reliable way to verify whether or not a symbol was missed.
MS-2 CCD Reader User’s Manual
4-9
Page 56
Serial Trigger Character (Delimited)
Serial Trigger Character (Delimited) Usage: Definition:
Serial Cmd: Default: Options:
Allows the user to define the trigger character and delimiters that start and stop the read cycle. A serial trigger is considered an online host command and requires the same command format as all host commands. It must be entered within angle bracket delimiters < > or, in the case of non-delimited triggers, it must define individual start and stop characters. <K201,serial trigger character> ^] (<GS>) Any single ASCII character, including control characters, except NUL (00 in hex), an existing host command character, or an on-line protocol character. Control characters entered on the command line are displayed in the menu as mnemonic characters. Note: Serial Data or Serial Data and Edge triggering mode must be enabled for Serial Trigger Character to take effect.
4-10
MS-2 CCD Reader User’s Manual
Page 57
Read Cycle
Start Trigger Character (Non-Delimited) Usage: Definition:
Serial Cmd: Default: Options:
Useful in applications where different characters are required to start a read cycle. A single ASCII host serial trigger character that starts the read cycle and is not enclosed by delimiters such as < and >. Non-delimited Start characters can be defined and will function according to the trigger event. When defining Start trigger characters, the following rules apply: • In External Edge the reader looks only for the Start trigger character and ignores any Stop trigger character that may be defined. • In External Level the Start trigger character begins the read cycle and the Stop trigger character ends it. Note that even after a symbol has been decoded and the symbol data transmitted, the reader remains in External Level trigger read cycle until a Stop character is received. • In Serial Data and Edge trigger mode, either a Start trigger character or a hardware trigger can start an edge trigger read cycle. <K229,start trigger character> NUL (00 in hex) (disabled) Two hex digits representing any ASCII character except XON and XOFF.
MS-2 CCD Reader User’s Manual
4-11
Page 58
Stop Trigger Character (Non-Delimited)
Stop Trigger Character (Non-Delimited) Usage: Definition:
Serial Cmd: Default: Options:
4-12
Useful in applications where different characters are required to end a read cycle. A single ASCII host serial trigger character that ends the read cycle and is not enclosed by delimiters such as < and >. Non-delimited Stop characters can be defined and will function according to the trigger event. When defining Stop trigger characters, the following rules apply: • In External Edge the reader looks only for the Start trigger character and ignores any Stop trigger character that may be defined. • In External Level the Start trigger character begins the read cycle and the Stop trigger character ends it. Note that even after a symbol has been decoded and the symbol data transmitted, the reader remains in External Level trigger read cycle until a Stop character is received. • In Serial Data and Edge trigger mode, either a Start trigger character or a hardware trigger can start an edge trigger read cycle. <K230,stop trigger character> NUL (00 in hex) (disabled) Two hex digits representing any ASCII character except XON and XOFF.
MS-2 CCD Reader User’s Manual
Page 59
Read Cycle
No Decode Timeout Usage: Definition: Serial Cmd: Default: Options:
Useful in tightly-controlled applications that require a read cycle to end before the next object appears. The duration of the read cycle as measured from either the beginning of the read cycle or from the last decode. <K220,0,seconds> Note: The first parameter is always a zero (0) and does not change. 2 1 - 4095
The MS-2 times out if no symbol is decoded during the No Decode Timeout period. If a symbol is decoded before the timeout expires, the timeout period starts again. No Decode Timeout only applies to Edge, Serial Data, and Serial Data and Edge trigger modes. In Single symbol mode (<K222,1>), No Decode Timeout is the maximum amount of time the reader will attempt to decode after a trigger is activated. In Multiple symbol mode (<K222,0>), No Decode Timeout is the time in which there are no decodes - not the time since the trigger was activated. That is, if the reader continues to decode symbols presented to it, it will not exit the read cycle unless the time between decodes exceeds the time set in No Decode Timeout. For Level trigger mode, the read cycle ends when the trigger falls, or when a symbol is decoded and output in Single symbol mode.
MS-2 CCD Reader User’s Manual
4-13
Page 60
Decodes Before Output
Decodes Before Output Note: When setting up, determine if the reader’s scan rate is capable of scanning your longest symbol the required number of times. Usage: Definition: Serial Cmd: Default: Options:
Useful for increasing the reliability of decodes for symbologies that do not have internal error checking. The number of decodes required per symbol before its data is sent. Requires the reader to decode a symbol successfully a designated number of times before data is output. <K221,decodes before output> 0 0 to 10 (0 = “Auto-adaptive”)
Note: If the reader is in Single symbol mode and it doesn’t achieve the number of required decodes during the read cycle, a No Read message will be output. Note: Higher Decodes Before Output settings will decrease the throughput rate.
4-14
MS-2 CCD Reader User’s Manual