Page 211

I/O Parameters

EZ Trax Output When EZ Trax output is enabled, the imager will attempt to output all enabled EZ Trax options. EZ Trax will not function unless enabled.

Comm Port Selects the communication port that will be used to transfer image files.

Image Mode

Determines which read cycle condition triggers image file output.

Image Format

Selects the format of the image output file.

MINI Hawk High Performance Imager User Manual

7-43

Page 212

Configuring EZ Trax Output

JPEG Quality Sets the amount of JPEG image compression. A value of 1 represents the highest compression and lowest image quality, and 100 represents the lowest compression and highest image quality.

Object Info Output

If a symbol is decoded during an image capture, the symbol type and other supplementary symbol information is also included.

Grade Output

Symbol quality information will be output in the form of an abbreviated identifier for each read cycle.

7-44

MINI Hawk High Performance Imager User Manual

Page 213

I/O Parameters

Configuring EZ Trax Output by Serial Command Usage: Definition: Serial Cmd: Default: Options:

Useful for when you want to configure EZ Trax output using serial commands instead of the EZ Trax graphic interface. When EZ Trax output is enabled, the imager will attempt to output all enabled EZ Trax options. EZ Trax will not function unless enabled. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Disabled 0 = Disabled 1 = Enabled

Comm Port Definition: Serial Cmd: Default: Options:

Selects the communication port that will be used to transfer image files. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Host 0 = Host 1 = Aux 2 = USB

Host Port Sends output using the current Host Port.

Aux Port Sends output using the Auxiliary Port.

USB Sends output using a USB connection.

Image Mode Definition: Serial Cmd: Default: Options:

Determines which read cycle condition triggers image file output. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Disabled 0 = Disabled 1 = Good Read 2 = No Read 3 = No Read and Good Read 4 = Mismatch

Good Read Outputs the first Good Read image in the read cycle. The image file will immediately follow the symbol data. MINI Hawk High Performance Imager User Manual

7-45

Page 214

Configuring EZ Trax Output

No Read Outputs the first No Read image in the read cycle. The image file will immediately follow the symbol data. Note: If two symbols are present in the same image capture and one of them is successfully decoded, no image will be output.

No Read and Good Read Outputs the first No Read or Good Read image in the read cycle.

Mismatch Outputs the image of the first mismatch condition.

Image Format Definition: Serial Cmd: Default: Options:

Selects the format of the image output file. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> Compressed 0 = Lossless 1 = Compressed

Lossless A “lossless” image is one that is in a high resolution format, and that will not lose resolution even when manipulated and saved multiple times. When Lossless is enabled, the JPEG Quality setting has no effect.

Compressed Outputs the image in JPEG format. Image quality is determined by the JPEG Quality setting.

JPEG Quality Definition: Serial Cmd: Default: Options:

7-46

Sets the amount of JPEG image compression. A value of 1 represents the highest compression and lowest image quality, and 100 represents the lowest compression and highest image quality. <K757,status,comm port,image mode,image format,JPEG quality,object info output, grade output> 90 1 to 100 (percent)

MINI Hawk High Performance Imager User Manual

Page 215

I/O Parameters

Object Info Output Definition: Serial Cmd: Default: Options:

If a symbol is decoded during an image capture, the symbol type and other supplementary symbol information is also included. <K757,comm port,image mode,image format,JPEG quality,object info output, grade output> Disabled 0 = Disabled 1 = Enabled

Grade Output Definition: Serial Cmd: Default: Options:

Symbol quality information will be output in the form of an abbreviated identifier for each read cycle. <K757,comm port,image mode,image format,JPEG quality,object info output, grade output> Disabled 0 = Disabled 1 = Enabled

MINI Hawk High Performance Imager User Manual

7-47

Page 216

Configuring EZ Trax Output

7-48

MINI Hawk High Performance Imager User Manual

Page 217

8 Symbol Quality Contents Symbol Quality by ESP................................................................................................................. 8-2 Symbol Quality Serial Commands ................................................................................................ 8-3 Overview of Symbol Quality.......................................................................................................... 8-4 Symbol Quality Separator/Data Matrix Output Mode.................................................................... 8-8 ISO/IEC 16022 Symbol Quality Output....................................................................................... 8-10 ISO/IEC 16022 Symbol Quality Output by ESP.......................................................................... 8-12 Omron Microscan Symbol Quality Output ..................................................................................8-13 Omron Microscan Symbol Quality Output by ESP .....................................................................8-16

This section describes parameters that, when enabled, will output detailed symbol quality evaluations.

MINI Hawk High Performance Imager User Manual

8-1

Page 218

Symbol Quality by ESP

Symbol Quality by ESP To open nested options, single-click the +. Click the Parameters button and then the Symbol Quality tab.

To change a setting, double-click the setting and use your cursor to scroll through the options.

8-2

MINI Hawk High Performance Imager User Manual

Page 219

Symbol Quality

Symbol Quality Serial Commands Symbol Quality Separator/ Data <K708,symbol quality separator,data matrix output mode> Matrix Output Mode ISO/IEC 16022 Symbol Quality <K709,symbol contrast,print growth,axial non-uniformity, unused ECC> Output Omron Microscan Symbol <K710,percent cell damage,total read time,capture time,locate time,decode time,pixels per element,ECC Quality Output level,matrix size,quiet zone,symbol angle> Static Validation Report <VAL>

MINI Hawk High Performance Imager User Manual

8-3

Page 220

Overview of Symbol Quality

Overview of Symbol Quality Information about symbol quality and timing can be appended to symbol data by enabling specific evaluation parameters. The order in which these values are appended corresponds directly to the order in which they appear in ESP‘s Symbol Quality tree control. Symbol Quality parameters are separated into ISO/IEC 16022 Symbol Quality Output and Omron Microscan Symbol Quality Output. Total Read Time, Symbol Quality Separator, and Output Mode are common to both ISO/IEC 16022 and Omron Microscan Grading Parameters.

The enabled Symbol Quality parameters are appended to symbol data output in the order that they appear in the tree control.

If you enable the outputs shown above, they will be returned in this order: Symbol Data

Locate Time (Omron Microscan Grading)

Symbol Contrast (ISO/IEC 16022)

Decode Time (Omron Microscan Grading)

0123456789abcdefghijklmnopqrstuv 256 A 30 210 3 17 Total Read Time

Pixels Per Element (Omron Microscan Grading)

Capture Time (Omron Microscan Grading)

8-4

MINI Hawk High Performance Imager User Manual