Page 201

LVS-7500 Operations Manual Version 20.2.X

For the “Reset Alarms” command, the <Binary Command Byte ID#> is equal to a binary 13 (0x0d hex), and the <ASCII String> is blank. The command format is then: <0x0d hex> The LVS® 7500 response when the command completes successfully would be: <0x0d hex> The LVS® 7500 response when the command fails would be: <0x0d hex><ASCII Error Message> NOTE: The LVS® 7500 system will send a TCP/IP packet containing the “Reset Available” binary command byte whenever the Reset button is available to the user. This occurs when an error is detected and the alarm is triggered. The “Reset Available” binary command byte ID is a binary 29 (0x1d hex). The “Reset Alarms” command will be ignored until the LVS® 7500 sends the “Reset Available” message via TCP/IP.

Exit To Main The “Exit To Main” command is valid only after a job has been loaded and is not currently running nor in Bypass or MakeReady mode. Sending this command will return the LVS® 7500 system to the main Welcome window. The command format is: <Binary Command Byte ID#><ASCII String> For the “Exit To Main” command, the <Binary Command Byte ID#> is equal to a binary 14 (0x0e hex), and the <ASCII String> is blank. The command format is then: <0x0e hex> The LVS® 7500 response when the command completes successfully would be: <0x0e hex> The LVS® 7500 response when the command fails would be: <0x0e hex><ASCII Error Message>

Close Out Job The “Close Out Job” command will close the currently loaded job on all connected LVS® 7500 systems. All job-related files are zipped up, removed from the Jobs folder and the jobname.zip file is move to the Archive folder. The command format is: <Binary Command Byte ID#><ASCII String>

LVS-7500 Operations Manual Version 20.2.X

Page 201 of 222

Page 202

LVS-7500 Operations Manual Version 20.2.X

For the “Close Out Job” command, the <Binary Command Byte ID#> is equal to a binary 1 (0x0f hex), and the <ASCII String> is blank. The command format is then: <0x0f hex> The LVS® 7500 response when the command completes successfully would be: <0x0f hex> The LVS® 7500 response when the command fails would be: <0x0f hex><ASCII Error Message>

Import Job The “Import Job” command will import the desired job on all connected LVS® 7500 systems. The command format is: <Binary Command Byte ID#><ASCII String> For the “Import Job” command, the <Binary Command Byte ID#> is equal to a binary 16 (0x10 hex), and the <ASCII String> designates the name of an existing job. The command format is then: <0x10 hex><Job Name to import> The LVS® 7500 response when the command completes successfully would be: <0x10 hex> The LVS® 7500 response when the command fails would be: <0x10 hex><ASCII Error Message>

Load Job From Archive The “Load Job from Archive” command will recall the desired job from the Archive folder onto all connected LVS® 7500 systems. The command format is: <Binary Command Byte ID#><ASCII String> For the “Load Job from Archive” command, the <Binary Command Byte ID#> is equal to a binary 17 (0x11 hex), and the <ASCII String> designates the name of a job in the Archive folder. The command format is then: <0x11 hex><Job Name from Archive> The LVS® 7500 response when the command completes successfully would be: <0x11 hex> The LVS® 7500 response when the command fails would be: <0x11 hex><ASCII Error Message>

LVS-7500 Operations Manual Version 20.2.X

Page 202 of 222

Page 203

LVS-7500 Operations Manual Version 20.2.X

Set Match To The “Set Match To” command sets the “Match this text” string of OCR, OCV, barcode grade or barcode read sectors. To use this feature, a field designator text string is entered into the “Match this text” setting in each sector. The field designator is a unique string identifier that is replaced at runtime when the “Set Match To” command is sent to the LVS® 7500 system. Multiple sectors may have the same field designator entered in the “Match this text” field. The command format is: <Binary Command Byte ID#><ASCII String> For the “Set Match To” command, the <Binary Command Byte ID#> is equal to a binary 28 (0x1c hex), and the <ASCII String> indicates the name of the field designator and the desired match to string separated by the vertical bar character “|”. The command format is then: <0x1c hex><Field Designator Name>|<Match To String> The LVS® 7500 response when the command completes successfully would be: <0x1c hex> The LVS® 7500 response when the command fails would be: <0x1c hex><ASCII Error Message> Example: The field designator “REF_NBR” has already been entered in the “Match this text” field for the desired sectors in the job.

To replace the field designator “REF_NBR” with “ABC123” at run time send the following “Set Match To” command: <0x01c hex><”REF_NBR|ABC123”> The field designator can be any string up to 25 characters as long as the Match To command contains the exact same string as the Field Designator Name. LVS-7500 Operations Manual Version 20.2.X

Page 203 of 222

Page 204

LVS-7500 Operations Manual Version 20.2.X

TCP/IP Output Data Protocol Output data from the LVS® 7500 is broadcast upon the occurrence of an event, such as an error in a sector inspection. The connected system can choose to use the data as it sees fit.

The TCP/IP data output from the LVS® 7500 have the following general format: <Binary Command Byte ID#><ASCII String> <Binary Command Byte ID#> is the unique command ID associated with the particular data contained in the output – and – <ASCII String> consists of command-specific data (if any).

Sector Error The “Sector Error” data output occurs for each sector that fails an inspection. The repeat number, timestamp and distance are sent along with the sector ID number, the special error marker character (ASCII 161) and error code, and any other sector-specific data. The command format is: <Binary Command Byte ID#><ASCII String> For the “Sector Error” command, the <Binary Command Byte ID#> is equal to a binary 7 (0x07 hex), and the <ASCII String> is given below. The data output format is then: <0x07 hex><ASCII String> Where the comma separated ASCII String = <RepeatNumber>,<Timestamp>,<Distance>,<SectorID>,<ErrorMarker><SectorR esultData> NOTE: There is no comma between the special error marker character (ASCII 161) and the sector results data.

Log Line The “Log Line” data output occurs once for each repeat and includes all of the data stored in the run file for that repeat, including all sector inspection results. The repeat number, timestamp and distance are sent along with the inspection results for each sector starting with sector #1 and continuing through the last sector present. The special error marker character (ASCII 161) is inserted immediately prior to the sector result data for any sector that fails an inspection. The command format is: <Binary Command Byte ID#><ASCII String> For the “Log Line” command, the <Binary Command Byte ID#> is equal to a binary 8 (0x08 hex), and the <ASCII String> is given below. The data output format is then: <0x08 hex><ASCII String> LVS-7500 Operations Manual Version 20.2.X

Page 204 of 222

Page 205

LVS-7500 Operations Manual Version 20.2.X

Where the comma separated ASCII String = <RepeatNumber>,<Timestamp>,<Distance>,<Sector 1 ResultData>,<Sector 1 ResultData>, ….. , <Sector X ResultData> NOTE: If one of the sectors failed an inspection, then the special error marker character (ASCII 161) is placed immediately prior to the sector results data and there is no comma between the special error marker character (ASCII 161) and the sector results data. The special warning marker (ASCII 191) is present in lieu of the error marker in the event that an inspection triggers a warning rather than a failure.

Reset Available The “Reset Available” data output occurs every time the “Reset Alarms” button becomes active. This happens when an inspection triggers an error for which the LVS® 7500 is configured to output an alarm. A connected system would know that when this message is received that the Reset Alarm command may now be sent to the LVS® 7500. The command format is: <Binary Command Byte ID#> For the “Reset Available” command, the <Binary Command Byte ID#> is equal to a binary 29 (0x1d hex). The data output format is then: <0x1d hex>

IP Status The “IP Status” output occurs 20 times a second while the LVS® 7500 is running a job. This information is usually unnecessary to remote systems. The information included under this binary command byte ID# is the current lines per second, feet per minute, camera speed, CPU usage, pass sector counter, fail sector count, total sector count, label repeat count, and display color. Each of these items is separated by the vertical bar character “|”. The command format is: <Binary Command Byte ID#><ASCII String> For the “IP Status” command, the <Binary Command Byte ID#> is equal to a binary 31 (0x1f hex). The data output format is then: <0x1f hex><ASCII String> Where the ASCII String is: <LinesPerSecond>|<FeetPerMinute>|<CameraSpeed>|<CPU>|<PassSectorCoun t>|<FailSectorCount>|<TotalSectorCount>|<LaberRepeatCount>|<DisplayColor>

LVS-7500 Operations Manual Version 20.2.X

Page 205 of 222

Page 206

LVS-7500 Operations Manual Version 20.2.X

Appendix H: LVS® 7500 Printronix Integrated System The information in this appendix is intended for users of the LVS® 7500 Printronix Integrated system used in conjunction with the Printronix T5000r thermal barcode printer.

Printing Stopped Error Message As the LVS® 7500 is validating labels, a failed label will be overstruck and reprinted. If the label fails a second time, the label will be overstruck and reprinted again. After three consecutive errors of the same type are detected (except Foreground and Background errors), the printer stops printing labels and the “Printing Stopped” message appears (see below).

“Printing Stopped” error message Note: The text appearing in the “Printing Stopped” message can be edited in the “BeforeAbort.txt” file. See the section further in this appendix entitled “Customizing the Printing Stopped and Manual Intervention Required Messages” for instructions. Options include:

Abort – Stops running the job. After clicking this button, Production mode users are prompted to login. After logging in, the “Manual Intervention Required” message appears (see below). The user must be granted the “Allow Abort” permission to allow the LVS® 7500 to stop running the job.

“Manual Intervention Required” message

LVS-7500 Operations Manual Version 20.2.X

Page 206 of 222

Page 207

LVS-7500 Operations Manual Version 20.2.X

Note: The text appearing in the “Manual Intervention Required” message can be edited in the “AfterAbort.txt” file. See the section entitled “Customizing Error Messages” for instructions. Important: After aborting a print job, there may be print data remaining in the printer’s memory buffer. If granted the “Allow Reset Printer” permission, a user can reset the printer to accept new print data. For detailed instructions, refer to the “Reset the Printer” section further in this appendix.

Ignore – Ignores the failed label and continues printing the next label in the job. After clicking this button, Production mode users are prompted to log in. The user must be granted the “Allow Ignore” permission to allow the LVS® 7500 to ignore the failed label and continue printing the next label in the job.

Retry – Voids the label and reprints the label. If the label fails, the “Printing Stopped” message appears. There is no user permission associated with the “Retry” option.

Foreground and Background Errors The three consecutive errors works differently for Foreground and Background errors. When a Foreground and Background error occur on the same label, a single counter is set by the larger of the two errors. Whichever of the Foreground or Background errors for that label repeat is larger will set/index the counter. This means that if there are both a Foreground and a Background error(s) on the same label, the bigger error (Foreground or Background) sets/indexes the counter. All errors are recorded but only the largest Foreground or Background error affects the counter. The result is three Foreground errors are allowed (i.e. will not stop the process) if a Background error accompanies any of the Foreground errors and the Background error is larger than the corresponding label’s Foreground error. Additionally, three Background errors are allowed (i.e. will not stop the process) if a Foreground error accompanies any of the Background errors and the Foreground error is larger than the corresponding label’s Background error.

Printing Timeout Error The following message appears when a printing timeout has occurred: “Queue up more labels to resume printing and the message will automatically go away. Or, push the “Stop” button to stop the run.”

LVS-7500 Operations Manual Version 20.2.X

Page 207 of 222

Page 208

LVS-7500 Operations Manual Version 20.2.X

Reset the Printer Sometimes the printer will stop printing for no apparent reason (like after aborting a print job) and there may be print data remaining in the printer’s memory buffer. To reset the printer back to a state where it is ready to accept new print data, follow the steps below. A user must be granted the “Allow Reset Printer” permission to reset the printer. 1. Using the Printronix printer’s console buttons, take the printer offline, clear the printer’s buffer, and place the printer back online. 2. The LVS® 7500 software must be at the “Ready to run” screen. 3. Hold down the [Alt] keyboard button and click the “Print Job” button in the top, right corner of the “Ready to run” screen. When the “Print Job” button is not active, ALT + R performs this function also. 4. Enter your login credentials. The “Confirm Reset Printer” message appears.

5. Click the “OK” button. The printer will reset and form feed a few label.

LVS-7500 Operations Manual Version 20.2.X

Page 208 of 222

Page 209

LVS-7500 Operations Manual Version 20.2.X

Customizing the “Printing Stopped” and “Manual Intervention Required” Messages To change the text appearing in the “Printing Stopped” and “Manual Intervention Required” error messages, follow the steps below. Windows Administrator access is required. 1. Open Windows Explorer and access the following path: C:\Program Files (x86)\Label Vision Systems\LVS 7500. 2. Copy the file(s) “AfterAbort.txt” or “BeforeAbort.txt” to your desktop (or another preferred location). The “AfterAbort.txt” file contains the text appearing in the “Manual Intervention Required” message. The “BeforeAbort.txt” file contains the text appearing in the “Printing Stopped” message.

3. Open the file in Notepad or another text editor program and make your changes. When changes are complete, save and then close the file. 4. Copy the files (“AfterAbort.txt” and/or “BeforeAbort.txt”) back to C:\Program Files (x86)\Label Vision Systems\LVS 7500. 5. Shut down and then restart the LVS® 7500 software for the error message changes to appear.

LVS-7500 Operations Manual Version 20.2.X

Page 209 of 222

Page 210

LVS-7500 Operations Manual Version 20.2.X

Form Feed Button The “Form Feed” button advances all labels under the LVS® 7500 readhead when printing is complete. ALT + F also performs this function. Labels fed from the printer after the “Form Feed” button is pressed are not inspected by the LVS® 7500. The ability to form feed labels is available only after the LVS® 7500 stops inspection (by pressing the “Stop” button). Form feeding labels is not available when a job is running or in MAKEREADY mode. To form feed labels: 1. Click the “Stop” button to stop label inspection. 2. Click the “Form Feed” button. The LVS® 7500 readhead advances all remaining labels.

LVS-7500 Operations Manual Version 20.2.X

Page 210 of 222