Page 161

LVS-7500 Operations Manual Version 20.2.X

Setting: Options: Default:

NumColorWBThreads none 5

NumColorWBThreads setting should only be configured by a qualified LVS® representative.

Setting: Options: Default:

NumGrabImages none 0

NumGrabImages setting should only be configured by a qualified LVS® representative.

Setting: Options: Default:

NumPacketThreads none 8

NumPacketThreads is the number of threads. It is set by how many processing CORES are available per PC.

Setting: Options:

OneBigBlemishSector 0=off 1=ON 0=off PDF417UEC none 0

OneBigBlemishSector makes the autosetup feature not separate label lanes.

Default:

Pharmacode 0=off 1=ON 0=off

Setting: Options: Default:

RelearnAuthorization none 0

Pharmacode indicates whether pharmacode is included in the automatic 1D bar code reading/discrimination. Pharmacode=0 means pharmacode is not included in the automatic 1D bar code reading/discrimination. Pharmacode=1 indicates pharmacode is included in the automatic 1D bar code reading/discrimination. RelearnAuthorization setting should only be configured by a qualified LVS® representative.

Setting: Options:

Default:

ReportUnusedCodes 0=off 1=ON 0=off ResyncMatchFileOnMismatch 0=off 1=ON 0=off

Setting: Options: Default:

SerialNumber none missing

Default: Setting: Options: Default: Setting: Options:

Default: Setting: Options:

LVS-7500 Operations Manual Version 20.2.X

PDF417UEC stops PDF from using ISO.

If enabled, a button entitled "Report unused codes" appears. This is used to generate a report of codes that were in the match file, but never seen by the system. If enabled, whenever a mismatch error occurs, the system will assume it is possibly not in the right location in the match file and try to find where it is again. This allows the system to recover instead of generating non-stop mismatch errors after an inserted or deleted label. SerialNumber is the LVS® 7500 serial number and should never be changed.

Page 161 of 222

Page 162

LVS-7500 Operations Manual Version 20.2.X

Setting: Options:

ShowAdditionalCounters 0=off 1=ON 0=off ShowBlemishMotion 0=off 1=ON 0=off ShowElapsedTime 0=off 1=ON 0=off ShowJobReportButton 0=off 1=ON 0=off ShowMod10CheckBox 0=off 1=ON 0=off ShowPrintJobButton 0=off 1=ON 0=off ShowReportLabel 0=off 1=ON 0=off ShowRuntimeGradeStats 0=off 1=ON 0=off ShowSnapshotButton 0=off 1=ON 0=off ShowSpeed 0=off 1=ON 0=off ShowStatusAlert 0=off 1=ON 0=off

ShowAdditionalCounters setting should only be configured by a qualified LVS® representative.

LVS-7500 Operations Manual Version 20.2.X

Page 162 of 222

Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default: Setting: Options: Default:

ShowBlemishMotion setting should only be configured by a qualified LVS® representative.

ShowElapsedTime setting should only be configured by a qualified LVS® representative.

When ShowJobReportButton setting is on the Job Report button is enabled on the edit mode Step 7 Save job to disk screen. ShowMod10CheckBox setting should only be configured by a qualified LVS® representative.

When ShowPrintJobButton setting is ON the Print Job button is enabled on the Ready to run screen. ShowReportLabel setting should only be configured by a qualified LVS® representative.

ShowRuntimeGradeStats setting should only be configured by a qualified LVS® representative. ShowSnapshotButton setting should only be configured by a qualified LVS® representative.

ShowSpeed setting should only be configured by a qualified LVS® representative.

ShowStatusAlert setting should only be configured by a qualified LVS® representative.

Page 163

LVS-7500 Operations Manual Version 20.2.X

Setting: Options: Default:

SkipAutoSetupShrink none 0

SkipAutoSetupShrink setting should only be configured by a qualified LVS® representative.

Setting: Options: Default:

Splash none none

Splash setting should only be configured by a qualified LVS® representative.

Setting: Options:

SpotVoidCheck 0=off 1=ON 0=off Stagger 0=off 1=ON 0=off StripParentheses 0=off 1=ON 0=off Strobe 0=Normal mode 1=Strobe mode 2=Advanced strobe mode 0= Normal mode TakeSecondPicture 0=off 1=ON 0=off ThreadLockTimeout none 0

SpotVoidCheck will error any spot anywhere within a barcode. Do not use unless instructed by LVS® technician.

TiltedRead attempts to read a short in height barcode by applying a slight rotation.

Default: Setting: Options: Default:

TiltedRead 0=off 1=ON 0=off TitleBar none none

Setting: Options: Default:

UsePhotoeyeSignal none 0

UsePhotoeyeSignal turns on the monitoring of the Photo optic signal Input line. Activated on LOW signal. This is special as it can use any available port via the integer used. Also, the polarity can be switched by making the integer a negative number. For example,

Default: Setting: Options: Default: Setting: Options: Default: Setting: Options:

Default: Setting: Options: Default: Setting: Options: Default: Setting: Options:

LVS-7500 Operations Manual Version 20.2.X

Stagger setting should only be configured by a qualified LVS® representative.

StripParentheses takes parentheses off decoded data strings from symbologies such as GS1 Databar or Data Matrix. Strobe allows the system to be used in viewing mode like a strobe light. There must be a consistent sync mark on the web to use this feature. Only use this setting if advised by an LVS® Technician. TakeSecondPicture setting should only be configured by a qualified LVS® representative.

ThreadLockTimeout setting should only be configured by a qualified LVS® representative.

TitleBar setting should only be configured by a qualified LVS® representative.

Page 163 of 222

Page 164

LVS-7500 Operations Manual Version 20.2.X

UsePhotoeyeSignal=-3 makes it respond to an active HIGH signal on input #3. Setting: Options: Default:

UseSectorTags none 0

UseSectorTags setting should only be configured by a qualified LVS® representative.

Setting: Options:

UseTrackingSignal 0=off 1=ON 0=off

UseTrackingSignal setting should only be configured by a qualified LVS® representative.

Default:

TCP/IP Setting: Options: Default:

Host none 0.0.0.0

Host setting should only be configured by a qualified LVS® representative. Leave as default value.

Setting: Options:

Default:

Mode off Remote XML off

Setting: Options: Default:

Port1 none 0

Mode indicates if the LVS® 7500 is connecting to another system. Use off if the LVS® 7500 is not connecting to another system. Use remote if the LVS® 7500 is connecting with a remote computer, such as the LVS® HMI Command Center or a non-LVS® system. Note that all remaining settings (such as Host, Port1, Port2, etc.) do not apply if the LVS® 7500 is not connecting to another system. Enter the port where the first external system will be listening for LVS® data.

Setting: Options: Default:

Port1Filter none none

Setting: Options: Default:

Port2 none 0

Setting: Options: Default:

Port2Filter none none

LVS-7500 Operations Manual Version 20.2.X

Specifies the type of information that the LVS®7500 system will broadcast on the port specified by Port1. This filter is for output from the LVS®7500 only. Leave this blank to receive all information from the LVS®7500 system. Enter a comma separated list of Binary Command Byte ID#’s to receive only the feedback information desired. Enter the port where the second external system will be listening for LVS® data. Specifies the type of information that the LVS®7500 system will broadcast on the port specified by Port2. This filter is for output from the LVS®7500 only. Leave this blank to receive all information from the LVS®7500 system. Enter a comma separated list of Binary Command Byte ID#’s to receive only the feedback information desired.

Page 164 of 222

Page 165

LVS-7500 Operations Manual Version 20.2.X

LVS-7500 Operations Manual Version 20.2.X

Page 165 of 222

Page 166

LVS-7500 Operations Manual Version 20.2.X

Appendix B: Epedigree

Systems with ePedigree enabled are using the normal LVS® 7500 software with enhanced tracking features. Below are a couple of steps that highlight the LVS® 7500 ePedigree process. 1. When drawing a sector over a 2D Data Matrix code, make sure that the center of the X within the sector is close to the center of the 2D code. This helps the software the software determine the position of the bar code.

LVS-7500 Operations Manual Version 20.2.X

Page 166 of 222

Page 167

LVS-7500 Operations Manual Version 20.2.X

2. Choose the Check for Duplicates option to make sure that the bar code’s encoded data is unique for this job or the sect. When running the job, a !DU error will show up if the software finds a duplicate of the bar code (see below).

LVS-7500 Operations Manual Version 20.2.X

Page 167 of 222

Page 168

LVS-7500 Operations Manual Version 20.2.X

3. Draw an OCR sector around the human readable characters and let it match its corresponding bar code. In this example, we have to start matching OCR at position 17 of the bar code, because the bar code is 28 characters long and the human readable characters match the last 12 characters of the bar code.

LVS-7500 Operations Manual Version 20.2.X

Page 168 of 222

Page 169

LVS-7500 Operations Manual Version 20.2.X

Appendix C: ImageSaver Instructions Saving Raw Images WITHOUT a Label Repeat 1. Create and name a target folder where images can be saved (see below).

2. Open the LVS® 7500 software and click “Settings” in the menu bar. a. Select the “Settings for this section” radio button and select “ImageSaver” from the dropdown list (see below). b. Double-click “Path” located in the “Setting” column (see below).

c. In the “Value” field, enter the path of the folder that was created to store images. Then, click “OK (save changes).”

LVS-7500 Operations Manual Version 20.2.X

Page 169 of 222

Page 170

LVS-7500 Operations Manual Version 20.2.X d. Double-click “SaveLabelRepeat.” e. Type “1” in the “Value” field and click “OK (save changes).” f. Close the “Settings” menu by clicking the red “X” in the top, right corner of the Configuration Editor screen. 3. Click “Yes” to the “Save Changes” confirmation.

4. Close and then restart the LVS® 7500 software. 5. Click “OK” at the ImageSaver warning. Clearing the ImageSaver after capturing the desired images is important as it will continue to save images causing the hard drive to fill up.

LVS-7500 Operations Manual Version 20.2.X

Page 170 of 222