Page 121

LVS-7500 Operations Manual Version 20.2.X

LVS-7500 Operations Manual Version 20.2.X

Page 121 of 222

Page 122

LVS-7500 Operations Manual Version 20.2.X

Photo Eye Wiring for Peel and Present Functionality on the LVS® 7500 Zebra Integrated Systems (Zebra Xi4 Series) This section applies to customers using the Peel and Present functionality on the LVS® 7500 Zebra integrated systems and applies only to the Zebra Xi4 series of printers. When integrated with a Zebra Xi4 model printer, the LVS® 7500 uses the signal from the Stop Motion Unit/Light Tower Unit to stop a label after printing. The wiring diagram below shows how to connect a DB9 Female connector to a cable that would connect to the terminal block CN7 on the Stop Motion circuit board. The DB9 Female connector must be plugged into a photo eye connection made available at the rear interface panel of the printer. CN7 is the photo eye connector for the Peel and Present functionality on the LVS® 7500 Zebra integrated systems.

LVS-7500 Operations Manual Version 20.2.X

Page 122 of 222

Page 123

LVS-7500 Operations Manual Version 20.2.X

Appendix A: User Configurable Settings Settings Configuration Editor The LVS® 7500 software is configurable through options that are user configurable. The “Settings” menu bar feature opens the LVS® 7500 Configuration Editor which allows you to configure the basic and advanced features and functionality of the LVS® 7500. A user must be assigned administrator rights to access the “Settings” menu bar. Refer to the “List of User Configurable Settings” section below for a description of the available settings and options.

LVS® 7500 Configuration Editor

The LVS® 7500 Configuration Editor offers the following options: Option

Description

Basic settings

Select this option to display the settings considered to be basic to LVS® 7500 configuration. Select this option to display any values that have been modified to a value other than the default value. Select this option to display all settings listed alphabetically by “Section.” Click the drop-down box to display all settings for the selected section. Enter a text string to search for settings containing the

Settings different from default All settings Settings for this section Settings containing

LVS-7500 Operations Manual Version 20.2.X

Page 123 of 222

Page 124

LVS-7500 Operations Manual Version 20.2.X

Option

Description this text entered text. For example, typing “camera” in the text field will display all settings containing the word “camera.”

Each setting is grouped by “Section,” “Setting,” and “Value” (see screenshot below). Double-click a setting row. The “LVS® 7500 Configuration Editor Individual Setting” window appears providing the Section, Setting, Default, Value, and Setting Description. The only editable field is the “Value” field. All other fields cannot be edited. Click “OK (save changes)” to save your changes or “Cancel (discard changes)” to discard any changes.

Setting description

LVS-7500 Operations Manual Version 20.2.X

Page 124 of 222

Page 125

LVS-7500 Operations Manual Version 20.2.X

List of User Configurable Settings Active Directory Setting: ActiveDirectoryAuthentication Options: 0=off 1=ON Default: 0=off Setting: ActiveDirectoryDomain Options: Text string Default: None

When ActiveDirectoryAuthentication is enabled the LVS® 7500 software will attempt to use Microsoft Server Active Directory for User Administration. Microsoft Server Active Directory Domain name. Example: your-company.com.

Setting: ActiveDirectoryLVSAllowAbort Options: Text string Default: None

When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Abort permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Accept Replace permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Administration permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Bypass MakeReady permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Calibration permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Create Edit permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Ignore permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Load Existing permission.

Setting: ActiveDirectoryLVSAllowAcceptReplace Options: Text string Default: None Setting: ActiveDirectoryLVSAllowAdministration Options: Text string Default: None Setting: ActiveDirectoryLVSAllowBypassMakeReady Options: Text string Default: None Setting: ActiveDirectoryLVSAllowCalibration Options: Text string Default: None Setting: ActiveDirectoryLVSAllowCreateEdit Options: Text string Default: None Setting: ActiveDirectoryLVSAllowIgnore Options: Text string Default: None Setting: ActiveDirectoryLVSAllowLoadExisting Options: Text string Default: None

LVS-7500 Operations Manual Version 20.2.X

Page 125 of 222

Page 126

LVS-7500 Operations Manual Version 20.2.X

Setting: ActiveDirectoryLVSAllowResetPrinter Options: Text string Default: None Setting: ActiveDirectoryLVSAllUsers Options: Text string Default: None

Setting: ActiveDirectoryLVSOrgUnit Options: Text string Default: None

When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that gives LVS® 7500 users the Allow Reset Printer permission. When ActiveDirectoryAuthentication is enabled, this is the Active Directory group that contains the list of all Active Directory users that will be integrated into the LVS® Operator permissions. When ActiveDirectoryAuthentication is enabled, this is the Active Directory Organizational Unit that contains Active Directory Groups that control access to the LVS® 7500 system.

Background - !BG Setting: Signal Duration Options: 10ms, 20ms, 100ms, 200ms, 500ms, 1 second, 5 seconds, hold Default: 100ms

Specifies the signal duration.

Setting: Signal Output Line Options: I/O line 1 I/O line 2 I/O line 3 I/O line 4 Default: I/O line 3

Sets the I/O line to activate for this error condition. The default value is I/O line 3, which corresponds to the red light.

Setting: StopMotion Options: do not stop immediately after 2 in a row after 3 in a row after 4 in a row after 5 in a row after 6 in a row after 7 in a row after 8 in a row after 9 in a row after 10 in a row Default: immediately

Specifies if this error condition should also trigger activation of the stop motion signal.

LVS-7500 Operations Manual Version 20.2.X

Page 126 of 222

Page 127

LVS-7500 Operations Manual Version 20.2.X

Checkdigit - !CD Setting: Signal Duration Options: 10ms 20ms 100ms 200ms 500ms 1 second 5 seconds hold Default: 100ms

Specifies the signal duration.

Setting: Signal Output Line Options: I/O line 1 I/O line 2 I/O line 3 I/O line 4 Default: I/O line 3

Sets the I/O line to activate for this error condition. The default value is I/O line 3, which corresponds to the red light.

Setting: StopMotion Options: do not stop immediately after 2 in a row after 3 in a row after 4 in a row after 5 in a row after 6 in a row after 7 in a row after 8 in a row after 9 in a row after 10 in a row Default: immediately

Specifies if this error condition should also trigger activation of the stop motion signal.

Die Cut) - !DC Setting: Signal Duration Options: 10ms 20ms 100ms 200ms 500ms 1 second 5 seconds hold Default: 100ms

LVS-7500 Operations Manual Version 20.2.X

Specifies the signal duration.

Page 127 of 222

Page 128

LVS-7500 Operations Manual Version 20.2.X

Setting: Signal Output Line Options: I/O line 1 I/O line 2 I/O line 3 I/O line 4 Default: I/O line 3

Sets the I/O line to activate for this error condition. The default value is I/O line 3, which corresponds to the red light.

Setting: StopMotion Options: do not stop immediately after 2 in a row after 3 in a row after 4 in a row after 5 in a row after 6 in a row after 7 in a row after 8 in a row after 9 in a row after 10 in a row Default: immediately

Specifies if this error condition should also trigger activation of the stop motion signal.

Alarm (Duplicate) Setting: Signal Duration Options: 10ms 20ms 100ms 200ms 500ms 1 second 5 seconds hold Default: 100ms

Specifies the signal duration.

Setting: Signal Output Line Options: I/O line 1 I/O line 2 I/O line 3 I/O line 4 Default: I/O line 3

Sets the I/O line to activate for this error condition. The default value is I/O line 3, which corresponds to the red light.

LVS-7500 Operations Manual Version 20.2.X

Page 128 of 222

Page 129

LVS-7500 Operations Manual Version 20.2.X

Setting: StopMotion Options: do not stop immediately after 2 in a row after 3 in a row after 4 in a row after 5 in a row after 6 in a row after 7 in a row after 8 in a row after 9 in a row after 10 in a row Default: immediately

Specifies if this error condition should also trigger activation of the stop motion signal.

Alarm (Foreground) Setting: Signal Duration Options: 10ms 20ms 100ms 200ms 500ms 1 second 5 seconds hold Default: 100ms

Specifies the signal duration.

Setting: Signal Output Line Options: I/O line 1 I/O line 2 I/O line 3 I/O line 4 Default: I/O line 3

Sets the I/O line to activate for this error condition. The default value is I/O line 3, which corresponds to the red light.

Setting: StopMotion Options: do not stop immediately after 2 in a row after 3 in a row after 4 in a row after 5 in a row after 6 in a row after 7 in a row after 8 in a row after 9 in a row after 10 in a row Default: immediately

Specifies if this error condition should also trigger activation of the stop motion signal.

LVS-7500 Operations Manual Version 20.2.X

Page 129 of 222

Page 130

LVS-7500 Operations Manual Version 20.2.X

Alarm (Gap) Setting: Signal Duration Options: 10ms 20ms 100ms 200ms 500ms 1 second 5 seconds hold Default: 100ms

This specifies the signal duration.

Setting: Signal Output Line Options: I/O line 1 I/O line 2 I/O line 3 I/O line 4 Default: I/O line 3

This sets the I/O line to activate for this error condition. The default value is I/O line 3, which corresponds to the red light.

Setting: StopMotion Options: do not stop immediately after 2 in a row after 3 in a row after 4 in a row after 5 in a row after 6 in a row after 7 in a row after 8 in a row after 9 in a row after 10 in a row Default: immediately

This specifies if this error condition should also trigger activation of the stop motion signal.

Alarm (Good Read) Setting: Signal Duration Options: 10ms 20ms 100ms 200ms 500ms 1 second 5 seconds hold Default: 100ms

LVS-7500 Operations Manual Version 20.2.X

This specifies the signal duration.

Page 130 of 222