Page 121
Calibration
Additional Notes about Calibration The following conditions apply to the imager’s calibration process. Some of these items are noted at various points throughout this section, or in other sections of the imager’s documentation. 1. The Mirrored Image parameter <K514> is not part of the calibration process. 2. Background Color <K451> is calibrated. 3. If Window of Interest Framing is enabled, the WOI will be set to full frame when calibration begins. If WOI Framing is disabled, the current WOI configuration will be used for the Search Pass. 4. IP Mode <K527> will not be altered during calibration. 5. Threshold Mode <K512> will be set to Adaptive during calibration. At the end of a successful calibration, Threshold Mode will remain set to Adaptive regardless of its previous setting. Otherwise, the value will revert to its original state. 6. If Symbol Type is enabled for calibration (Autodiscriminate), then Interleaved 2 of 5 Range Mode <K472> will be enabled. This allows variable-length Interleaved 2 of 5 symbols to be decoded. If an Interleaved 2 of 5 symbol is decoded during calibration, then code length # 1 will be set to the decoded symbol length at the end of calibration. Otherwise, the symbol lengths will be restored to their original configurations. 7. Pharmacode is not calibrated. 8. The symbol size, dimension, and orientation parameters of the No Symbol / Bad Symbol and Symbol Qualification command <K718> will be updated if calibration is completed successfully. 9. All symbol types that were enabled before calibration will still be enabled after calibration. For example, if Data Matrix ECC 200 was enabled before calibration, and the calibration routine was performed on a Code 128 symbol, then after calibration is completed both Data Matrix ECC 200 and Code 128 will be enabled. 10. The calibration process does not modify the global Composite status <K453>. The global Composite status must be configured properly before calibration. 11. If the user requires calibration of a stacked symbology <K482>, <K483>, or <K484>, that symbology must be enabled and configured appropriately before calibration. 12. The Search process will use the configurable Window of Interest for image captures. However, when the search process is completed, the WOI will be reduced to include only the symbol of interest and some additional boundary area.
EZ Match User Manual
5-17
Page 122
Additional Notes about Calibration
5-18
EZ Match User Manual
Page 123
6 Read Cycle Contents Read Cycle by ESP ...................................................................................................................... 6-2 Read Cycle Serial Commands...................................................................................................... 6-3 Read Cycle Setup......................................................................................................................... 6-4 Multisymbol................................................................................................................................... 6-5 Trigger Mode and Duration........................................................................................................... 6-7 External Trigger Polarity ............................................................................................................. 6-12 Serial Trigger .............................................................................................................................. 6-13 Start Trigger Character (Non-Delimited)..................................................................................... 6-14 Stop Trigger Character (Non-Delimited) ..................................................................................... 6-15 End of Read Cycle...................................................................................................................... 6-16 Capture Mode ............................................................................................................................. 6-18 Capture Timing ........................................................................................................................... 6-23 Image Storage ............................................................................................................................ 6-25 Minimum Good Reads ................................................................................................................ 6-28
After you’ve established communications with the imager, you will need to address the spatial and timing parameters associated with your application. This section explains those parameters.
EZ Match User Manual
6-1
Page 124
Read Cycle by ESP
Read Cycle by ESP
Click the Parameters button and then the Read Cycle tab.
To open nested options, single-click the +.
6-2
To change a setting, double-click the setting and use your cursor to scroll through the options.
EZ Match User Manual
Page 125
Read Cycle
Read Cycle Serial Commands Trigger Mode/Duration Serial Trigger Character External Trigger State End of Read Cycle Minimum Good Reads Multisymbol Start Trigger Character Stop Trigger Character Capture Mode Capture Timing Image Storage
EZ Match User Manual
<K200,trigger mode,trigger filter duration> <K201,serial trigger character> <K202,external trigger state> <K220,end of read cycle,read cycle timeout> <K221,minimum good reads> <K222,number of symbols,multisymbol separator> <K229,start character> <K230,stop character> <K241,capture mode,number of captures,rapid capture mode, pre-processing> <K242,time before first capture,time between first and second captures,,,,,,,,time between seventh and eighth captures> <K244,image storage type,image store mode>
6-3
Page 126
Read Cycle Setup
Read Cycle Setup Setting up Read Cycle and triggering parameters will involve a series of decisions based on your particular application, as follows: 1. Select the number of symbols to be read in a single cycle. The EZ Match Imager can read multiple symbols in a single image frame. 2. Decide on the trigger type to be used: if serial, choose a serial character; if external, choose either Level or Edge. 3. Designate how the read cycle should end (Timeout, New Trigger, Last Frame). 4. Select Capture Mode, Continuous Mode, or Rapid Capture Mode. 5. Select Number of Captures (if in Rapid Capture Mode). 6. Set the Time Before First Capture and Time Between Captures, if any. Note: The imager’s camera captures images at a rate of about 16/second (based on a full-size 1280 x 1024 monochrome image). The capture rate increases as frame sizes decrease.
6-4
EZ Match User Manual
Page 127
Read Cycle
Multisymbol Usage: Definition: Conditions:
Multisymbol is commonly used in shipping applications where a shipping symbol contains individual symbols for part number, quantity, etc. This feature allows on trigger to pick up all the symbols. Multisymbol allows the user to define up to 100 symbols that can be read in a single read cycle. The following conditions apply: • Each symbol must be different to be read, unless in Rapid Capture Mode, configured for “triggered capture”. • The maximum number of characters in a read cycle is 3,000 for all symbols. • All No Read messages are posted at the end of the data string, unless output filtering is enabled. • If more than one symbol is within the field of view at the same time, symbol data may not be displayed in the order of appearance. • If Matchcode Type is set to Sequential or if Trigger is set to Continuous Read 1 Output, the imager will behave as if Number of Symbols were set to 1, regardless of the user-defined configuration.
Number of Symbols Definition:
Serial Cmd: Default: Options:
This parameter has no function when a Multi-Operation read cycle is active. Each operation index (1 through 3) requires one qualified output for each active operation. Example: If tube and cap detection are both enabled for operation index 1, the required number of symbols for that index would be 2. Important: The same symbol, tube, or cap can be processed and qualified by two different operations in the same read cycle. <K222,number of symbols, multisymbol separator> 1 1 to 100
EZ Match User Manual
6-5
Page 128
Multisymbol
Multisymbol Separator Usage: Definition: Serial Cmd: Default: Options:
Used to delimit or separate data fields with a user defined character. Any valid ASCII character, inserted between each symbol read when Multisymbol is set to any number greater than 1. <K222,number of symbols,multisymbol separator> , (comma) Any available ASCII character. Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K222h,,3C> For >: <K222h,,3E> For , : <K222h,,2C> See ASCII Character Entry Modifier on page 4-33 for a more detailed explanation of how to enter ASCII characters as hex values. Note: If No Read messages are disabled and there are No Reads occurring, separators will only be inserted between symbol data outputs.
6-6
EZ Match User Manual
Page 129
Read Cycle
Trigger Mode and Duration Trigger Definition:
The event that initiates the read cycle.
Note: When calibrating the imager or testing Read Rate, the current trigger setting will be disregarded.
Trigger Mode Serial Cmd: Default: Options:
<K200,trigger mode,trigger filter duration> Continuous Read 0 = Continuous Read 1 = Continuous Read 1 Output 2 = External Level 3 = External Edge 4 = Serial Data 5 = Serial Data and Edge
Continuous Read Usage: Definition:
Serial Cmd:
Continuous Read is useful in testing symbol readability or reader functions. It is not recommended for normal operations. In Continuous Read, trigger input options are disabled, the imager is always in the read cycle, and it will attempt to decode and transmit every capture. If a single symbol stays within read range for multiple read cycles, its data will be transmitted repeatedly until it leaves the read range. The imager sends replies to serial commands that require responses when symbol data is transmitted, or read cycle timeout is enabled and a timeout occurs and at least one captured image has been processed. Depending on the combination of enabled symbologies and the Threshold Mode setting, the imager may take longer than the timeout to process a captured image. Note: When to Output and No Read options have no affect on Continuous Read. <K200,0>
EZ Match User Manual
6-7
Page 130
Trigger Mode and Duration
Continuous Read 1 Output Usage:
Definition:
Serial Cmd:
Continuous Read 1 Output can be useful in applications where it is not feasible to use a trigger and all succeeding symbols contain different information. It is also effective in applications where the objects are presented by hand. In Continuous Read 1 Output the imager self-triggers whenever it decodes a new symbol or a timeout occurs. If End of Read Cycle is set to Timeout and the symbol doesn’t change, the output is repeated at the end of each timeout period. For example, if Timeout is set to one second, the imager sends the symbol data immediately and repeats the output at intervals of one second for as long as it continues to capture the symbol. If End of Read Cycle is set to New Trigger, the imager will send the current symbol data immediately, but only once. A new symbol appearing in the imager’s range will be read and sent immediately, provided it is not identical to the previous symbol. <K200,1> Caution: In automated environments, Continuous Read 1 Output is not recommended because there is typically no reliable way to verify that a symbol was missed. Note: If Trigger Mode is set to Continuous Read 1 Output, the imager will behave as if Number of Symbols were set to 1, regardless of the user-defined configuration.
6-8
EZ Match User Manual