Page 111
Calibration
Disabled When this feature is disabled, the Window of Interest is not modified after the calibration process is complete.
Row and Column If the calibration process is successful, the Window of Interest will be modified to frame the symbol as well as an additional margin area around the symbol, determined by the WOI Margin parameter. Refer to the diagram below:
Column If the calibration process is successful, the Window of Interest columns will be modified to vertically frame the symbol, plus an additional margin area around the symbol, determined by the WOI Margin parameter. Refer to the diagram below:
Row If the calibration process is successful, the Window of Interest rows will be modified to horizontally frame the symbol, plus an additional margin area around the symbol, determined by the WOI Margin parameter. Refer to the diagram below:
EZ Match User Manual
5-7
Page 112
Calibration Options
Straight Line This feature is intended for use with linear symbologies. If the calibration process is successful, the orientation of the symbol is determined and the Window of Interest is modified according to the symbol orientation. The scan line orientation is determined to be vertical if the symbol tilt is between 225° and 315°, or between 45° and 135°. Otherwise, the scan line will be horizontal. If the symbol is vertical, the image column size will be set by the scan height parameter, and will be configured for full row resolution. If the symbol is horizontal, the image row size will be set by the scan height parameter, and will be configured for full column resolution. The scan line will be centered in the middle of the symbol. If the symbol is tilted such that the scan line will not pass completely through the symbol, the scan width will be adjusted to include the entire symbol. Refer to the diagram below:
Straight Line Framed This parameter is the same as “Straight Line”, except that the Window of Interest will also frame the scan line on the symbol length as well. The scan line includes the symbol plus an additional margin area determined by the WOI Margin parameter.
5-8
EZ Match User Manual
Page 113
Calibration
Window of Interest (WOI) Margin Definition: Serial Cmd: Default: Options:
Sets the margin size that is applied to the calibrated symbol. This parameter is expressed in number of pixels. If the margin causes the image to exceed the maximum image size, it will be reduced accordingly. <K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> 75 (pixels) 20 to 1280
Line Scan Height Definition: Serial Cmd: Default: Options:
This parameter is only used with the Straight Line modes. It sets the scan height of the straight-line image, and it is expressed in number of pixels. <K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> 64 (pixels) 3 to 1024
Processing Definition: Serial Cmd: Default: Options:
This setting defines the amount of time and effort the imager will spend attempting to decode a symbol for each parameter configuration. <K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> Medium 0 = Low 1 = Medium 2 = High 3 = Definable
Low The imager will spend a low amount of effort attempting to decode the given symbol for each parameter configuration.
Medium The imager will spend a medium amount of effort attempting to decode the given symbol for each parameter configuration.
High The imager will spend a high amount of effort attempting to decode the given symbol for each parameter configuration.
Definable The processing time for each image frame is defined by the Image Processing Timeout parameter <K245>. EZ Match User Manual
5-9
Page 114
Calibration by ESP
Calibration by ESP ESP’s Calibration view incorporates all the functionality of the <K529> (Calibration Options) command in a single easy-to-use, intuitive interface. This calibration process allows the user much finer control of individual parameters than the calibration routine in EZ Mode or on the Video tab.
Before / During Calibration Before and After tabs give the user an authentic representation of the difference in what the imager “sees” before and after the calibration routine.
Focus, Shutter, Gain, and Image Quality values are all tracked in real-time during the calibration process. Gain and Shutter Speed are configurable in the Calibration interface.
Determines the amount of time and effort the imager will spend attempting to decode a symbol for each parameter configuration. (Low, Medium, High, Definable).
Dropdown menu of 1D and 2D symbologies, and a check box for enabling or disabling Composite symbologies.
Initiates the calibration routine. Cancels the calibration process, if necessary.
The Window of Interest section of the Calibration interface allows the user to make precise adjustments to WOI Framing, WOI Margin (in pixels), and the scan height of the straight-line image (in pixels).
5-10
EZ Match User Manual
Page 115
Calibration
After Calibration Notice the improved image resolution after the calibration process is complete. The Focus, Shutter, Gain, and Quality indicator bars remain stationary at the end of calibration to provide a visual reference for the final values of these four parameters.
The calibration progress bar indicates that the process has been completed.
Saves the optimal configuration parameters determined by calibration results.
EZ Match User Manual
5-11
Page 116
Initiating Calibration
Initiating Calibration Calibration can be initiated by serial command, the EZ Button, or from the Calibration interface in ESP. In ESP, the Calibrate button starts the calibration routine.
To initiate calibration by serial command, enter the serial command<@CAL> from a terminal, such as ESP’s Terminal interface. <@CAL> is a calibration routine that emphasizes contrast and readability.
See Calibrate by EZ Button on page 1-11 for the EZ Button calibration procedure.
5-12
EZ Match User Manual
Page 117
Calibration
Calibration Progress Indicators
During the calibration process, the imager has a number of ways to communicate its progress to the user. The following three types of feedback are the primary ways that the imager communicates calibration progress.
Read Rate LEDs When calibration is initiated through a serial command or the EZ Button, the imager’s Read Rate LEDs indicate calibration progress as follows: LEDs 20% 20%, 40%, 60% 20%, 40%, 60%, 80% 20%, 40%, 60%, 80%, 100%
Calibration Progress Search Pass in progress. Medium Pass in progress. Fine-Tune Pass in progress. Calibrated Successfully (holds for green flash duration).
Calibrated Successfully Fine-Tune Pass Medium Pass
Search Pass
Beeper / Green Flash The beeper and green LED flash are used to indicate calibration status as shown in the table below: Indicator 2 quick beeps 5 quick beeps 1 long beep 1 sec. Green Flash
EZ Match User Manual
Status Calibration has started. Calibration has failed. Calibration has been completed successfully. Calibration has been completed successfully.
5-13
Page 118
Initiating Calibration
Calibration Progress Output to Terminal
When initiated by serial command or EZ Button, the calibration operation will output progress and process information to a terminal if the system is configured to do so.
Terminal Depending upon how the calibration options are configured, the calibration process can have up to 5 steps (Search Pass, Medium Pass, Fine-Tune Pass). Calibration data is shown in a simple table format under 8 category headings, each of which are defined below. Heading Prog Rating Decode Quality Shutter Gain Locate
Definition Progress indicator with a range of 0 - 100. Indicates percent complete. Rating indicator; higher ratings are more favorable. Number of successful decodes for the pass. Quality indicator; higher ratings are more favorable. Shutter setting under operation. Gain setting under operation. Time required to locate symbol, expressed in milliseconds.
Calibration Progress Output Examples (Terminal) Search Pass
The purpose of Search Pass is to decode a symbol in as few configurations as possible during calibration. This process is designed to identify symbology type, and to identify any other image processing parameters that are already calibrated. Running Search Pass
5-14
Prog
Rating
Decode
Quality
Focus
Shutter
Gain
Locate
12 14 16 18
0 0 0 2174
0 0 0 1
0 0 0 72
313 313 313 313
250 250 250 250
0 9 18 27
0 0 0 24
EZ Match User Manual
Page 119
Calibration
Medium Pass
The theory behind Medium Pass is to develop a readable range from which to dial in a decoded symbol. Running Medium Pass Prog
Rating
Decode
Quality
Focus
Shutter
Gain
Locate
51 53 55 57 59 61 63 65 67 69 71 73 75
4216 8436 6336 8448 8456 8464 8472 8488 8504 8512 8528 8584 8644
2 4 3 4 4 4 4 4 4 4 4 4 4
8 20 24 32 40 48 56 72 88 96 112 168 228
386 386 386 386 386 386 386 386 386 386 386 386 386
250 250 250 250 250 250 250 250 250 250 250 250 250
0 4 8 12 16 20 24 28 32 36 40 44 48
5 4 4 4 5 4 5 5 5 5 6 5 5
Rating List (Best): Rating = 8606; Focus = 386; Shutter = 250; Gain = 48 Rating = 8581; Focus = 386; Shutter = 250; Gain = 44 Rating = 8553; Focus = 386; Shutter = 250; Gain = 40 Rating = 8527; Focus = 386; Shutter = 250; Gain = 36 Rating = 8505; Focus = 386; Shutter = 250; Gain = 32 Rating = 8488; Focus = 386; Shutter = 250; Gain = 28 Rating = 8476; Focus = 386; Shutter = 250; Gain = 24
EZ Match User Manual
5-15
Page 120
Initiating Calibration
Fine-Tune Pass
The theory behind Fine-Tune Pass is to dial in a decoded symbol based on the range supplied by the Medium Pass. Running Fine Tune Pass Prog
Rating
Decode
Quality
Focus
Shutter
Gain
Locate
77 79 81 84 86 88 90 93 95 97 100
6416 6424 6424 6432 6448 6464 6476 6488 6506 6524 6538
3 3 3 3 3 3 3 3 3 3 3
104 112 112 120 136 152 164 176 194 212 226
386 386 386 386 386 386 386 386 386 386 386
250 250 250 250 250 250 250 250 250 250 250
38 39 40 41 42 43 44 45 46 47 48
7 6 6 6 5 5 5 5 6 5 6
Rating List (Best): Rating = 6527; Focus = 386; Shutter = 250; Gain = 48 Rating = 6517; Focus = 386; Shutter = 250; Gain = 47 Rating = 6505; Focus = 386; Shutter = 250; Gain = 46 Rating = 6491; Focus = 386; Shutter = 250; Gain = 45 Rating = 6476; Focus = 386; Shutter = 250; Gain = 44 Rating = 6462; Focus = 386; Shutter = 250; Gain = 43 Rating = 6449; Focus = 386; Shutter = 250; Gain = 42 Calibration PASSED.
5-16
EZ Match User Manual