Page 101

Communications

ASCII Character Entry Modifier Commands that require ASCII text fields, such as Preamble and Postamble commands, can be sent to the imager as hex pairs (see Appendix E, ASCII Table, for conversions). Serial Cmd Format: <Knnnh,00-FF> To enter ASCII fields as hex values (00 to FF), add a lower-case h directly after the command’s K number, and then enter the hex value that corresponds with the desired ASCII character. Example: Consider the Postamble command: Serial Cmd:

<K142,status,postamble character(s)>

Imagine that your application requires the ASCII character > to be the postamble in your symbol decode output. The ASCII characters <, >, and , can only be entered as hex pairs. So, to make > the postamble in your symbol decode output, enter the Postamble command as follows: <K142h,,3E> Note that the “status” field contains only a , . This is because the only field that is being changed is the “postamble character(s)” field. (See Serial Configuration Command Conventions on page A-13 for a more detailed explanation of this command shortcut.)

EZ Match User Manual

4-33

Page 102

Preamble

Preamble Preamble Status Usage: Definition: Serial Cmd: Default: Options:

Useful for identifying and controlling incoming data. For example, defining the preamble as a carriage return and a line feed causes each decoded message to be displayed on its own line. Defines a one to four character data string that can be added to the front of the decoded data. <K141,status,preamble character(s)> Disabled 0 = Disabled 1 = Enabled (within any protocol)

Preamble Characters Serial Cmd: Default: Options:

<K141,status,preamble character(s)> ^M corresponds to: carriage return. To enter control characters within a serial command, hold down the Ctrl key while typing the desired character. Example: <K141,1,CNTL-m> to enter the control character ^M.

Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K141h,,3C> For >: <K141h,,3E> For , : <K141h,,2C> See ASCII Character Entry Modifier on page 4-33 for a detailed explanation of how to enter ASCII characters as hex values.

4-34

EZ Match User Manual

Page 103

Communications

Postamble Postamble Status Usage: Definition: Serial Cmd: Default: Options:

Useful for identifying and controlling incoming data. For example, defining the postamble as a carriage return and a line feed causes each decoded message to be displayed on its own line. Allows the user to enable or disable up to four postamble characters that can be added to the end of the decoded data. <K142,status,postamble character(s)> Enabled 0 = Disabled 1 = Enabled (within any protocol)

Postamble Characters Serial Cmd: Default: Options:

<K142,status,postamble character(s)> ^M^J corresponds to: carriage return/line feed. To enter control characters within a serial command, hold down the control key while typing the desired character. Example: <K142,1,CNTL-m CNTL-j> to enter ^M^J.

Important: The ASCII characters <, >, and , can only be entered as hex pairs: For <: <K142h,,3C> For >: <K142h,,3E> For , : <K142h,,2C> See ASCII Character Entry Modifier on page 4-33 for a detailed explanation of how to enter ASCII characters as hex values.

EZ Match User Manual

4-35

Page 104

Postamble

4-36

EZ Match User Manual

Page 105

5 Calibration Contents Calibration Serial Commands ....................................................................................................... 5-2 Calibration Overview..................................................................................................................... 5-2 Calibration Options ....................................................................................................................... 5-3 Calibration by ESP..................................................................................................................... 5-10 Initiating Calibration ................................................................................................................... 5-12 Additional Notes about Calibration............................................................................................. 5-17

This section shows the EZ Match Imager’s calibration options, and explains the different ways that those options can be configured.

EZ Match User Manual

5-1

Page 106

Calibration Serial Commands

Calibration Serial Commands Calibration Options Calibration

<K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> <@CAL>

Calibration Overview Calibration is one of the most powerful features of the EZ Match Imager. The calibration process can be initiated by serial command, EZ Button, or ESP user interface. When the imager enters calibration, it runs through an optimization cycle that sets the ideal parameters for reading symbols at the highest possible level of performance. Calibration can be specially configured to optimize specific parameters, such as gain, shutter speed, and symbol type. The calibration process consists of five separate phases: search pass, medium pass, and fine-tune pass. During this process the imager quickly locates configuration setup, defines a calibration range for the configuration parameters, and then tunes those parameters for optimal performance.

5-2

EZ Match User Manual

Page 107

Calibration

Calibration Options This command specifies the operation of the imager’s calibration feature. The default configuration is set up to perform calibration on gain and symbol type. The default shutter speed setting is disabled, because not all applications require shutter speed, and when it is required, the application-specific settings tend to vary widely. Note: Background Color (page 11-25) is part of the calibration process regardless of the calibration setup. After successful completion of the calibration routine, the background color will be set to the appropriate value, depending on the symbol in the field of view. The Mirrored Image parameter (page 11-26) is not part of the calibration process, and it must be configured appropriately prior to calibration.

Gain Definition: Serial Cmd: Default: Options:

When enabled, gain will be calibrated to provide the best available image quality and performance. When disabled, gain is fixed and is not part of the calibration process. <K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> Enabled 0 = Disabled 1 = Enabled

Shutter Speed Unless an application is a static setup, the shutter speed setting should be configured by the user based on the application. For dynamic applications, the user should configure the shutter speed setting so the calibration process can optimize the gain for that setting. Serial Cmd: <K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> Default: Disabled Options: 0 = Disabled 1 = Enabled 2 = Fast Shutter Note: The following table shows general guidelines for shutter speed configurations based on various line speeds. The configurations shown depend on the imager’s optical configuration, and on symbol element size. Definition:

Shutter Speed 0 - 250 250 - 750 750 - 1500 1500 - 2000 2000 - 2500

EZ Match User Manual

Line Speed static 5” / sec 10” / sec 15” / sec 20” / sec

5-3

Page 108

Calibration Options

Disabled (Shutter Speed) When this feature is disabled, the shutter speed is fixed and is not part of the calibration process.

Enabled (Shutter Speed) When this feature is enabled, shutter speed will be calibrated to provide the best possible image quality and performance.

Fast Shutter The calibration process will concentrate on achieving the fastest possible shutter setting that will still provide good performance. The image quality or contrast may not be as good as what would be achieved with the Enabled setting. The calibration process is not designed to choose the fastest shutter speed that can decode a symbol, but rather to optimize for the fastest shutter speed that still provides good image quality.

5-4

EZ Match User Manual

Page 109

Calibration

Symbol Type Serial Cmd: Default: Options:

<K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> Enabled 0 = Disabled 1 = Enabled

Disabled When this feature is disabled, only the currently-enabled symbologies will be considered during the calibration process.

Enabled When this feature is enabled, autodiscrimination will be in effect during the calibration process. All symbologies that are readable by the imager except PDF417 and Pharmacode will be considered during calibration. Any new symbologies successfully decoded during calibration will remain enabled at the end of the process. All enabled symbologies will remain enabled. For example, assume that only Code 39 is enabled at the beginning of calibration. If a Code 128 symbol is read during calibration, then Code 128, as well as Code 39, will be enabled after the calibration process is complete.

EZ Match User Manual

5-5

Page 110

Calibration Options

Window of Interest (WOI) Framing Definition:

Serial Cmd: Default: Options:

Configured WOI

If a WOI framing mode is enabled, it will set the camera WOI to a full-size image when calibration begins. Once a symbol is decoded, the camera WOI will be zoomed-in both vertically and horizontally (regardless of which WOI mode is enabled) to include the symbol plus an additional margin area. This is done to accelerate the calibration process. When a successful calibration process is completed, the camera WOI will be adjusted according to the mode enabled. Otherwise, the original WOI configuration will be retained. <K529,gain,shutter speed,symbol type,WOI framing,WOI margin,line scan height,processing> Disabled 0 = Disabled 1 = Row and Column 2 = Row 3 = Column 4 = Straight Line 5 = Straight Line Framed

Starting WOI

After Decode

Calibration Passed

Calibration Failed

If WOI framing is not enabled, the current WOI configuration will be used until a symbol has been decoded. After a symbol is decoded, the WOI is framed exactly as it would be if a WOI framing mode was enabled. When the calibration process is complete, the original WOI configuration is restored. Refer to the diagram below:

Configured WOI

5-6

After Decode

Calibration Passed

Calibration Failed

EZ Match User Manual