Page 61
Read Cycle
Trigger Note: When doing calibration or read rate testing, the current trigger setting will be disregarded. Definition:
The type of trigger event that will initiate the read cycle.
Trigger Mode Serial Cmd:
<K200,trigger mode,trigger filter duration>
Default:
External Edge
Options:
0 = Continuous Read 2 = External Level 4 = Serial Data
1 = Continuous Read 1 Output 3 = External Edge 5 = Serial Data & External Edge
Continuous Read Usage:
Continuous Read is useful in testing symbol readability or reader functions. It is not recommended for normal operations.
Definition:
In Continuous Read, trigger input options are disabled, the reader is always in the read cycle, and it will attempt to decode and transmit every capture. When To Output and Noread options have no affect on Continuous Read.
Serial Cmd:
<K200,0>
Quadrus EZ Reader User’s Manual
3-7
Page 62
Trigger
Continuous Read 1 Output Usage:
Continuous Read 1 Output can be useful in applications where it is not feasible to use a trigger and all symbols contain different information. It is also effective in applications where the objects are presented by hand.
Definition:
In Continuous Read 1 Output the reader self-triggers whenever it decodes a new symbol or a timeout occurs. If End Of Read Cycle is set to Timeout and the symbol doesn't change, the output is repeated at the end of each timeout period. For example, if Timeout is set to one second, the reader sends the symbol data immediately and repeats the output at intervals of one second for as long as the symbol continues to be captured. If End Of Read Cycle is set to New Trigger, the reader will send the current symbol data immediately, but send it only once. A new symbol appearing in the reader’s range will be read and sent immediately provided it is not identical to the previous symbol read.
Serial Cmd:
<K200,1>
Caution: In automated environments, Continuous Read 1 Output is not recommended because there is no one to verify that a symbol was missed. Note: If Trigger Mode is set to Continuous Read 1 Output, Number of Symbols will default to 1 (if set to any number greater than 1).
3-8
Quadrus EZ Reader User’s Manual
Page 63
Read Cycle
External Trigger Level Usage:
This mode is effective in an application where the speeds of the conveying apparatus are variable and the time the reader spends reading each object is not predictable. It also allows the user to determine if a noread has occurred.
Definition:
External Trigger Level allows the read cycle (active state) to begin when a trigger (change of state) from an external sensing device is received. The read cycle endures until the object moves out of the sensor range and the active trigger state changes again.
Serial Cmd:
<K200,2>
Important: Level and Edge apply to the active logic state (Active Off (IOFF) or Active On (ION)) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object. Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle. Associated waveforms assume External Trigger State is set to Active On.
End Read Cycle: The same object, moving out of the detector beam, causes another change in the trigger state, ending the read cycle.
Trigger Level
Quadrus EZ Reader User’s Manual
3-9
Page 64
Trigger
External Trigger Edge Usage:
This mode is highly recommended in any application where conveying speed is constant or if spacing, object size, or timeouts are consistent.
Definition:
External Trigger Edge, as with Level, allows the read cycle (active state) to begin when a trigger (change of state) from an external sensing device is received. However, the passing of an object out of sensor range does not end the read cycle. The read cycle ends with a good read output or, depending on End of Read Cycle setting, a timeout or new trigger occurs.
Serial Cmd:
<K200,3>
Important: Level and Edge apply to the active logic state (Active Off (IOFF) or Active On (ION)) that exists while the object is in a read cycle, between the rising edge and falling edge. Rising edge is the trigger signal associated with the appearance of an object. Falling edge is the trigger signal associated with the subsequent disappearance of the object. Initiate Read Cycle: Object # 1, moving in front of the detector beam, causes a change in the trigger state, initiating the read cycle. Associated waveforms assume External Trigger State is set to Active On.
Initiate Second Read Cycle: Object # 2, moving in front of the detector beam, causes another change in the trigger state. This signal initiates a new read cycle and ends the previous read cycle unless Timeout is enabled and a good read or timeout has not occurred.
Trigger Edge
3-10
Quadrus EZ Reader User’s Manual
Page 65
Read Cycle
Serial Data Usage:
Serial Data is effective in a highly controlled environment where the host knows precisely when the object is in the field of view. It is also useful in determining if a noread has occurred.
Definition:
In Serial Data, the reader accepts an ASCII character from the host or controlling device as a trigger to start a read cycle. A Serial Data trigger behaves the same as an External Edge trigger. Serial commands are entered inside corner brackets, such as <t>.
Serial Cmd:
<K200,4>
Note: In Serial Data, sending a non-delimited start serial character will start a read cycle; however a non-delimited stop serial character has no effect.
Serial Data or External Edge
Usage:
Serial Data or External Edge is seldom used but can be useful in an application that primarily uses an external sensing device but occasionally needs to be manually triggered. An auxiliary terminal can be connected to the aux port so the user can send the serial trigger character through the reader to the host.
Definition:
In this mode the reader accepts either a serial ASCII character or an external trigger pulse to start the read cycle.
Serial Cmd:
<K200,5>
Note: In Serial Data or External Edge, sending a non-delimited start serial character will start a read cycle; however a non-delimited stop serial character has no effect.
Quadrus EZ Reader User’s Manual
3-11
Page 66
Trigger
Trigger Filter Duration Usage:
Trigger Filter Duration can be used to help the unit trigger more consistently with an unstable external trigger.
Definition:
Trigger Filter Duration can prevent trigger bounce from falsely triggering the reader by limiting the time during which trigger pulses can be received.
Serial Cmd:
<K200,trigger mode,trigger filter duration>
Default:
313 (x 32µs = 10.24mS)
Options: 0 to 65535 (corresponding to 0 to 2.097seconds in 32µs steps) If the unit is in External Edge trigger mode, then Trigger Filter Duration specifies the time following an edge that the reader will not consider another edge as a valid trigger. If the unit is in External Level mode, then Trigger Filter Duration specifies the time following the active edge that the trigger must be sampled once again as active before considering it a valid level trigger.
External Trigger State Usage:
Allows users to select the trigger state that will operate with their systems. (If using an object detector, use Active Off.)
Definition:
When enabled for Active On (ION) the triggering device imposes a current on the optoisolator to activate the read cycle. When enabled for Active Off (IOFF) the triggering device interrupts the current to the optoisolator to activate the read cycle.
Serial Cmd:
<K202,external trigger state>
Default:
Active On
Options: 0 = Active Off 1 = Active On Note: External Level, External Edge, or Serial Data or Edge triggering mode must be enabled for External Trigger to take effect.
3-12
Quadrus EZ Reader User’s Manual
Page 67
Read Cycle
Serial Trigger Usage:
Allows the user to define the trigger character and delimiters that start and stop the read cycle.
Definition:
A serial trigger is considered an online host command and requires the same command format as all host commands: that is, to be entered within angle brackets delimiters < > or in the case of non-delimited triggers, define individual start and stop characters.
Serial Trigger Character (delimited) Usage:
Allows the user to define the trigger character that initiates the read cycle.
Definition:
A single ASCII host serial trigger character that initiates the read cycle. A delimited trigger character is one that either starts or ends the read cycle and is enclosed by delimiters such as < and >.
Serial Cmd:
<K201,serial trigger character>
Default:
Spacebar (corresponds to <SP> displayed in the embedded menu)
Any single ASCII character, including control characters, except NUL (00H), an existing host command character, or an on-line protocol character. Control Options: characters entered on the command line are displayed in the menu as mnemonic characters. Note: Serial Data or Serial Data or Edge triggering mode must be enabled for Serial Trigger Character to take effect.
Start and Stop Trigger Characters (non-delimited) Usage:
It is useful in applications where different characters are required to start and end a read cycle.
Definition:
A non-delimited trigger character is one that either starts or ends the read cycle and is NOT enclosed by delimiters such as < and >. Both Start and Stop non-delimited characters can be defined and will function according to the trigger event, as follows: When defining Start and Stop trigger characters, the following rules apply: • In External Edge the reader looks only for the start trigger character and ignores any end trigger character that may be defined. • In External Level, the start trigger character begins the read cycle and end trigger character ends it. Note that even after a symbol has been decoded and the symbol data transmitted, the reader remains in External Level trigger read cycle until a Stop character is received. • In Serial Data & Edge trigger mode, command, either a start trigger character or a hardware trigger can start an edge trigger read cycle.
Quadrus EZ Reader User’s Manual
3-13
Page 68
Serial Trigger
Start Character (non-delimited) Definition:
A single ASCII host serial trigger character that starts the read cycle and is not enclosed by delimiters such as < and >.
Serial Cmd:
<K229,start character>
Default:
NULL (00 in hex) (disabled)
Options:
Two hex digits representing an ASCII character except <, >, XON, and XOFF.
Stop Character (non-delimited) Usage:
It is useful in applications where different characters are required to start and end a read cycle.
Definition:
A single ASCII host serial trigger character that ends the read cycle and is not enclosed by delimiters such as < and >.
Serial Cmd:
<K230,stop character>
Default:
NULL (00 in hex) (disabled)
Options:
Two hex digits representing an ASCII character except <, >, XON, and XOFF.
3-14
Quadrus EZ Reader User’s Manual
Page 69
Read Cycle
End of Read Cycle Definition:
The read cycle is the time during which the reader will attempt to capture and decode a symbol. A read cycle can be ended by a timeout, a new trigger, or by the last frame in a capture sequence or a combination of the above.
End of Read Cycle Mode Note: When operating in Continuous Read or Continuous Read 1 Output, the reader is always in the read cycle. Serial Cmd:
<K220,end of read cycle,read cycle timeout>
Default:
Timeout
Options:
0 = Timeout 2 = Timeout or New Trigger 4 = Last Frame or New Trigger
1 = New Trigger 3 = Last Frame
Timeout
Usage:
Typically used with Serial Data or Edge Trigger and Continuous One Output. It is effective in highly controlled applications when the maximum length of time between objects can be predicted. It assures that a read cycle ends before the next bar-coded object appears, giving the system extra time to decode and transmit the data to the host.
Definition:
Timeout ends the read cycle, causing the reader to stop reading symbols and send the symbol data or noread message when the time set in Timeout elapses (times out), if When to Output is set to End of Read Cycle. If in Continuous Read 1 Output, a timeout initiates a new read cycle and allows the same symbol to be read again. With either External Edge, Serial Data, or Serial Data & Edge enabled, a timeout ends the read cycle and symbol data or a noread message is sent to the host. With External Level enabled, the read cycle does not end until the falling edge trigger or a timeout occurs. The next read cycle does not begin until the next rising edge trigger.
Quadrus EZ Reader User’s Manual
3-15
Page 70
End of Read Cycle
New Trigger Usage:
New Trigger is an effective way to end a read cycle when objects move past the reader at irregular intervals (not timing dependent).
Definition:
New Trigger ends the current read cycle and initiates a new one when a new trigger occurs. New Trigger refers only to a “rising edge” trigger. With either External Edge, Serial, or Serial or Edge enabled, an edge or serial trigger ends a read cycle and initiates the next read cycle. In the case of External Level, a falling edge trigger ends the read cycle but the next read cycle does not begin until the occurrence of the next rising edge trigger.
Timeout or New Trigger Usage:
Useful in applications that require an alternative way to end the read cycle. For example, if an assembly line should stop completely or the intervals between objects are highly irregular.
Definition:
Timeout or New Trigger is identical to Timeout except that a timeout or a new trigger (whichever occurs first) ends the read cycle.
Last Frame Usage:
Useful in applications in which the number of captures needed can be defined but the timeout duration varies.
Definition:
Last Frame only applies to Rapid Capture mode.
Last Frame or New Trigger Usage:
Useful in applications in which line speeds are irregular and a new labeled object could appear before the last frame in a Rapid Capture sequence.
Definition:
New Trigger or Last Frame is identical to New Trigger except that a new trigger or last frame (whichever occurs first) ends the read cycle.
Read Cycle Timeout Definition:
Read Cycle Timeout is the duration of the read cycle.
Serial Cmd:
<K220, end of read cycle,read cycle timeout>
Default:
100 (x 10mS)
Options:
0 to 65,535
3-16
Quadrus EZ Reader User’s Manual