The text of a manual · part 25 of 37
Quadrus EZ Reader User's Manual, pages 241 to 250
Pages 241 to 250 of 366
The original document is the PDF (2.4 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
Filed in the catalog under the Microscan EZ 6700 EX
Page 241
IP Database
Threshold Value Serial Cmd:
<K250,CCD database index,shutter speed,gain,threshold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
0 to 255
Processing Mode Serial Cmd:
<K250,CCD database index,shutter speed,gain,threshold mode,threshold value,processing mode,background color,narrow margins>
Default:
Standard
Options:
0 = Standard 3 = Mode 3
1 = Mode 1 4 = Mode 4
2 = Mode 2 5 = Fast Linear Mode
Background Color Serial Cmd:
<K250,CCD database index,shutter speed,gain,threshold mode,threshold value,processing mode,background color,narrow margins>
Default:
White
Options:
0 = White
1 = Black
Narrow Margin Status Serial Cmd:
<K250,CCD database index,shutter speed,gain,threshold mode,threshold value,processing mode,background color,narrow margins>
Default:
Disabled
Options:
0 = Disabled
Quadrus EZ Reader User’s Manual
1 = Enabled
10-9
Page 242
CMOS Image Sensor
CMOS Image Sensor The database indexes (arrays) of camera/image processing settings that are applied when one or more indexes are made active in IP Database Size. Note: The following commands apply to the CMOS reader only.
CMOS Database Index Definition:
Defines the specific database index (1 to 10) of settings that will be applied during operations. Each index (1 to 10) will have its own settings, for example <K251,1...>, <K251,2...>, etc.
Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threshold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
1 to 10
Shutter Speed Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
50 to 50,000
Gain Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
0 to 63
10-10
Quadrus EZ Reader User’s Manual
Page 243
IP Database
Contrast Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
0 to 7
Offset Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
0 to 63
Threshold Mode Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
Adaptive
Options:
0 = Adaptive 1 = Fixed
Threshold Value Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
0
Options:
0 to 255
Processing Mode Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
Standard
Options:
0 = Standard 3 = Mode 3
Quadrus EZ Reader User’s Manual
1 = Mode 1 4 = Mode 4
2 = Mode 2 5 = Fast Linear Mode
10-11
Page 244
CMOS Image Sensor
Background Color Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
White
Options:
0 = White
1 = Black
Narrow Margins Serial Cmd:
<K251,CMOS database index,shutter speed,gain,contrast,offset,threhold mode,threshold value,processing mode,background color,narrow margins>
Default:
Disabled
Options:
0 = Disabled
10-12
1 = Enabled
Quadrus EZ Reader User’s Manual
Page 245
IP Database
Save Current Settings to Database Saves settings currently enabled to the designated database index.
CCD Serial Cmd:
<K250+,database index>
CMOS Serial Cmd:
<K251+,database index>
Load Current Settings from Database Loads selected database index settings into current reader settings.
CCD Serial Cmd:
<K250-,database index>
CMOS Serial Cmd: <K251-,database index> Example: <K251-,5> loads settings from database index # 5.
Quadrus EZ Reader User’s Manual
10-13
Page 246
Request Database Settings
Request Database Settings Returns settings for selected database index.
CCD Serial Cmd:
<K250?,database index>
CMOS Serial Cmd:
<K251?,database index>
Request All Database Settings Returns settings for the entire database.
CCD Serial Cmd:
<K250?>
CMOS Serial Cmd:
10-14
<K251?>
Quadrus EZ Reader User’s Manual
Page 247
11 Terminal Contents Terminal Window ........................................................................................................................11-2 Find Function ..............................................................................................................................11-3 Macros ........................................................................................................................................11-4 Terminal Window Menus.............................................................................................................11-5
This section describes the Terminal window and macro functions. Note: The characters NULL, <, >, and , can only be entered through embedded menus, not through ESP or serial commands. Note: You can learn the current setting of any parameter by inserting a question mark after the number, as in <K100?>. To see all “K” commands, send <K?>.
Quadrus EZ Reader User’s Manual
11-1
Page 248
Terminal Window
Terminal Window To use ESP’s Terminal, click on the Terminal button:
You will see this window: Clear screen
Macros listed on this bar
Copy, paste, save
Type text here to find matching text in the Terminal window.
Terminal screen
Click on Macros arrow to Add Macro, Remove Macro, or Edit Macro.
Type serial commands here.
Click on desired Macro to run.
The Terminal screen allows you to send serial commands to the reader by using Macros, by copying and pasting, or by typing commands in the “Send” text field. The Terminal screen also displays symbol data or information from the reader. You can also right click on the Terminal screen to bring up a menu of further options. 11-2
Quadrus EZ Reader User’s Manual
Page 249
Terminal
Find Function The Find box allows you to enter text strings to be searched for in the Terminal window. For example, a series of symbols have been read into the Terminal view and you want to determine if a particular code starting with “ABC” has been read. 1. Type ABC into the Find box.
2. Press Enter. The first instance of ABC will be highlighted in the Terminal window. 3. Press the F3 key to search again for the next instance of ABC. 4. Press Shift-F3 to search for the previous instance of ABC.
Quadrus EZ Reader User’s Manual
11-3
Page 250
Macros
Macros Macros can be stored in a macro selection bar, edited in a separate window, and executed by clicking on the macro name.
Click on Next Row to see the next row of macros
Click on the first arrow here to see Add Macro or Default Macros. When you default macros, the entire macros set is restored to their original macro commands.
Click on subsequent arrows to edit macros.
When you click on the macro name, the macro is executed in the Terminal window. If this is a command, it is sent to the reader at the same time that it is displayed.
Editing a Macro When you click the arrow next to any macro and select Edit, the following appears:
You can edit an existing macro or type in the Macro Name text field and define it in the Macro Value text field. Click OK.
11-4
Quadrus EZ Reader User’s Manual