Page 71

Protocol Selected Protocol If selecting one of the options from 0 to 4 (Point-to-Point, Point-to-Point with RTS/CTS, Point-to-Point with XON/XOFF, Point-to-Point with RTS/CTS and XON/XOFF, or Polling Mode D), use this format: Format: <Kfprotocol> protocol: 0 = Point-to-Point 1 = Point-to-Point with XON/XOFF 3 = Point-to-Point with RTS/CTS and XON/XOFF

4 = Polling Mode D 5 = Multidrop (requires address) 6 = User Defined 7 = User Defined Multidrop

If selecting Multidrop (5), you must define an address and append it to the command string. Format: <Kf5,address> protocol: 5 = Multidrop address: Any number from 1 to 50

To change the Protocol to Multidrop with an address of 33, enter <Kf5,33>. If selecting User Defined (6) or User Defined Multidrop (7), complete the format by either choosing new parameters or concatenating unchanged data fields (separate by commas). Note: For User Defined, first select Point-to-Point <Kf0> and then User Defined <Kf6...>. For user Defined Multidrop, first select Multidrop <Kf5>, then User Defined Multidrop <Kf7...>. Format: <Kfprotocol,RES,address,REQ,EOT,STX,ETX,ACK,NAK> protocol: 6 = User Defined; From Host: 0 = disabled 1 = enabled To select an unpolled ACK/NAK User Defined protocol with LRC disabled, enter <Kf0><Kf6,,,,,,,^F,^U><Kc0>. ACK and NAK will be displayed in the menu.<Kf0> nulls the address and <Kc0> disables LRC. Address, can be assigned any ASCII character except a null. Control characters are used to define RES through NAK (except Address).Table 3-1 lists the control characters Quadrus 2D Code Reader User’s Manual

3-7

3–Serial Commands

To change the Protocol to Point-to-Point with XON/XOFF, enter <Kf2>

Page 72

Chapter 3 used for these data fields. (Refer to Table A-7, “ASCII Table with Control Characters,” on page A-12 for more information.) Format: <Kfprotocol,RES,address,REQ,EOT,STX,ETX,ACK,NAK> protocol: 7 = User Defined Multidrop To select an unpolled ACK/NAK User Defined protocol with LRC disabled, enter <Kf0><Kf6,,,,,,,^F,^U><Kc0>. <Kf0> nulls the address and <Kc0> disables LRC. ACK and NAK will be displayed in the menu. 1 Note: Address, can be assigned any ASCII character except a null. Control characters are used to define RES through NAK (except Address). Table 3-1 lists the control characters used for these data fields. (Refer to Table A-7, “ASCII Table with Control Characters,” on page A-12 for more information.)

3–Serial Commands

Table 3-1 Protocol Commands

Protocol Command (Mnemonic displayed on Microscan menu)

Control Characters (Entered in menu or serial command)

Effect of Command

RES

^D

Reset

REQ

^E

Request

EOT

^D

Reset

STX

^B

Start of Text

ETX

^C

End of Text

ACK

^F

Acknowledge

NAK

^U

Negative Acknowledge

Preamble Format: <Kdstatus,preamble character(s)> status: 0 = Disabled 1 = Enabled preamble character(s): Enter one to four preamble characters Default is CR

1. A control character, although conventionally represented here and in the ASCII table on page A-12 as two characters (^F or ^U, etc.), is actually a single ASCII character that is entered on the keyboard by holding down the control key while pressing the desired letter.

3-8

Quadrus 2D Code Reader User’s Manual

Page 73

Postamble <Kestatus,postamble character(s)> status: 0 = Disabled 1 = Enabled postamble character(s): Enter one to four postamble characters Default is CR, LF

Response Timeout Format: <KAtimeout setting> Any number from 0 to 255 in 10 ms increments Default is 4 (40 ms) Note: A zero creates an indefinite wait. Allows you to set the time the reader will wait before timing out if ACK, NAK, and ETX are enabled, and a host response is expected.

Format: <KBtime interval> Any number from 0 to 255 in 1ms increments. Default is 0 Allows you to set the time interval in milliseconds between individual characters transmitted from the Quadrus™ to the host computer.

Longitudinal Redundancy Check (LRC) Format: <Kcstatus> 0 = Disabled 1 = Enabled An error-checking routine that verifies the accuracy of transmissions.

Communications/Protocol Status Request Format: <KT?> Returns status of each command in the group.

Quadrus 2D Code Reader User’s Manual

3-9

3–Serial Commands

Intercharacter Delay

Page 74

Chapter 3

Operations Trigger Format: <Kgmode,trigger filter time,single shot,multi-capture, # captures after trigger> mode: 0 = Continuous Read 1 = Continuous Read 1 Output 2 = External Level single shot: 0 = Disabled 1 = Enabled

3 = External Edge multi-capture: 0 = Disabled 1 = Enabled

3–Serial Commands

4 = Serial Data 5 = Serial Data & Edge trigger filter time: Any number between 0 - 65,535 Default is 217

# captures after trigger: Any number between 1 - 10 Default is 1

Mode Allows you to establish the type of trigger event that initiates or ends the read cycle.

Trigger Filter Time Any number between 0 and 65,535 (3 seconds) in 46 µs increments. Allows you to set the duration of the trigger debounce time.

Edge Single Shot When enabled the read cycle ends after a single captured image, whether or not it is successfully decoded. This is valid only if External Edge trigger is enabled.

Multi-Capture If enabled, multiple captures can proceed from a single trigger.

# of Captures after Trigger Number of captures that can occur after a trigger if Multi-Capture is enabled.

3-10

Quadrus 2D Code Reader User’s Manual

Page 75

Serial Trigger Character Format: <Kitrigger character> Default is ^] Allows you to define a single ASCII character as the host serial trigger character that initiates the read cycle.

Trigger Variance Format: <KJtrigger variance mode> 0 = Fixed 1 = Auto Applies only if Two Triggers is selected. You can choose Fixed to apply the same Trigger Delay Time to each read cycle or you can choose Auto to have Quadrus™ compare subsequent edge triggers and automatically adjust Trigger Delay Time dynamically. (See Important Note under Trigger Source.)

Format: <Kjtrigger polarity> 0 = Negative 1 = Positive Allows you to determine whether a positive or negative transition will initiate the read cycle.

Trigger Source Format: <KItrigger source> 0 = One Trigger 1 = Two Trigger Allows you to select one or two trigger sources.

Quadrus 2D Code Reader User’s Manual

3-11

3–Serial Commands

Trigger Polarity

Page 76

Chapter 3 Trigger Delay Format: <KGtrigger delay type,trigger delay time,# pulses> trigger delay type: trigger delay time: # pulses: Any number from 0 to 65,535 Any number from 0 = Time (in 46 µs Increments) 0 to 255. 1 = Pulse Default is 0 Default is 0 If you select Time for the Trigger Delay Type, at the occurrence of an Edge trigger, the Quadrus™ will wait the designated time before the read cycle is activated. If you select Pulse, the Quadrus™ will wait the designated number of pulses before the read cycle is activated.

Read Cycle

3–Serial Commands

Format: <Khend of read cycle,timeout> end of read cycle: 0 = Timeout 1 = New Trigger 2 = Timeout & New Trigger timeout: Any number from 0 to 65,535 Default is 100 (in 10 ms Increments)

Allows you select the conditions for ending the read cycle and the timeout duration, if a timeout is selected.

Noread Message Format: <Kkstatus,noread message> status: 0 = Disabled 1 = Enabled noread message: Any ASCII string up to 7 digits. Default NOREAD

Allows you to enable (or disable) and define the noread message.

Multilabel Format: <KLnumber of labels,multilabel separator> number of labels: Any number between 1 and 6 Default is 1 multilabel separator: Any valid ASCII character Default is ,

Allows you to define the number of different labels that the reader will read during a single trigger event. The labels can be a mix of all enabled bar code symbologies and more than one label can be decoded per frame.

3-12

Quadrus 2D Code Reader User’s Manual

Page 77

Match Code Format: <Kntype,sequential matching,match start position,match length,wild card character,sequence on noread,sequence on every mismatch> type: 0 = Disabled 1 = Enabled 2 = Sequential match length: 1 to Max length of allowed. Default is 1a sequence on noread: 0 = Disabled 1 = Enabled wild card character: Any ASCII character Default is * sequence on every mismatch: 0 = Disabled 1 = Enabled

3 = Wild Card sequential matching: 0 = Increment 1 = Decrement match start position: 0 to Max length allowed. Default is 0a

When set to other than Disabled, you can receive a master label into the reader’s memory to be compared with subsequently scanned labels. The results of this comparison determine the data display and output pulses, if enabled.

Master Label Database Size Format: <KMnumber of master labels> You can select and define from 1 to 10 master labels with this command or by entering or reading in master labels directly in the Master Label dialogue box in “Master Label Data Base,” on page 4-12.

New Master Pin Format: <Kzstatus> 0 = Disabled 1 = Enabled When Match code is enabled, enabling New Master Pin will allow you to momentarily connect the New Master Pin to ground.

Quadrus 2D Code Reader User’s Manual

3-13

3–Serial Commands a. See “Master Label Data Base” on page 4-12 .

Page 78

Chapter 3 Reader Output Format: <Klbar code output,when to output> bar code output: 0 = Disabled 1 = Match 2 = Mismatch 3 = Good Read when to output: 0 = As soon as possible 1 = End of Read Cycle

Allows you to choose whether or not to output a good decode (label data) or a noread to the host. When disabled, a label is decoded and the read cycle transpires as usual, but neither label data nor the noread message is transmitted to the host. The read cycle counters are updated, and the current number of good reads or noreads can be obtained via operational commands.

Beeper Status 3–Serial Commands

Format: <Kustatus> 0 = Disabled 1 = On Good 2 = On Noread

3-14

Quadrus 2D Code Reader User’s Manual

Page 79

Symbol Symbol Type Format: <KQECC 200 status,ECC 000 status,ECC 050 status,ECC 080 status, ECC 100 status,ECC 140 status> status: (for each of the six matrix symbols identified above) 0 = Disabled 1 = Enabled default is ECC 200

Narrow Margins/Symbology Identifier Format: <Konarrow margins status,symbology identifier> symbology identifier status: 0 = Disabled 1 = Enabled

3–Serial Commands narrow margins status (not used): 0 = Disabled 1 = Enabled

If Symbology Identifier is enabled, it adds a prefix to the decoded data.

Quadrus 2D Code Reader User’s Manual

3-15

Page 80

Chapter 3

In/Out Allows you to choose the conditions that will allow the Quadrus™ to receive or output pulses and data to the host computer.

Output Filter Format: <KEstatus,start position,length> status: 0 = Disabled 1 = Enabled start position: 1 to 1556 Default is 1 length: 1 to 1556 Default is 1556

Input # 1

3–Serial Commands

Format: <KKmode,polarity> mode: 0 = Disabled 1 = Reset Counts 2 = Unlatch Output 3 = Set Trigger Delay polarity: 0 = Negative 1 = Positive

Input #2 Format: <KCmode,polarity> Functions the same as Input # 1, except that mode 3 option is not available.

Output # 1 Format: <Kvoutput on,polarity,pulse width,output mode trend analysis mode,trigger evaluation period,# to output on> output on:

polarity:

pulse width:

0 = Mismatch or Noread 1 = Good Match 2 = Mismatch

0 = Negative 1 = Positive

Any number from 0 - 255 (in 10 ms incr.) Default is 5

3 = Noread 4 = Trend Analysis

3-16

Quadrus 2D Code Reader User’s Manual