The text of a manual · part 14 of 17

Quadrus 2D Code Reader User's Manual, pages 131 to 140

Pages 131 to 140 of 166

The original document is the PDF (1.4 MB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.

Filed in the catalog under the Microscan 6500

Page 131

After your analysis, you can return to the Calibration menu, make changes in Illumination, then check read rates and resulting histograms and line scans. Be sure to save your settings to file when you arrive at the optimum reads.

Quadrus 2D Code Reader User’s Manual

6-7

6–Evaluation

Action

Page 132

6–Evaluation

Chapter 6

6-8

Quadrus 2D Code Reader User’s Manual

Page 133

A

Appendices Appendices

Contents Appendix A, Quadrus™ Specifications ............................................. A-2 Appendix B, Electrical Specifications ..............................................A-3 Appendix C, Transferring Files ........................................................A-5 Appendix D, Maximum Symbol Sizes ...................................................A-7 Appendix E, Optional Equipment ..................................................A-10 Appendix F, ASCII Table ................................................................A-12 Appendix G, Data Matrix Symbology ...............................................A-13 Appendix H, Histograms ......................................................................A-15 Appendix I, Multidrop Communications ....................................................A-17 Appendix J, Glossary of Terms .....................................................A-20

Quadrus 2D Reader User’s Manual

A-1

Page 134

Appendix A, Quadrus™ Specifications Mechanical Weight: 1 lbs. (0.5 kg)

2.96" 7.52 cm

Housing: IP65 Operating temperature: 0 to 40°C Humidity: 5% to 90% (non-condensing)

Light Source

Appendices

2.27" 5.77cm

Type: High Output LEDs

Software-adjustable strobe time

Light Collection CCD array, 659 x 494 pixels

M4 (4 pls) mounting*

1.00" 2.54 cm

Progressive scan, square pixel

Interface Protocols RS-232, RS-422, RS-485

0.97" 2.5 cm

Communication Protocols Point-to-point, Point-to-point w/RTS/CTS, Point-topoint w/XON/XOFF, Point-to-point w/RTS/CTS & XON/XOFF, Polling Mode D, Multidrop, User Defined, User Defined Multidrop

Electrical

5.14" 13.06 cm

Power requirement: 10 to 28 VDC, max. 3.5 W, max.

2.97" 7.54 cm

2A inrush current

Power Supplies:

97-100004-07 120 VAC version 97-100004-08 240 VAC version

0.81" 2.07 cm

Safety & Compliance Certifications Designed for: FCC, TüV, CEd, UL/cUL ISO 9001/Cert. No. 06-1080

Status Lights 3

2

1

Yellow Power LED: On power-on

2

Green Status LED: Normal operations: illuminates

4 25

1 14

13

1 whenever a data matrix label is decoded and remains on until a new trigger occurs. Continuous Read or Continuous Read 1 Output: flashes for each good read.

*Outputs sink 8 mA, source 400 µA **Inputs VIL = 0.8 V, VIH = 2.0 V (see 74 LS for further details) *** VIL = 2.0V, VIH = 3.0V

*Caution: Maximum penetration depth of screws is 0.200" (5.08 mm).

Figure A-1 Quadrus™ Dimensions

A-2

Quadrus 2D Reader User’s Manual

Page 135

Appendix B, Electrical Specifications Host Connector

Power Connector

Mates with 25-pin D-Subminiature socket.

Table A-1 Host Connector 1

Signal ground

2

Transmit data RS-232 (out)

3

Receive data RS-232 (in)

Mates with Microscan power pack PN 97-100001-14 or 97-100001-15, or with Switchcraft EN3C2F16K

Table A-2 Power Connector 1

Power 10 to 28 VDC (in)

2

Power ground

4

Request-to-send (out)

5

Clear-to-send (in)

6

Output-1 (out) (TTL level)*

7

Signal ground

8

Output-2 (out) (TTL level)*

1

9

Strobe (out)

2

+12 VDC (out) @50 mA maximum

10

Trigger Input (in) (max. 24 V)***

3

Ground

11

Default (in) (TTL level)**

4

Strobe (out)

12

Input-1 (in) (TTL level)**

13

RXD 485 + (in)

14

TXD 485 – (out)

Appendices

Trigger Connector Mates with Microscan Object Detector PN 99440002-01 or Switchcraft EN3C4MK

Table A-3 Trigger Connector

15

Output-3 (out) (TTL level)

16

RXD 485 – (in)

17

Power ground

18

Power 10 to 28 VDC (in)

19

TXD 485 + (out)

20

Aux transmit data RS-232 (out)

21

Input-2 (in) (TTL level)**

22

Ground

23

Aux receive data RS-232 (in)

24

+12VDC (out) @50 mA maximum

25

New master (in) (TTL level)**

*Outputs sink 8 mA, source 400 µA **Inputs VIL = 0.8 V, VIH = 2.0 V (see 74 LS for further details) *** VIL = 2.0V, VIH = 3.0V

Trigger (in)

(Power/Trigger Connector Kit 98-200025-03) Important Note: When using two triggers (see “Trigger Source” on page 2-17). You can wire two trigger devices to a single input (either pin 1 of the trigger connector or pin 10 of the host connector) provided that you are using open collector devices (“wire-or”); or you can wire two triggers inputs independently, the first to either pin 1 of the trigger connector or pin 10 of the host connector and the second to pin 12 of the host connector.

External object detector

Inputs to/outputs from internal signals

Quadrus

Host

To power supply Figure A-2 System Diagram

Quadrus 2D Reader User’s Manual

A-3

Page 136

COM

I/O

A-4

AC Source

Chassis Signal

VIN +

Scanner

Ground Loops

VIN +

Ground loops, signal degradation due to different ground potentials in communicating devices, can be eliminated or minimized by ensuring that both the host, scanner, and their power supplies are connected to a common earth ground.

+ –

+ –

AC Source

+ –

AC Source

RX– RX+ TX– TX+

RS-232 signals have a common signal ground (pin 4). Pin 4 is normally connected to pin 13 (chassis ground) in the scanner; however, under certain conditions (e.g., when potential differences exist between power outlet grounds) signal and chassis grounds can be isolated from each other inside the scanner by Microscan technicians.

VIN +

Host

< Control Signal >

Proper grounding is necessary for operator safety, noise reduction, and the protection of equipment from voltage transients. Buildings, including any steelwork, all circuits, and all junction boxes must be grounded directly to an earth ground in compliance with local and national electrical codes.

RX– RX+ TX– TX+

Appendices

Ground and Shield Considerations

Scanner

Figure A-3 Grounding Diagram, Scanner-Host

Quadrus 2D Reader User’s Manual

Page 137

Transferring Files

Appendix C, Transferring Files Receiving Settings from Quadrus™ If you want your host computer to receive the current settings from Quadrus™, click RECEIVE FROM READER.

from a File If you want to load the settings from a saved settings file to your host computer, open the desired file and then click SEND/NO SAVE and follow this with RECEIVE FROM READER. (Clicking RECEIVE FROM READER in this context ensures that the reader’s current settings will appear in your host settings list.)

Saving Current Configuration to Quadrus™ You can save in two ways: 1. Click Apply in the current menu page. Note: Selecting OK on these menu pages will only update the items in the Current Settings window. 2. From within the SEND/RECV box, click SEND/NO SAVE to initialize settings and reset counters only, or click SEND/SAVE to initialize, reset counters, and save these settings to the Quadrus™’s memory for power up.

to File You can save a file by selecting SAVE AS (or SAVE) from the File pull-down menu. The settings are saved only on the computer’s hard drive or other storage media. This is particularly useful for backing up files or for configuring multiple configurations.

Creating a New Configuration File To create a new file of configuration settings, first be sure you’ve saved the current configuration file to the computer’s hard drive or other storage media. Next pull down the File menu in the ESP™ and select New. A new Untitled window will open with a listing of the factory default settings. Change these settings to suit your requirements and save this file to the computer’s hard drive under a new name. You can also load, edit, and save an existing configuration file.

Quadrus 2D Reader User’s Manual

A-5

Appendices

If receiving scan data and you are in Continuous Read, symbols being scanned may interfere with your data request.

Page 138

Restoring Factory Default Settings Note: Be certain that you’ve saved any other configuration files to the host computer that you may want to reload later.

Defaulting by Menu Commands Default settings can be reset for individual menus pages or for all menus at once. To restore factory defaults for individual menus (Communications, Operations, Code Types, or Reader Settings),

Appendices

1. Click the DEFAULT button on the specific menu, 2. Click SEND/NO SAVE, and then RECEIVE FROM READER. To restore factory defaults for ALL menus, select the DEFAULT button on the Reset page of the Utilities menu.

Defaulting by New 1. Close and save your current file. 2. Select New in the File pull-down menu. The resulting File will have your original factory default settings. Give it a new name and save to it as you make changes. 3. Click SEND/NO SAVE, and then RECEIVE FROM READER. See “Software Reset,” on page 4-14 for a complete discussion of transfer options.

A-6

Quadrus 2D Reader User’s Manual

Page 139

Maximum Symbol Sizes

Appendix D, Maximum Symbol Sizes Each of the three Quadrus™ models - low density, high density, and ultra high density (UHD) - have finite fields of view that limit symbol size as well as the number of characters per element dimension. A symbol does not exceed the maximum symbol size if its diagonal dimension plus 25% is not greater than the Y dimension of the Quadrus™’s field of view. For example the diagonal dimension plus 25% of a 30 mil symbol with 22 x 22 elements is equal to 2.0" which is less than 1.5".

Appendices

Low Density Symbol Sizes Low Density F ield of V iew = 1.67" x 1.25"

Field of View X

Y

2.0" 1.5" 5.08 cm 3.81 cm

D iagon al (inch es) plu s 2 5%

D iagon al (inch es)

Figure A-4 Low Density Symbol Size Limitations

Table A-4 Maximum Symbol Size by Element (Low Density) Element

Square Symbols Rectangular Symbols

Row x Col. in elements

13 mil

15 mil

20 mil

25 mil

30 MIL

52 x 52

44 x 44

32 x 32

26 x 26

22 x 22

Diagonal symbol size plus 1.195" 1.167" 1.131" 1.149" 1.167" 25% 3.04 cm 2.96 cm 2.87 cm 2.92 cm 2.96 cm Row x Col. in elements

16 x 48

16 x 48

16 x 16

16 x 36

6 x 12

Diagonal symbol size plus 0.822" .949" 1.265" 1.231" 1.074" 25% 2.09 cm 2.41 cm 3.21 cm 3.13 cm 2.73 cm

Quadrus 2D Reader User’s Manual

A-7

Page 140

High Density Symbol Sizes High Density F ield of V iew = 1.0" x 0.75"

Field of View Y 0.75" 1.9 cm

D iagon al (inch es) plu s 2 5%

Appendices

X 1.0" 2.54 cm

D iagon al (inch es)

Figure A-5 High Density Size Limitations

Table A-5 Maximum Symbol Size by Element (High Density) Element 8 mil Row x Col. in elements Square Symbols

Row x Col. in elements Rectangular Symbols

A-8

25 mil

30 mil

52 x 52 40 x 40 32 x 32 26 x 26 20 x 20 16 x 16

14 x14

Diagonal symbol size .735" plus 25% 1.87 cm

10 mil .707" 1.80 cm

13 mil .735" 1.87 cm

15 mil .689" 1.75 cm

20 mil .707" 1.80 cm

16 x 48 16 x 48 16 x 36 16 x 36 12 x 26

Diagonal symbol size .506" plus 25% 1.29 cm

.632" 1.61 cm

.640" 1.63 cm

.739" 1.88 cm

Quadrus 2D Reader User’s Manual

.716" 1.82 cm

.707" .742" 1.80 1.88 cm cm 8 x 12

8 x 12

.451 .541 1.15 1.37 cm cm