Page 41
Operations Menu Allows you to choose when bar code data is sent to the host. Selecting:
Has this effect:
As Soon As Possible
Causes bar code data (good reads) to be transmitted immediately upon a good decode.
End of Read Cycle
Causes bar code data output to be delayed until the end of the read cycle.
Default: 1 Options: 1 to 31 Allows you to select the number of good reads (from 1 to 31) required per label before a good decode output. Note: Be sure to set the value within the determined scan rate for the scanning setup so that the decoder is capable of scanning a label the required number of times.
Match Code Default: Disabled Options: Disabled, Enabled Note: A triggered mode (page 2-14 to page 2-15) must be enabled for Match Code to take effect. Note: If Match Code is enabled with Continuous Read, the decoder defaults to Continuous Read 1 Output mode, and the label data must change before the decoder will output data again, unless a timeout, if enabled, occurs. Note: Enabling Match Code when Number of Labels is set to any number greater than one will cause Number of Labels to default back to one. Allows you to enter a master label into the decoder’s memory to be compared with subsequently scanned labels.
MS-3000 Single Head Decoder User’s Manual
2-19
2–Menu Config.
Number of Reads Before a Good Decode
Page 42
Chapter 2
Menu Configuration
2–Menu Config.
With Match Code enabled, a master label can be entered in three ways: 1. With New Master Pin enabled (see page 2-31), toggling pin 25 to ground (pin 7) enables the next good read to be the master label. 2. Sending serial command <G> enables the next good read to be the master label. 3. Sending serial command <)XXXX)> downloads data as master label. (Master label data is entered in place of the Xs.) Figure 2-5 shows the sequence of operation (and reference) for setting up and entering master labels. Enable Match Code: 1. by menu selection, 2. by operational command <E>, or 3. by serial configuration command <Kn1>.
* See the option/outputs available in the relay driver you have selected.
Enable New Master Pin: 1. by menu selection, or 2. by serial configuration command <Kz1>.
Enter the next label scanned as the master label: 1. by operational command <G>.
Download a new master label directly into memory: 1. by operational command <)XXXX)>.
Figure 2-5 Match Code Logic Diagram
2-20
MS-3000 Single Head Decoder User’s Manual
Toggle internal pin 25 to GND (pin 7) to enter next label scanned as the
Read in the master label.
Compare the master label with subsequent labels*
Page 43
Operations Menu
Number of Labels
Field Separator Default: , (comma) Options: Any available ASCII character, except NUL. Allows you to choose the separator character to be inserted between labels.
MS-3000 Single Head Decoder User’s Manual
2-21
2–Menu Config.
Default: 1 Options: 1 to 6 Note: If Number of Labels is set to any number greater than one while Match Code or Continuous Read 1 Output is enabled, Number of Labels will default back to one. Allows you to choose the number of different labels that will be read in a single trigger event. The labels can be a mix of any of the enabled bar code symbologies and more than one label can be decoded per scan line. The following conditions apply: 1. Each label must be different to be read. 2. The maximum number of characters in any one label is 31. 3. The maximum number of characters in a single scan line is 62. 4. The maximum number of characters for all labels is 128, including preamble and postamble and all spaces and commas. 5. All noread messages are posted at the end of the data string. 6. If more than one label is within the scan beam at the same time, label data may not be displayed in the order of appearance.
Page 44
Chapter 2
Menu Configuration
Code Types Menu The Code Types menu allows you to choose among five bar code types and define their parameters. – Bold text represents default settings.
Code Types Narrow Margins Status – Disabled
2–Menu Config.
– Enabled
Code 39 Status – Enabled – Disabled
Check Digit
Check Digit Output
– Disabled – Enabled
Code Length
Large Interchar Fixed Code Length Gap
– 10
– Disabled
– Disabled
– Disabled
– User Definable
Start & Stop Match
Start & Stop Output
Large Interchar Gap
Fixed Code Length
Code Length
– Enabled
– Enabled
– Enabled
– Enabled
(1 to 31)
– Enabled
Codabar Status – Disabled – Enabled
– Disabled
– Disabled
– Disabled – Enabled
– 10
– User
– Disabled
Definable (1 to 31)
– Enabled
I 2 of 5 Status – Disabled – Enabled
Check Digit – Disabled – Enabled
Check Digit Output – Disabled – Enabled
Code Length #1 – 10
– User Definable
(even, 0 to 30)
Code Length #2 – 6
– User Definable
(even, 0 to 30)
UPC Status – Disabled – Enabled
EAN
Supplementals
– Disabled – Enabled
– Disabled – Enabled
– Required
Separator – Enabled – Disabled
Code 128 Status – Disabled – Enabled
Fixed Code Length – Disabled – Enabled
Code Length – 10
– User Definable
(1 to 31)
Figure 2-6 Code Types Menu Structure
2-22
MS-3000 Single Head Decoder User’s Manual
Separator (character) – , (comma)
– User Definable
(ASCII char.)
Page 45
Code Types Menu Note: Options listed in the command line do not always appear in the same order as those posted in the first column of the menu. Options can be defined for any bar code type whether or not the bar code itself is enabled at the time.
Narrow Margins Status
Code 39 Status Default: Enabled Options: Enabled, Disabled
Check Digit Default: Disabled Options: Disabled, Enabled Code 39 is self-checking and does not normally require a check digit. However, for additional data integrity, a Modulus 43 check digit can be added to the bar code message. With Check Digit and an External or Serial trigger option enabled (see “Triggering Mode” on page 2-14), an invalid check digit calculation will cause a noread message to be transmitted at the end of the read cycle.
Check Digit Output Default: Disabled Options: Disabled, Enabled When enabled, the check digit character is sent along with the label data. When disabled, label data is sent without the check digit.
MS-3000 Single Head Decoder User’s Manual
2-23
2–Menu Config.
Default: Disabled Options: Disabled, Enabled Allows the decoder to read bar codes with quiet zones less than 10 times the narrow-bar-width. Quiet zone is a term used to describe the minimum space at the leading and trailing ends of a label. Each quiet zone can be as small as five times the narrow bar element when Narrow Margins is enabled.
Page 46
Chapter 2
Menu Configuration
Large Intercharacter Gap Default: Disabled Options: Disabled, Enabled Allows the decoder to read labels with gaps between bar code characters exceeding three times the narrow element width.
Fixed Code Length 2–Menu Config.
Default: Disabled Options: Disabled, Enabled Used to increase data integrity by ensuring that only one label length will be accepted.
Code Length Default: 10 Options: 1 to 31 Note: Fixed Code Length (page 2-24) must be enabled for Code Length to take effect. Allows you to specify the exact number of characters that the decoder will recognize (this does not include start and stop). The decoder will ignore any code not having the specified length.
Codabar Status Default: Disabled Options: Disabled, Enabled
Start & Stop Match Default: Enabled Options: Enabled, Disabled Requires the Codabar start and stop characters (a, b, c, or d) to match before a valid read can occur.
Start & Stop Output Default: Enabled Options: Enabled, Disabled Allows the start and stop characters to be transmitted with bar code data.
2-24
MS-3000 Single Head Decoder User’s Manual
Page 47
Code Types Menu
Large Intercharacter Gap Default: Disabled Options: Disabled, Enabled Allows the decoder to read labels with gaps between bar code characters exceeding three times the narrow element width.
Fixed Code Length
Code Length Default: 10 Options: 1 to 31 Note: Fixed Code Length (page 2-25) must be enabled for Code Length to take effect. Allows you to specify the exact number of characters that the decoder will recognize. The decoder will ignore any code not having the specified length.
Interleaved 2 of 5 Status Default: Disabled Options: Disabled, Enabled Because I 2 of 5 is a continuous code, it is prone to substitution errors. Hence, a code length must be defined and a bar code label containing an even number of digits must be used. It is also recommended that a Modulus 10 check digit be used to ensure the best possible data integrity.
Check Digit Default: Disabled Options: Disabled, Enabled I 2 of 5 uses a Modulus 10 check digit.
MS-3000 Single Head Decoder User’s Manual
2-25
2–Menu Config.
Default: Disabled Options: Disabled, Enabled Used to increase data integrity by ensuring that only label length will be accepted.
Page 48
Chapter 2
Menu Configuration
Check Digit Output Default: Disabled Options: Disabled, Enabled When enabled, the check digit character is sent along with the label data. When disabled, label data is sent without the check digit.
Code Length #1 2–Menu Config.
Default: 10 Options: 0 to 30, even. If you enter an odd number the decoder will use the next lower number. With I 2 of 5, two code lengths can be defined. When using only one label length in an application, we recommend setting Code Length #2 to 0 to ensure data integrity. If a check digit is used, it must be included in the code length count.
Code Length #2 Default: 6 Options: 0 to 30, even. If you enter an odd number the decoder will use the next lower number. If using a second label, you may also specify a zero or any even code length from 2 to 30. If not using a second label, set Code Length #2 to 0 to ensure data integrity.
UPC Status Default: Disabled Options: Disabled, Enabled When enabled, the decoder will read UPC version A and UPC version E only.
EAN Default: Disabled Options: Disabled, Enabled Note: UPC must be enabled for EAN to take effect. When EAN is enabled, the decoder will read UPC version A, UPC version E, EAN 13, and EAN 8. It will also append a leading zero to UPC version A label information and transmit 13 digits. If you do not want to transmit 13 digits when reading UPC version A labels, disable EAN.
2-26
MS-3000 Single Head Decoder User’s Manual
Page 49
Code Types Menu
Supplementals
Separator Default: Enabled Options: Enabled, Disabled Allows you to insert a character between the standard UPC or EAN code and the supplemental code.
Separator (character) Default: , (comma) Options: Any ASCII character. Allows you to change the separator character from a comma to a new value.
MS-3000 Single Head Decoder User’s Manual
2-27
2–Menu Config.
Default: Disabled Options: Disabled, Enabled, Required Note: Narrow Margins must be enabled if the gap between the standard code and the supplemental code is between 5:1 and 10:1 (in relation to narrow-bar-width). Allows the decoder to read supplemental bar code data that has been appended to the standard UPC or EAN codes. When set to Required, the decoder treats the supplemental data and the bar code label as a single label. Also, supplemental data must be found or a noread will result. When set to Enabled, the decoder treats the supplemental data and the bar code label as separate labels.
Page 50
Chapter 2
Menu Configuration
Code 128 Status Default: Disabled Options: Disabled, Enabled
Fixed Code Length 2–Menu Config.
Default: Disabled Options: Disabled, Enabled Allows you to increase data integrity by ensuring that only one label length will be accepted.
Code Length Default: 10 Options: 1 to 31 Note: Fixed Code Length must be enabled for Code Length to take effect. Allows you to specify the exact number of characters that the decoder will recognize. The decoder will ignore any code not having the specified length.
2-28
MS-3000 Single Head Decoder User’s Manual