Page 91
Operational Commands
Program Management Commands
<A> Initialize <A> Software Reset. Initializes all serial configuration commands in RAM and RAM resets all counters and operating parameROM NOVRAM ters. Changes for Baud Rate, Stop Bits, Code Length, etc. do not take effect until this command is sent. Note: Software Reset will cause the numeric counters in use to lose their count; record all data that you wish to save prior to sending this command. <D> Enter Configuration Mode. Enters the menu configuration program. See Chapter 2, “Menu Configuration.”1 Initialize <Z> <Z> Save Configuration for Power-on. Saves the current configuration to nonvolaRAM tile memory for availability on power-on. ROM NOVRAM The values of numeric counters are not saved by this command. Note: The <Z> Save Configuration command can be executed 10,000 times. In normal usage this will exceed the life of the decoder. If frequent changes to the operating parameters are required, the <Z> command should be used only when the current configuration has been changed and the changes are to be permanent.
Device Control Commands <Bdata> Echo Data to Aux Monitor. Echoes data from host to auxiliary monitor (see Appendix H, “Auxiliary Monitor,” on page A-13 for more detail). <H> Enable Laser Scanning (Laser On). <I> Disable Laser Scanning (Laser Off). This feature is useful during extended periods of time when no bar code labels are being scanned. Disabling laser scanning will not affect any downloaded commands to the decoder. The decoder remains active during this period. <L> Host Relay Driver. Allows you to send a pulse (at any time regardless of Match Code or Relay Driver status) to pin 2 of the trigger connector.
Code Type Commands
1. The <D> command is the only serial command that the MS-3000 decoder will recognize from an auxiliary terminal. MS-3000 Single Head Decoder User’s Manual
5-3
4–Operations
<P> Autodiscriminate All Codes. Enables the decoder to decode all available bar code types without changing decoder configuration settings.
Page 92
Chapter 5
Operational Commands NOTE: For maximum scanning speed, enable only those bar code symbologies used in the application. <Q> Enable Code 39 Only. Allows only Code 39 labels to be read. <R> Enable Codabar Only. Allows only Codabar labels to be read. <S> Enable I 2 of 5 Only. Allows only Interleaved 2 of 5 labels to be read.
Counter Commands
4–Operations
The Xs in all counter commands denote a numeric value from 00000 to 65,535. After reaching the maximum numeric limit of 65,535, you will receive an error message and the counter will automatically rollover and start counting again at 00000. To obtain the cumulative total of counts after the rollover has occurred, add 65,536 per each rollover (the decoder does not keep track of the number of rollovers) to the current count. Note: You will lose all counter values if power to the decoder is cycled, when sending the <A> command, or upon entering and exiting configuration setup menus. Note: If you activate the counter command during a read cycle, the decoder will not output the count until the read cycle ends. <N> Noread Counter. The message N/XXXXX displays the total number of noreads that have occurred since power-on or the last Noread Counter Reset command. <O> Noread Counter Reset. Sets Noread Counter to 00000. <T> Trigger Counter. The message T/XXXXX displays the total number of triggers since power-on or the last Trigger Counter Reset command. <U> Trigger Counter Reset. Sets the trigger counter to 00000. <V> Match Counter (or Good Read Counter). The message V/XXXXX displays the total number of good reads matching the master label since power-on or the last Match Counter Reset command. This counter is always enabled, but will only work as a match count when Match Code option is enabled. If the Match Code option is not enabled, this counter adds the number of good reads, or decodes. This count can be requested at any time. 1 <W> Match Counter Reset. Sets the Match Counter to 00000. <X> Mismatch Counter. The message X/XXXXX displays the number of labels successfully read that do not match the master label since power-on or the last Mismatch Counter command. <Y> Mismatch Counter Reset. Sets the Mismatch Counter to zero.
1. Can also be used as a good read counter when Match Code is not enabled.
5-4
MS-3000 Single Head Decoder User’s Manual
Page 93
Operational Commands
Test Commands <C> Enter Read Rate Test. Instructs the decoder to output the percentage of scans decoded. The read rate can vary dramatiRead Rate cally due to the angle and location of the Percentage label in relation to the scan beam. This test Label is very useful in aligning and positioning the Data scan head during installation. <m> Enter Scan Rate Test. Displays the current number of scans per second produced by the spinning mirror. <J> Exit Read Rate or Scan Rate Test. Ends the read rate test or scan rate test and returns to read or ready mode.
Status Commands <#> Display Software Part Number. Displays software part number. <!> Display Checksum of EPROM. Displays a four-digit hex number (corresponding to a given firmware version) used to verify a decoder’s EPROM.
Master Label Commands
MS-3000 Single Head Decoder User’s Manual
5-5
4–Operations
<E> Enable Match Code Option. Identical to Match Code command on page 2-19. Instructs the decoder to compare bar code labels being scanned with a master label that has been entered in nonvolatile or volatile RAM, and may under certain conditions send a relay driver signal to pin 2 of the trigger connector. If no master label has been entered, every decoded label will be a “mismatch” and will increment the mismatch counter by one. Enable Match Code Option is intended for use when the decoder is in a triggered mode. If the Match Code option is enabled in the Continuous Read mode, the decoder defaults to Continuous Read 1 Output mode, and the label data must change before the decoder will output data again, unless a timeout, if enabled, occurs. See “Match Code” on page 2-19 and “Relay Driver” on page 2-30. <F> Disable Match Code Option. Disables Match Code. <G> Store Next Label Scanned as Master Label. Causes the decoder to use the next bar code label read as the master label if Match Code option has been enabled. All subsequently decoded labels are compared against the master label information stored in RAM. (See “Match Code” on page 2-19.) <)XXXX)> Download Master Label Information. Downloads master label information from the host or a terminal. The master label information can be
Page 94
Chapter 5
Operational Commands
4–Operations downloaded at any time, and can be saved in nonvolatile memory with a <Z> command. A stored master label will not affect standard operation unless Match Code option is enabled. The Xs denote alphanumeric data, from 1 to 31 characters. <)> Request Master Label Information. Immediately sends the master label information to the host. To prevent conflicts with outputting label data, first send the <I> command (Disable Laser Scanning (Laser Off)). <))> Delete Master Label Information. Deletes master label information that has previously been loaded by either <)XXXX)> Download Master Label Information Command or <G> Store Next Label as Master Label command. Note: If the master label information has previously been stored in nonvolatile RAM (by a <Z> command), sending an <A> Reset or cycling the power will restore that information.
5-6
MS-3000 Single Head Decoder User’s Manual
Page 95
Contents Appendix A, Decoder Specifications ...............................................A-2 Appendix B, ASCII Table.................................................................A-3 Appendix C, Defaulting the Decoder................................................A-4 Appendix D, Troubleshooting...........................................................A-6 Appendix E, Interfacing with the MS-90 Scan Head .......................A-9 Appendix F, Bar Code Symbology .................................................A-11 Appendix G, Interface Standards ...................................................A-12 Appendix H, Auxiliary Monitor ........................................................A-13 Appendix I, Multidrop Communications...............................................A-17 Appendix J, Glossary of Terms ......................................................A-21
MS-3000 Single Head Decoder User’s Manual
A-1
Appendices
Appendices
Page 96
Page 97
Page 98
Appendices Appendices
Appendix C, Defaulting the Decoder Defaulting the decoder resets the decoder configuration parameters to their original default values. Defaulting is necessary if: • You wish to quickly restore default settings to the configuration program after making some temporary changes. 1 • Communications between the decoder and another device are interrupted because of incompatible settings (for example, a terminal is set to communicate at 9600 baud, but the decoder is configured at 38.4K baud). • The decoder has been assigned a polling address and you wish to access the decoder’s menu. Access to the configuration menus of a decoder that is in a polled mode can be forced by use of a profile card (see mode 0 in Chapter 4, “ Profile Card Configuration”), or by sending the <D> command from an auxiliary terminal via the auxiliary RS-232 port.
Using the Profile Card
OFF
LOAD
TERMINAL
2. Connect end of the profile card labeled “DECODER” to the decoder’s host connector.
ON
DECODER
Procedure: 1. Turn power to the decoder ON.
PROFILE CARD O N 1
2 3 4 5 6 7 8 O 1 2 3 4
DATA
N
MODE
Figure A-3 Profile Card Default Setting
3. Set all four “MODE” switches to OFF and set number 3 “DATA” switch to ON (as shown in figure A-3). 4. Press the “LOAD” button. 5. Listen for two beeps in succession (the second louder than the first). If the beeps do not occur as described, repeat the default procedure.
1. There are no menu options or host commands for resetting the configuration program.
A-4
MS-3000 Single Head Decoder User’s Manual
Page 99
Defaulting the Decoder
If a profile card is not accessible, it is necessary to default the decoder by shorting pins 7 and 11. Caution: Be certain that the correct pins are shorted. Shorting the wrong pins can cause serious damage to the unit.
Procedure: 1. Turn power to the decoder ON. 2. Locate pins 7 and 11 on the host connector (figure A-4) and mark with a pen. (Use a small length of light wire approx. 4 inches, 18 to 26 gauge for shorting.) 3. Momentarily short pins 11 and 7. Listen for a series of short beeps. Within 3 seconds, short pins 11 and 7 again. A longer beep should be heard. If not, repeat the process. Note: Using switching power supplies as a power source is not recommended due to excessive ripple characteristics.
Pin 11
Pin 7
Figure A-4 Host Connector Default Pins
MS-3000 Single Head Decoder User’s Manual
A-5
Appendices
Shorting Pins 7 and 11
Page 100
Appendices Appendices
Appendix D, Troubleshooting The MS-3000 decoder, when used correctly with a properly adjusted scan head, should produce good reads (see scan head user’s manual). If not, A-3 lists some of the more common problems associated with setting up and using the decoder. If you are unable to locate or correct the problem, call your Microscan representative. Table A-3 Troubleshooting Table
Problem
Possible Cause
Power indicator does not Power supply problem illuminate when decoder is ON Menu does not display when configuration command is sent
Decoder does not transmit or decode labels (decoder indicator lights on but no data is displayed on screen)
A-6
Solution Have technician measure AC input voltage and power supply DC output voltages. Replace if power supply is not supplying correct voltages.
Host cable defective or not Check cable connections and wiring wired properly (see “Attach Cabling,” on page 1-3). Wrong configuration command (or lower case d) sent
Verify that the <D> serial command (with an upper case D) is being sent.
Decoder configuration set- Reset decoder to default and/or match host tings do not match host's settings with decoder's. Host cable defective
Ensure the cable is correctly wired (see table 1-5 on page 1-8).
Wrong label type (different label type being scanned than that enabled in firmware)
Check label configuration settings to be certain that they match the label type being scanned. If label type is in doubt, enter <P> (the Autodiscriminate All Codes command). Ensure that Fixed Code Length and Checksums, if enabled, are set correctly.
Poor label quality
Scan a label that is known to be good.
Decoder not configured properly
Reset decoder to default and/or match host settings with decoder’s.
Excessive ambient light, sunlight, or strobes
Shield the bar code and/or scan head to verify that excessive light is not the problem.
Faulty scan head
Try another scan head that is known to be good.
Faulty cable to scan head
Try another scan head cable.
MS-3000 Single Head Decoder User’s Manual