For the Metrologic IS6520
Page 28
TROUBLESHOOTING GUIDE (CONTINUED) RS-232 only
IS6520/MS6520 Series Troubleshooting Guide (continued) SYMPTOMS
POSSIBLE CAUSE(S)
SOLUTION
Powers up OK and scans OK but does not communicate properly to the host
Com port at the host is not working or configured properly
Check to make sure that the baud rate, data bits, stops bits and parity of the scanner and the communication port match and the program is looking for "RS232" data
Powers up OK and scans OK but does not communicate properly to the host
Cable not connected to the proper com port
Check to make sure that the baud rate, data bits, stops bits and parity of the scanner and the communication port match and the program is looking for "RS232" data
Powers up OK and scans OK but does not communicate properly to the host
Com port not operating properly
Check to make sure that the baud rate, data bits, stops bits and parity of the scanner and the communication port match and the program is looking for "RS232" data
The host is receiving data but the data does not look correct
The scanner and host may not be configured for the same interface font
Check that the scanner and the host are configured for the same interface font
Characters are being dropped
Intercharacter delay needs to be added to the transmitted output
Add some Intercharacter delay to the transmitted output by using the MetroSelect ® Programming Guide MLPN 2407
22
Page 29
APPENDIX A Design Specifications Operational Light Source:
VLD 650 ± 10 nm, 0.681 milliwatts (PEAK)
Depth of Field:
0 mm to 176 mm (0” to 7”) for 0.33 mm (13 mil) bar code at default setting (programmable)
Scan Speed:
1000 scan lines/second
Scan Pattern:
5 fields of 4 parallel lines (omnidirectional)
Scan Lines:
20
Min Bar Width:
0.145 mm (5.7 mil)
Decode capability:
Autodiscriminates all standard bar codes; for other symbologies call Metrologic
System Interfaces:
RS-232, Light Pen Emulation, Keyboard Wedge, Stand Alone Keyboard
Print Contrast:
35% minimum reflectance difference
Roll, Pitch, Yaw:
360°, 60° ,60°
Indicators:
Green = laser on, ready to scan; Red = good read, decoding
Beeper Operation:
7 tones or no beep
Mechanical Length:
102 mm (4.0")
Width:
83 mm (3.25")
Height (front edge):
70 mm (2.75")
Height (back edge):
55 mm (2.25")
Weight:
298 g (10.5 oz)
Termination:
10 Position Modular RJ45
Mounting Holes:
3 mm threaded; maximum depth of 6 mm (0.24”)
23
Page 30
APPENDIX A (CONTINUED) Design Specifications Electrical Power:
1.0 W
Input Voltage:
5 VDC ± 0.25 V
Operating Current:
200 mA typical @5VDC
DC Transformers:
Class 2; 5.2 VDC @ 650 mA
EMC:
FCC, ICES-003 & EN 55022 Class A
Laser Class:
CDRH Class IIa EN 60825-1:1994/A11: 1996 CLASS 1
Environmental Operating Temperature:
-20°C to 40°C (-4°F to 104°F)
Storage Temperature:
-40°C to 60°C (-40°F to 140°F)
Humidity:
5% to 95% relative humidity, non-condensing
Light Levels:
Up to 4840 Lux (450 footcandles)
Contaminants:
Sealed to resist airborne particulate contaminants
Ventilation:
None required
Specifications subject to change without notice.
24
Page 31
APPENDIX B Default Settings Many functions of the scanner can be "programmed" - that is, enabled or disabled. The scanner is shipped from the factory programmed to a set of default conditions. The default parameter of the scanner has an asterisk (*) in the charts on the following pages. If an asterisk is not in the default column then the default setting is Off or Disabled. Every communication does not support every parameter. If the communication supports a parameter listed in the charts on the following pages, a check mark will appear. Default
RS-232*
Light Pen
UPC/EAN
* a a
Code 128
* a a
Code 93
* a a a a a a
MOD 10 Check on ITF a a
Code 11 a a a a
Full ASCII Code 39 a a
MOD 43 Check on Code 39 a a
MSI-Plessey a a
MSI-Plessey, 10/10 Check Digit a a a a
Parameter
Codabar Interleaved 2 of 5 (ITF)
Code 39
MSI-Plessey, MOD 10 Check Digit
*
*
*
Paraf Support a a
Variable a a
4 a a
Symbol Length Lock
None a a
Bars High as Code 39
*
ITF Symbol Lengths Minimum Symbol Length a
Spaces High as Code 39 a
Bars High as Scanned a
Spaces High as Scanned a
DTS/SIEMENS
25
Page 32
APPENDIX B (CONTINUED) Parameter DTS/NIXDORF
Default
RS-232*
Light Pen
*
NCR F NCR S Poll Light Pen Source a
Beeper Tone
Normal a a
Beep/Transmit Sequence
Before Transmit a a
Communication Timeout
None a a
Razzberry Tone on Timeout a a
Three Beeps on Timeout a a
No Beeps on Timeout
* a a
Enter Power Save Mode
10 mins.
a a a a a a
Same Symbol Rescan Timeout: 1250 msecs a a
Same Symbol Rescan Timeout: 2000 msecs a a
Same Symbol Rescan Timeout: 200 msecs Same Symbol Rescan Timeout: 500 msecs Programmable in 50 msec steps (MAX 6.35 seconds)
*
1 msecs 10 msecs in KBW a
Number of Scan Buffers
1 a a
Transmit UPC-A Check Digit
* a a
Transmit UPC-E Check Digit a a
Expand UPC-E a a
Convert UPC-A to EAN-13 a
Transmit Lead Zero on UPC-E a
Convert EAN-8 to EAN-13 a
Intercharacter Delay Programmable in 1 msec steps(MAX 255 msecs)
Transmit UPC-A Number System 26
* a a a
Page 33
APPENDIX B (CONTINUED) Parameter
Default
RS-232*
Light Pen
Transmit UPC-A Manufacturer ID#
* * a a a a
Transmit UPC-A Item ID# Transmit Codabar Start/Stop Characters a
CLSI Editing a
Transmit Mod 43 Check Digit on Code 39 a
Transmit Code 39 Stop/Start Characters a
Transmit Mod 10/ITF Transmit MSI-Plessey Check Characters Parity
Space
Baud Rate
9600 a a
8 Data Bits 7 Data Bits
*
Transmit Sanyo ID Characters Nixdorf ID LRC Enabled UPC Prefix UPC Suffix Transmit AIM ID Characters STX Prefix ETX Suffix Carriage Return Line Feed
* *
Tab Prefix Tab Suffix “DE” Disable Command “FL” Laser Enable Command DTR Handshaking Support RTS/CTS Handshaking Character RTS/CTS
* a a a a a a a a a a a a a a a a a a a a a 27
Page 34
APPENDIX B (CONTINUED) Parameter
Default
RS-232*
Light Pen
Message RTS/CTS a
XON/XOFF Handshaking a
ACK/NA K a
Two Digit Suppliments a as code 39
Five Digit Suppliments a as code 39
Bookland a as code 39
977 (2 digit) Supplemental Requirement a a a a a a a a a a a as code 39 a
Suppliments are not Required Two Digit Redundancy
* *
Five Digit Redundancy 100 msec to Find Supplement Programmable in 100 msec steps (MAX 800 msec)
*
Coupon Code 128 Programmable Code Lengths
7 aval.
a
Programmable Prefix Characters
10 avail.
a
Suffix Characters a
Prefixes for individual Code Types Editing Inter Scan-Code Delay Programmable (100 msec steps)
a a
800 msec
Function/Control Key Support Minimum Element Width Programmable in 5.6 µsec steps
1 msec a
Depth of Field Optimal High Depth of Field
* a a
Normal Depth of Field a a
Optimal Low Depth of Field a a
Close Depth of Field a a
Far Depth of Field a a
28
Page 35
APPENDIX C Scanner Pinout Connections The IS6520/MS6520 scanner interfaces terminate to a 10-pin modular jack. The serial # label indicates the interface enabled when the scanner is shipped from the factory.
IS6520-47 / MS6520-47 Keyboard Wedge Pin Function 1 Ground 2 RS-232 Transmit Output 3 RS-232 Receive Input 4 PC Data 5 PC Clock 6 KB Clock 7 PC +5V 8 KB Data 9 +5VDC 10 Shield Ground
1
10
IS6520-41 / MS6520-41 RS-232/LTPN Pin Function 1 Ground 2 RS-232 Transmit Output 3 RS-232 Receive Input 4 RTS Output 5 CTS Input 6 DTR Input/LTPN Source 7 Reserved 8 LTPN Data 9 +5VDC 10 Shield Ground
Cable Connector Configurations “Standard” PowerLink cable (MLPN 53000 or 54000) 9-pin D-type female connector to the PC Pin
Function
1 2 3 4 5 6 7 8 9
Shield Ground RS-232 Transmit Output RS-232 Receive Input DTR Input Power/Signal Ground Light Pen Data CTS Input RTS Output +5VDC*
9
5
6
1
9-Pin D-Type Connector
* If a PowerLink power supply is plugged into the PowerLink cable, +5V will NOT be available on this pin. This pin is used when the host is supplying +5V to the scanner. 29
Page 36
APPENDIX C (CONTINUED) Keyboard Wedge Powerlink and Adapter Cable (MLPN 54-54008) The Keyboard Wedge PowerLink cable is terminated with a 5-pin DIN female connector on one end, and a 6-pin mini DIN male on the other. 4
2
5
2
3
4
1
1 3 6 5
Keyboard Wedge PowerLink Cable
5-Pin DIN, Female
6-Pin DIN, Male
Metrologic will supply an adapter cable with a 5-pin DIN male connector on one end and a 6-pin mini DIN female connector on the other. 5 3
2
4
2
1
1
3
4 5 6
5-Pin Din, Male
Adapter Cable 6-pin Mini Din, Female
According to the termination required, connect the appropriate end of the adapter cable to the PowerLink cable, leaving the necessary termination exposed for connecting to the keyboard and the keyboard port on the PC. The pin assignments are as follows:
Pin 1 2 3 4 5
PowerLink Cable
Adapter Cable
5-pin Female DIN
5-pin Male DIN
Function Keyboard Clock Keyboard Data No Connect Power Ground +5 Volts DC
Pin 1 2 3 4 5
6-pin Male Mini-DIN Pin 1 2 3 4 5 6 30
Function PC Data No Connect Power Ground +5 Volts DC PC Clock No Connect
Function PC Clock PC Data No Connect Power Ground +5 Volts DC
6-pin Female Mini-DIN Pin 1 2 3 4 5 6
Function Keyboard Data No Connect Power Ground +5 Volts DC Keyboard Clock No Connect