Page 41
6-5 Correcting Measurement Error (Calibration)
Note
• Calibrate the LC by shifting the target only within the measuring range. • If the calibration measurement is unacceptable, the error message "C-Err" (Calibration Error) appears on the display when the key is pressed. If this occurs, repeat the calibration.
Note
The calibration range is as follows: Value to be entered in step 8
0.9 ≤
Measurement obtained with factory settings
≤ 1.1
Calibration by Entering a Numerical Value
The LC can be calibrated simply by entering previously recorded calibration values. Check and record the calibration values by following the procedures shown below. The LC can then be calibrated simply by entering this value.
Checking Calibration Value
Check the calibration value obtained from the calibration procedure using a target. 1. Calibrate the LC using a target. Reference: Calibration by Target Position (P.32)
2. To interrupt the laser beam, close the laser beam shield or insert a lightshielding object at the closest possible position to the sensor head, between the sensor head and target. The message "dar" appears on the display panel. Note
When interrupting the laser beam, be careful not to touch the glass window of the sensor head. 3. Press the
LIMIT OFFSET
PRM
CAL key repeatedly until the CAL LED lights.
INT. ANALOG BAUD LIMIT RATIO RATE
DIGIT
PRM
4. Press the key. The LC is ready to accept a new calibration value, and the calibration value appears on the display. Record the value. 5. Press the key again. The LC returns to the measuring mode.
34 CHAPTER 6 Setting Up
Page 42
6-5 Correcting Measurement Error (Calibration) 6-6 Offset Values
Entering Calibration Value
Enter the calibration value by following the procedures shown below. 1. Follow the same procedures used for checking the calibration value. 2. Change the displayed value to the recorded calibration value. Reference: Entering Numerical Values (P.28)
Note
Press the PRM key first, then the key as many times as required to return to the measuring mode. The entered value and calibration mode are canceled. 3 Press the key. Calibration is complete. The LC returns to the measuring mode.
Note
To calibrate the LC through the RS-232C interface, enter the following calibration coefficient:
Calibration coefficient =
Desired preset value Factory setting value
6-6 Offset Values Offset values are added to or subtracted from a measurement before displaying or outputting it. Offset values are used for displaying or outputting the target height or thickness as an absolute value. This function is also used to tune the LC to the input requirements of the equipment that is connected to it. The maximum and minimum offset values available depend on the sensor head as shown below. Sensor head Offset value range Max. output range LC-2420 +199.99 µm ±327.66 µm LC-2430 +499.98 µm ±655.32 µm LC-2440 +2.9998 mm ±6.5532 mm LC-2450 +7.9995 mm ±16.3830 mm Factory setting: 0 If a displayed value with a offset becomes beyond the maximum output range is specified, "- - - - -" will appear. 1. Press the PRM key repeatedly until the OFFSET LED lights. The current offset value appears on the display panel and the LC is ready to accept a new offset value. LIMIT OFFSET
CAL
INT. ANALOG BAUD LIMIT RATIO RATE
DIGIT
PRM
2. Enter an offset value. Note
To cancel entry of the offset value, press the PRM key first, then the key as many times as required to return to the measuring mode. The entered value and offset mode are canceled.
CHAPTER 6 Setting Up
35
Page 43
6-7 Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) 6-8 Reducing the Effect of Target Surface Irregularities (Setting Low-pass Filter Value) (Setting Low-pass Filter Value)
3. Press the key. The offset value is stored, and the LC will returns the measurement mode.
6-7 Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) The LC has a function for creating an average measurement value from a series of measurements. The more variation occurring in measurements, the higher the number of measurements should be taken in order to attain a stable, average measurement. The number of measurements can be selected from 18 different settings ranging from 1 to 131072. This averaging function and the filtering function explained in Section 6-8 cannot be used at the same time. Factory setting: 2048 Reference:
Number of averaging measurements vs. response frequency (P.83) Averaging method (P.91)
1. Press the AVERAGE key. The LED above the AVERAGE key lights, and the current number of averaging measurements appears on the display panel. 2. Press the or key to change the value. Every time the key is pressed, the number of measurements increases or decreases to twice or half the original value (1, 2, 4, 8, 16, 32, 64, 128 ..... 131072). Note
To cancel entry of the number of measurements, press AVERAGE key. This mode is canceled and the LC returns to the measuring mode. 3. Press the key. The number of measurements chosen is set, and the LC returns to the measuring mode. The LED above the AVERAGE key remains lit during measurements.
6-8 Reducing Effect of Target Surface Irregularities (Setting Low-pass Filter Value) A low-pass filter is used to attenuate measurements at frequencies higher than the preset limit. The preset frequency is referred to as the cutoff frequency. This is useful for targets with treated surfaces (e.g. hairline finish). When measuring these targets, the sensor detects surface irregularities (picks up noise), preventing accurate measurement. By setting a filter value, the LC can more accurately measure the displacement of the target without distortion from the texture of the target surface. Factory setting: FILTER OFF, cutoff frequency: 100
36 CHAPTER 6 Setting Up
Page 44
6-8 Reducing Effect of Target Surface Irregularities (Setting Low-pass Filter Value) 6-9 Setting Range for Light Intensity (INT. LIMIT)
• A digital low-pass filter is used. • The filtering function cannot be used when using the control unit independently. • This function and the averaging function explained in the previous section cannot be used at the same time. 1. Press the FILTER key. The LED above the key flashes and the current cutoff frequency appears on the display. 2. Set the cutoff frequency The cutoff frequency can be set between 0100 and 4999 Hz. Note
If this frequency is set below 0100, the error message "f-Err" appears.
Note
To cancel entry of the cutoff frequency, press the FILTER key. The setting is canceled, and the LC returns to the measuring mode. 3. Press the key. The new cutoff frequency is set, and the LC returns to the measuring mode. The LED above the FILTER key remains lit during measurements.
6-9 Setting Range for Light Intensity (INT. LIMIT) The range for light intensity can be set. The upper/lower limits are referred to as the "HIGH-INTENSITY LIMIT" and "LOW-INTENSITY LIMIT", respectively. Also, the measurement of displacement just before the light intensity exceeds the HIGH-INTENSITY LIMIT or falls below the LOW-INTENSITY LIMIT can be retained. The duration to retain the measurement is referred to as a "level-cut time". This can eliminate inaccurate measurements resulting from sudden changes in the light intensity or presence of dents or grooves on the target surface. Factory settings: LOW-INTENSITY LIMIT: 50 (selectable between 00000 and 49998) HIGH-INTENSITY LIMIT: 49750 (selectable between 00000 and 49998) Level-cut time: 0000.0 ms (selectable between 0000.0 and 2999.9 ms or 9999.9 [infinity]) • A level-cut time must be set each for the LOW-INTENSITY LIMIT and HIGH-INTENSITY LIMIT. • The HIGH-INTENSITY LIMIT must be greater than the LOW-INTENSITY LIMIT. 1. Measure the light intensity of the irregular portion of the surface and record the value. Reference: Displaying Light Intensity (P.29)
CHAPTER 6 Setting Up
37
Page 45
6-9 Settings Range for Light Intensity (INT. LIMIT)
2. Press the PRM key repeatedly until the INT. LIMIT LED lights. The HIGH LED lights and the LC switches to the entry mode. Also, the current HIGH-INTENSITY LIMIT appears on the display panel.
LIMIT OFFSET
HIGH
CAL
INT. ANALOG BAUD LIMIT RATIO RATE
DIGIT
PRM
GO LOW
3. Change the displayed value to the desired HIGH-INTENSITY LIMIT. Note
To use the current HIGH-INTENSITY LIMIT, press the key. The setting remains unchanged, and the LC will proceed to the setting of the HIGH-INTENSITY LIMIT (Step 5). 4. Press the key. Simultaneously, the HIGH and GO LEDs light and the LC is ready for setting the INTENSITY LIMIT TIME. The current INTENSITY LIMIT TIME appears on the display panel.
HIGH GO LOW
5. Change the displayed value to the desired INTENSITY LIMIT TIME. Note key. The To use the current INTENSITY LIMIT TIME, press the setting remains unchanged, and the LC will proceed to the setting of the LOW-INTENSITY LIMIT (Step 7). 6. Press the key. The new INTENSITY LIMIT TIME is set and the HIGH and GO LEDs go off. Simultaneously, the LOW LED lights and the LC is ready to accept a new LOW-INTENSITY LIMIT. The display indicates the current LOWINTENSITY LIMIT. HIGH GO LOW
7. Change the displayed value to the desired LOW-INTENSITY LIMIT. Note
To use the current LOW-INTENSITY LIMIT, press the key. The setting remains unchanged, and the LC will proceed to setting of the levelcut time for LOW-INTENSITY LIMIT (Step 9). 8. Press the key. The new LOW-INTENSITY LIMIT is set. Simultaneously, the LOW and GO LEDs light and the system is switched to the setting mode. The display indicates the current INTENSITY LIMIT TIME.
38 CHAPTER 6 Setting Up
Page 46
6-9 Setting Range for Light Intensity (INT. LIMIT) 6-10 Setting Upper/Lower Limits for Measurements (Setting Tolerance)
HIGH GO LOW
9. Change the displayed value to the desired INTENSITY LIMIT TIME. Note
To use the current INTENSITY LIMIT TIME, press the key. The setting remains unchanged, and the LC will return to the measuring mode. 10. Press the key. The new INTENSITY LIMIT TIME is set, the LOW and GO LEDs go off, and the LC will return to the measuring mode.
Note
To cancel the entry of light intensity parameters, press the PRM key first, then the key as many times as required to return to the measuring mode. The entered value will also be canceled. The light intensity parameters are set only after the LOW-INTENSITY LIMIT TIME has been entered and the key pressed. If you stop before entering all the parameters, the parameters entered will be canceled.
6-10 Setting Upper/Lower Limits for Measurements (Setting Tolerance) The LC can set upper/lower limits for measurements in each measurement mode. Measurements outside the preset range is indicated by the display or output signal. The upper/lower limits range differs, depending on the sensor head as shown below. Numbers in parentheses are factory settings. LC-2420: +299.99 µm (±200.00 µm) LC-2430: +599.98 µm (±500.00 µm) LC-2440: +5.9998 mm (±3.0000 mm) LC-2450: +15.9995 mm (±8.0000 mm) Note
Upper and lower measurement limits can be set for each measurement mode. 1. Choose the measurement mode and press the key.
2. Press the PRM key. The LIMIT LED lights. The HIGH LED lights and the LC is ready to accept a new upper limit. The display indicates the current upper limit.
LIMIT OFFSET
CAL
INT. ANALOG BAUD LIMIT RATIO RATE
DIGIT
PRM
3. Change the displayed value to the desired upper limit.
CHAPTER 6 Setting Up
39
Page 47
6-10 Setting Upper/Lower Limits for Measurements (Setting Tolerance) 6-11 Setting Number of Displayed Digits
4 .Press the key. The new upper limit is set. The LOW LED lights and the LC is ready to accept a new lower limit. The display now indicates the current lower limit. HIGH GO LOW
5. Change the displayed value to the desired lower limit. 6. Press the key. The new lower limit is set, and the LC returns to the measuring mode. The upper/lower limits are set only after the key has been pressed. If you stop before entering the lower limit, the new upper limit will be canceled.
6-11 Setting Number of Displayed Digits Follow the procedure below to set the number of displayed digits. Factory setting: All digits 1. Press the PRM key repeatedly until the DIGIT LED lights. The display starts flashing. LIMIT OFFSET
CAL
INT. ANALOG BAUD LIMIT RATIO RATE
DIGIT
PRM
2. Press the or key to set the number of displayed digits. Every time the key is pressed, the number of digits on the display increases or decreases by one, thus activating or deactivating the last four digits. Press the key to increase the number of displayed digits; key to decrease it. press the If you choose not to display integer digits, "0" will be displayed in these digits. 3. Press the key. Measurements will be displayed as specified. Note
40 CHAPTER 6 Setting Up
To cancel entry of the number of displayed digits, press the PRM key first, then the key as many times as required to return to the measuring mode.
Page 48
6-12 Storing and Loading Settings (Programming)
6-12 Storing and Loading Settings (Programming) A set of parameters, including an offset value and upper and lower measurement limits, can be stored as a program. Up to five different programs can be stored. By storing programs prepared for each target, the need for setting individual parameters is eliminated. This is useful when frequent product changeover is required. Factory setting: See the factory setting of each parameter.
Storing Settings (SAVE)
1. Set the parameters required for measurement, then switch the LC to the measuring mode. 2. Press the SAVE key. The LED above the key starts flashing and one of the five PROGRAM LEDs lights. 1
2
3. Press the Note
3
4 or
5
PROGRAM key to select a program number.
• To cancel the saving of a program, press the
SAVE key.
• When overwriting a previously saved program (corresponding PROGRAM indicator is lit) be sure that this program does not contain settings that should be retained. 4. Press the key. The message "rEAdy" appears on the display panel. Press the key again to store the parameter settings. The LC is ready to measure using these settings. 5. Press the
Loading Stored Settings (LOAD)
Follow the steps below to load the stored program. 1. Press the LOAD key. The LED above the key starts flashing and one of the five PROGRAM LEDs lights. 1
2
2. Press the 1
Note key.
2
3
4 or 3
5
PROGRAM key to select a program number. 4
5
PROGRAM
To cancel the loading of a program, press the LOAD key. The selected program will be canceled and the LC will return to the measurement mode.
CHAPTER 6 Setting Up
41
Page 49
6-12 Storing and Loading Settings (Programming)
3. Press the key. The setting is loaded, and the LC switches to the measurement mode. The LED above the LOAD key goes off, but the selected PROGRAM LED remains lit. At this time, if any parameter is changed the LED will go off.
1
42 CHAPTER 6 Setting Up
2
3
4
5
PROGRAM
Page 50
CHAPTER 7
MEASUREMENT PROCEDURE 7-1
Measuring with One Sensor Head ................................. 44
7-2
Measuring with Two Sensor Heads ............................... 45