Page 51

7-1 Measuring with One Sensor Head

The versatile LC-2400 Series can measure height or eccentricity as well as displacement. With an optional board installed, the LC can measure thickness or height deviation using two sensor heads. Stored parameters can be quickly called up by using the LOAD function. Note

• The LC requires 60 min. to warm up. Turn on the controller and wait at least 60 min. before starting measurements. • If you find displayed values or output measurements incorrect, check the sensor head position as well as parameters. Reference: Storing and Loading Settings (Programming) (P.41) Error Message List (P.82)

7-1 Measuring with One Sensor Head Measuring Target Height

Measure the target height with respect to the reference surface by following the procedures shown below. [LC-2400A] 1. Turn the key-operated power switch to the ON position to turn on the LC controller. 2. Position the sensor head. Keep in mind the measuring range and angle of the sensor heads.

LC-2400A OFF

ON

POWER

[LC-2400W] 1. To start the LC controller, turn on the power switch at the rear of the controller and then turn the key-operated switch to the ON position . 2. Position the sensor head. Keep in mind the measuring range and angle of the sensor heads.

LC-2400W STAND-BY

ON

POWER

Reference: Adjusting Sensor Head Position (P.25) Note

When two sensor heads are connected, select either one by pressing the 1CH or 2CH key. The LED above the key lights. 3. Press the NORMAL key to switch the LC to the NORMAL measuring mode. 4. Calibrate the LC. Reference: Correcting Measurement Error (Calibration) (P. 32)

5. Set the number of averaging measurements. Reference: Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) (P.36)

6. Measure the reference surface of the target. 7. Press the ZERO key. The displayed value and output are set to zero. Now the displayed value and output indicate the deviation in height from the reference value. Note

To determine the target shape, move the sensor head or target in parallel with respect to the other.

44 CHAPTER 7 Measurement Procedure

Page 52

7-1 Measuring with One Sensor Head 7-2 Measuring with Two Sensor Heads

Measuring Runout and Eccentricity

Measure the runout and eccentricity of a revolving target by following the procedures shown below.

1. Position the sensor head. Be careful about the measuring range and angle of the sensor head. Reference: Adjusting Sensor Head Position (P.25)

2. Turn the key-operated power switch to turn the system on. LC-2400W

LC-2400A

STAND-BY

OFF

ON

ON

POWER

POWER

3. Calibrate the LC. Reference: Correcting Measurement Error (Calibration) (P.32)

4. Set the number of averaging measurements. Reference: Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) (P.36) P–P 5. Press the key to switch the LC to the NORMAL mode. Now the displayed value and output indicate the runout and eccentricity (max. and min. values).

6. To measure again, press the

P–P key to reset the measurement.

7-2 Measuring with Two Sensor Heads The LC-2400W series does not support this feature.

By installing the optional LC-B3 board, the LC-2400 can measure a target using two sensor heads. With two sensor heads used, thickness or height deviation can be measured by adding together the measurements from the two sensor heads or by subtracting one measurement from the other. Even while two sensor heads are connected, the LC can still be used for measurements that requires only one sensor head. Note

Be sure to use sensor heads of the same model when measuring in the ADD or SUB mode.

CHAPTER 7 Measurement Procedure

45

Page 53

7-2 Measuring with Two Sensor Heads

Measuring Thickness (ADD)

Measure the thickness of a target by following the procedures shown below.

1. Position the sensor heads and the reference target. Keep in mind the measuring range and angle of the sensor heads.

Equal distances

Reference: Adjusting Sensor Head Position (P.25)

2. Turn the key-operated power switch to turn on the LC. LC-2400A OFF

ON

POWER

3. Press the

1CH key. The LED above the key lights.

4. Calibrate the 1CH sensor head. Reference: Correcting Measurement Error (Calibration) (P.32)

5. Press the

2CH key. The LED above the key lights.

6. Calibrate the 2CH sensor head. 7. Press the

ADD key. The LED above the key lights.

8. Set the number of averaging measurements as in Step 4. Reference: Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) (P.36)

9. Press the NORMAL key. The LED above the key lights. The LC is now in the NORMAL mode. 10. Press the ZERO key. The displayed value and output are set to zero. Now the displayed value and output indicate the deviation in thickness from the reference value. Skip this step and follow the steps described below to display and output the thickness.

46 CHAPTER 7 Measurement Procedure

Page 54

7-2 Measuring with Two Sensor Heads

11. Change the current offset value to the thickness of the target. When entry of the offset value is complete, the display will show the reference thickness. LIMIT OFFSET

CAL

INT. ANALOG BAUD LIMIT RATIO RATE

DIGIT

PRM

The following shows the maximum displayable thicknesses using the LC sensor heads: Sensor head Max. value LC-2420 +327.66 µm LC-2430 +655.32 µm LC-2440 +6.5532 mm LC-2450 +16.3830 mm If the measurement that contains the offset value exceeds the display range, the LC holds the measurement at the maximum value. Reference: Offset Values (P.35)

12. Start measurement. Thickness of the target is displayed and output. If the automatic zero setting function was executed in Step 10, deviation from the reference value will be displayed and output.

Measuring Height Deviation (SUB)

Measure height deviation by following the procedures shown below.

1. Position the sensor heads and the target. Be careful about the measuring range and the angle of the sensor heads. Reference: Adjusting Sensor Head Position (P.25)

2. Turn the key-operated power switch to turn the system on. LC-2400A OFF

ON

POWER

3. Press the

1CH key. The LED above the key lights

4. Calibrate the 1CH sensor head. Reference: Correcting Measurement Error (Calibration) (P.32)

5. Press the

2CH key. The LED above the key lights.

CHAPTER 7 Measurement Procedure

47

Page 55

7-2 Measuring with Two Sensor Heads

6. Calibrate the 2CH sensor head. 7. Press the

SUB key. The LED above the key lights.

8. Set the number of averaging measurements. Reference: Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) (P.36)

9. Press the NORMAL key. The LED above the key lights. The LC is now in the NORMAL mode. 10. Press the ZERO key. The displayed value and output are set to zero. Now the displayed value and output indicate the deviation in thickness from the reference value. Skip this step and follow the steps described below to display and output the thickness. 11. Change the current offset value to the height deviation of the reference target. When entry of the offset value is complete, the display will show the reference height deviation. LIMIT OFFSET

CAL

INT. ANALOG BAUD LIMIT RATIO RATE

DIGIT

PRM

The following shows the range of height deviation values that can be displayed using the LC sensor heads: Sensor head LC-2420 LC-2430 LC-2440 LC-2450

Range +327.66 µm to -327.67 µm +655.32 µm to -655.34 µm +6.5532 mm to -6.5534 mm +16.3830 mm to -16.3835 mm

If the measurement that contains the offset value exceeds the display range, the LC holds the measurement at the corresponding value.

12. Start measurement. The target height difference is displayed and output as an absolute value. When you start measurements, after pressing the auto-zero key as shown in step 10, the deviation in height difference from the reference value is displayed and output.

48 CHAPTER 7 Measurement Procedure

Page 56

CHAPTER 8

EXTERNAL INPUT/OUTPUT 8-1

RS-232C Interface ......................................................... 50

8-2

Control I/O (Standard Equipment on Display Unit) ........ 71

8-3

Monitor Output (Standard Equipment) ........................... 73

8-4

Digital I/O (Standard Equipment on Control Unit) .......... 74 For information on the optional GP-IB (LC-B1) or DIGITAL I/O (LC-B2) boards, please refer to Chapter 9, section 9-9.

Page 57

8-1 RS-232C Interface (Standard Equipment)

The LC-2400 series is equipped with an RS-232C serial interface, control I/O port (when the display unit is used), monitor output, and a digital I/O board. Using these I/O connections, the LC can transmit measurement data and alarm signals to external equipment and receive parameter settings from external equipment.

8-1 RS-232C Interface (Standard Equipment) The LC-2400 series can communicate with a computer using the RS-232C interface. It can transmit measurement data to the computer and receive control signals from the computer. When the control unit is used separately, the parameters of each function can be set by the computer. The ferrite cores are packaged with the LC-2400W series. When using the RS-232C or the CONTROL I/O connecting cable, attach the appropriate ferrite cores at both ends of the cable. Otherwise, noise interference may occur.

CAUTION

When using the RS-232C cable

Specifications

This serial interface conforms to EIA (Electronic Industries Association) RS232C specifications.

Communication Parameters

Duplex Full Synchronization Start-stop Data format ASCII Data bit length 7/8 bit* Stop bit length 1/2 bit* Parity check Even/odd/none* Baud rate 75/150/300/600/1200/2400/4800/9600/19200 bps* * Parameters are selectable.

Connecting Cable

Use a standard serial cable for connection to a computer. For alternative connection configurations, see the following. When using a personal computer: [Example 1]

50 CHAPTER 8 External Input/Output

[Example 2]

Personal computer

LC-2400A/ LC-2400W

Personal computer

LC-2400A/ LC-2400W

1

1

2

2

3

3

5

5

1 2 3 7

6

6

7

7

1 2 3 7 4 6 8 5 J

8

8

J

J

Page 58

8-1 RS-232C Interface (Standard Equipment)

Pin Assignment

14 1

Note

Setting Communication Parameters

Note

Setting Communication Parameters When Using the Control Unit Together with the Display Unit

15 2

16 3

17

18

4

5

19 6

20 7

Pin No. 1

RS-232C signals Shield

2

SD (TXD)

3

RD (RXD)

5

CS (CTS)

6

DR (DSR)

7 8

SG (GND) CD (DCD)

20

ER (DTR)

21 8

22 9

23 10

24 11

25 12

13

Description of signals Input/output Shield (common to signal ground) Receives data from exterInput nal equipment. Transmits measurement Output data to external equipment. ON (High) when the power Output is turned on ON (High) when the power Output is turned on Signal ground ON (High) when the power Output is turned on Enables transmission of measurement data to Input external equipment when the signal is ON (High).

Pin 20 is connected to the power supply through a pull-up resistor inside the LC controller. Even if this terminal is not connected to external equipment, measurement data can be output. Initial settings Baud rate: Data bit length: Parity check: Stop bit length:

1200 bps 8 bits None 1 bit

Be sure that the communication parameters of the LC are the same as those of the external device.

When the control unit is installed in the display unit, communication parameters can be set using the display unit keys. 1. Press the PRM key until the BAUD RATE indicator lights. The current baud rate is displayed on the display panel. mm

NORMAL

P±P

µm AUTO

1

2

3

4

GAIN

HIGH LIMIT OFFSET

CAL

GO LOW LASER ON

1

2

3

INT. ANALOG BAUD LIMIT RATIO RATE

4

5

DIGIT

GAIN

1CH

PRM

AVERAGE

PRM

PROGRAM LOCK

LOAD

SAVE PROGRAM

CHAPTER 8 External Input/Output

51

Page 59

8-1 RS-232C Interface (Standard Equipment)

2. Press the or key to select the desired baud rate. The following baud rates can be selected. 75/150/300/600/1200/2400/4800/9600/19200 3. Press the key. The remaining parameters are displayed on the display panel. mm µm AUTO

1

2

3

4

GAIN

HIGH LIMIT OFFSET

CAL

GO LOW

1

2

3

INT. ANALOG BAUD LIMIT RATIO RATE

4

5

LASER ON

Date bit length

DIGIT

PRM

PROGRAM LOCK

Parity check

Stop bit length

: bit

: Even

: 1 bit

: bit

: Odd

: 2 bit

: None

4. Press the or key to select the desired parameter. The following data formats can be selected.

Note

To cancel entry of communication parameters, press the PRM key and then press the key several times until the LC returns to the measurement mode. Any changes are ignored, and the LC is in the measurement mode. 5. Press the key. Once the communication parameters are entered, the LC will return to the measurement mode.

52 CHAPTER 8 External Input/Output

Page 60

8-1 RS-232C Interface (Standard Equipment)

When the control unit is used separately, adjust the baud rate using the baud rate dial, located on the side of the control unit.

2 3

5 6

Baud rate setting dial

7 8

4

Control unit side view

0 1

Setting Communication Parameters with the Separate Control Unit

Dial numbers and baud rates correspond as shown below: Dial 0 1 2 3 4 5 6 7 8 9 number Baud 75 150 300 600 1200 2400 4800 9600 19200 19200 rate Note

• When the control unit is used separately, parameters other than baud rate are fixed at the following values: Data bit length: 8 bits, Parity check: None, Stop bit length: 1 bit • When the baud rate is changed using the display unit keys after installing the control unit in the display unit, the baud rate set with the baud rate dial is ignored.

Changing and Verifying Parameter Settings using External Equipment

By transmitting the proper character or ASCII code from an external device to the LC, the LC parameter settings can be changed or verified. Procedure for Changing and Verifying Parameter Settings Follow the procedure below to change or verify parameter settings. 1

External equipment

2

LC

3

1. Transmit the proper character or ASCII code from the external device to the LC. 2. When the character is received by the LC, the message "PASS CR" is sent to the external device. The parameter is changed to the setting corresponding to the character. Note

When a character for setting measurement output or information output is received, the LC does not output the "PASS CR" message.

CHAPTER 8 External Input/Output

53