The text of a manual · part 4 of 13
Ultra High Accuracy Laser Displacement Meter LC-2400 Series Instruction Manual, pages 31 to 40
Pages 31 to 40 of 121
The original document is the PDF (805 KB). This is its text as extracted, unedited; tables and drawings read best in the PDF, and each page heading opens it at that page.
For the Keyence LC-2400A
Page 31
5-4 Installing LC-B3 Board (2CH Sensor Head Board) 5-5 Installing GP-IB or Digital I/O Board
5-4 Installing LC-B3 Board (2CH Sensor Head Board) Cannot be used with the LC-2400W series. To use two sensor heads with the LC controller, the LC-B3 board must be inserted into the LC-B3 board slot located at the rear of the control unit. (The LC-B3 board is optional. Contact KEYENCE for details.)
CAUTION
Turn OFF the LC and unplug the power cable from the wall outlet before starting the following procedure. 1. Remove the two screws securing the LC-B3 board slot cover on the rear of the control unit.
2. Align the LC-B3 board with the grooves of the LC-B3 board slot, insert the LC-B3 board all the way into the slot, and secure the board using the two screws removed in step (1).
4-5 Installing GP-IB or Digital I/O Board (Options) Cannot be used with the LC-2400W series.
CAUTION
Turn OFF the controller and unplug the power cable from the wall outlet. Insert the GP-IB or Digital I/O Board into the expansion slot on the display unit as shown in the following figure. (See chapter 9.) 1. Remove the two screws securing the expansion slot cover at the rear of the display unit.
24 CHAPTER 5 Connections
Page 32
5-5 Installing GP-IB or Digital I/O Board 5-6 Adjusting Sensor Head Position
2. Align the board with the grooves of the expansion slot, insert the GP-IB or DIGITAL I/O board all the way into the slot, and secure the board using the two screws removed in step (1).
Expansion slot cover
Rear view of the controller
5-6 Adjusting Sensor Head Position 1. Align the sensor head using the following three positions as alignment points: (1) The three mounting holes on each side of the sensor head (2) The sides of the sensor head (3) The front (laser-emitting surface) of the sensor head (1)
(1)
LC±2440
LC±2420
(2)
(2)
Parallel
(3)
(3)
Right angle
Right angle
Target
Target
When there is a significant change in target color, material, or height, measurement errors may result depending on the orientation of the sensor head. Proper orientation is shown below.
Incorrect
Correct
Incorrect
Correct
Recess Protrusion Color change Rotating shaft
Rotating shaft
CHAPTER 5 Connections
25
Page 33
5-6 Adjusting Sensor Head Position
2. Turn on the LC controller and adjust the sensor head position so that the laser ON alarm LED located on the sensor head lights orange. The sensor head is at the "reference distance" from the target when the above LED lights orange.
Checking for Reference Distance
When the distance between the target and sensor head is within the reference distance range, the laser ON alarm LED lights orange. The sensor head can be easily positioned even when the control unit is used without the display unit.
Laser ON alarm LED Measuring range
NEAR
FAR
Reference distance
Green
Sensor head LC-2420 LC-2430 LC-2440 LC-2450
26 CHAPTER 5 Connections
Orange
Range the LED lights orange ±0.02 mm ±0.05 mm ±0.3 mm ±0.8 mm
Green
Page 34
CHAPTER 6
SETTING UP 6-1
Basic Operation of Controller ........................................ 28
6-2
Selecting Measurement Mode ....................................... 30
6-3
Calibrating Sensitivity (Gain Selection) ......................... 30
6-4
Selecting Display/Output Mode ..................................... 31
6-5
Correcting Measurement Error (Calibration) ................. 32
6-6
Offset Values ................................................................. 35
6-7
Reducing Variation in Measurement Data (Setting Number of Averaging Measurements) ............. 36
6-8
Reducing Effect of Target Surface Irregularities (Setting Low-pass Filter Value) ..................................... 36
6-9
Setting Range for Light Intensity (INT. LIMIT) ............... 37
6-10 Setting Upper/Lower Limits for Measurements (Setting Tolerance) ........................................................ 39 6-11 Setting Number of Displayed Digits ............................... 40 6-12 Storing and Loading Settings (Programming) ............... 41
Page 35
6-1 Basic Operation of Controller
6-1 Basic Operation of Controller This section describes the basic operation procedures of the controller.
Selecting a Parameter ( PRM key)
1. Press the PRM key. One of the PARAMETER LEDs lights, and the current value for the parameter appears on the display panel. The LC is now in the setting mode. 2. Press the PRM key repeatedly. The PARAMETER LEDs go on and off one after another as shown below. The illuminated LED means that the corresponding parameter is selected.
LIMIT
PRM
OFF SET PRM
PRM
ANALOG RATIO
INT. LIMIT
CAL PRM
PRM
BAUD RATE PRM
DIGIT PRM
PRM
Entering Numerical , , Values ( and keys)
,
1. Select the desired parameter. The LC is ready to accept a numerical value. 2. Select a digit. Press the key. The selected digit starts flashing. Press the key to shift the selection to the right, and press the key to shift it to the left.
mm µm to the left to the right
3. Change the digit. Press the or key while the digit is flashing. Press the to increase the value. Press the key to decrease it.
0
1
2
3
4
5
6
7
8
9
4. Confirm the setting. Press the key to confirm the new setting.
28 CHAPTER 6 Setting Up key
Page 36
6-1 Basic Operation of Controller
Displaying Light Intensity
A function is available to display the intensity of reflected light from the target. The light intensity is displayed and output within a range of 1 to 50000. If possible, the target should have a light intensity of 00050 or greater for accurate displacement measurement.
( INTENSITY key)
1. Position a target for measurement. 2. Press the INTENSITY key. The LED above the INTENSITY key lights, and the light intensity appears on the display panel. Note
An analog voltage equivalent to the light intensity is always output. 3. Press the INTENSITY key again. The LED above the INTENSITY key goes off, and the system will return to the measuring mode.
Automatic Zero Function ( ZERO key)
This function sets the current measurement and displayed values to +0.0000 mm in the NORMAL, BOTTOM, or PEAK mode. After the function is executed, the LC will display and output displacement relative to the new reference (zero) point. 1. Execute the automatic zero function. Press the ZERO key. The LED above the ZERO key lights and the displayed value turns to +0.0000 mm. Now the display will show the displacement relative to the new reference point.
mm µm
2. Cancel the automatic zero setting function. Press the ZERO key again. The display restores the value before the automatic zero function was executed. Note
• This function can be operated even when the ADD or SUB key is used. • The automatic zero function can also be operated via the CONTROL I/O port.
Hold Function ( HOLD key)
The hold function is used to retain and output the last measured value. This function enables reading and recording of a specific analog value at a given moment. 1 Press the HOLD key. The LED above the HOLD key lights and the last measured value is displayed as well as output. 2 Press the HOLD key again. The hold function is deactivated and the LC-2400 series returns to the measurement mode.
CHAPTER 6 Setting Up
29
Page 37
6-2 Selecting Measurement Mode 6-3 Calibrating Sensitivity (Gain Selection)
Note
The HOLD function can also be operated via the CONTROL I/O port. Activating or deactivating the hold function in the P-P, PEAK, or BOTTOM mode will cause the LC to return to the measurement mode and reset the measurements.
6-2 Selecting Measurement Mode Four measurement modes are available. Measurements from each mode can be displayed as well as output to external equipment. Measurement mode NORMAL P-P (Peak-to-Peak) PEAK BOTTOM
Function Measures displacement from reference point Measures displacement between maximum and minimum values Measures maximum value from reference point Measures minimum value from reference point
Factory setting: NORMAL P–P Press the mode selector key ( NORMAL , , PEAK , or BOTTOM key). The LED of the selected mode lights, confirming that the controller is now in that measurement mode. Note that the measurements in any of the P-P, PEAK, or BOTTOM mode can be reset by pressing the same mode selector key again.
NORMAL
Note
P±P
PEAK
The measurements can also be reset by pressing
BOTTOM
HOLD twice.
6-3 Calibrating Sensitivity (Gain Selection) Light sensitivity of the sensor head can be adjusted through gain selection. The LC calculates a measurement based on the intensity of reflected light from the target. Because the light intensity varies with the reflectance of the target surface, the sensitivity of the sensor head must be optimized accordingly. The system offers four levels of sensitivity, which can be identified by the GAIN LEDs on the display unit. The correlation between the GAIN LEDs and light sensitivity of the sensor head is as shown below. AUTO
1
2
3
4
GAIN
Reflectance of target High
Low
Factory setting: AUTO Reference: Displaying Light Intensity (P. 29)
30 CHAPTER 6 Setting Up
Page 38
6-3 Calibrating Sensitivity (Gain Selection) 6-4 Selecting the Display/Output Mode
1 Press the GAIN key. The AUTO LED and one of the four GAIN LEDs will light. When measurements are performed with these two LEDs on, the sensitivity will change automatically according to the light intensity received. AUTO
1
2 Press the Note
2
3
GAIN
4
GAIN key again to select the lit LED.
For a target whose reflectance varies greatly, the AUTO function can be selected. However, the measurement may show some variation when sensitivity is changed.
6-4 Selecting Display/Output Mode The 2CH mode cannot be used with the LC-2400W series. Any of the following modes can be selected when two sensor heads are used during measurements. Mode 1 CH 2 CH ADD
Function Displays/outputs the 1CH measurement. Displays/outputs the 2CH measurement. Adds the 1CH and 2CH measurements and displays/ outputs the negative of this result. – (l 1 + l 2) Subtracts the 2CH measurement from the 1CH measurement. (l 1 – l 2)
SUB
Factory setting: 1CH • Measurement example in the ADD mode
Thickness
1CH head CONTROL I/O l1
MONITOR
HEAD
No.
100~240V AC l2
INTENSITY
RS-232C
DISPLACEMENT
BATTERY
2CH head
LASER ON
LASER REMOTE
F.G.
LC±C1 FUSE 3.15A
• Measurement example in the SUB mode
1CH head l1
2CH head Height difference l2
CONTROL I/O
MONITOR
HEAD
No.
100~240V AC INTENSITY
RS-232C
DISPLACEMENT
BATTERY
LASER ON
LASER REMOTE
F.G.
LC±C1 FUSE 3.15A
CHAPTER 6 Setting Up
31
Page 39
6-4 Selecting Display/Output Mode 6-5 Correcting Measurement Error (Calibration)
• Only one of the four modes can be selected at a time. • None of these modes can be selected when only one sensor head is used. • In the ADD or SUB mode the 1CH intensity is displayed/output. To enter any of the above modes, press the desired key ( 1CH , 2CH , ADD , or SUB key) and the LED above the key pressed will light.
6-5 Correcting Measurement Error (Calibration) Calibrating the LC corrects the difference between the actual displacement and that measured by the LC. Always calibrate the LC after having moved the sensor head or having changed over the target types. There are two ways to calibrate the LC; calibration with a target (target positioning), and calibration without a target (entering a numerical value). Calibrate the LC using one of these methods. If frequent product changeover occurs, it is useful to record the calibration value for each product after calibration. The LC can then be re-calibrated simply by entering the calibration value recorded. Factory settings:
Calibration by Target Position
LC-2420: 200.00 µm LC-2440: 3.0000 mm
LC-2430: 500.00 µm LC-2450: 8.0000 mm
The LC is calibrated based on measurements of the target position. 1 Position the target within the measuring range.
Sensor head
Measuring range
2 Press the
Target
NORMAL
Actual displacement range key. The LED above the key will light.
3 Press the PRM key repeatedly until the CAL LED lights. The LED lights and the LC continues to measure.
LIMIT OFFSET
32 CHAPTER 6 Setting Up
CAL
INT. ANALOG BAUD LIMIT RATIO RATE
DIGIT
PRM
Page 40
6-5 Correcting Measurement Error (Calibration)
4 Shift the target to either end of the actual displacement range (closest or farthest from the sensor head). Sensor head
Shift the target to either end
Actual displacement range
Measuring range
• Be careful not to position the target outside of the measuring range. • Instead of the target, the sensor head may be shifted during calibration. 5. Press the ZERO key. The displayed value changes to zero. 6. Shift the target to the other end of the displacement range. Again, the sensor head may be shifted instead of the target. Shift the target to the other end of the displacement range.
Shift the target to the other end of the displacement range
Be careful not to position the target outside of the measuring range. key. 7 Press the The digits on the display start flashing and the LC switches to the calibration mode. Note
During calibration, pressing the PRM key, then the key will cancel calibration. When the calibration mode is canceled, the LC will return to the measurement mode. 8 Change the value on the display panel to the actual distance the target was moved. Reference: Entering Numerical Values (P.28)
9 Press the key. Calibration is complete. The LC returns to the measuring mode.
CHAPTER 6 Setting Up
33